← Product Code [NUH](/productcode/NUH) · K140650

# HIVOX ELECTRIC STIMULATOR OTC TENS (K140650)

_Hivox Biotek, Inc. · NUH · Jun 27, 2014 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K140650

## Device Facts

- **Applicant:** Hivox Biotek, Inc.
- **Product Code:** [NUH](/productcode/NUH.md)
- **Decision Date:** Jun 27, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5890
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The HIVOX Electric Stimulator OTC TENS, Rapid ReliefTM Pennypad PP-904, is indicated for temporary relief of pain associated with sore and aching muscles in the upper and lower extremities (arm and/or leg), and lower back due to strain from exercise or normal household and work activities.

## Device Story

HIVOX Electric Stimulator OTC TENS (Rapid Relief Pennypad PP-904) is a transcutaneous electrical nerve stimulator for OTC use. Device generates small electrical pulses delivered through skin to activate underlying nerves for pain relief. User operates device to select pulse strength (0-15 stages) and fixed treatment time (20 minutes). Device features fixed pulse rates (2, 5, 40 Hz) and pulse width (200µS). Output is delivered via electrode pads. Intended for home use by patients to manage muscle soreness/aches in arms, legs, and lower back. Clinical benefit is temporary pain relief. Device is powered by a CR2032 battery.

## Clinical Evidence

Bench testing only. Performance testing included compliance with IEC/EN 60601-1 (safety), IEC/EN 60601-1-2 (EMC), and IEC/EN 60601-2-10 (nerve and muscle stimulators). No clinical data presented.

## Technological Characteristics

Transcutaneous electrical nerve stimulator; battery-powered (CR2032); fixed pulse width 200µS; fixed pulse rates 2, 5, 40Hz; 15-stage adjustable pulse strength; max output 57.6 Vp-p (@500 Ohm); dimensions 113x70x10mm. Complies with IEC 60601-1, 60601-1-2, and 60601-2-10.

## Regulatory Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

## Predicate Devices

- HIVOX Electric Stimulator OTC TENS, Pennypad 904, 907, and 909 ([K112392](/device/K112392.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "HIVOX" in a bold, sans-serif font. To the right of the word is a symbol that looks like four squares connected to each other. The symbol is slightly smaller than the letters in the word "HIVOX".

5F., No. 123, Shingde Road, Sanchong District, New Taipei City, Taiwan Phone: +886 2 8511 2668 Fax: +886 2 8511 2669

# 5. 510(k) SUMMARY (According to 21 CFR 807.92)

| • | 510(K) OWNER'S<br>NAME | HIVOX BIOTEK INC.<br>5F, No.123, Shingde Road, Sanchong Dist.,<br>New Taipei City, 24158, TAIWAN, R.O.C.<br>TEL: +886-2-85112668 FAX:+886-2-85112669                                                                                                                                                           |
|---|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| • | Contact Person         | Dr. JEN, KE-MIN<br>TEL: 886-2-85112668 FAX:886-3-5209783<br>Email: ceirs.jen@msa.hinet.net                                                                                                                                                                                                                     |
| • | Date of Submission:    | May 23, 2014                                                                                                                                                                                                                                                                                                   |
| • | Trade Name             | HIVOX Electric Stimulator OTC TENS, Rapid ReliefTM<br>Pennypad PP-904                                                                                                                                                                                                                                          |
| • | Predicate Devices      | K112392<br>HIVOX Electric Stimulator OTC TENS, Pennypad 904, 907,<br>and 909                                                                                                                                                                                                                                   |
| • | Common Name            | Electric Stimulator OTC TENS                                                                                                                                                                                                                                                                                   |
| • | Classification Name    | Stimulator, Nerve, Transcutaneous, Over-The-Counter<br>(21 CFR 890.5890, Product Code NUH)                                                                                                                                                                                                                     |
| • | Panel                  | Neurology                                                                                                                                                                                                                                                                                                      |
| • | Intended Use           | The HIVOX Electric Stimulator OTC TENS, Rapid ReliefTM<br>Pennypad PP-904, is indicated for temporary relief of pain<br>associated with sore and aching muscles in the upper and lower<br>extremities (arm and/or leg), and lower back due to strain from<br>exercise or normal household and work activities. |
| • | Device Design          | The HIVOX Electric Stimulator OTC TENS, Rapid ReliefTM<br>Pennypad PP-904, can generate small pulses of electrical<br>current and delivered these pulses pass through the skin and<br>activated underlying nerves.                                                                                             |

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the word "HIVOX" in all caps. To the right of the word is a symbol that looks like a plus sign made of squares. The letters are all the same size and font.

5F., No. 123, Shingde Road, Sanchong District, New Taipei City, Taiwan Phone: +886 2 8511 2668 Fax: +886 2 8511 2669

Product Specifications: Power CR2032 x 1 Pulse rate : 2, 5, and 40Hz (Fixed) Pulse width: 200uS(fixed) Electric impedance of the device: 4.80~5.20 Mohms Output voltage Max. : 57.6 Vp-p, based on 500 Ohm load ± 10% Treatment time: 20 minutes fixed Pulse strength: 0 ~ 15 stages adjustable Operation environment : 10 ~ 40° C, 30 ~ 85% RH Storage environment: -10 ~ 50° C, 10 ~ 95% RH Transport environment: -10 ~ 50° C, 35 ~ 85% RH Size: 113 x 70 x 10 mm

Applicable Electrical Range for the Electrode Pads: Pulse rate: 1~150 Hz Pulse strength: 1~150 V Pulse width: 100µS ~ 500 mS

· Performance Tests Submitted:

The relevant standards including:

- IEC/EN 60601-1 : Medical electrical equipment Part 1. General 1. requirements for safety, 1996.
- 2. IEC/EN 60601-1-2 : Medical electrical equipment, Part 2. Electromagnetic compatibility - Requirements and tests, 2007.
- 3. IEC/EN 60601-2-10 : Medical electrical equipment, Part 2-10: Particular requirements for safety of nerve and muscle stimulators, 2001.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the word "HIVOX" in a bold, sans-serif font. To the right of the word is a graphic that looks like four squares arranged in a plus sign shape. The squares are not solid, but rather have a textured or pixelated appearance. The overall impression is of a logo or brand name.

5F., No. 123, Shingde Road, Sanchong District, New Taipei City, Taiwan Phone: +886 2 8511 2668 Fax: +886 2 8511 2669

| Parameter                                       |                         | HIVOX<br>(K11-2392)                              | HIVOX<br>Subject Device           |
|-------------------------------------------------|-------------------------|--------------------------------------------------|-----------------------------------|
| Mode or Program Name                            |                         | PP909<br>PP907<br>PP904                          | Rapid ReliefTM<br>Pennypad PP-904 |
| Indication for Applied Area for pain<br>relieve |                         | Lower<br>Back<br>Arm and<br>Leg<br>Lower<br>Back | Arm, Leg, and<br>Lower Back       |
| dimensions                                      |                         | 113L * 70W * 9.7H mm                             |                                   |
| Waveform                                        |                         | Symmetrical Biphasic                             |                                   |
| Shape                                           |                         | Rectangular                                      |                                   |
| Maximum Output Voltage<br>(Volts) (±20%)        | @500Ω                   | 40.0Vpp                                          | 57.6Vpp                           |
|                                                 | @2KΩ                    | 84.0Vpp                                          | 89.6Vpp                           |
|                                                 | @10KΩ                   | 92.0Vpp                                          | 96.0Vpp                           |
| Maximum Output<br>Current(±20%)                 | @500Ω                   | 80.0mApp                                         | 115.2mApp                         |
|                                                 | @2KΩ                    | 42.0mApp                                         | 44.8mApp                          |
|                                                 | @10KΩ                   | 9.2mApp                                          | 9.6mApp                           |
| Duration of primary (depolarizing) phase (µSec) |                         | NA                                               |                                   |
| Pulse Duration (µSec)                           |                         | 200µSec (fixed)                                  |                                   |
| Frequency (Hz)                                  |                         | 35                                               | 2, 5, and 40                      |
| Net Charge (µC) per pulse                       | @500Ω                   | 0.3200                                           | 0.2304                            |
| Maximum Charge (µC)                             | @500Ω                   | 16.0                                             | 23.04                             |
| Maximum Current Density<br>(mA/cm², r.m.s.)     | @500Ω                   | 1.964                                            | 2.828                             |
| Maximum Average Power<br>Density (W/cm²)        | @500Ω                   | 0.078                                            | 0.163                             |
| Burst Mode                                      | a. Pulse per burst      | 1                                                | NA                                |
|                                                 | b. Burst per second     | 4                                                | NA                                |
|                                                 | c. Burst duration (sec) | 2                                                | NA                                |
|                                                 | d. Duty Cycle           | 8                                                | NA                                |
| ON Time(sec)                                    |                         | 120                                              |                                   |
| Off Time(sec)                                   |                         | 0                                                |                                   |
| Additional Features                             |                         | NA                                               |                                   |

# Compared to Legally Marketed Predicate Devices

Note: Only the subject device Pennypad 909 has the "Burst Mode".

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "HIVOX" in bold, black letters. To the right of the "X" is a small graphic of four squares arranged in a plus sign shape. The squares are also black and have a slightly textured appearance.

5F., No. 123, Shingde Road, Sanchong District, New Taipei City, Taiwan Phone: +886 2 8511 2668 Fax: +886 2 8511 2669

## DISCUSSION:

The predicate devices, HIVOX Electric Stimulator OTC TENS, Pennypad 904, Pennypad 907, and Pennypad 909, have the same visional appearance, software, and dimensions ( 113 * 70 * 9.7 mm ) as the subject device.

The main differences among the predicated devices are the Indication for Applied Area for pain relieve ..

- Pennypad -904 and Pennypad 909 OTC TENS - Lower Back pain relieve
- · Pennypad 907 OTC TENS Arm and Leg pain relieve
- Only Pennypad 909 has the "Burst Mode". ●

The subject device, HIVOX Electric Stimulator OTC TENS, Rapid Relief 11M Pennypad PP-904, is identical to the predicate devices and only extends the Indication for Applied Area for pain relieve to include the Arm and Leg pain relief.

### CONCLUSION:

The subject device, HIVOX Electric Stimulator OTC TENS Rapid Relief ™ Pennypad PP-904, applied for the Arm, Leg, and Lower Back pain relieve is as safe and effective as, and functions in a manner equivalent to the predicate devices. The conclusions drawn from the non-clinical tests demonstrate that the device is as safe, as effective, and performs as well as the legally marketed device identified in the submission. Thus the subject device is substantially equivalent to the predicate devices.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like symbol with three curved lines representing its wings or feathers. The symbol is enclosed within a circular border, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the border of the circle.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 27, 2014

HIVOX BIOTEK INC. Dr. Jen. Ke-Min 8F, No. 98, Shingde Road, Sanchong District, Taipei City. Taiwan, ROC

Rc: K140650

Trade/Device Name: Pennypad 904 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator Regulatory Class: Class II Product Code: NUH Dated: May 24, 2014 Received: May 30, 2014

Dear Dr. Jen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into cither class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

{5}------------------------------------------------

Page 2 - Dr. Jen, Ke-Min

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Felipe Aquel -S

for

Carlos L. Peña, Ph.D., M.S. Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

# DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below

5 10(k) Number (if known)

#### K140650

Device Name

HIVOX Electric Stimulator OTC TENS, Rapid Relief TN Pennypad PP-904

#### Indications for Use (Describe)

The Rapid Relief 711 Pennypad PP-904 is indicated for temporary relief of pain associated with sore and aching muscles in the upper and lower extremities (arm and/or leg), and lower back due to strain from exercise or normal household and work activities.

Type of Use (Select one or both, as applicable)

[[] Prescription Use (Part 21 CFR 801 Subpart D)

0 Over-The-Counter Use (21 CFR 801 Subpan C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

| Felipe   | Date:            |
|----------|------------------|
|          | 2014.06.27       |
| Aguel -S | 16:33:16 -04'00' |

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAH. ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response. including the time to review instructions. search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collections for reducing this burden. 10:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStoffi@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB manber."

FORM FDA 3881 (1/14)

Page I of 1

EF. PSC Publishing Services (301) 443-6740

---

**Source:** [https://fda.innolitics.com/device/K140650](https://fda.innolitics.com/device/K140650)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
