← Product Code [MED](/productcode/MED) · K140703

# OPA 30 SOLUTION (K140703)

_Ciden Technologies, LLC · MED · Aug 19, 2014 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K140703

## Device Facts

- **Applicant:** Ciden Technologies, LLC
- **Product Code:** [MED](/productcode/MED.md)
- **Decision Date:** Aug 19, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6885
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

OPA 30 Solution is a high level disinfectant for reprocessing heat-sensitive semi-critical medical devices for which sterilization is not suitable, and when used according to the Directions for Use. OPA 30 Solution may be used or reused at or above its Minimum Recommended Concentration (MRC) of 0.3%, as determined by Opaciden OPA Reagent Strips, in manual reprocessing with an immersion time of at least 12 minutes at a minimum of 20°C (68°F), for a reuse period not to exceed 30 days. OPA 30 Solution may also be used or reused in a legally marketed automatic endoscope reprocessor (that can be set to a minimum of 25°C), at or above its Minimum Recommended Concentration (MRC), as determined by Opaciden OPA Reagent Strips, with an immersion time of at least 5 minutes at a minimum of 25°C (77°F), for a reuse period not to exceed 14 days.

## Device Story

OPA 30 Solution is a liquid chemical high-level disinfectant containing 0.60% ortho-Phthalaldehyde (OPA) in an aqueous base with buffers, chelating agents, and corrosion inhibitors. It is used to reprocess heat-sensitive semi-critical medical devices. The solution is applied via manual immersion or in legally marketed automatic endoscope reprocessors. Users verify the Minimum Recommended Concentration (MRC) of 0.3% using Opaciden OPA Reagent Strips before use. The device is intended for clinical settings where medical instruments require disinfection. By effectively eliminating vegetative organisms, fungi, viruses, and spores, the solution ensures instrument safety for subsequent patient use, reducing the risk of cross-contamination.

## Clinical Evidence

No clinical data included in this submission. Performance is supported by non-clinical bench testing for stability, reuse, and material compatibility.

## Technological Characteristics

Clear, pale blue liquid, pH 7.5. Active ingredient: 0.60% ortho-Phthalaldehyde. Contains buffers, chelating agents, and corrosion inhibitors. Compatible with Olympus and Pentax endoscopes and Acrylonitrile Butadiene Styrene (ABS). Stability: 24-month shelf life. Manual reuse: 30 days; Automated reuse: 14 days. Verification via Opaciden OPA Reagent Strips.

## Regulatory Identification

A liquid chemical sterilant/high level disinfectant is a germicide that is intended for use as the terminal step in processing critical and semicritical medical devices prior to patient use. Critical devices make contact with normally sterile tissue or body spaces during use. Semicritical devices make contact during use with mucous membranes or nonintact skin.

## Special Controls

*Classification.* Class II (special controls). Guidance on the Content and Format of Premarket Notification (510(k)) Submissions for Liquid Chemical Sterilants/High Level Disinfectants, and user information and training.

## Predicate Devices

- Opaciden® Solution ([K070627](/device/K070627.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-000

August 19, 2014

CIDEN Technologies, LLC c/o Courtney Clark, CBA, RAC Regulatory Affairs Manager of Submissions Sybron Dental Specialties 1717 W. Collins Ave Orange, CA 92867

Re: K140703

> Trade/Device Name: OPA 30 Solution Regulation Number: 21 CFR 880.6885 Regulation Name: Liquid chemical sterilant/high level disinfectants Regulatory Class: II Product Code: MED Dated: July 22, 2014 Received: July 23, 2014

Dear Ms. Clark,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital. Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K140703

Device Name OPA 30 Solution

### Indications for Use (Describe)

OPA 30 Solution is a high level disinfectant for reprocessing heat sensitive semi-critical medical devices for which sterilization is not suitable, and when used according to the Directions for Use. OPA 30 Solution may be used of reused at or above its Minimum Recommended Concentration (MRC) of 0.3%, as determined by Opaciden OPA Reagent Strips, in manual reprocessing with an immersion time of at least 12 minutes at a minimum of 20°C (68°F), for a reuse period not to exceed 30 days. OPA 30 Solution may also be used or reused in a legally marketed automatic endoscope reprocessor (that can be set to a minimum of 25°C), at or above its Minimum Recommended Concentration (MRC), as determined by Opaciden OPA Reagent Strips, with an immersion time of at least 5 minutes at a minimum of 25℃ (77ºF), for a reuse period not to exceed 14 days.

### Type of Use (Select one or both, as applicable)

- Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## SECTION 5. 510(k) SUMMARY for OPA 30 Solution (K140703)

- 1. Submitter Information: Ciden Technologies, LLC 360 Cold Spring Avenue West Springfield, MA 01089
Contact Person: Courtney Clark Telephone Number: 714-516-7426 Fax Number: 714-516-7472

Date Prepared: 22 July 2014

2. Device Name:

| Proprietary Name:      | OPA 30 Solution                                   |
|------------------------|---------------------------------------------------|
| Classification Name:   | Liquid chemical germicide/high level disinfectant |
| CFR Number:            | 21 CFR 880.6885                                   |
| Device Class:          | II                                                |
| Product Code:          | MED                                               |
| FDA review panel code: | INCB                                              |

#### 3. Predicate Device:

The legally marketed predicate device is Opaciden® Solution (K070627), Product Code MED, manufactured by Ciden Technologies LLC and determined to be substantially equivalent to a legally marketed device on August 2, 2007.

#### 4. Description of Device:

The OPA 30 Solution is a clear, pale blue liquid with a pH of 7.5. It contains 0.60% ortho-Phthalaldehyde in an aqueous base containing buffers, chelating agents and a corrosion inhibitor. It is stable at 15-30° C (59-86° F) for its labeled shelf life of 24 OPA 30 Solution may be used or reused for manual (up to 30 days) or months. automated reprocessing (up to 14 days), according to the Directions for Use. OPA 30 Solution must be used at or above its Minimum Recommended Concentration (MRC), as determined by Opaciden OPA Reagent Strips, with the immersion time and temperature specified for the disinfection method.

- న. Statement of Intended Use:
OPA 30 Solution is a high level disinfectant for reprocessing heat-sensitive semi-critical medical devices for which sterilization is not suitable, and when used according to the Directions for Use. OPA 30 Solution may be used or reused at or above its Minimum Recommended Concentration (MRC) of 0.3%, as determined by Opaciden OPA Reagent Strips, in manual reprocessing with an immersion time of at least 12 minutes at a minimum of 20° C (68° F), for a reuse period not to exceed 30 days. OPA 30 Solution may also be used or reused in a legally marketed automatic endoscope reprocessor (that can be set to a minimum of 25℃), at or above its Minimum Recommended Concentration (MRC), as determined by Opaciden OPA Reagent Strips, with an 

{4}------------------------------------------------

immersion time of at least 5 minutes at a minimum of 25°C (7°P), for a reuse period not to exceed 14 days.

#### 6. Description of Safety and Substantial Equivalence:

## Technological Characteristics

The chemical formulation and technical characteristics of the proposed OPA 30 Solution are identical to those of the predicate Opaciden® Solution (K070627). The indications for use, material compatibility and shelf life have been expanded for the proposed OPA 30 Solution due to the results of additional performance testing. There are no substantial technical or functional differences between the proposed OPA 30 Solution and the predicate (K070627) device. Based on this information and the results of the performance testing, it is expected that the proposed OPA 30 Solution will be as safe and effective as the predicate Opaciden® Solution (K070627) when used as directed.

| ELEMENT                                    | PREDICATE DEVICE:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | PROPOSED DEVICE:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                            | Opaciden® Solution<br>(K070627)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | OPA 30 Solution                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| CLASSIFICATION NAME<br>(CFR; Product code) | Liquid chemical germicide/high<br>level disinfectant (21 CFR §<br>880.6885, Product Code MED)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Liquid chemical germicide/high<br>level disinfectant (21 CFR §<br>880.6885, Product Code MED)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| INDICATIONS FOR USE                        | Opaciden® Solution (K070627)<br>is a high level disinfectant for<br>reprocessing heat-sensitive semi-<br>critical medical devices for<br>which sterilization is not suitable,<br>and when used according to the<br>Directions for Use. Opaciden®<br>Solution may be used or reused<br>at or above its Minimum<br>Recommended Concentration<br>(MRC) of 0.3%, as determined<br>by Opaciden OPA Reagent<br>Strips, in manual reprocessing<br>with an immersion time of at<br>least 12 minutes at a minimum of<br>20°C (68°F), for a reuse period<br>not to exceed 14 days.<br>Opaciden® Solution may also<br>be used or reused in a legally<br>marketed automatic endoscope<br>reprocessor (that can be set to a<br>minimum 25°C), at or above its<br>Minimum Recommended<br>Concentration (MRC), as<br>determined by Opaciden OPA<br>Reagent Strips, with an<br>immersion time of at least 5<br>minutes at a minimum of 25°C<br>(77°F), for a reuse period not to<br>exceed 14 days. | OPA 30 Solution is a high level<br>disinfectant for reprocessing<br>heat-sensitive semi-critical<br>medical devices for which<br>sterilization is not suitable, and<br>when used according to the<br>Directions for Use. OPA 30<br>Solution may be used or reused<br>at or above its Minimum<br>Recommended Concentration<br>(MRC) of 0.3%, as determined<br>by Opaciden OPA Reagent<br>Strips, in manual reprocessing<br>with an immersion time of at<br>least 12 minutes at a minimum of<br>20°C (68°F), for a reuse period<br>not to exceed 30 days. OPA 30<br>Solution may also be used or<br>reused in a legally marketed<br>automatic endoscope reprocessor<br>(that can be set to a minimum of<br>25°C), at or above its Minimum<br>Recommended Concentration<br>(MRC), as determined by<br>Opaciden OPA Reagent Strips,<br>with an immersion time of at<br>least 5 minutes at a minimum of<br>25°C (77°F), for a reuse period<br>not to exceed 14 days. |
| ACTIVE INGREDIENT                          | ortho-Phthalaldehyde 0.60%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | ortho-Phthalaldehyde 0.60%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| ELEMENT                                    | PREDICATE DEVICE:<br>Opaciden® Solution<br>(K070627)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | PROPOSED DEVICE:<br>OPA 30 Solution                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| INERT INGREDIENTS                          | 99.40%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 99.40%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| PHYSICAL<br>PROPERTIES                     | Clear, pale blue liquid with a pH<br>of 7.5. Contains 0.60% OPA in<br>an aqueous base containing<br>buffers, chelating agents and a<br>corrosion inhibitor. It is stable at<br>15-30° C (59-86° F) for its<br>labeled shelf life.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Clear, pale blue liquid with a pH<br>of 7.5. Contains 0.60% OPA in<br>an aqueous base containing<br>buffers, chelating agents and a<br>corrosion inhibitor. It is stable at<br>15-30° C (59-86° F) for its<br>labeled shelf life.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| MINIMUM EFFECTIVE<br>CONCENTRATION (MEC)   | 0.3%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 0.3%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| REUSE PERIOD                               | 14 Days Manual, 14 days<br>automated                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 30 Days Manual, 14 days<br>automated                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| MICROBIOLOGY                               | Effective against:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Effective against:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Vegetative organisms                       | Staphylococcus aureus,<br>Salmonella choleraesuis,<br>Pseudomonas aeruginosa,<br>Mycobacterium bovis                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Staphylococcus aureus,<br>Salmonella choleraesuis,<br>Pseudomonas aeruginosa,<br>Mycobacterium bovis                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Fungi                                      | Trichophyton mentagrophytes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Trichophyton mentagrophytes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Viruses, non-enveloped                     | Poliovirus Type 1<br>Rhinovirus Type 42<br>Hepatitis A (AER)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Poliovirus Type 1<br>Rhinovirus Type 42<br>Hepatitis A (AER)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Viruses, enveloped                         | Avian influenza<br>Hepatitis B<br>Influenza Type A (Hong Kong)<br>HIV-1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Avian influenza<br>Hepatitis B<br>Influenza Type A (Hong Kong)<br>HIV-1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Spores                                     | Herpes simplex Type 1<br>Clostridium sporogenes<br>Bacillus subtilis                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Herpes simplex Type 1<br>Clostridium sporogenes<br>Bacillus subtilis                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| TOXICITY                                   | Positive for acute oral toxicity,<br>moderate skin irritation.<br>Negative for dermal and<br>cytotoxicity. Probable eye<br>irritant.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Positive for acute oral toxicity,<br>moderate skin irritation.<br>Negative for dermal and<br>cytotoxicity. Probable eye<br>irritant.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| REUSABLE DEVICE<br>COMPATIBILITY           | Olympus and Pentax endoscopes<br>are compatible                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Olympus and Pentax endoscopes<br>are compatible                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| CHEMICAL INDICATOR                         | Opaciden OPA Reagent Strips                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Opaciden OPA Reagent Strips                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| SHELF LIFE                                 | 15 month shelf life for a closed<br>container.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 24 month shelf life for a closed<br>container.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| MATERIAL<br>COMPATIBILITY                  | Materials previously cleared in<br>the predicate Opaciden®<br>Solution<br>(K070627)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Materials previously cleared in<br>the predicate Opaciden®<br>Solution<br>(K070627)<br>and<br>Acrylonitrile Butadiene<br>Styrene (ABS)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |

## Device Comparison Table

{5}------------------------------------------------

{6}------------------------------------------------

## Non-Clinical Performance Data

The OPA 30 Solution has been tested for stability, re-use for manual processing and material compatibility following:

- . Guidance for Industry and FDA Reviewers: "Content and Format of Premarket Notification [510(k)] Submission for Liquid Chemical Sterilants/High Level Disinfectants" Section III
Due to the fact that the chemical formulation of the proposed OPA 30 Solution and the predicate Opaciden® Solution (K070627) is identical, additional testing requirements described in the FDA guidance "Content and Format of Premarket Notification [510(k)] Submission for Liquid Chemical Sterilants/High Level Disinfectants" have not been repeated. The required testing was previously submitted, reviewed and cleared in K070627. Results for the applicable tests were acceptable and demonstrated performance in accordance with the proposed OPA 30 Solution product labeling.

## Clinical Performance Data

There is no clinical data included in this submission.

## Conclusion as to Substantial Equivalence

The OPA 30 Solution is as safe, as effective, and performs as well as the predicate Opaciden® Solution (K070627), and is therefore determined to be substantially equivalent to Opaciden® Solution (K070627).

---

**Source:** [https://fda.innolitics.com/device/K140703](https://fda.innolitics.com/device/K140703)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
