K140579 · Edan Instruments, Inc. · KNG · Feb 6, 2015 · Obstetrics/Gynecology
Device Facts
Record ID
K140579
Device Name
ULTRASONIC POCKET DOPPLER
Applicant
Edan Instruments, Inc.
Product Code
KNG · Obstetrics/Gynecology
Decision Date
Feb 6, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.2660
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K140579 · Feb 6, 2015
ULTRASONIC POCKET DOPPLER
Edan Instruments, Inc.
Hospital clinical performance evaluation
The device's clinical performance (fetal heart rate detection) was evaluated across 6 hospitals to confirm effectiveness at specific gestational ages (9-10 weeks and 11-13 weeks) in a clinical setting.
Clinical performance evaluation in hospital settings
Pregnant women (9-13 weeks gestation); Sample Size: 261 total tests (162 for 2 MHz probe; 99 for 3 MHz probe); Number of Sites: 6 hospitals
Not applicable for this study
Detection of fetal heart rate (pass/fail)
Indications for Use
The SD3 Series Ultrasonic Pocket Dopplers (hereinafter called "the Doppler") are intended to be used by health care professionals including registered nurses, midwives, ultrasound technicians, and physician assistants, by prescription from licensed physicians in hospitals, clinics and private offices. The 2 MHz and/or 3 MHz waterproof probes are indicated for the detection of fetal heart rate from early gestation thru delivery and as a general indication of fetal well being. The 3 MHz waterproof probe is used for more than 9-week gestation and the 2 MHz is used for 12-week gestation. They can also be used to verify fetal heart viability. The 4 MHz, 5 MHz and/or 8 MHz waterproof vascular probes are indicated for the detection of blood flow in veins and arteries for assisting in the detection of peripheral vascular disease.
Device Story
Hand-held ultrasonic pocket Doppler; uses 2/3 MHz probes for fetal heart rate (FHR) and 4/5/8 MHz probes for vascular blood flow. Ultrasound waves transmitted to body; reflected signals from moving objects (fetal heart/blood) received by probe; converted to electric signals. CPU performs digital processing; output displayed on OLED screen; audio signal routed to speaker. Used by clinicians (nurses, midwives, technicians) in hospitals/clinics. Provides real-time FHR or vascular flow data to assist clinical assessment of fetal well-being or peripheral vascular disease. Battery-powered; non-invasive.
Clinical Evidence
Clinical study conducted at 6 hospitals. 162 tests (11-13 weeks gestation) using 2 MHz probe; 157 passes (96.9%). 99 tests (9-10+ weeks gestation) using 3 MHz probe; 95 passes (96%). Pass rate >95% requirement met. Bench testing confirmed FHR accuracy ±2 BPM (50-240 BPM range) and repeatability.
Technological Characteristics
Hand-held ultrasonic Doppler; 2/3/4/5/8 MHz waterproof probes. OLED display. Internally powered (AA batteries or 18650 Lithium). Complies with IEC 60601-1, IEC 60601-2-37, IEC 60601-1-2, ISO 10993-1. IPX8 probe rating. Digital signal processing via CPU.
Indications for Use
Indicated for fetal heart rate detection (from early gestation to delivery) and fetal well-being assessment in pregnant women; also indicated for blood flow detection in veins and arteries to assist in peripheral vascular disease diagnosis. 3 MHz probe for >9-week gestation; 2 MHz probe for >12-week gestation.
Regulatory Classification
Identification
A fetal ultrasonic monitor is a device designed to transmit and receive ultrasonic energy into and from the pregnant woman, usually by means of continuous wave (doppler) echoscopy. The device is used to represent some physiological condition or characteristic in a measured value over a period of time (e.g., perinatal monitoring during labor) or in an immediately perceptible form (e.g., use of the ultrasonic stethoscope). This generic type of device may include the following accessories: signal analysis and display equipment, electronic interfaces for other equipment, patient and equipment supports, and component parts. This generic type of device does not include devices used to image some relatively unchanging physiological structure or interpret a physiological condition, but does include devices which may be set to alarm automatically at a predetermined threshold value.
Predicate Devices
Sonotrax Series Ultrasonic Pocket Doppler (K101960)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure composed of three overlapping profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird-like figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 9, 2015
Edan Instruments, Inc Queena Chen Certification Engineer 3/F - B, Nanshan Medical Equipments Park Nanhai Road 1019# Shekou. Nanshan Shenzhen 518067 P.R. China
Re: K140579 Trade/Device Name: SD3 Series Ultrasonic Pocket Doppler- SD3 LITE, SD3, SD3 PLUS, SD3 PRO, SD3 VASCULAR Regulation Number: 21 CFR 884.2660 Regulation Name: Fetal ultrasonic monitor and accessories Regulatory Class: II Product Code: KNG, DPW Dated: January 4, 2015 Received: January 7, 2015
Dear Queena Chen,
This letter corrects our substantially equivalent letter of February 6, 2015.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Joyce M. Whang -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use Statement
510(k) Number (if known): K140579
Device Name: SD3 Series Ultrasonic Pocket Doppler SD3 LITE, SD3, SD3 PLUS, SD3 PRO, SD3 VASCULAR
#### Intended Use:
The SD3 Series Ultrasonic Pocket Dopplers (hereinafter called "the Doppler") are intended to be used by health care professionals including registered nurses, midwives, ultrasound technicians, and physician assistants, by prescription from licensed physicians in hospitals, clinics and private offices.
The 2 MHz and/or 3 MHz waterproof probes are indicated for the detection of fetal heart rate from early gestation thru delivery and as a general indication of fetal well being. The 3 MHz waterproof probe is used for more than 9-week gestation and the 2 MHz is used for 12-week gestation. They can also be used to verify fetal heart viability.
The 4 MHz, 5 MHz and/or 8 MHz waterproof vascular probes are indicated for the detection of blood flow in veins and arteries for assisting in the detection of peripheral vascular disease.
n Prescription Use Or Over the Counter Use
(21 CFR Part 801 Subpart D)
Over the Counter Use **_**
(21 CFR Part 801 Subpart C)
#### IPI FASE DO NOT WRITE RFI OW THIS I INF-C.ONTINUIF ON ANOTHER PAGE IFNFFDFD)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
{3}------------------------------------------------
## 510(k) Summary
### Prepared in accordance with the requirements of 21 CFR Part 807.92
| 1. Submitter: | Edan Instruments, Inc. |
|--------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 3/F-B, Nanshan Medical Equipments Park, Nanhai Rd 1019#, Shekou, Nanshan Shenzhen, 518067 P.R. China |
| | Tel.: (0755) 26856469 |
| | Fax: (0755) 26882223 |
| Contact Person: | Queena Chen |
| Prepare date: | 2015-02-04 |
| 2. Device name and classification: | Device Name: SD3 series Ultrasonic Pocket Doppler |
| | Model: SD3 LITE, SD3, SD3 PLUS, SD3 PRO, SD3 VASCULAR, hereinafter called SD3 series |
| Classification Name: | |
| | 21 CFR 884.2660 Fetal ultrasonic monitor and accessories |
| | 21 CFR 870.2100 Cardiovascular blood flowmeter |
| | Product code: KNG, DPW |
| | Regulatory Class: Class II |
| 3.Premarket Notification Class III Certification and Summary | Not applicable, the subject device is Class II. |
| 4. Predicate Device(s): | Sonotrax Series Ultrasonic Pocket Doppler / K101960 / Edan Instruments, Inc. |
| 5. Device Description: | The SD3 series Ultrasonic Pocket Doppler is a hand-held device for non-invasive measurement and display of fetal heart rate and blood flow velocity utilizing the principle of Doppler shift of an ultrasound. The ultrasound is transmitted from the probe to patient body (maternal abdominal wall), and moves through biophysical objects. The acoustic ultrasound is reflected by blood and moving objects such as the fetal heart. The reflected ultrasound is received by the probe and is converted into electric signals. The waveform data are applied to the CPU for all the digital processing on OLED Display, operation keys. The audio signal is taken off for the routing to the speaker to generate the analogue signals before digital processing. The following probes are supplied |
Page 1 / 5
{4}------------------------------------------------
with the SD3 series Ultrasonic Pocket Doppler: 1. 2 MHz for fetal heart rate. 2. 3 MHz for fetal heart rate 3. 4 MHz for detections of arterial and venous blood flow velocity. 4. 5 MHz for detections of arterial and venous blood flow velocity. 5. 8 MHz for detections of arterial and venous blood flow velocity. The subject device is not a sterilized product, so there is no detailed information about its sterilization status and the shelf life included in the submission. FHR Performance: Sensitivity: 9 weeks gestation (3 MHz) FHR Measuring Range: 50 bpm ~ 240 bpm Resolution: 1 bpm ±2 bpm Accuracy: The SD3 Series Ultrasonic Pocket Dopplers (hereinafter called "the 6. Indications for Doppler") are intended to be used by health care professionals including Use: registered nurses, practical nurses, midwives, ultrasound technicians, and physician assistants, by prescription from licensed physicians in hospitals, clinics and private offices. The 2 MHz and/or 3 MHz waterproof probes are indicated for the detection of fetal heart rate from early gestation thru delivery and as a general indication of fetal well being. The 3 MHz waterproof probe is used for more than 9-week gestation and the 2 MHz is used for 12-week gestation. They can also be used to verify fetal heart viability.
The 4 MHz, 5 MHz and/or 8 MHz waterproof vascular probes are indicated for the detection of blood flow in veins and arteries for assisting in the detection of peripheral vascular disease.
#### 7. Predicate Device Comparison
Comparison to the predicate devices, the subject device has same intended use, similar product design, same performance effectiveness, performance safety as the predicate device.
The differences between the subject device and predicate device include physical specifications, display type. Please refer to following table. All above differences do not affect the basic design principle, usage, effectiveness and safety of the subject device. And no question is raised regarding to effectiveness and safety.
| Item | SONOTRAX | SD3 |
|-------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer/K# | EDAN Instruments/ K101960 | EDAN Instruments/ K140579 |
| Intended Use | The Sonotrax series of Ultrasonic Pocket<br>Doppler are intended for use by health care<br>professionals including registered nurses,<br>practical nurses, midwives, ultrasound<br>technicians, and physicians assistants, by<br>prescription from licensed physicians in | The SD3 Series Ultrasonic Pocket<br>Dopplers (hereinafter called “the<br>Doppler”) are intended to be used by<br>health care professionals including<br>registered nurses, practical nurses,<br>midwives, ultrasound technicians, and |
| | hospitals, clinics and private offices. | physician assistants, by prescription from |
| | | licensed physicians in hospitals, clinics and |
| | | private offices. |
| | The 2 MHz and/ or 3 MHz obstetrical<br>probes are indicated for the detection of fetal<br>heart rate from early gestation thru delivery<br>and as a general indication of fetal well<br>being. They can also be used to verify fetal<br>heart viability following patient trauma. | The 2 MHz and/or 3 MHz waterproof<br>probes are indicated for the detection of<br>fetal heart rate from early gestation thru<br>delivery and as a general indication of fetal<br>well being. The 3 MHz waterproof probe is<br>used for more than 9-week gestation and<br>the 2 MHz is used for 12-week gestation.<br>They can also be used to verify fetal heart<br>viability. |
| | The 4 MHz, 5 MHz and/or 8 MHz vascular<br>probes are indicated for the detection of<br>blood flow in veins and arteries for assisting<br>in the detection of peripheral vascular<br>disease. | The 4 MHz, 5 MHz and/or 8 MHz<br>waterproof vascular probes are indicated<br>for the detection of blood flow in veins and<br>arteries for assisting in the detection of<br>peripheral vascular disease. |
| Installation<br>and<br>use | hand-held | hand-held |
| Mode of operation | Continuous operation | Continuous operation |
| Dimensions | 138mm (W) × 85 mm (L) × 32 mm (H) | 168mm (W)×67 mm (L) × 31 mm (H) |
| Weight | 290g | <350 g (including the battery) |
| Operating: | 5°C ~40°C | 5°C ~40°C |
| Transport/<br>Storage: | -20°C~55°C | -20~55°C |
| | 25%~85% RH | 25% ~ 80%RH |
| Operating: | (no condensation) | (no condensation) |
| Transport/ | 25%~93% RH | 25% ~ 93%RH |
| Storage: | (no condensation) | (no condensation) |
| Operating: | 860hPa~1060hPa | 860hPa ~1060hPa |
| Transport/<br>Storage: | 700hPa~1060hPa | 700hPa ~1060hPa |
| Type of protection<br>against<br>electric<br>shock | Internally powered equipment | Internally powered equipment |
| The<br>degree<br>of<br>protection against<br>electric shock | Type B | Type B |
| The<br>degree<br>of | The main unit : IPX0, | The main unit : IPX0, |
| protection against<br>harmful ingress of | probes : IPX8 | probes : IPX8 |
| liquid<br>The<br>degree<br>of | | |
| safety<br>of<br>application in<br>the<br>presence<br>of<br>a<br>flammable gas | Equipment not suitable for use in presence<br>of flammable gases | Equipment not suitable for use in presence<br>of flammable gases |
| The degree of RF | Group 1, Class B | Group 1, Class B |
| The<br>electrical, | Complies with the standard: | Complies with the standard: |
| mechanical<br>and | IEC 60601-1:2005 | IEC 60601-1:2005 |
| thermal<br>safety | | |
| Evaluation | IEC 60601-2-37:2007 | IEC 60601-2-37:2007 |
| EMC Evaluation | Complies with the standard:<br>IEC<br>60601-1-2:2007 | Complies with the standard:<br>IEC<br>60601-1-2:2007 |
| Software<br>life | Complies with the standard:<br>IEC | Complies with the standard:<br>IEC |
| cycle processes | 62304:2006 | 62304:2006 |
| Acoustic output<br>Evaluation | Complies with the standard: | Complies with the standard: |
| | IEC 61157:2007 | IEC 61157:2007 |
| | IEC 60601-2-37:2007 | IEC 60601-2-37:2007 |
| | NEMA UD 2-2004 | NEMA UD 2-2004 |
| Biocompatibility<br>Evaluation | Complies with the standard: ISO 109931-1. | Complies with the standard: ISO<br>109931-1. |
| | | |
| | Clean the equipment case, probe, etc. as<br>above, and then wipe the probe with an | Clean the equipment case, probe, etc. as<br>above, and then wipe the probe with an |
| Disinfection | alcohol | alcohol |
| | impregnated wipe (ethanol 75%,<br>isopropanol alcohol ≤ 70% or<br>glutaraldehyde ≤3.6%). | impregnated wipe (ethanol 75%,<br>isopropanol alcohol ≤ 70% or<br>glutaraldehyde ≤3.6%). |
| | Wipe the probe with a clean, dry cloth to<br>remove any remaining moisture. | Wipe the probe with a clean, dry cloth to<br>remove any remaining moisture. |
| Display | 45mm*25mm LCD display | 0.96'' OELD double color screen |
| Battery Supply | 2*LR6AA | 3*LR6AA |
| | | 18650 Lithium Battery |
| Auto Shut down | 1 minute after no signal or operation, auto<br>shut down | 1 minute after no signal or operation, auto<br>shut down |
| | | Probe replacement, auto shut down |
| | | |
| Resolution | 1 bpm | 1 bpm |
| Accuracy | ±3bpm | ±2bpm |
| Sensitivity | 10 weeks gestation (3MHz)/<br>105.4dB in 200mm distance | 9 weeks gestation (3MHz)/<br>130dB in 200mm distance |
| | | |
| FHR Measuring<br>Range | 50bpm ~ 240bpm | 50bpm ~ 240bpm |
| | | |
| Recording Length | 240s | 240s |
Page 2 / 5
{5}------------------------------------------------
{6}------------------------------------------------
#### 8. Performance Testing:
#### Non-clinical test:
The following safety standards are conducted on the subject device:
(1) IEC 60601-1:2005 Safety requirements for medical electrical systems
(2) IEC 60601-1-2:2007 Electromagnetic compatibility requirements and tests
(3) AIUM/NEMA UD 2-2004 Acoustic output measurement standard for diagnostic ultrasound equipment
(4) IEC 60601-2-37:2007 Particular requirements for the safety of ultrasonic medical diagnostic and monitoring equipment
(5) ISO 10993-1:2009, ISO 10993-5:2009 and ISO 10993-10:2010 Biological evaluation of medical devices
(6) Y Y0449:2009 Ultrasonic Doppler fetal monitor (a Chinese standard)-Performance requirements and methods of measurement and reporting
(7) IEC61266:1994 Ultrasonics Hand-held probe Doppler fetal heartbeat
detectors-Performance requirements and methods of measurement and reporting
Accuracy of FHR for the SD3 devices was tested with the 2 MHz and 3 MHz probes using an FHR simulator for FHR between 50 and 240 BPM. Testing showed the accuracy met the pre-specified criteria of ±2 BPM.
Page 4 / 5
{7}------------------------------------------------
Repeatability of FHR measurements for the SD3 devices was shown by taking multiple measurements and comparing with set values on the FHR simulator. A students T test analysis showed that measurements were repeatable to a 95% confidence.
#### Clinical test:
Physicians at 6 hospitals used the 2 MHz or the 3 MHz probe depending on the gestational age. There were 162 tests for 11-week to 13-week pregnant women and 99 tests for 9-week to 10(+)-week pregnant women. The test was considered a "pass" if the FHR could be detected and a "fail" if no FHR was detected. A pass rate of 95% was required to support use at a particular gestational age.
Of the 162 cases for the 2 MHz probe from 11-week to 13-week , 157 were considered a "pass." This met the requirements of the test.
Of the 99 cases for the 3 MHz probe for 9-week to 10(+)-week pregnant women, 95 were considered a "pass." This met the requirements of the test.
The subject device passed all testing. The tests were selected to show substantial equivalence between the subject device and the predicate, and since the testing passed, substantial equivalence is shown.
#### 9. Conclusion:
Verification and validation testing was conducted on the SD3 series Ultrasonic Pocket Doppler SD3 series and all testing passed prespectied criteria. This premarket notification submission demonstrates that the Diagnostic Ultrasonic System is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.