K140606 · Resonance Innovations, LLC · MOS · May 12, 2014 · Radiology
Device Facts
Record ID
K140606
Device Name
1.5T PROCURE ARRAY COIL, 3.0T PROCURE ARRAY COIL
Applicant
Resonance Innovations, LLC
Product Code
MOS · Radiology
Decision Date
May 12, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The intended use of this PROCURE Array coil family is to provide high-quality images of the reproductive and urological anatomies in an easy-to-position, wearable, and very flexible design.
Device Story
PROCURE Array Coil Family consists of wearable, flexible, multi-channel phased array MRI coils designed for 1.5T or 3.0T GE MRI scanners. Device features SemiFlex design for close-proximity positioning to reproductive and urological anatomies; enclosure uses liquid-impermeable, biocompatible materials with disposable USP Class 6 PET liners. Coils are receive-only; they interface with the system Body Coil and do not transmit energy. Operated by MRI technicians in clinical settings; images are interpreted by physicians to assist in diagnosis. Device accommodates external biopsy systems. Benefits include improved image quality through closer antenna geometry and ease of patient positioning.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing including SNR and uniformity analysis, blocking analysis, heat testing, and risk management (FMEA). Imaging performance demonstrated via analysis in sagittal, coronal, and transverse planes.
Technological Characteristics
Multi-channel receive-only phased array MRI coil. Materials: flame retardant EVA foam, nylon fabric, biocompatible liquid-impermeable enclosure. Dimensions: SemiFlex form factor. Connectivity: interfaces with GE 1.5T/3.0T MRI systems. Energy source: passive (receive-only, powered by MRI system). Safety standards: IEC60601-1 3rd Edition. Biocompatibility: USP Class 6 PET liners.
Indications for Use
Indicated for use by physician order as an accessory to GE 1.5T or 3.0T MRI scanners for general human anatomy imaging to assist in medical diagnosis.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
**MAY 12 2014**
Image /page/0/Picture/1 description: The image shows the logo for ScanMed. The logo is in black and white, with the word "ScanMed" in a bold, sans-serif font. Below the word "ScanMed" is the tagline "The Image is Everything" in a smaller, serif font. The logo is simple and modern, and it conveys the company's focus on imaging.
Resonance Innovations, LLC 9840 S. 140" St., Suite 8 Omaha, NE 68138 (0) 402.934.2650 (F) 402.778.9699 www.scanmed.com
| Section 8. | 510(k) Summary, K140606 |
|-------------------------------|--------------------------------------------------|
| Submitter's Name: | Resonance Innovations LLC |
| Submitter's Address: | 9840 South 140th St., Suite 8<br>Omaha, NE 68138 |
| Submitter's Telephone: | 402-934-2650 |
| Submitter's Contact: | Randall Jones, President |
| Date 510(k) Summary prepared: | April 21, 2014 |
| Proprietary Name: | PROCURE™ Array Coil Family |
| Common or Usual Name: | MRI coil(s) |
| Classification Name: | Coil, Magnetic Resonance, Specialty |
| Classification Code: | MOS |
| Predicate Device: | 1.5T ScanMed® PV Array, K022395 |
# Description of the Device
ﻴﺔ ﺍﻟ
The PROCURE™ Coil Family interfaces with a 1.5T or 3.0T 8-channel GE MRI scanner and provides high-quality images of the reproductive and urological anatomies in an easy-to-position, wearable, and very flexible design. This lightweight SemiFlex™ design facilitates effortless and accurate positioning similar to wearing a diaper and positions the multiple antenna elements as close as possible to the target anatomies regardless of patient size. The enclosure for the antenna set is made of flexible liquid impermeable, biocompatible materials, and the coil is accompanied by disposable liners (USP Class 6 PET) easily changed between patients, should scanning without clothing be desired. The two coils in the product family are identical in design, construction, materials, and operation, with the exception of resonant frequency. The designs and materials used for manufacture of both coils are identical to standard MRI coil technology that has existed for several years. The coils use similar blocking networks and impedance matching circuits, and they do not transmit energy into the patient, neither predicate nor current submission.
This phased array coil also accommodates popular biopsy systems, both delivered via an opening near the anus and/or vagina, neither included in the scope of this submission as they are independently operated and cleared by the FDA
40606
page 1 of 3
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for ScanMed. The text "ScanMed" is written in a bold, sans-serif font, with the words "The Image is Everything" written in a smaller font below. The logo is simple and modern, and it conveys the company's focus on medical imaging.
Resonance Innovations, LLC 9840 S. 140" St., Suite 8 Omaha, NE 68138 (0) 402.934.2650 (F) 402.778.9699 www.scanmed.com
| Device Model Number | Device Description |
|---------------------|-------------------------|
| 508GE1501 | 1.5T PROCURE Array Coil |
| 508GE3001 | 3.0T PROCURE Array Coil |
### Intended Use
The intended use of this PROCURE Array coil family is to provide high-quality images of the reproductive and urological anatomies in an easy-to-position, wearable, and very flexible design.
#### Indications for Use
The coil is indicated for use by the order of a physician to be used as an accessory to a General Electric 1.5T or 3.0T magnetic resonance scanner for general human anatomy imaging as supported by the scanner. These images, when interpreted by a trained physician, may assist in medical diagnosis.
# Technological Characteristics
The comparison between the predicate and the current submission is described, below.
- 1. Design. This submission is for a dedicated coil that may give diagnostic quality images of the reproductive and urological anatomies. It is a multi-channel coil with all channels designed to work at once, receive-only, in conjunction with the system Body Coil. The predicate multi-channel, PV Array covers very similar ranges of anatomy, also working as receive-only to the Body Coil. Both predicate and modified coils have semi-flexible anterior and posterior sections, with an antenna section between the legs. The modified device is more form-fitted to allow the antenna geometry to physically closer to the patient's anatomy. All design principles employed are mature and well-known throughout the industry.
- 2. Principles of operation: The scientific principles of operation (magnetic resonance) are identical between the predicate and modified devices. Theory of operation is very well understood throughout the industry.
- 3. Materials. The same materials are used in the construction of the predicate and modified devices. All internal circuitry is encapsulated in flame retardant EVA foam, then completely covered with a nylon fabric or compressed EVA. This nylon fabric has been successfully used for over 10 years in the predicate device with no reported biocompatibility issues. Both modified and predicate devices have been tested for mechanical and electrical safety using IEC60601-1 3rd Edition.
- 4. Chemical Composition. Both predicate and modified devices have a successful biocompatibility track record, as demonstrated by cytotoxicity testing and by their history of use in previously cleared devices.
{2}------------------------------------------------
K140606
Page 3 of 3
Resonance Innovations, LLC 9840 S. 140th St., Suite 8 Omaha, NE 68138 (0) 402.934.2650 (F) 402.778.9699 www.scanmed.com
- 5. Energy Source. Both of these products are receive-only coils not generating their own power, but rather controlled by the MRI system as the energy source.
# Non-Clinical Tests
The coils have similar dimension in the head area and have induced similar fields by the transmit coil (Body Coil) that stimulates them. The predicate and current submission have been subject to similar risk management studies, as listed below, and determined to be substantially equivalent.
- 1. Blocking Analysis
- 2. SNR and uniformity analysis
- 3. Risk management (including hazard analysis and FMEA)
- 4. Heat Testing
- 5. Compliance testing to IEC60601-1 3rd Edition
## Clinical Tests
Analyses in all 3 planes (sagittal, coronal and transverse) were run on the PROCURE Array product to show that the anatomies of the submitted and predicate coils have substantial equivalence; the predicate device images the thorax, abdomen, pelvis and hip regions, and the current submission is designed for imaging the reproductive and urological anatomies.
# Substantial Equivalence Decision
As described in this summary, the modified device is substantially equivalent to the predicate device based on the analysis herein. The modified device raises no new concerns of safety or efficacy. Note that the predicate device was made by Medic, Inc., which is a historic company name. Resonance Innovations is under the same leadership, and for both the dba is ScanMed®
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the words "Public Health Service" in a bold, sans-serif font. The text is arranged on a single line. The words are clearly legible and the image appears to be a close-up of the text.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 12, 2014
Resonance Innovations, LLC % Mr. Randall Jones President & CEO 9840 South 140th Street OMAHA NE 68138-3693
Re: K140606
Trade/Device Name: 1.5T PROCURE™ Array Coil; 3.0T PROCURE™ Array Coil 2 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: MOS Dated: March 28, 2014 Received: April 22, 2014
Dear Mr. Jones:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2-Mr. Jones
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Smh.7)
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K140606
Device Name PROCURE Array Family
Indications for Use (Describe)
The coil is indicated for use by the order of a physician to be used as an accessory to a General Electric 1.5T or 3.0T magnetic resonance scanner for general human anatomy imaging as supported by the scanner. These images, when interpreted by a trained physician. may assist in medical diagnosis.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signalure)
Michael D. O'Hara
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.