← Product Code [HRS](/productcode/HRS) · K140724

# MSP METATARSAL SHORTENING SYSTEM (K140724)

_Nextremity Solutions, Inc. · HRS · May 14, 2014 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K140724

## Device Facts

- **Applicant:** Nextremity Solutions, Inc.
- **Product Code:** [HRS](/productcode/HRS.md)
- **Decision Date:** May 14, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Nextremity Solutions MSP™ System is indicated for fixation of fractures, osteotomies, nonunions, malunions and fusions of small bones and small bone segments, particularly in osteopenic bone. Examples include, but not limited to, the hand, foot (shortening of the lesser metatarsal) and ankle.

## Device Story

MSP Metatarsal Shortening System consists of titanium bone plates and cortical/locking screws; includes surgical site preparation and insertion instruments. Used by surgeons for internal fixation of small bone fractures, osteotomies, non-unions, malunions, and fusions. Provides mechanical stability to bone segments during healing; particularly suited for osteopenic bone. Operates via standard orthopedic fixation techniques.

## Clinical Evidence

Bench testing only. Mechanical performance validated via 4-point bending (static and dynamic) per ASTM F-382 and screw torque-to-failure per ASTM F-543. Axial push-out testing performed to compare screw designs.

## Technological Characteristics

Materials: Medical grade titanium. Components: Bone plates, cortical screws, locking screws. Mechanical testing: ASTM F-382 (4-point bending), ASTM F-543 (torque to failure).

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Modular Mini Fragment LCP System ([K063049](/device/K063049.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# MAY 1 4 2014

K140724 page 1 of 2

# 510(k) Summary (Per 21 CFR 807.92)

General Company Information:

Nextremity Solutions, Inc. Jorge A. Montoya Director, Product Development 54 Broad Street, Suite 200 Red Bank, NJ 07701 Phone: (732) 383-7901 Fax: (732) 865-7632

#### Date Prepared

February 20, 2014

# General Device Information

Product Name:

Classification:

MSPTM Metatarsal Shortening System

Single/multiple component metallic bone fixation appliances and accessories 21 CFR 888.3030 Product code: HRS

Smooth or Threaded metallic bone fixation fastener and accessories 21 CFR 888.3040 Product code: HWC Class II device

### Predicate Devices

Synthes, Inc.

Modular Mini Fragment LCP System (Marketed as Modular Mini Fragment LCP System) [510(k) K063049]

#### Description

The Nextremity Solutions MSP™ Metatarsal Shortening System is a set, consisting of:

- 1. A bone plate.
- 2. Specific length, cortical and locking screws.
- Necessary surgical site preparation and insertion instruments (as a procedure pack). 3.

The cortical and locking screw(s) are used in conjunction with the bone plate and are individually packaged.

{1}------------------------------------------------

The plate and screws are fabricated from medical grade Titanium and the design allows for fixation of fractures, osteotomies, non-unions, malunions and fusions of small bones and small bone segments found in the foot, particularly in osteopenic bone.

# Intended Use (Indications)

The Nextremity Solutions MSP™ System is indicated for fixation of fractures, osteotomies, nonunions, malunions and fusions of small bones and small bone segments, particularly in osteopenic bone. Examples include, but not limited to, the hand, foot (shortening of the lesser metatarsal) and ankle.

#### Substantial Equivalence

The Nextremity Solutions, MSP™ Metatarsal Shortening System possesses the same technological characteristics of the predicate devices. These characteristics include the intended use, basic design, material, size and fundamental technology.

# Performance Data

Mechanical testing was performed as described in relevant recognized standards, including 4 point bending (static and dynamic) for the MSP plate per ASTM F-382 and torque to failure for the screws per ASTM F-543. An axial push-out test was implemented to properly compare the predicate device to the proposed MSP screw designs given the short length of the socies.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the department's name around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure embracing another figure. The symbol is composed of three curved lines.

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G66 Silver Spring, MD 20993-0002

May 14, 2014

Nextremity Solutions, Incorporated Mr. Jorge A. Montoya Director, Product Development 54 Broad Street, Suite 200 Red Bank, New Jersey 07701

Re: K140724

Trade/Device Name: MSP™ Metatarsal Shortening System Regulation Number: 21 CFR 888.3030 Regulation Name: Single multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: March 27, 2014 Received: March 28, 2014

Dear Mr. Montoya:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. Jorge A. Montoya

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/dcfault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Ronald P. Jean -S for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### K140724 page 1 of 1

# Indications for Use Statement

510(k) Number (if known); K140724

Device Name: MSPTM Metatarsal Shortening System

Indications For Use:

The Nextremity Solutions MSP™ System is indicated for fixation of fractures, osteotomies, nonunions, malunions and fusions of small bones and small bone segments, particularly in osteopenic bone. Examples include, but not limited to, the hand, foot (shortening of the lesser metatarsal) and ankle.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

### Concurrence of CDRH, Office of Device Evaluation (ODE)

Elizabeth L鼎ank -S

Division of Orthopedic Devices

---

**Source:** [https://fda.innolitics.com/device/K140724](https://fda.innolitics.com/device/K140724)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
