K140730 · Medimop Medical Projects, Ltd. · LHI · Jun 12, 2014 · General Hospital
Device Facts
Record ID
K140730
Device Name
VIAL2BAG DIRECT CONNECT
Applicant
Medimop Medical Projects, Ltd.
Product Code
LHI · General Hospital
Decision Date
Jun 12, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Intended Use
The Vial2Bag™ Direct Connect is indicated to serve as a connecting part between the IV bag and an external IV line. The integrated vial adapter is used to reconstitute and admix drugs from a vial into the infusion solution.
Device Story
Vial2Bag™ Direct Connect is a mechanical accessory for IV administration; consists of a piercing spike, cover, twist-off connector, and integrated vial adapter. Used in healthcare facilities or home settings by caregivers to facilitate drug reconstitution and admixture from a vial into an infusion solution. Device connects an IV bag to an external IV line. No medicinal substances contained within device. Operation involves physical attachment to vial and IV bag to enable fluid transfer. Benefits include simplified, closed-system drug preparation and administration, reducing potential for contamination or handling errors during reconstitution.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via air leakage, detachment force, breaking torque, penetration force, and spike damage tests. Biocompatibility testing performed per ISO standards for external communicating devices with prolonged indirect blood contact (24 hours to 30 days), including cytotoxicity, sensitization, irritation, hemocompatibility, acute systemic toxicity, and pyrogenicity testing. All tests passed acceptance criteria.
Technological Characteristics
Mechanical device; no electronics or software. Materials classified for prolonged indirect blood contact (24 hours to 30 days). Sterilization via double ethylene oxide (EtO) process. Components include piercing spike, twist-off connector, and integrated vial adapter.
Indications for Use
Indicated for use in healthcare facilities or home environments by caregivers to connect IV bags to external IV lines and to reconstitute/admix drugs from vials into infusion solutions.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
K170095 — Vial2Bag Direct Connect · Medimop Medical Project , Ltd. · Mar 8, 2017
K072759 — VIAL2BAG · Medimop Medical Projects, Ltd. · Dec 13, 2007
K201415 — Vial2Bag Advanced 20mm Admixture Device · West Pharmaceutical Services Az, Inc. · Oct 21, 2020
K240940 — Vial2Bag Advanced® 20mm Admixture Device · West Pharmaceutical Services, Inc. · May 3, 2024
K212525 — Advcare Vial Direct to Bag (VDB) Needle-free Admixture Devices · Advcare Medical · Aug 19, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
#### 510(K) SUMMARY 5
#### Vial2Bag™ Direct Connect Device:
## Company Name:
Medimop Medical Projects Ltd 17 Hatidhar St Ra'anana 4366519 Israel Phone: 972-9-7778666 Fax: 972-9-7485916
# Contact person:
Ilanit Goldgraber Director of Regulatory Phone: 972-9-7778666 Fax: 972-9-7485916 E-mail: ilanit.goldgraber@westpharma.com
# US Agent:
Kevin Lentz Director of Regulatory/PDS group West Pharmaceutical Services, Inc 530 Herman O. West Drive Exton, PA 19341 Phone: (610) 594-4353 E-mail: kevin.lentz@westpharma.com
....
Preparation date: March 17, 2014
{1}------------------------------------------------
## Classification:
Classification name: IV Administration Set Trade name: Vial2Bag™ Direct Connect Common/usual name: Vial2Bag" Direct Connect Product Code: LHI Regulation No.: 880.5440 Class: II Panel identification: General Hospital Panel
Predicate Devices:
Vial2Bag" from Medimop Medical Projects, Ltd, cleared by 510(k) number K072759
Mixject Dispensing Pin (Vial Adapter) from Medimop Medical Projects, Ltd cleared by 510(k) number K001293
#### Device Description:
The proposed device, Vial2Bag" Direct Connect, consists of the Vial2Bag" piercing spike and cover, the twist-off connector and an integrated Vial Adapter for access to the drug/solution vial. It does not contain any medicinal substances and can be used with standard drug vials. It is intended for use in healthcare facilities or in home environment by a care-giver to aid and support prescribed treatment and therapy.
#### Indications for use:
The proposed device, Vial2Bag™ Direct Connect is indicated to serve as a connecting part between the IV bag and an external IV line. The integrated vial adapter is used to reconstitute and admix drugs from a vial into the infusion solution.
#### Substantial equivalence
The proposed device. Vial2Bag™ Direct Connect has the same intended use, materials of construction and principle of operation as the predicate devices. Vial2Bag™ device (K072759) and the Mixject Dispensing Pin (Vial Adapter) (K001293) manufactured by Medimop Medical Projects Ltd. Therefore, the Vial2Bag™ Direct Connect is substantially equivalent to the predicate devices.
17 March 2014
{2}------------------------------------------------
# Vial2Bag™ Direct Connect
### Performance Testing
Bench testing was performed to verify design requirements and to mitigate any potential risks. All bench testing was performed following double EtO sterilization to ensure that the sterilization process does not have a detrimental effect on the proposed device, Vial2Bag™ Direct Connect performance. The following tests were completed and all of them passed the established acceptance criteria.
- Air leakage test ●
- Vial adapter detachment force from Vial2Bag™ body .
- Vial adapter breaking torque �
- Vial adapter total vial penetration force .
- Vial adapter detachment force from vial �
- Vial adapter spike damage
### Biocompatibility Tests
The proposed device. Vial2Bag™ Direct Connect, is classified as an external communicating device with prolonged indirect blood contact (24 hours to 30 days). The following tests were performed. All the tests passed the established acceptance criteria.
- Cytotoxicity (MEM Elution -ISO) .
- Sensitization (ISO Guinea Pig Maximization Sensitization Test) .
- . Irritation / Intracutaneous (ISO Intracutaneous Study in Rabbits)
- Hemocompatibility (In Vitro Hemolysis Study Extraction Method) .
- Acute Systemic Toxicity (ISO Systemic Toxicity Study in Mice) ●
- · Material Mediated Pyrogenicity (USP Rabbit Pyrogen Extraction Method)
- Bacterial Pyrogenicity (LAL Test Method) .
#### Conclusion
ﺭ
Evaluation of the proposed device. Vial2Bag™ Direct Connect does not raise any additional concerns regarding safety and effectiveness. Therefore, it is considered substantially equivalent to the predicate devices, the Vial2Bag" device (K072759) and the Mixject Dispensing Pin (Vial Adapter) (K001293).
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged in a circular fashion around an emblem. The emblem is a stylized representation of an eagle or bird-like figure, with three curved lines forming its body and wings. The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 12, 2014
Medimop Medical Projects Limited Mr. Ilanit Goldgraber 17 Hatidhar Street Ra'anana 4366519 ISRAEL
Re: K140730
Trade/Device Name: Vial2Bag Direct™ Connect Regulation Number: 21 CFR 880.5440 Regulation Name: Vial2Bag Direct™ Connect Regulatory Class: II Product Code: LHI Dated: March 17, 2014 Received: March 24, 2014
Dear Mr. Goldgraber:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Goldgraber
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21.CFR Part.801); medical device reporting (reporting of medical . device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Mary-S.Runner-S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K140730
#### Device Name
-Vial2Bag™ Direct Connect-
#### Indications for Use (Describe)
The Vial2Bag™ Direct Connect is indicated to serve as a connecting part between the IV line. The integrated vial adapter makes it possible to reconstitute and admix drugs from a vial into the infusion solution.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ FOR FDA USE ONLY ... . . . .
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/5/Picture/13 description: The image shows the letters FDA in a stylized font. The letters are bold and have a textured appearance. The letters are arranged horizontally, with the F on the left, followed by the D, and then the A on the right.
Digitally signed by Richard C. Chapman -S Date: 2014.06.12 14:06:42 -04'00'
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.