← Product Code [FLL](/productcode/FLL) · K140681

# BELTER INFRA-RED EAR THERMOMETER (K140681)

_Shenzhen Belter Health Measurement and Analysis TE · FLL · Aug 21, 2014 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K140681

## Device Facts

- **Applicant:** Shenzhen Belter Health Measurement and Analysis TE
- **Product Code:** [FLL](/productcode/FLL.md)
- **Decision Date:** Aug 21, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.2910
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Pediatric

## Indications for Use

The Belter Infra-red Ear Thermometer is an electronic thermometer using an infrared sensor to detect human body temperature from the ear canal on people of all ages for home and professional use.

## Device Story

The Belter Infra-red Ear Thermometer is a handheld electronic device used to measure human body temperature via the ear canal. It utilizes an infrared sensor to detect infrared energy emitted from the eardrum tissue. The device is operated by clinicians or patients in home or professional settings. The user positions the probe in the ear canal; the device captures the infrared signal and converts it into a temperature reading displayed on an LCD screen. It includes features such as a fever alarm, buzzer, and memory for 10 sets of readings. The device provides rapid temperature assessment, aiding healthcare providers or patients in monitoring health status and identifying potential fever, facilitating clinical decision-making regarding patient care.

## Clinical Evidence

Bench testing and clinical repeatability data provided. Laboratory accuracy reported as 0.01–0.19°C (meets ISO 80601-2-56 requirement of ≤0.2°C). Clinical repeatability reported as 0.07–0.11°C (meets ISO 80601-2-56 requirement of ≤0.3°C). Structural integrity verified via shock and broad-band random vibration testing.

## Technological Characteristics

Infrared sensor-based electronic thermometer. Powered by 2x AAA batteries. Features LCD display, fever alarm, and 10-set memory. Conforms to ISO 80601-2-56 (performance), EN 60601-1 (safety), IEC 60601-1-2 (EMC), and ISO 10993-1, 5, 10 (biocompatibility).

## Regulatory Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

## Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; 
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and 
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

## Predicate Devices

- Infrared Ear Thermometer THP Series ([K111637](/device/K111637.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is an image of three stylized human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 21, 2014

Shenzhen Belter Health Measurement and Analysis Technology Company, Limited Pang Ming RA Manager 702, 704, Block C, Tsinghua Unis Science Park, Langshan Rd, Hi-Tech Industrial Park (north), Nanshan District, Shenzhen, Gangdong CHINA 518057

Re: K140681 Trade/Device Name: Belter Infra-red Ear Thermometer, Models TE-66, TE-68, TE-79, TE-90, and TE-91 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical electronic thermometer Product Code: FLL Regulatory Class: II Dated: July 15, 2014 Received: July 21, 2014

Dear Mr. Ming:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Mr. Ming

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mary S. Runner -S

Erin Keith, M.S. Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K140681

Device Name Belter Infra-red Ear Thermometer Model: TE-66,TE-68, TE-79, TE-90, TE-91

#### Indications for Use (Describe)

The Belter Infra-red Ear thermometer is an electronic thermometer using an infrared sensor to detect human body temperature from the ear canal on people of all ages for home and professional use.

Type of Use (Select one or both, as applicable)

- Prescription Use (Part 21 CFR 801 Subpart D)

> Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for eBelter. The logo consists of the letter 'e' in orange, followed by the word 'BELTER' in blue. The 'e' is stylized with a circular shape, and the word 'BELTER' is in a bold, sans-serif font.

# Section 5

# 510 (K) Summary

## Date: 2014,1,15 Submitter Information

Name : Shenzhen Belter Health Measurement and Analysis Technology Co.,Itd Address: 702、704,block C ,Tsinghua Unis science park,No.13 Langshan Rd,Hi-Tech Industrial park(north),Nanshan District,518057 shenzhen. People's Republic of China Tel: +86-755-61869839 Fax: +86-755-61869122 Contact person: pang minq E-mail: panqm@ebelter.com

## Name of device

Trade name: Belter Infra-red Ear Thermometer Model: TE-66,TE-68, TE-79, TE-90, TE-91 Common name: Belter Infra-red Ear Thermometer Classification name: Thermometer, Electronic, Clinical Production regulation: 21 CFR 880.2910 Product code: FLL

## Predicate Device

Infrared Ear Thermometer THP Series (Radiant Innovation Inc,FLL,K111637)

## Description

The Belter Infra-red Ear Thermometer is characterized by measuring human body temperature from the ear canal. It utilizes infrared technology to measure infrared energy emitted from eardrum tissue when making a temperature measurement.

### Intended use

The Belter Infra-red Ear Thermometer is an electronic thermometer using an infrared sensor to detect human body temperature from the ear canal on people of all ages for home and professional use.

## Summary of technological characteristics of device compared to the predicate devices (K111637), see the table1

| Table 1     |                     |                  |         |
|-------------|---------------------|------------------|---------|
| SE          | Subject device      | Predicate device | comment |
| Comparisons | Present application | K111637          | nt      |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "BELTER" in blue font with an orange "e" symbol to the left of the word. The "e" symbol is a stylized letter "e" inside of a circle. The word "BELTER" is in all capital letters and has a slightly rounded font. The image is likely a logo or brand name.

| Intended Use&<br>Indication for use  | The Belter Infra-red Ear<br>Thermometer is an<br>electronic thermometer<br>using an infrared sensor<br>to detect human body<br>temperature from the<br>ear canal on people of<br>all ages for home and<br>professional use. | The device is an electronic clinical<br>thermometer using an infrared<br>sensor to detect body<br>temperature from the auditory<br>canal in the neonatal, pediatric<br>and adult population used in the<br>home setting. | Similar |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Measurement<br>Range                 | 32.0 °C -42.2℃ (89.6 °F<br>-108.0°F)                                                                                                                                                                                        | 34.0°C ~42.2°C<br>(93.2°F ~108.0°F)                                                                                                                                                                                      | Similar |
| Accuracy                             | ±0.2℃ (36℃≤t<39) ;<br>±0.3°Cother range                                                                                                                                                                                     | ±0.2°C (0.4°F) within<br>35.5~42°C (95.9~107.6°F),<br>±0.3°C(0.5°F) other range                                                                                                                                          | Similar |
| Response time                        | 1 sec.                                                                                                                                                                                                                      | 1 sec.                                                                                                                                                                                                                   | same    |
| Measurement<br>place                 | ear                                                                                                                                                                                                                         | ear                                                                                                                                                                                                                      | same    |
| Scale selection                      | °C /°F                                                                                                                                                                                                                      | C /°F                                                                                                                                                                                                                    | same    |
| Display screen                       | LCD                                                                                                                                                                                                                         | LCD                                                                                                                                                                                                                      | same    |
| Memory                               | 10 sets                                                                                                                                                                                                                     | 25 sets                                                                                                                                                                                                                  | Similar |
| Fever alarm                          | Yes                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                      | same    |
| Disposable<br>probe cover            | No                                                                                                                                                                                                                          | MOLD RING W/PE FILM<br>COVER                                                                                                                                                                                             | ----    |
| Backlight                            | optional                                                                                                                                                                                                                    | optional                                                                                                                                                                                                                 | Same    |
| Buzzer                               | Yes                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                      | same    |
| Auto power-off<br>while<br>operation | no Yes                                                                                                                                                                                                                      | Yes                                                                                                                                                                                                                      | Same    |
| Power supply                         | 2 x AAA                                                                                                                                                                                                                     | CR2032 *1pcs                                                                                                                                                                                                             | ---     |
| Conformance<br>standard              | ISO80601-2-56(performance),<br>EN60601-1(Safety),<br>IEC60601-1-2(EMC)<br>ISO<br>10993-1,5,10(Biocompatibility)                                                                                                             | ISO80601-2-56(performance),<br>EN60601-1(Safety),<br>IEC60601-1-2(EMC)<br>ISO<br>10993-1,5,10(Biocompatibility)                                                                                                          | Same    |

{5}------------------------------------------------

Image /page/5/Picture/9 description: The image shows the logo for Belter. The logo consists of an orange circle with a white "c" inside of it, followed by the word "BELTER" in blue. The font is sans-serif and the letters are bolded.

#### Performance Data

Compliance to applicable standards includes ISO80601-2-56,as well as IEC 60601-1 , IEC 60601-1-2 and IEC60601-1-11 requirements.

The bench test report of this submission shows the laboratory accuracy is 0.01~0.19℃, which is not greater than 0.2 ℃, and the clinical repeatability is 0.07 ~ 0.11℃,which is less than 0.3℃,so the accuracy of the device ,both bench and clinical, meets the requirements of ISO80601-2-56.

About the structural integrity of device item, the device has be done shock tests and broad-band random vibration test ,the result in bench test of this submission showed the device meets the requirement.

### Conclusion

The Belter Infra-red Ear Thermometer, has the similar intended use and characteristics as the cleared device in K111637. Moreover, bench testing contained in this submission supplied demonstrate that the modification of Belter units, do not raise any new questions of safety or effectiveness. Thus, the subject device is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K140681](https://fda.innolitics.com/device/K140681)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
