K140681 · Shenzhen Belter Health Measurement and Analysis TE · FLL · Aug 21, 2014 · General Hospital
Device Facts
Record ID
K140681
Device Name
BELTER INFRA-RED EAR THERMOMETER
Applicant
Shenzhen Belter Health Measurement and Analysis TE
Product Code
FLL · General Hospital
Decision Date
Aug 21, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Belter Infra-red Ear Thermometer is an electronic thermometer using an infrared sensor to detect human body temperature from the ear canal on people of all ages for home and professional use.
Device Story
The Belter Infra-red Ear Thermometer is a handheld electronic device used to measure human body temperature via the ear canal. It utilizes an infrared sensor to detect infrared energy emitted from the eardrum tissue. The device is operated by clinicians or patients in home or professional settings. The user positions the probe in the ear canal; the device captures the infrared signal and converts it into a temperature reading displayed on an LCD screen. It includes features such as a fever alarm, buzzer, and memory for 10 sets of readings. The device provides rapid temperature assessment, aiding healthcare providers or patients in monitoring health status and identifying potential fever, facilitating clinical decision-making regarding patient care.
Clinical Evidence
Bench testing and clinical repeatability data provided. Laboratory accuracy reported as 0.01–0.19°C (meets ISO 80601-2-56 requirement of ≤0.2°C). Clinical repeatability reported as 0.07–0.11°C (meets ISO 80601-2-56 requirement of ≤0.3°C). Structural integrity verified via shock and broad-band random vibration testing.
Technological Characteristics
Infrared sensor-based electronic thermometer. Powered by 2x AAA batteries. Features LCD display, fever alarm, and 10-set memory. Conforms to ISO 80601-2-56 (performance), EN 60601-1 (safety), IEC 60601-1-2 (EMC), and ISO 10993-1, 5, 10 (biocompatibility).
Indications for Use
Indicated for temperature measurement in the ear canal for all ages, including neonates, pediatric, and adult populations, in home and professional settings.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is an image of three stylized human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 21, 2014
Shenzhen Belter Health Measurement and Analysis Technology Company, Limited Pang Ming RA Manager 702, 704, Block C, Tsinghua Unis Science Park, Langshan Rd, Hi-Tech Industrial Park (north), Nanshan District, Shenzhen, Gangdong CHINA 518057
Re: K140681 Trade/Device Name: Belter Infra-red Ear Thermometer, Models TE-66, TE-68, TE-79, TE-90, and TE-91 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical electronic thermometer Product Code: FLL Regulatory Class: II Dated: July 15, 2014 Received: July 21, 2014
Dear Mr. Ming:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Page 2 - Mr. Ming
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mary S. Runner -S
Erin Keith, M.S. Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K140681
Device Name Belter Infra-red Ear Thermometer Model: TE-66,TE-68, TE-79, TE-90, TE-91
#### Indications for Use (Describe)
The Belter Infra-red Ear thermometer is an electronic thermometer using an infrared sensor to detect human body temperature from the ear canal on people of all ages for home and professional use.
Type of Use (Select one or both, as applicable)
- Prescription Use (Part 21 CFR 801 Subpart D)
> Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for eBelter. The logo consists of the letter 'e' in orange, followed by the word 'BELTER' in blue. The 'e' is stylized with a circular shape, and the word 'BELTER' is in a bold, sans-serif font.
# Section 5
# 510 (K) Summary
## Date: 2014,1,15 Submitter Information
Name : Shenzhen Belter Health Measurement and Analysis Technology Co.,Itd Address: 702、704,block C ,Tsinghua Unis science park,No.13 Langshan Rd,Hi-Tech Industrial park(north),Nanshan District,518057 shenzhen. People's Republic of China Tel: +86-755-61869839 Fax: +86-755-61869122 Contact person: pang minq E-mail: panqm@ebelter.com
## Name of device
Trade name: Belter Infra-red Ear Thermometer Model: TE-66,TE-68, TE-79, TE-90, TE-91 Common name: Belter Infra-red Ear Thermometer Classification name: Thermometer, Electronic, Clinical Production regulation: 21 CFR 880.2910 Product code: FLL
## Predicate Device
Infrared Ear Thermometer THP Series (Radiant Innovation Inc,FLL,K111637)
## Description
The Belter Infra-red Ear Thermometer is characterized by measuring human body temperature from the ear canal. It utilizes infrared technology to measure infrared energy emitted from eardrum tissue when making a temperature measurement.
### Intended use
The Belter Infra-red Ear Thermometer is an electronic thermometer using an infrared sensor to detect human body temperature from the ear canal on people of all ages for home and professional use.
## Summary of technological characteristics of device compared to the predicate devices (K111637), see the table1
| Table 1 | | | |
|-------------|---------------------|------------------|---------|
| SE | Subject device | Predicate device | comment |
| Comparisons | Present application | K111637 | nt |
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "BELTER" in blue font with an orange "e" symbol to the left of the word. The "e" symbol is a stylized letter "e" inside of a circle. The word "BELTER" is in all capital letters and has a slightly rounded font. The image is likely a logo or brand name.
| Intended Use&<br>Indication for use | The Belter Infra-red Ear<br>Thermometer is an<br>electronic thermometer<br>using an infrared sensor<br>to detect human body<br>temperature from the<br>ear canal on people of<br>all ages for home and<br>professional use. | The device is an electronic clinical<br>thermometer using an infrared<br>sensor to detect body<br>temperature from the auditory<br>canal in the neonatal, pediatric<br>and adult population used in the<br>home setting. | Similar |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Measurement<br>Range | 32.0 °C -42.2℃ (89.6 °F<br>-108.0°F) | 34.0°C ~42.2°C<br>(93.2°F ~108.0°F) | Similar |
| Accuracy | ±0.2℃ (36℃≤t<39) ;<br>±0.3°Cother range | ±0.2°C (0.4°F) within<br>35.5~42°C (95.9~107.6°F),<br>±0.3°C(0.5°F) other range | Similar |
| Response time | 1 sec. | 1 sec. | same |
| Measurement<br>place | ear | ear | same |
| Scale selection | °C /°F | C /°F | same |
| Display screen | LCD | LCD | same |
| Memory | 10 sets | 25 sets | Similar |
| Fever alarm | Yes | Yes | same |
| Disposable<br>probe cover | No | MOLD RING W/PE FILM<br>COVER | ---- |
| Backlight | optional | optional | Same |
| Buzzer | Yes | Yes | same |
| Auto power-off<br>while<br>operation | no Yes | Yes | Same |
| Power supply | 2 x AAA | CR2032 *1pcs | --- |
| Conformance<br>standard | ISO80601-2-56(performance),<br>EN60601-1(Safety),<br>IEC60601-1-2(EMC)<br>ISO<br>10993-1,5,10(Biocompatibility) | ISO80601-2-56(performance),<br>EN60601-1(Safety),<br>IEC60601-1-2(EMC)<br>ISO<br>10993-1,5,10(Biocompatibility) | Same |
{6}------------------------------------------------
Image /page/5/Picture/9 description: The image shows the logo for Belter. The logo consists of an orange circle with a white "c" inside of it, followed by the word "BELTER" in blue. The font is sans-serif and the letters are bolded.
#### Performance Data
Compliance to applicable standards includes ISO80601-2-56,as well as IEC 60601-1 , IEC 60601-1-2 and IEC60601-1-11 requirements.
The bench test report of this submission shows the laboratory accuracy is 0.01~0.19℃, which is not greater than 0.2 ℃, and the clinical repeatability is 0.07 ~ 0.11℃,which is less than 0.3℃,so the accuracy of the device ,both bench and clinical, meets the requirements of ISO80601-2-56.
About the structural integrity of device item, the device has be done shock tests and broad-band random vibration test ,the result in bench test of this submission showed the device meets the requirement.
### Conclusion
The Belter Infra-red Ear Thermometer, has the similar intended use and characteristics as the cleared device in K111637. Moreover, bench testing contained in this submission supplied demonstrate that the modification of Belter units, do not raise any new questions of safety or effectiveness. Thus, the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.