K140497 · Cardioinsight Technologies, Inc. · DQK · Nov 19, 2014 · Cardiovascular
Device Facts
Record ID
K140497
Device Name
ECVUE MAPPING SYSTEM
Applicant
Cardioinsight Technologies, Inc.
Product Code
DQK · Cardiovascular
Decision Date
Nov 19, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1425
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K140497 · Nov 19, 2014
ECVUE MAPPING SYSTEM
Cardioinsight Technologies, Inc.
Routine clinical practice records
Clinical experience from 60 patients in routine EP clinical practice was used to complement GLP animal study results and provide supporting evidence of safety and intended performance.
Routine clinical practice; Clinical experience; EP clinical practice
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective analysis of routine clinical practice
60 patients undergoing EP procedures; Sample Size: 60; Number of Sites: 2
Not applicable for this study
Safety and performance in routine clinical practice
Indications for Use
The CardioInsight ECVUE Mapping System is intended for acquisition, analysis, display and storage of cardiac electrophysiological data and maps for analysis by a physician.
Device Story
ECVUE is a non-invasive 3D cardiac mapping system. It acquires real-time chest ECG signals via a torso-mounted sensor array and integrates them with patient CT scan data. The system processes these inputs to generate virtual electrograms and interactive 3D color maps, including potential, activation, voltage, propagation, and phase maps. Used in clinical settings like the EP lab or at the patient's bedside, the system is operated by healthcare professionals. Physicians use the displayed maps and signal analyses to characterize cardiac rhythms, localize origins, and determine activation times. This information assists in clinical decision-making during electrophysiological studies, potentially benefiting patients by providing non-invasive diagnostic insights into cardiac rhythm patterns.
Clinical Evidence
Evidence includes bench testing (verification/validation, electrical safety, EMC), a GLP porcine animal study, and clinical experience from 60 patients across two European centers. The animal study validated the system's ability to acquire ECG signals, perform signal analysis, and display 3D maps to localize rhythms and characterize cycle lengths/activation times compared to standard EP study methods. Clinical data from routine practice supported the safety and intended performance of the system. No adverse events were reported in either the animal or clinical studies.
Technological Characteristics
Non-invasive 3D cardiac mapping system. Components include a torso-mounted sensor array for ECG signal acquisition and a programmable diagnostic computer. Integrates external CT scan data for anatomical mapping. Software provides signal analysis and 3D visualization (potential, activation, voltage, propagation, phase maps). Mobile form factor for bedside or EP lab use. Verification testing confirms compliance with electrical safety and electromagnetic compatibility standards.
Indications for Use
Indicated for use by physicians for the acquisition, analysis, display, and storage of cardiac electrophysiological data and maps in patients undergoing cardiac electrophysiological evaluation.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized image of three faces in profile, stacked on top of each other. The faces are black and are oriented to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
November 19, 2014
CardioInsight Technologies, Inc. Ms. Christina V. Vacca Vice President Quality, Regulatory Affairs and Operations 11000 Cedar Ave., Suite 210 Cleveland, OH 44106
Re: K140497
> Trade/Device Names: CardioInsight ECVUE Mapping System Regulatory Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II (Two) Product Code: DOK Dated: October 21, 2014 Received: October 22, 2014
Dear Ms. Vacca:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Page 2 - Ms. Christina V. Vacca
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Melissa A. Torres -S
For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
**Indications for Use**
| Form Approved: OMB No. 0910-0120 |
|-----------------------------------|
| Expiration Date: January 31, 2017 |
| See PRA Statement on last page. |
510(k) Number (if known)
Device Name ECVUE
Indications for Use (Describe)
The CardioInsight ECVUE Mapping System is intended for acquisition, and storage of cardiac electrophysiological data and maps for analysis by a physician.
Type of Use (Select one or both, as applicable)
🇿 Prescription Use (Part 21 CFR 801 Subpart D)
[ ] Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
FORM FDA 3881 (1/14)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{4}------------------------------------------------
| Submitter: | CardioInsight Technologies, Inc.<br>11000 Cedar Ave, Suite 210<br>Cleveland, OH 44106<br>chris@cardioinsight.com<br>Office: 216-453-5950 ext 105<br>Mobile: 440-315-6973 | |
|--------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| Contact Person: | Christina V. Vacca<br>Vice President Quality, Regulatory Affairs and Operations | |
| Date Prepared: | October 21, 2014 | |
| Name and<br>Classification | ECVUE™ | |
| | Electrophysiological cardiac mapping system:<br>Programmable diagnostic computer | 21 CFR<br>870.1425, DQK |
| Predicate Device(s): | The legally marketed device to which we are claiming equivalence is<br>Bard LabSystem PRO EP Recording System (K113811) | |
| Summary | ECVUE is a non-invasive mapping system for beat-by-beat, multi-<br>chamber, 3D mapping of the heart. The system displays cardiac maps<br>and virtual electrograms from real-time chest ECG signals (measured<br>by a Sensor Array placed on the torso) and CT scan data. The<br>ECVUE software provides various cardiac signal analyses and<br>displays interactive 3D color maps including potential, activation,<br>voltage, propagation, and phase maps. The system is mobile and can<br>be used for mapping at the patient's bedside or in the EP lab. | |
| Intended/Indication for<br>Use: | The CardioInsight ECVUE Mapping System is intended for<br>acquisition, analysis, display and storage of cardiac<br>electrophysiological data and maps for analysis by a physician. | |
| Device Characteristics<br>Compared to the<br>Predicate | Performance Testing: Bench<br>To assure reliable design and performance, the Company performed<br>in vitro testing including software and hardware verification and<br>validation, electrical safety testing, and electromagnetic compatibility<br>testing. The test results demonstrate that the ECVUE mapping system<br>meets the applicable standards and specifications. | |
| | Performance Testing: Animal<br>A GLP study was conducted to demonstrate the performance of the<br>ECVUE system in identifying origin and pattern of various rhythms<br>as compared to a standard electrophysiology (EP) study in a<br>controlled porcine model. The ability of the ECVUE system to<br>acquire ECG signals, perform signal analyses, and display 3D maps<br>to localize various rhythms and characterize cycle length and | |
| | activation times was established in this study. There were no adverse<br>events. The study clearly demonstrated the substantial equivalence of<br>the ECVUE system to the legally marketed predicate devices. | |
| | Performance Testing: Clinical Experience with commercially<br>available device<br>Clinical experience from 60 patients, representing the use of the<br>ECVUE system in routine EP clinical practice from two European<br>centers were included to complement the results of the GLP animal<br>study. There have been no adverse events reported with the use of<br>ECVUE. These clinical data provide strong supporting information<br>indicating that in routine EP clinical practice, the non-invasive<br>ECVUE system is safe and performs as intended. | |
| Conclusion: | CardioInsight believes the ECVUE mapping system is substantially<br>equivalent to the predicate devices. | |
| | The conclusions drawn from in-vitro testing, GLP animal testing and<br>clinical experience (discussed above) demonstrate that the device is<br>equivalent and is as safe and effective as the predicate devices. | |
{5}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.