← Product Code [JCT](/productcode/JCT) · K140472

# BONASTENT TRACHEAL / BRONCHIAL (K140472)

_Endochoice, Inc. · JCT · Oct 17, 2014 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K140472

## Device Facts

- **Applicant:** Endochoice, Inc.
- **Product Code:** [JCT](/productcode/JCT.md)
- **Decision Date:** Oct 17, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.3720
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The BONASTENT® Tracheal/Bronchial Stent System is indicated for the treatment of tracheobronchial strictures caused by malignant neoplasms.

## Device Story

Self-expanding tubular prosthesis; maintains patency of tracheobronchial strictures caused by malignant tumors. Composed of Nitinol wire in hook & cross weave; covered with silicone membrane to prevent migration. Available in diameters 10-20mm and lengths 50-80mm. Deployed via Y-Shape (≤10Fr) or I-Shape (>10Fr) handle delivery devices. Used by clinicians in medical settings to treat airway obstructions. Stent provides radial force to keep airway open; silicone cover prevents tissue ingrowth/migration. Benefits include restored airway patency and improved patient breathing.

## Clinical Evidence

Bench testing only. Performance validated via deployment testing, expansion force, compression force, dimensional testing, corrosion testing, and tensile strength testing. Biocompatibility established via reliance on reference devices (Bonastent Esophageal/Biliary) for cytotoxicity, sensitization, systemic toxicity, pyrogenicity, intracutaneous reactivity, mutagenicity, and intramuscular implantation.

## Technological Characteristics

Nitinol wire stent; hook & cross weave construction; silicone membrane covering. Self-expanding. Dimensions: 10-20mm diameter, 50-80mm length. Delivery systems: Y-Shape (≤10Fr) and I-Shape (>10Fr). No software or electronic components.

## Regulatory Identification

The tracheal prosthesis is a rigid, flexible, or expandable tubular device made of a silicone, metal, or polymeric material that is intended to be implanted to restore the structure and/or function of the trachea or trachealbronchial tree. It may be unbranched or contain one or two branches. The metal tracheal prosthesis may be uncovered or covered with a polymeric material. This device may also include a device delivery system.

## Special Controls

The special control for this device is FDA's “Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses.”

## Predicate Devices

- AERO Tracheobronchial Stent System ([K062511](/device/K062511.md))

## Reference Devices

- Bonastent™ Esophageal ([K092144](/device/K092144.md))
- Bonastent™ Biliary ([K093003](/device/K093003.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized design of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a sense of depth and unity. The seal is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 17, 2014

EndoChoice, Incorporated C/O Ms. Bosmat Friedman Regulatory Consultant Push-Med LLC/MJ RAC 1208-12 Rockford Road Toronto, ON M2R3A2 CANADA

Re: K140472

Trade/Device Name: BONASTENT® Tracheal/Bronchial Regulation Number: 878.3720 Regulation Name: Prosthesis, Tracheal, Expandable Regulatory Class: Class II Product Code: JCT Dated: September 15, 2014 Received: September 17, 2014

Dear Ms. Friedman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D. Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

K140472

### Device Name

BONASTENT® Tracheal/Bronchial

Indications for Use (Describe)

The BONASTENT® Tracheal/Bronchial Stent System is indicated for the treatment of tracheobronchial strictures caused by malignant neoplasms.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Form Approved: OMB No. 0910-0120

Expiration Date: December 31, 2013

{3}------------------------------------------------

Section 5

# 510(k) Summary

{4}------------------------------------------------

# 510(K) SUMMARY [as required by section 807.92(c)] BONASTENT® Tracheal/Bronchial 510(k) Number K140472

#### SUBMITTER 1.

## Applicant's Name:

EndoChoice, Inc. 18110 Wills Road, Suite 100 Alpharetta, GA 30009 Phone: 678-534-6021 Fax: 770-410-9008

## Contact Person:

Bosmat Friedman Regulatory Affairs Consultant 647-975-3974 bosmat@pushmed.com

### Date Prepared (revised):

October 15, 2014

#### 2. DEVICE

Trade Name: BONASTENT® Tracheal/Bronchial

# Common or Usual Name:

Tracheal Prosthesis

### Classification:

Name: Prosthesis, Tracheal, Expandable Product Code: JCT Regulation No: 878.3720 Class: 2 Classification Panel: General & Plastic Surgery

#### PREDICATE DEVICES 3.

### Main Predicate:

· AERO Tracheobronchial Stent System by Alveolus, Inc; K062511. Reference devices:

- · Bonastent™ Esophageal by EndoChoice Inc.; K092144.
- · Bonastent™ Biliary by EndoChoice Inc .; K093003.

{5}------------------------------------------------

#### 4. DEVICE DESCRIPTION

The BONASTENT® Tracheal/Bronchial (BTB) is a self-expanding tubular prosthesis designed to maintain patency of tracheobronchial strictures caused by malignant tumors. The stent is made of Nitinol wire and is weaved using a hook & cross wire construction; a silicon membrane covers the stent and is designed to prevent stent migration. Stent sizes range in diameter from 10mm to 20mm and in length from 50mm to 80mm.

The BONASTENT® Tracheal/Bronchial is available on two types of delivery devices. The Y-Shape Handle Delivery Device is used for stents with an outer diameter of 10Fr and less. The I-Shape Handle Delivery Device which is used for stents with an outer diameter larger than 10F.

#### 5. INDICATIONS FOR USE

The BONASTENT® Tracheal/Bronchial Stent System is indicated for the treatment of tracheobronchial strictures caused by malignant neoplasms.

#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE 6. PREDICATE DEVICE

The stent weaving and construction is identical to the weaving and construction of both Bonastent predicates. The Y-shape Handle delivery device is identical to the delivery device of the BONASTENT Biliary cleared under K093003. The I-Shape Handle delivery device is identical to the delivery device of the Bonastent Esophageal cleared under K092144.

The stent part was tested for compression and expansion and compared to the AERO stent. The results of the tests support the company's claim of substantial equivalence.

#### 7. PERFORMANCE DATA

# Bench Testing

The following performance tests were conducted in accordance with FDA Guidance for Industry - Guidance for the Content of Premarket Notifications for Esophageal and Tracheal Prostheses:

| # | Name of Test                 | Purpose                                                                                                                       |
|---|------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| 1 | Deployment<br>Testing        | To validate the accuracy and repeatability of the delivery system                                                             |
| 2 | Expansion Force              | To measure the force exerted by the metal self-expanding stent during<br>expansion (compared to predicate)                    |
| 3 | Compression<br>Force Testing | To measure the force required to compress the metal self-expanding<br>types of stents after expansion (compared to predicate) |
| 4 | Dimensional<br>Testing       | To verify the reproducibility of the metal self-expanding types of<br>stents length and diameter after deployment             |
| 5 | Corrosion<br>Testing         | To establish the compatibility of the stent materials with the corrosive<br>environment in the tracheobronchial tree          |
| 6 | Tensile strength<br>tests    | Performed for any deployment system that includes components that<br>are bonded or welded                                     |

{6}------------------------------------------------

The in-vitro tests that were performed confirm that all components, subassemblies, and/or full devices met the required specifications.

The comparative tests that were performed (expansion and compression) support our substantial equivalence claim to the AERO Tracheobronchial Stent System (K062511).

# Biocompatibility

Due to the similarities in materials and manufacturing techniques between the Bonastent Tracheal/Bronchial stent and its reference predicate devices the company relied on the previous results that were obtained. The following tests were conducted on the reference devices:

# Stent Part

- Cytotoxicity i
- Maximization sensitization ।
- -Acute systemic toxicity
- Rabbit Pyrogen Study ।
- -Intracutaneous reactivity
- Bacterial reverse mutation assay -
- Intramuscular implantation study (13 and 26 week). -

# Deliverv Device

- Cytotoxicity ।
- Maximization sensitization -
- Rabbit Pyrogen Study ।
- Intracutaneous reactivity ।

#### SUBSTANTIAL EQUIVALENCY TABLE 8.

The following table illustartes the similarties and differences bewteen the subject device and the main predicate, the AERO stent:

| Product name        | BONASTENT®<br>Tracheal/Bronchial Stent<br>System                                                                                            | AERO Tracheobronchial<br>Stent System                                                                                                                  |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) No.          | K140472                                                                                                                                     | K062511                                                                                                                                                |
| Intended Use        | The BONASTENT® Tracheal/Bronchial Stent System is indicated for the treatment of tracheobronchial strictures caused by malignant neoplasms. | The Alveolus AERO™ Tracheobronchial Stent System is indicated for use in the treatment of tracheobronchial strictures produced by malignant neoplasms. |
| Stent material      | Nitinol                                                                                                                                     | Nitinol                                                                                                                                                |
| Covering material   | Silicone                                                                                                                                    | polyurethane membrane                                                                                                                                  |
| Stent diameter (mm) | 10, 12, 14, 16, 18 and 20                                                                                                                   | 10, 12, 14, 16, 18 and 20                                                                                                                              |
| Stent length (mm)   | 20-80                                                                                                                                       | 20-80                                                                                                                                                  |
| Delivery diameter   | 8F - 12F                                                                                                                                    | 16F and 22F                                                                                                                                            |
| Deployment time     | Non-aged stents average 16.97 sec<br>Aged stents 16.4 sec                                                                                   | Unknown                                                                                                                                                |

{7}------------------------------------------------

| Expansion Force                              |                                                                   |                                      |
|----------------------------------------------|-------------------------------------------------------------------|--------------------------------------|
|                                              | 10X20 0.367 lbs.<br>20X80 0.748 lbs.                              | 10X20 0.390 lbs.<br>20X80 0.763 lbs. |
| Compression force                            |                                                                   |                                      |
|                                              | 10X20 0.517 lbs.<br>20X80 1.026 lbs.                              | 10X20 0.519 lbs.<br>20X80 1.026 lbs. |
| Corrosion (in<br>simulated gastric<br>fluid) | Resistant to corrosion for<br>a duration equivalent to 3<br>years | Unknown                              |

# Conclusion:

The Bonastent Tracheal/Bronchial has the same intended use and indication as the AERO stent. The two main differences between the Bonastent Tracheal/Bronchial and the AERO are: 1) different covering material and 2) different delivery device diameter. The results of the comparative testing (expansion and compression testing) as well as the biocompatibility results support our claim that the Bonastent Tracheal/Bronchial stent is substantially equivalent to the AERO stent.

Furthermore, the Bonastent Tracheal/Bronchial is substantially equivalent in technological characteristics to the two reference devices, Bonastent™ Esophageal and Bonastent™ Biliary.

Consequently, it is clear that the Bonastent Tracheal/Bronchial is substantially equivalent to its predicate and reference devices.

---

**Source:** [https://fda.innolitics.com/device/K140472](https://fda.innolitics.com/device/K140472)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
