← Product Code [MAV](/productcode/MAV) · K140745

# ENCORE 26 ADVANTAGE KIT (K140745)

_Boston Scientific Corp · MAV · Apr 17, 2014 · Cardiovascular · SESK_

**Canonical URL:** https://fda.innolitics.com/device/K140745

## Device Facts

- **Applicant:** Boston Scientific Corp
- **Product Code:** [MAV](/productcode/MAV.md)
- **Decision Date:** Apr 17, 2014
- **Decision:** SESK
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1650
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Encore™ 26 Advantage Kits are intended for use as accessories for percutaneous coronary angiography (PTCA) procedures. They create and monitor balloon inflation and facilitate wire introduction and control. Individual Device Intended Use: Encore™ 26 Inflation Device: used with balloon dilatation catheters to create and monitor pressure in the balloon, and to deflate the balloon. Gateway™ Plus Y-Adaptor: used for providing hemostasis around guidewires, balloon dilatation catheters, and other therapeutic devices. Torque Device: used for guidewire manipulation. Insertion Tool: used to facilitate the introduction of a guidewire.

## Device Story

Encore™ 26 Advantage Kit is a sterile, disposable procedural kit for PTCA and general intravascular procedures. Kit includes: inflation device, Y-adaptor, torque device, and insertion tool. Inflation device creates, monitors, and releases balloon pressure; Y-adaptor provides hemostasis; torque device aids guidewire manipulation; insertion tool facilitates guidewire introduction. Used in clinical settings (e.g., cath lab) by physicians or trained clinical staff. Device provides mechanical control and monitoring during interventional procedures; output is manual pressure regulation and physical guidance of catheters/wires. Benefits include procedural efficiency and maintenance of vascular access during coronary interventions.

## Clinical Evidence

Bench testing only. Design verification and validation testing confirmed performance and usability of the modified insertion tool are equivalent to the predicate. Sterilization, packaging, and biocompatibility testing verified overall substantial equivalence.

## Technological Characteristics

Kit of sterile, disposable mechanical accessories. Components: inflation device (pressure monitoring/actuation), Y-adaptor (hemostasis), torque device (guidewire manipulation), insertion tool (guidewire introduction). Materials: equivalent to predicates (including sheath and hub colorant). Sterilization: same method and SAL as predicates. No electronic or software components.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- Encore™ 26 Advantage Kit ([K123214](/device/K123214.md))
- SCIMED AVENUE Coronary Guidewire Insertion Tool ([K922410](/device/K922410.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K140745

Special 510(k) Submission

i

APR 1 7 2014

## 510(k) Summary

per 21 CFR §807.92

| Submitter's Address<br>Contact Name and<br>Information | Boston Scientific Corporation<br>One Scimed Place<br>Maple Grove, MN 55311<br>Lisa Mee<br>Senior Regulatory Affairs Specialist<br>Phone: 763-494-1185<br>Fax:<br>763-494-2222<br>e-mail: lisa.mee@bsci.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |  |  |
|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
|                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |  |  |
| Date Prepared                                          | 24 Mar 2014                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |  |  |
| Proprietary Name                                       | Encore™ 26 Advantage Kit                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |  |  |
| Common Name                                            | Balloon Inflation Kit<br>Comon name of the Kit components:<br>Inflation device, Insertion Tool, Y-Adaptor and Torque Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |  |  |
| Product Code                                           | MAV - Syringe, Balloon Inflation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |  |  |
| Classification                                         | Class II, 21 CFR Part 870.1650 - Cardiovascular                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |  |  |
| Predicate Device(s)                                    | Encore™ 26 Advantage Kit<br>K123214<br>Nov 13, 2012                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |  |  |
|                                                        | SCIMED AVENUE Coronary<br>K922410<br>Jul 23, 1992                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |  |  |
|                                                        | Guidewire Insertion Tool                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |  |  |
| Device Description                                     | The Encore™ 26 Advantage Kit is a Kit of sterile disposable<br>devices intended for use as accessories and percutaneous<br>coronary angiography (PTCA) procedures. They allow for<br>balloon inflation and wire control.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |  |  |
| Intended Use                                           | The Encore™ 26 Advantage Kits are intended for use as<br>accessories for percutaneous coronary angiography (PTCA)<br>procedures. They create and monitor balloon inflation and<br>facilitate wire introduction and control.<br>Individual Device Intended Use:<br>Encore™ 26 Inflation Device: used with balloon<br>●<br>dilatation catheters to create and monitor pressure in the<br>balloon, and to deflate the balloon.<br>Gateway™ Plus Y-Adaptor: used for providing<br>hemostasis around guidewires, balloon dilatation<br>catheters, and other therapeutic devices.<br>Torque Device: used for guidewire manipulation.<br>Insertion Tool: used to facilitate the introduction of a<br>guidewire. |  |  |
|                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |  |  |

{1}------------------------------------------------

| Indications for Use                               | Individual Device Indications for Use:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|---------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison of<br>Technological<br>Characteristics | The Encore™ 26 Inflation devices are recommended for<br>use with balloon dilatation catheters to create and<br>monitor pressure in the balloon, and to deflate the<br>balloon.  The GateWay™ PLUS Y-Adapter is recommended for<br>providing homeostasis around balloon dilatation<br>catheters, guidewires, and other therapeutic devices<br>during general intravascular procedures.  The Torque Device is used for guidewire manipulation<br>during general intravascular procedures.  The Insertion Tool is used to facilitate the introduction of<br>a guidewire during general intravascular procedures. |
|                                                   | The proposed Encore™ 26 Advantage Kit Inflation device, Y-<br>Adaptor and Torque device components incorporate<br>substantially equivalent design, fundamental technology,<br>manufacturing processes, packaging, sterilization and intended<br>use as the Inflation device, Y-Adaptor and Torque device<br>components featured in the Boston Scientific predicate Encore™<br>26 Advantage Kit (K123214).                                                                                                                                                                                                     |

The proposed Encore " 26 Advantage Kit Insertion Tool ce component incorporates substantially equivalent design, fundamental technology, manufacturing processes, packaging, sterilization and intended use as the SCIMED AVENUE Coronary Guidewire Insertion Tool device cleared for marketing under K922410.

### Comparison to Predicates:

.

| Characteristic           | Proposed compared to<br>Predicates                                                |
|--------------------------|-----------------------------------------------------------------------------------|
| Mechanism of Action      | Same mechanism of action.                                                         |
| Components               | Same components,<br>configuration, design and<br>function.                        |
| Materials                | Modified Insertion Tool has<br>equivalent materials (sheath<br>and hub colorant). |
| Packaging                | Same packaging materials and<br>packaging configuration.                          |
| Sterilization Method/SAL | Same method and level of<br>sterility assurance.                                  |
| Device Compatibility     | Same compatibility.                                                               |
| Device Dimensions        | Modified Insertion Tool has<br>equivalent dimensions.                             |
| Biocompatibility         | Same biocompatibility.                                                            |

{2}------------------------------------------------

Encore™ 26 Advantage Kit

Special 510(k) Submission

Performance Data

Design Verification and Design Validation Testing was performed to verify that the performance and usability of the modified Insertion Tool remains substantially equivalent to the predicate device via K140673. In addition Sterilization, Packaging and Biocompatibility testing verifies the overall substantial equivalence to the kit predicate. No additional testing was required for kit inclusion.

No new safety or performance issues were raised during the device testing. Therefore, these devices may be considered substantially equivalent to the predicate device.

#### Conclusion

Based on the indications for use, technological characteristics, safety and performance testing, the proposed Encore™ 26 Advantage Kit has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the Encore™ 26 Advantage Kit as submitted in K123214 and the SCIMED AVENUE Coronary Guidewire Insertion Tool as submitted in K922410.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.

Public Health Scrvice

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

## April 17, 2014

Boston Scientific Corporation c/o Lisa Mee Senior Regulatory Affairs Specialist One Scimed Place Maple Grove, MN 55311

Re: K140745

Trade/Device Name: Encore™ 26 Advantage Kit Regulation Number: 21 CFR 870.1650 Regulation Name: Balloon Inflation Kit Regulatory Class: II Product Code: MAV Dated: March 24. 2014 Received: March 25, 2014

Dear Ms. Mee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be

{4}------------------------------------------------

Page 2 - Ms. Lisa Mee

found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Bram D. Zuckerman -S

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K140745

Device Name: Encore™ 26 Advantage Kit

Indications for Use:

-

The Encore 26 Advantage Kit is a kit of sterile disposable devices intended of ruse as accessories for percutaneous coronary angiography (PTCA) procedures. They allow for balloon inflation and wire control.

- Encore™ 26 Inflation Device: indicated for use with balloon dilatation . catheters to create and monitor pressure in the balloon, and to deflate the balloon.
- Gateway™ Plus Y-Adaptor: used for providing hemostasis around balloon . dilatation catheters, guidewires and other therapeutic devices during general intravascular procedures.
- Torque Device: used for guidewire manipulation during general ◆ intravascular procedures.
- Insertion Tool: used to facilitate the introduction of a quidewire during . general intravascular procedures.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR .

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

# Bram D. Zuckerman -S 2014.04.17 17:31:34 -04'00'

Boston Scientific Corporation

. Attachment 1

---

**Source:** [https://fda.innolitics.com/device/K140745](https://fda.innolitics.com/device/K140745)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
