K140542 · Rhinosystems, Inc. · KMA · Jun 27, 2014 · Ear, Nose, Throat
Device Facts
Record ID
K140542
Device Name
NAVAGE NASAL IRRIGATOR, NAVAGE NOSE CLEANER
Applicant
Rhinosystems, Inc.
Product Code
KMA · Ear, Nose, Throat
Decision Date
Jun 27, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.5550
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The Naväge Nose Cleaner is intended to help relieve nasal and/or sinus congestion and stuffiness by washing and moisturizing the nasal cavity with a pressure-controlled stream of irrigant rinse.
Device Story
Naväge Nose Cleaner is a handheld, battery-powered nasal irrigator for OTC home use. Device uses dual-action pressure: positive pressure (gravity) to introduce irrigant into nasal cavity and negative pressure (powered suction) to aspirate effluent into a removable collection tank. Miniature battery-powered pump creates pressure differential to maintain flow and balance pressures, minimizing sensation within nasal cavity. Device provides cleaner, more convenient irrigation experience compared to traditional gravity-only systems. User operates device to wash and moisturize nasal passages, potentially relieving congestion and stuffiness.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing of flow rate and pressure performance, and a comprehension study of the Instructions for Use to validate consumer understandability and ease of use.
Technological Characteristics
Handheld, battery-powered (3V DC) device. Materials include plastic, electrical pump, silicone, wiring, and power button. Operates via combined positive (gravity) and negative (vacuum pump) pressure. Flow rate 0.25-1.50 LPM; max suction 23.5 inches water. Not sterile. Complies with IEC 60601 for EMC and electrical safety; ISO 10993 parts 5 and 10 for biocompatibility.
Indications for Use
Indicated for individuals age 12 and older to relieve nasal and/or sinus congestion and stuffiness by washing and moisturizing the nasal cavity.
Regulatory Classification
Identification
A powered nasal irrigator is an AC-powered device intended to wash the nasal cavity by means of a pressure-controlled pulsating stream of water. The device consists of a control unit and pump connected to a spray tube and nozzle.
Predicate Devices
Respironics RinoFlow E.N.T. Nasal Wash System (K973875)
Ubimed Cleanoz Nasal Aspirator Kit
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the text "K140542" in a bold, sans-serif font. The text is black and appears to be part of a larger document or label. Above the text, there are a few small, dark marks that could be smudges or part of a design.
JUN 2 7 2014
# 510(k) Summary (as required by section 807.92(c))
Submitted by: RhinoSystems, Inc. 5399 Lancaster Dr., Unit 6 Brooklyn Heights, Ohio 44131 Contact: Martin R. Hoke, President Office phone: 216 351-6262 Email: mrhoke@cox.net Date First Submitted:
February 24, 2014 Date of Revised Submission: June 25, 2014
#### Trade Name:
Naväge Nose Cleaner Common Name: Powered nasal irrigator Classifications: 21 CFR 874.5550 Powered Nasal Irrigator Product Code KMA 21 CFR 880.6740 Vacuum Powered Body Fluid Suction Apparatus Product Code KDQ
# Predicate Devices:
Respironics RinoFlow E.N.T. Nasal Wash System (K973875) Ubimed Cleanoz Nasal Aspirator Kit
# Device Description:
The Naväge Nose Cleaner ("Naväge") is a powered nasal irrigator intended for OTC use to wash and moisturize the nasal cavity. There are many nasal irrigation devices in commercial distribution in the US, and Naväge is substantially identical to those devices in terms of intended use and indications for use. It is similar in terms of mechanical functionality with the single exception that in addition to using positive pressure, Naväge simultaneously uses negative pressure (suction). That is, the device uses a combination of positive pressure (gravity) to introduce irrigant rinse into the nasal cavity, and negative pressure (powered suction) to aspirate the rinse out of the nasal cavity. The simultaneous use of positive and negative pressure makes it possible for the device to be self-contained so that after washing through the nasal cavity, the irrigant rinse can
{1}------------------------------------------------
flow into a removable collection tank attached to the device. This provides an improved nasal irrigation experience resulting from functional simplicity, superior ergonomics, and less mess than currently available devices.
Head pressure, the gravitational force resulting from the distance between the upper tank fill-line and the entrance nostril, is a positive pressure that decreases to zero as the irrigant runs out of the upper tank. Head pressure exerts its greatest influence at the beginning of the irrigation cycle when it helps initiate irrigant flow.
Negative pressure generated by a miniature, battery-powered pump serves two purposes. First, it evacuates air from the lower tank which allows the irrigant to flow into it. Second, it creates a small pressure differential that draws the irrigant out of the upper tank, through the nasal cavity, and into the lower tank, thereby making up for the loss of head pressure as the irrigant flows out of the upper tank. Naväge is designed so that the positive and negative pressures are essentially kept in balance which results in the user feeling little or no pressure within the nasal cavity during the cycle, thus providing a superior nasal irrigation experience. It also makes it possible for the effluent irrigant to be collected in the lower tank, resulting in a neater and more convenient overall experience.
## Indications for Use:
The Naväge Nose Cleaner is intended to help relieve nasal and/or sinus congestion and stuffiness by washing and moisturizing the nasal cavity with a pressure-controlled stream of irrigant rinse.
### Substantial Equivalence Statement:
The Naväge Nose Cleaner is substantially equivalent to the Respironics RinoFlow E.N.T. Nasal Wash System ("RinoFlow") cleared under K973875 as a 21 CFR 874.5550 Powered Nasal Irrigator. Naväge and RinoFlow are powered nasal irrigation devices that have the same intended use and indications for use, function in a similar manner, are constructed from the same basic materials, and share the same basic operational principles and technical characteristics. Naväge differs from RinoFlow in that in addition to irrigating, Naväge simultaneously aspirates.
With respect to aspiration, Naväge is substantially equivalent to the Ubimed Cleanoz Nasal Aspirator Kit ("Cleanoz") listed under 21 CFR 880.6740 Vacuum Powered Body Fluid Suction Apparatus. Therefore, in accordance with FDA guidance concerning how to identify a predicate when the subject device has two features not previously combined in a single predicate, both predicates are identified as substantially equivalent.
Performance testing data for flow rate of irrigant and pressure is submitted to demonstrate substantial equivalence with respect to operational characteristics. Both predicates are over-the-counter, and a comprehension study of the subject device's Instructions for Use is submitted to demonstrate substantial equivalence with respect to over-the-counter characteristics for consumer understandability and ease of use,
{2}------------------------------------------------
| Substantial Equivalence Comparison Table: | | | |
|-----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Attribute | Subject Device<br>Naväge Nose<br>Cleaner | Predicate Device<br>Rinoflow Micronized<br>E.N.T. Wash System | Predicate Device<br>Cleanoz Nasal<br>Aspirator Kit |
| 510(k) Number | K140542 | K973875 | N/A |
| Regulation | TBD | 21 CFR 874.5550 | 21 CFR 880.6740 |
| Name | TBD | Powered Nasal<br>Irrigator | Vacuum Powered<br>Body Fluid Suction<br>Apparatus |
| Product Code | TBD | KMA | KDQ |
| Classification | TBD | Class I | Class II |
| Premarket<br>Notification | TBD | Exempt | Exempt |
| Classification<br>Description | TBD | A powered nasal<br>irrigator is an AC-<br>powered device<br>intended to wash the<br>nasal cavity by means<br>of a pressure-<br>controlled pulsating<br>stream of water. The<br>device consists of a<br>control unit and pump<br>connected to a spray<br>tube and nozzle. | A vacuum-powered<br>body fluid suction<br>apparatus is a<br>device used to<br>aspirate, remove, or<br>sample body fluids.<br>The device is<br>powered by an<br>external source of<br>vacuum. This<br>generic type of<br>device includes<br>vacuum regulators,<br>vacuum collection<br>bottles, suction<br>catheters and tips,<br>connecting flexible<br>aspirating tubes,<br>rigid suction tips,<br>specimen traps,<br>noninvasive tubing,<br>and suction<br>regulators (with<br>gauge). |
| Panel | TBD | Ear, Nose, and Throat | General Hospital<br>and Personal Use |
| Intended use | The Naväge Nose<br>Cleaner is intended<br>to help relieve<br>nasal and/or sinus<br>congestion and<br>stuffiness by<br>washing and<br>moisturizing the<br>nasal cavity with a<br>pressure-controlled | "Nasal and sinus<br>irrigation and<br>humidification of the<br>upper respiratory<br>tract" | From labeling:<br>"To clean a stuffy or<br>runny nose" |
| | stream of irrigant | | |
| | rinse. | | |
| | | | |
| Indications for<br>use | The Naväge Nose<br>Cleaner is intended<br>to help relieve<br>nasal and/or sinus<br>congestion and<br>stuffiness by<br>washing and<br>moisturizing the<br>nasal cavity with a<br>pressure-controlled<br>stream of irrigant<br>rinse. | "To treat conditions<br>and disorders of the<br>upper respiratory tract<br>where homeostasis of<br>the nasal mucosa is<br>disturbed, resulting in<br>symptoms such as<br>catarrh, and<br>mucopurulent or<br>crusty secretions.<br>Such conditions and<br>disorders include:<br>• Rhinitis<br>• Both acute and<br>chronic sinusitis." | From labeling:<br>1. Cleanoz is<br>designed for<br>household use, only<br>to aspirate baby's<br>nasal secretions.<br>2. To improve the<br>efficiency of<br>Cleanoz, it is<br>recommended to<br>irrigate nasal<br>passages with saline<br>solution before<br>aspirating<br>secretions. |
| Intended Age<br>Group | Age 12 and older | For adults and<br>"children over three<br>years of age" | Labeling specific to<br>pediatric use |
| Anatomical sites | Nasal and sinus<br>cavities | Nasal and sinus<br>cavities | Nasal and sinus<br>cavities |
| OTC or Rx | OTC | OTC | OTC |
| | | | |
| Human factors | Device is designed<br>for personal use by<br>consumers at | Device is designed for<br>personal use by<br>consumers at home. | Device is designed<br>for personal use by<br>consumers at home. |
| | home. | | |
| Power | 3 volt DC | 110 volt AC | 3 volt DC |
| Design | Handheld, self-<br>contained battery-<br>powered unit is<br>designed to irrigate<br>the nasal cavity,<br>and to remove and<br>hold the irrigant<br>effluent in a<br>removable attached<br>container. A small<br>vacuum pump is<br>used to accomplish<br>this. | Same except that this<br>device uses positive<br>pressure only; is AC<br>powered; and has a<br>handheld nasal<br>interface that is<br>attached by tubes to a<br>table-top control unit. | Same except that<br>this device is used<br>for aspiration only. |
| Performance<br>(As specified) | Flow of 0.25 to 1.50<br>LPM; maximum<br>suction of 23.5<br>inches water. | Flow of from 2 to 9.5<br>LPM; maximum<br>pressure of 27.7<br>inches water. | Not published |
| | corresponding<br>restrictions to flow<br>of 0% to 75%;<br>maximum suction<br>of 21.4 inches<br>H2O. | pressure of 18.1<br>inches H2O. | |
| | | | |
| Standards met | ISO 10993 parts 5<br>and 10 for<br>biocompatibility;<br>IEC60601 for EMC. | Same (presumed,<br>since clearance<br>issued) | Same (presumed) |
| Materials | The device uses<br>common consumer<br>product materials<br>including plastic, an<br>electrical pump,<br>silicone, batteries,<br>a power button,<br>and wiring. | Same or similar | Same or similar |
| Biocompatibility | Biocompatible | Same (presumed;<br>clearance issued) | Same (presumed) |
| Compatibility<br>with<br>environment and<br>other devices | Compatible | Same (presumed,<br>since clearance<br>issued) | Same (presumed) |
| Sterility | Not provided sterile | Same | Same |
| Electrical safety | Complies with IEC-60601-1, including<br>EMC requirements | Same (presumed,<br>since clearance<br>issued) | Same (presumed) |
| Mechanical<br>safety | N/A | Same | Same |
| Chemical safety | N/A | Same | Same |
| Thermal safety | N/A | Same | Same |
| Radiation safety | N/A | Same | Same |
.
:
s
{3}------------------------------------------------
{4}------------------------------------------------
# Substantial Equivalence Conclusion:
The Naväge Nose Cleaner shares the same or similar intended use, indications for use, intended users, device operation, overall technical and functional capabilities, and technological characteristics and performance with RinoFlow as a Powered Nasal Irrigator and with Cleanoz as a Vacuum Powered Body Fluid Suction Apparatus. Therefore it is substantially equivalent to its predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an image of an eagle with its wings spread, with three lines representing the feathers. The logo is black and white.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
June 27, 2014
RhinoSystems, Inc. c/o Mr. Martin R. Hoke President 5399 Lancaster Drive. Unit 6 Brooklyn Heights. OH 44131
Re: K140542
Trade/Device Name: Navage Nose Cleaner Regulation Number: 21 CFR 874.5550 Regulation Name: Powered Nasal Irrigator Regulatory Class: Class II Product Code: KMA, KDQ Dated: May 22, 2014 Received: May 27, 2014
Dear Mr. Hoke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misloading.
11 your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Mr. Martin R. Hoke
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Deborah L. Falls -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
510(k) Premarket Notification
Naväge Nasal Irrigator
.
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Orug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date. December 31, 2013 See PRA Statement on last page.
510(k) Number (if known) K 140542
Device Name
RhinoSystems. Inc. Naväge Nose Cleaner
Indications for Use (Describe)
The Newlye Nose Cleaner is interest to help relieve rains congestion and stuffiness by washing and moistantialize the nasal cavity with a pressure-controlled stream of irrigard rinse.
| Type of Use (Select one or both, as applicable) | <table><tr><td><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</td></tr><tr><td><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</td></tr></table> | <input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | | | |
| <input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) | | | |
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
| Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) | Vasant G. Malshet -S |
|------------------------------------------------------------------------------|----------------------|
|------------------------------------------------------------------------------|----------------------|
| FORM FDA 3881 (9/13) | Page 1 of 2 | |
|----------------------|-------------|--|
|----------------------|-------------|--|
{8}------------------------------------------------
Naväge Nasal Irrigator
510(k) Premarket Notification
This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the lime to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
*An agency may not conduct or sponsor, and a person is not required to respond to. a collection of intormation unless it displays a currently valid OMB number
FORM FDA 3881 (9/13)
Page 2 of 2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.