XENMATRIX SURGICAL GRAFT

K140501 · C. R. Bard · FTM · Apr 28, 2014 · General, Plastic Surgery

Device Facts

Record IDK140501
Device NameXENMATRIX SURGICAL GRAFT
ApplicantC. R. Bard
Product CodeFTM · General, Plastic Surgery
Decision DateApr 28, 2014
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.3300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Intended for implantation to reinforce soft tissue where weakness exists and for surgical repair of damaged or ruptured soft tissue, including: abdominal plastic and reconstructive surgery; muscle flap reinforcement; hernia repair including abdominal, femoral, diaphragmatic, scrotal, umbilical, and incisional hernias.

Device Story

XenMatrix™ Surgical Graft is an acellular, sterile, non-pyrogenic porcine dermal matrix; supplied hydrated in sterile saline. Used by surgeons for soft tissue reinforcement and repair of damaged/ruptured tissue; specifically indicated for hernia repairs and plastic/reconstructive procedures. Device provides a scaffold for tissue integration. Clinical benefit derived from reinforcing weakened soft tissue sites. No software or electronic components.

Clinical Evidence

Bench testing only. Included burst strength, tensile strength, suture retention strength, and packaging seal/ship testing. Pre-clinical testing included histology examination. No clinical data presented.

Technological Characteristics

Acellular porcine dermal matrix; sterile; non-pyrogenic; hydrated in saline. No electronic components, software, or energy sources. Sterilization method not specified.

Indications for Use

Indicated for patients requiring soft tissue reinforcement or repair of damaged/ruptured soft tissue, including abdominal plastic/reconstructive surgery, muscle flap reinforcement, and repair of abdominal, femoral, diaphragmatic, scrotal, umbilical, and incisional hernias.

Regulatory Classification

Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Davol Inc. Subsidiary of C.R. Bard, Inc. 100 Crossings Blvd. Warwick, RI 02886 (401) 463-7000 Image /page/0/Picture/2 description: The image shows the logo for BARD DAVOL INC. The word "BARD" is in large, bold, sans-serif font, with each letter connected to the next. Below "BARD" is the text "DAVOL INC.", in a smaller, sans-serif font. The logo is black and white. TECHNOLOGY TECHNIQUE TRAINING ពេជ្យនេះ ## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS APR 2 8 2014 This 510(k) Summary is provided per the requirements of section 807.92(c). #### Submitter Information: | Submitter's Name: | Tony John, MS<br>Regulatory Affairs Specialist | |------------------------|------------------------------------------------| | Company Name: | Davol Inc., Subsidiary of C. R. Bard, Inc | | Company Address: | 100 Crossings Boulevard<br>Warwick, RI 02886 | | Telephone: | (401) 825-8692 | | Fax: | (401) 825-8763 | | Email: | tony.john@crbard.com | | Date Summary Prepared: | April 8, 2014 | | Device Name: | | | Trade Name: | XenMatrix™ Surgical Graft | | Product Code: | FTM | | Classification Name: | Mesh, Surgical | | Regulation Number: | 21 CFR 878.3300 | | Device Class: | II | #### Predicate Device: Panel: K081272 "Porcine Dermal Matrix Surgical Mesh" FDA cleared on 31 July 2008 from Brennen Medical, LLC . Davol, Inc. acquired the 510(k) product line in June 2009 and listed it as XenMatrix™ Surgical Graft. General & Plastic Surgery #### Device Descriptions: XenMatrix™ Surgical Graft is an acellular, sterile, non-pyrogenic porcine dermal matrix packed hydrated in sterile saline for use in the reconstruction of soft tissue deficiencies. #### Intended Use: . Intended for implantation to reinforce soft tissue where weakness exists and for surgical repair of damaged or ruptured soft tissue, including: abdominal plastic and reconstructive surgery; muscle flap reinforcement; hernia repair including abdominal, femoral, diaphragmatic, scrotal, umbilical, and incisional hernias. #### PREMARKET NOTIFICATION FOR XENMATRIX™ SURGICAL GRAFT K140501 PAGE 1 OF 2 11 {2}------------------------------------------------ ## Summary of Similarities and Differences in Technological Characteristics, Performance and Intended Uses: The XenMatrix™ Surgical Graft has the same intended use and fundamental scientific technology as the predicate device. The following technological characteristics of the XenMatrix™ Surgical Graft are the same as the predicate device: material, biocompatibility, and sterilization. The difference between the predicate device and the proposed device is the proposed device has larger sizes and smaller sizes than the currently cleared predicate. Performance testing demonstrates that these differences do not adversely affect the safety and effectiveness of the proposed device. Davol has also conducted a risk analysis to address the potential clinical risks that are associated with meshes that are larger than our previously cleared sizes. Based on our analysis and the results of our validation testing Davol does not expect any different clinical risks due to the larger mesh sizes being proposed. #### Summary of Non-Clinical Testing: Bench and pre-clinical testing was performed on the XenMatrix™ Surgical Graft to support substantial equivalence. Bench testing on the proposed device included burst strength, tensile strength, and suture retention strength. Packaging testing included seal strength and ship testing. Pre-clinical testing includes histology examination. Results of testing demonstrate that the XenMatrix™ Surgical Graft of the expanded sizes meets the same product specifications and intended uses. #### Statement of Equivalence: The test results and risk analysis provided in this submission support the safety and effectiveness of the device for its intended use and demonstrate that the proposed device is substantially equivalent to its predicate devices. #### PREMARKET NOTIFICATION FOR XENMATRIX™ SURGICAL GRAFT K140501 PAGE 2 OF 2 12 {3}------------------------------------------------ Image /page/2/Picture/0 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular in shape, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the circle is a stylized symbol that resembles an abstract bird or human figure. The symbol is composed of three curved lines that converge at the bottom. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002 April 28, 2014 C.R. Bard Mr. Tony John Regulatory Affairs Specialist 100 Crossings Boulevard Warwick, Rhode Island 02886 Re: K140501 Trade/Device Name: XenMatrix" Surgical Graft Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTM, OWV, OXH Dated: April 9, 2014 Received: April 10, 2014 Dear Mr. John: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (2 1 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {4}------------------------------------------------ Page 2 – Mr. Tony John forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # David Krause -S for - Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation . Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## INDICATION FOR USE STATEMENT 510(k) Number (if known): K140501 Device Name: XenMatrix™ Surgical Graft Indications for Use: Intended for implantation to reinforce soft tissue where weakness exists and for surgical repair of damaged or ruptured soft tissue, including: abdominal plastic and reconstructive surgery; muscle flap reinforcement; hernia repair including abdominal, femoral, diaphragmatic, scrotal, umbilical, and incisional hernias. Prescription Use (Part 2) CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page __ of_ Peter L. Hudson -S
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...