K140516 · Kofu Factory of Terumo Corp. · FMI · Aug 5, 2014 · General Hospital
Device Facts
Record ID
K140516
Device Name
Terumo Pen Injector Needle 34
Applicant
Kofu Factory of Terumo Corp.
Product Code
FMI · General Hospital
Decision Date
Aug 5, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Terumo® Pen Injector Needle 34 is intended for use with a pen injector device for the subcutaneous injection of drugs, including insulin.
Device Story
Terumo® Pen Injector Needle 34 is a manual, single-use, disposable hypodermic needle; designed for subcutaneous drug delivery via pen injector. Device features double-tapered stainless steel cannula; 29G at cartridge interface tapering to 34G at patient interface; 4 mm insertion length. Patient interface end utilizes 3-bevel asymmetrical tip. Device includes plastic hub, inner protector cap, and outer case. Operated manually by patient or clinician; user attaches needle to pen injector, removes caps, performs injection, and uses outer case for safe removal/disposal. Device facilitates drug administration; provides patient benefit through fine-gauge needle design for subcutaneous access.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including ISO 11608-2 performance standards (surface finish, dimensions, stiffness, flow rate, bond strength, needle points), biocompatibility (ISO 10993), and sterilization validation (ISO 11137-1).
Technological Characteristics
Cannula: Stainless steel; Lubricant: Silicone oil; Adhesive: Polyacrylate; Hub/Outer Case: Polypropylene; Protector Cap: Polyethylene. Double-tapered design (29G to 34G). Manual operation. Sterilization: Electron beam (SAL 10^-6). Complies with ISO 11608-2:2012, ISO 9626, ISO 10993-1, and ISO 11137-1.
Indications for Use
Indicated for subcutaneous injection of drugs, including insulin, using a pen injector device. Suitable for general use.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
Predicate Devices
BD 31G and 32G Extra Thin Wall (XTW) Pen Needles with PentaPoint™ (K123300)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of a human figure in profile, with three faces overlapping each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 5, 2014
Mr. Phillip Michael Lester Regulatory Affairs Specialist Terumo Medical Corporation 950 Elkton Boulevard ELKTON MD 21921
Re: K140516
Trade/Device Name: Terumo® Pen Injector Needle 34 Regulation Number: 21 CFR 880.5570 Regulation Name: Needle, Hypodermic, Single Lumen Regulatory Class: II Product Code: FMI Dated: July 8, 2014 Received: July 9, 2014
Dear Mr. Lester:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Lester
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mary S. Runner -S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K140516
Device Name The Terumo® Pen Injector Needle 34
Indications for Use (Describe)
The Terumo® Pen Injector Needle 34 is intended for use with a pen injector device for the subcutaneous injection of drugs, including insulin.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
× Over-The-Counter Use (21 CFR 801 Subpart C)
Digitally signed by Richard C.
Date: 2014.08.01 14:09:47 -04'00'
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/2/Picture/12 description: The image shows the text "FORM FDA 3881 (1/14)". The text is in bold font and is located at the bottom left of the image. There are two horizontal lines above the text. The text indicates that the image is a form from the Food and Drug Administration.
Chapman -S
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### SECTION 5 - 510(k) SUMMARY
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
# Summary of Safety and Effectiveness
| A. Submitter Information. | 39 |
|----------------------------------|----|
| B. Device Name. | 40 |
| C. Predicate Device. | 40 |
| D. Reason for 510(k) Submission. | 40 |
| E. Device Description. | 40 |
| F. Intended Use. | 41 |
| G. Substantial Equivalence. | 41 |
| H. Nonclinical Tests. | 45 |
| I. Clinical Tests. | 47 |
| J. Conclusion. | 47 |
{5}------------------------------------------------
TERUMO CORPORATION Terumo® Pen Injector Needle 34 Section 5. 510(k) Summary
#### 510(k) SUMMARY
## A. SUBMITTER INFORMATION (807.92(a)(1))
| Prepared by: | Phillip Michael Lester |
|--------------|-------------------------------|
| | Regulatory Affairs Specialist |
| | Terumo Medical Corporation |
| | Tel. (410) 392-7321 |
| | Fax (410) 398-6079 |
| Prepared for: | Owner/Operator<br>Terumo Corporation<br>44-1, 2-Chome, Hatagaya<br>Shibuya-Ku, Tokyo, Japan 151-0072<br>Registration Number: 8010026 |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Manufacturer and Sterilization Facility (Applicant)<br>Kofu Factory of Terumo Corporation<br>1727-1, Tsuijiarai, Showa-Cho<br>Nakakoma-Gun, Yamanashi, Japan 409-3853<br>Registration Number: 9681835 |
| Contact Person: | Phillip Michael Lester<br>Regulatory Affairs Specialist<br>Terumo Medical Corporation<br>950 Elkton Boulevard<br>Elkton, MD 21921<br>Tel. (410) 392-7321<br>Fax (410) 398-6079<br>E-mail: philmichael.lester@terumomedical.com |
Date prepared: April 4, 2014
{6}------------------------------------------------
## B. DEVICE NAME (807.92(a)(2))
| Proprietary Name: | Terumo® Pen Injector Needle 34 (or similar) |
|-----------------------|---------------------------------------------|
| Common Name: | Hypodermic single lumen needle |
| Classification Name: | Needle, Hypodermic, Single Lumen |
| Classification Panel: | General Hospital |
| Regulation: | 21 CFR 880.5570 |
| Product Code: | FMI |
| Classification: | Class II |
## C. PREDICATE DEVICE (807.92(a)(3))
The legally marketed device(s) to which substantial equivalence is claimed is/are:
- 1. K123300 BD 31G and 32G Extra Thin Wall (XTW) Pen Needles with PentaPoint™, manufactured by Becton, Dickinson and Company, New Jersey.
- 2. K052561 Terumo® Micro Tapered Pen Needle, manufactured by Kofu Factory of Terumo Corporation, Japan.
# D. REASON FOR 510(K) SUBMISSION
This 510(k) is being submitted due to design modifications to the previously cleared device (K052561) that would prompt a new submission: 1) new needle gauge size, 34G (0.18 mm); 2) new needle insertion length of 4 mm (from previous 5 mm, cleared under K052561); 3) change in sterilization method (from gamma to electron beam); and 4) a change in the indications for use from the previously cleared device. Therefore, this Traditional 510(k) is being submitted for modifications to the current device (K052561) that do not qualify for Special 510(k).
# E. DEVICE DESCRIPTION (807.92(a)(4))
The Terumo® Pen Injector Needle 34 is comprised of a double-tapered stainless steel needle cannula that is pointed at both the cartridge interface end, to puncture the cartridge for drug delivery, and the patient interface end, for injection into the patient's body tissue. The patient interface end features a 3-bevel, asymmetrical needle edge.
The cannula tapers from a 29G (0.33 mm) outer diameter at the cartridge interface
{7}------------------------------------------------
end to a 34G (0.18 mm) outer diameter at the needle tip (patient interface end). It is normal-walled (min. 0.089 mm inner diameter) at the needle tip, and the internal diameter tapers like the outer diameter (wall thickness is uniform throughout the needle length). The double-pointed needle is attached to a plastic hub which screws on to a compatible pen injector (not supplied with this device).
The needle insertion length (from the glue mount on the hub to the tip of the patient interface end) is 4 mm (5/32"). The length of the needle cartridge side is 6.5 mm (1/4"). The patient interface end (needle tip) is covered by a white plastic needle protector cap (inner needle cap) which is then covered by a clear plastic outer case (outer needle case). The cartridge interface end is then covered with a sealing film.
Just prior to use, the outer needle case is removed and retained for recapping once the injection is completed. The inner needle cap is then removed to expose the needle and the injection administered. After the injection, the user inserts the used needle into the open end of the outer needle case so it can be safely removed from the pen injector and disposed of immediately.
This pen needle device is individually packaged and sterilized by electron beam. It is a manually operated, disposable device intended for single use only.
#### F. INTENDED USE (807.92(a)(5))
The Terumo® Pen Injector Needle 34 is intended for use with a pen injector device for the subcutaneous injection of drugs, including insulin.
## G. SUBSTANTIAL EQUIVALENCE COMPARISON (807.92(a)(6))
The Terumo® Pen Injector Needle 34, subject of this Traditional 510(k), is substantially equivalent in its intended use, technology/principal of operation, materials, and performance to:
- 1. K123300 BD 31G and 32G Extra Thin Wall (XTW) Pen Needles with PentaPoint™, manufactured by Becton, Dickinson and Company, New Jersey.
- 2. K052561 Terumo® Micro Tapered Pen Needle, manufactured by Kofu Factory of Terumo Corporation, Japan.
{8}------------------------------------------------
A comparison of the intended use/indications for use and technological characteristics is summarized in the table on the following page. The minor differences of indications for use and the technological differences do not impact the safety and effectiveness of the device.
Note: A statement of substantial equivalence to another product is required by 21CFR807.87, and relates only to whether the present product can be marketed without prior reclassification or clinical approval. The present submission is, therefore, not related to the coverage of any patent, and is not to be interpreted as an admission or used as evidence in a patent infringement lawsuit. As the Commissioner of the FDA has stated, " ...a determination of substantial equivalence under the federal Food, Drug, and Cosmetic Act relates to the fact that the product can lawfully be marketed without pre-market approval or reclassification. This determination is not intended to have any bearing whatever on the resolution of patent infringement suits" 42 Fed. Reg. 42,520, et seq. (1977)
{9}------------------------------------------------
| Device<br>Characteristic | New Device: Terumo® Pen Injector<br>Needle 34 | Predicate: TERUMO® Micro<br>Tapered Pen Needle (K052561) | Predicate: BD 31G and 32G Extra<br>Thin Wall (XTW) Pen Needles with<br>PentaPoint™ (K123300) | |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| Manufacturer | Kofu Factory of Terumo Corporation | Kofu Factory of Terumo Corporation | Becton, Dickinson and Company | |
| Intended Use | The Terumo® Pen Injector Needle<br>34 is intended for use with a pen<br>injector device for the subcutaneous<br>injection of drugs, including insulin. | The TERUMO® Micro tapered<br>pen needle is intended for use with<br>a pen injector device for the<br>subcutaneous injection of insulin.<br>It is indicated for general use and<br>for pediatric patients. | BD Pen Needle is intended for use<br>with pen injector device for<br>subcutaneous injection of drugs,<br>including insulin and exenatide. | |
| Operation<br>Principle | Manual | Manual | Manual | |
| Design /<br>Construction | • Needle assembly (cannula, needle<br>hub, protector cap)<br>• Designed to fit specified pen<br>injectors (new ISO 11608-2:2012<br>compatibility requirements) | • Needle assembly (cannula, needle<br>hub, protector cap)<br>• Designed to fit Type A pen<br>injectors (ISO 11608-2:2000) | • Needle assembly (cannula, needle<br>hub, protector cap)<br>• Designed to fit Type A pen<br>injectors | |
| Materials | • Cannula – Stainless Steel<br>• Lubricant - Silicone Oil<br>• Adhesive – Polyacrylate<br>• Needle Hub – Polypropylene<br>• Protector Cap – Polyethylene<br>• Outer Case – Polypropylene<br>• Sealing Film – Kraft Paper | • Cannula – Stainless Steel<br>• Lubricant - Silicone Oil<br>• Adhesive – Polyacrylate<br>• Needle Hub – Polypropylene<br>• Protector Cap – Polyethylene<br>• Outer Case – Polypropylene<br>• Sealing Film – Kraft Paper | • Cannula – Stainless Steel<br>• Lubricant - Silicone Based<br>Note: this is the information publicly<br>available for the device. | |
| Package | • Plastic outer case<br>• Sealing Film<br>• Shelf box | • Plastic outer case<br>• Sealing Film<br>• Shelf box | • Plastic outer case<br>• Sealing Film<br>• Shelf box | |
| Device<br>Characteristic | | New Device: TerumoⓇ Pen Injector<br>Needle 34 | Predicate: TERUMO® Micro<br>Tapered Pen Needle (K052561) | Predicate: BD 31G and 32G Extra<br>Thin Wall (XTW) Pen Needles with<br>PentaPoint™ (K123300) |
| Specifications | Taper | Double Tapered Design; 29G (0.33 mm) O.D. > 34G (0.18 mm) O.D. | Double Tapered Design; 28G (0.35 mm) O.D. > 33G (0.20 mm) O.D. | None; remains straight throughout |
| | Needle<br>Length | 4 mm (5/32") – Patient Interface<br>6.5 mm (1/4") – Cartridge Interface | 5 mm (3/16") – Patient Interface<br>5.5 mm (7/32") – Cartridge Interface | 4mm for 32G; 5mm and 8mm for<br>31G – Patient Interface |
| | Effective<br>Gauge | 34G | 33G | 31 and 32G |
| | Tip Configuration | Patient-side: Asymmetrical 3-bevel<br>Cartridge-side: Bent (Eccentric) | Patient-side: Asymmetrical 3-bevel<br>Cartridge-side: Bent (Eccentric) | Patient-side: 5-bevel |
| | Wall Type | Normal Wall (min. 0.089 mm) at<br>distal tip (patient interface end)* | Normal Wall (min. 0.089 mm) at<br>distal tip (patient interface end) | Extra Thin Wall |
| | Sterilization | E-beam radiation (validated in<br>accordance with ISO 11137-1 to<br>achieve SAL 10⁻⁶) | Gamma radiation (validated in<br>accordance with ISO 11137-1 to<br>achieve SAL 10⁻⁶) | Gamma radiation |
{10}------------------------------------------------
*The I.D. tapers like the O.D. of the device. The I.D. is the distal portion of the needle. The wall thickness of the cannula remains uniform throughout.
The proposed device and the predicate devices are classified under 21 CFR 880.5570, which states: "A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the surface of the skin."
The proposed device and the predicate devices use statements similar to the excerpt from the regulation. The intended use of the devices are identical (the injection of fluids subcutaneously), and the indications only differ slightly between the proposed device and the predicates. The proposed device only differs from K 123300 in that exenatide is not included. However, the Terumo device is otherwise identical in indications for use to the BD Pen Needle (K123300).
{11}------------------------------------------------
# H. NON CLINICAL TESTS (807.92(b)(1))
## Performance
Performance testing was conducted to ensure the safety and effectiveness of the Terumo® Pen Injector Needle 34 throughout the shelf life, verify conformity to the applicable parts of ISO standards, and demonstrate substantial equivalence to the predicate device. No new issues of safety and effectiveness were raised with the testing performed. The following performance tests were performed:
| Test | Standard | Result |
|----------------------------------------------------------------------------------|-----------------------------------------------|--------------------|
| Surface finish | ISO 11608-2: 2012,<br>ISO 9626: 1991, A1 2001 | Meets<br>standard. |
| Cleanliness | | |
| Limits for acidity and alkalinity | | |
| Dimensions | | |
| Stiffness | | |
| Resistance to breakage | | |
| Resistance to corrosion | | |
| Dimensions for needles | ISO 11608-2: 2012 | Meets<br>standard. |
| Determination of flow rate through the<br>needle | | |
| Bond between hub and needle tube | | |
| Needle points | | |
| Freedom from defects | | |
| Lubrication | | |
| Dislocation of measuring point at patient<br>end | | |
| Determination of functional compatibility<br>with needle-based injection systems | | |
| Ease of assembly and disassembly | | |
| Sterility | ISO 11608-2: 2012,<br>ISO 11137-1: 2006 | Meets<br>standard. |
No deviations from recognized consensus ISO standards were identified in the testing to standards, except where the design of the needle resulted in a modified method or acceptance criterion.
{12}------------------------------------------------
Additionally, performance testing other than to the above ISO Standards was performed on the device. The device complies with the acceptance criteria established based on the predicates:
| Performance Test | Results |
|----------------------------|---------------------------|
| Penetration resistance | Meets acceptance criteria |
| Seal film fitting strength | Meets acceptance criteria |
| Cap-hub fitting strength | Meets acceptance criteria |
Performance testing demonstrates that the Terumo® Pen Injector Needle 34 conforms to the recognized consensus ISO standards, is substantially equivalent to the predicate devices, and is acceptable for clinical use throughout the shelf life.
## Biocompatibility
In accordance with ISO 10993-1, the Terumo® Pen Injector Needle 34 is classified as: Externally Communicating Device, Blood Path Indirect, Short Term (<24 hours) Use, as the cannula is immediately withdrawn after injection into the body. The finished device's blood/body contacting parts were tested in accordance with the tests recommended in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" and Draft Guidance for Industry and Food and Drug Administration Staff - Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing." Screening tests were performed on accelerated aged devices to show that the biocompatibility is maintained throughout the shelf life of the product. Results of the testing demonstrate that the device is biocompatible throughout the shelf life of the product.
| Non-aged, sterile, whole device | |
|----------------------------------------------|--|
| Cytotoxicity | |
| Sensitization | |
| Intracutaneous reactivity (acute) | |
| Systemic toxicity (acute) | |
| Pyrogenicity | |
| Hemolysis | |
| Physicochemical Profile (USP <661> and FTIR) | |
{13}------------------------------------------------
| Accelerated-aged (5 years), sterile, whole device |
|---------------------------------------------------|
| Cytotoxicity |
| Hemolysis |
| Physicochemical Profile (USP <661> and FTIR) |
## Sterilization
The sterility of the device is assured using a sterilization method validated in accordance with ANSI/AAMI/ISO 11137 - Medical Devices - Validation and Routine Control of Radiation Sterilization. Though the sterilization method has changed from gamma irradiation to electron beam irradiation for the modified device, the Terumo® Pen Injector Needle 34 is sterilized to provide a Sterility Assurance Level (SAL) of 10 . Performance testing demonstrated that the device functionality is substantially equivalent, and biocompatibility testing confirmed that the device is still biocompatible.
## Risk Analysis
A Product Risk Analysis was conducted in accordance with ISO 14971, taking into account the modifications to the previous device, and it was determined that any new risks were adequately captured and mitigated.
## I. CLINICAL TESTS (807.92(b)(2))
This 510(k) does not include data from clinical tests.
## J. CONCLUSION (807.92(b)(3))
In summary, the Terumo® Pen Injector Needle 34, subject of this 510(k), is substantially equivalent in its intended use, technology/principal of operation, materials, and performance to:
- 1. K123300 BD 31G and 32G Extra Thin Wall (XTW) Pen Needles with PentaPoint™, manufactured by Becton, Dickinson and Company, New Jersey.
- 2. K052561 Terumo® Micro Tapered Pen Needle, manufactured by Kofu Factory of Terumo Corporation, Japan.
There is no significant difference that raises any new issues of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.