← Product Code [NDK](/productcode/NDK) · K140649

# VECTRIX EXTERNAL FIXATOR (K140649)

_3D Medical Concepts, LLC · NDK · Dec 29, 2014 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K140649

## Device Facts

- **Applicant:** 3D Medical Concepts, LLC
- **Product Code:** [NDK](/productcode/NDK.md)
- **Decision Date:** Dec 29, 2014
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The Vectrix External Fixator is indicated for use in cases in which external fixation is necessary for the treatment of long- and short-bone trauma and reconstruction in adult and pediatric patients, including limb lengthening, pseudarthroses, infected fractures, distraction, arthrodesis, and severe open fractures.

## Device Story

Vectrix External Fixator is a modular external fixation system for bone trauma and reconstruction. System components include 6AL-4V titanium or 304SS/316L stainless steel spline rods, self-tapping threaded fixation pins, titanium pin clamps, ball collets, collet caps, drive knobs, and drive screws. Surgeon percutaneously inserts pins into bone segments; pins are secured to spline rods via pin clamps. Ball and collet design enables extensive pin rotation and placement freedom. Drive screws and knobs allow for bone compression and distraction. Used by surgeons in clinical settings to stabilize fractures or facilitate limb reconstruction. Benefits include customizable configuration for patient-specific anatomical needs.

## Clinical Evidence

Bench testing only. Non-clinical interconnection testing was performed per ASTM F1541-02 (Standard Specification and Test Methods for External Skeletal Fixation Devices), specifically A2.8.4 Pin/Clamp Eccentric Loading, to confirm equivalence to the predicate device.

## Technological Characteristics

Modular external fixator constructed from 6AL-4V titanium, 304SS, and 316L stainless steel. Conforms to ASTM F136 (Wrought Ti-6Al-4V ELI), ASTM F138 (Wrought 18Cr-14Ni-2.5Mo Stainless Steel), and ASTM F86 (Surface Preparation). Features ball collet pin clamp system, spline rods (0.125, 0.25, 0.375 inch diameters), and self-tapping threaded pins (3.5mm, 4.0mm, 5.0mm diameters).

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Torus External Fixation System ([K925635](/device/K925635.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 29, 2014

3D Medical Concepts, LLC Ms. Beverly Laird 1061 Morgan Park Road Pelham, Alabama 35242

Re: K140649

Trade/Device Name: Vectrix External Fixator Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: NDK Dated: October 20, 2014 Received: October 27, 2014

Dear Ms. Laird:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

510(k) Number (if known)

#### K140649

Device Name Vectrix External Fixator

Indications for Use (Describe)

The Vectrix External Fixator is indicated for use in which external fixation is necessary for the treatment of long- and shortbone trauma and reconstruction in adult and pediatic patients, including limb lengthening, pseudarthroses, infected fractures, distraction, arthrodesis, and severe open fractures.

Type of Use (Select one or both, as applicable)

|× Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

## 510(k) Summary

Company Name:

3D Medical Concepts, LLC 1061 Morgan Park Rd. Pelham, AL 35124 Phone: (205) 987-0935 Fax: (205) 987-0936

Contact Name: Date Prepared: Trade Name: Common Name: Classification:

Product Code: Predicate Device: Device Description: Dr. Beverly Laird October 20, 2014 Vectrix External Fixator External Fixation Device Class II (21 CFR 888.3040) Smooth or Threaded Metallic Bone Fixation Fastener NDK Torus External Fixation System (K925635) The Vectrix External Fixator is a comprehensive and modular external fixator, the design of which is based on a ball collet pin clamp system and consists of:

- 6AL-4V titanium and/or 304SS stainless steel spline . rod (0.125, 0.25, and 0.375 inches in diameter)
- . 6AL-4V titanium and/or 316L stainless steel selftapping, threaded fixation pins (5.0 mm and 4.0 mm thread diameter pins with thread lengths from 25 mm to 75 mm, varying in 5 mm increments; and 3.5 mm thread diameter pins with thread lengths of 15 mm)
- Titanium pin clamps
- Titanium ball collets ●
- Titanium collet caps
- Titanium drive knob ●
- Titanium drive screw ●

Fixation pins are percutaneously inserted into the bone segments and secured with pin clamps to the external spline rod. The choice of thread diameter and length is made by the surgeon on a case-by-case basis. The ball and collet design allows extensive pin rotation and, thus, pin placement freedom.

The system can be assembled in three sizes, determined by the spline rod size. Furthermore, elements of the system may be combined in a wide range of configurations as needed to achieve the external fixation structure best suited for each

{4}------------------------------------------------

|                                                 | patient. The drive screw and drive knob can also be added to the spline rod constructs of each size to mediate bone compression and distraction for limb reconstruction.                                                                                                                                                                                                                                                                                                                     |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use:                            | The Vectrix External Fixator is indicated for use in cases in which external fixation is necessary for the treatment of long- and short-bone trauma and reconstruction in adult and pediatric patients, including limb lengthening, pseudarthroses, infected fractures, distraction, arthrodesis, and severe open fractures.                                                                                                                                                                 |
| Substantial Equivalence:                        | The design, indications, materials, and characteristics of the Vectrix External Fixator are substantially equivalent to the Torus External Fixation System (K925635).                                                                                                                                                                                                                                                                                                                        |
| Technological<br>Characteristics<br>Comparison: | The Vectrix External Fixator was designed in conformance with ASTM F136 (Standard Specification for Wrought Titanium-6 Aluminum-4 Vanadium ELI Alloy for Surgical Implant Applications), ASTM F138 (Standard Specification for Wrought 18Chromium-14Nickel-2.5Molybdenum Stainless Steel Bar and Wire for Surgical Implants (UNS S31673), and ASTM F86 (Standard Practice for Surface Preparation and Marking of Metallic Surgical Implants). All requirements were met for these standards. |
| Performance Data:                               | Non-clinical interconnection testing per ASTM 1541-02, Standard Specification and Test Methods for External Skeletal Fixation Devices (A2.8.4 Pin/Clamp Eccentric Loading), was conducted to confirm that the connection mechanisms of the Vectrix External Fixator are equivalent to those of the Torus External Fixation System (K925635).                                                                                                                                                 |

---

**Source:** [https://fda.innolitics.com/device/K140649](https://fda.innolitics.com/device/K140649)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
