DIOTECH LASER FIBER

K140458 · Diotech CO · GEX · Aug 26, 2014 · General, Plastic Surgery

Device Facts

Record IDK140458
Device NameDIOTECH LASER FIBER
ApplicantDiotech CO
Product CodeGEX · General, Plastic Surgery
Decision DateAug 26, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Diotech Laser Fiber is indicated for endovascular coagulation of the great saphenous vein in patients with superficial vein reflux, for the treatment of varicose veins and varicosities associated with superficial reflux of the great saphenous vein, and for the treatment of incompetence and reflux of superficial veins of the lower extremity. The Diotech Laser Fiber used with CW laser with wavelength range 810~1470nm, a power range 5~15W, using an SMA905 connector.

Device Story

Diotech Laser Fiber is a free-beam delivery device for endovascular laser coagulation. It consists of an optical fiber (silica core/cladding) and a catheter tube (PE, PP, polycarbonate). The device connects to a CW laser (810-1470nm, 5-15W) via an SMA905 connector. Used by physicians in clinical settings to treat lower extremity venous reflux; the fiber is inserted through the catheter into the vein. The fiber tip delivers laser energy diagonally (70-80° angle) to the vein wall to induce coagulation. Catheter markings (every 10mm) assist in depth control. The device is single-use and EO-sterilized. It benefits patients by providing a minimally invasive treatment for varicose veins.

Clinical Evidence

Bench testing only. Testing included biocompatibility (ISO 10993) and functional performance comparisons against the predicate device. No clinical data provided.

Technological Characteristics

Materials: Silica (fiber core/cladding), PE (catheter tube), PP (hub), Polycarbonate (outer inlet), Stainless steel (connector). Sensing/Actuation: Free-beam laser delivery. Energy: CW laser (810-1470nm, 5-15W). Dimensions: 2900mm total length, 850mm catheter length, 0.6mm fiber diameter. Connectivity: SMA905 connector. Sterilization: EO gas. Single-use.

Indications for Use

Indicated for patients with superficial vein reflux, varicose veins, varicosities associated with superficial reflux of the great saphenous vein, and incompetence/reflux of superficial veins of the lower extremity.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol consists of a series of overlapping human profiles, creating an abstract representation of people. The profiles are arranged in a way that suggests movement or progress. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 26, 2014 Plus Global Mr. Peter Chung 300 Atwood Street Pittsburgh, Pennsylvania 15213 Re: K140458 Trade/Device Name: Diotech Laser Fiber Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: July 15, 2014 Received: July 28, 2014 Dear Mr. Chung: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may; therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {2}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # David Krause -S - Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K140458 Device Name Diotech Laser Fibers #### Indications for Use (Describe) Diotech Laser Fibers (400um, 600um) is indicated for endovascular coagulation of the great saphenous vein in patients with superficial vein reflux, for the treatment of varicose veins and varicosties associated with superficial reflux of the great saphenous vein, and for the treatment of incompetence and reflux of superficial veins of the lower extremity. The Diotech Laser Fibers (400um, 600um) may be used with CW laser with wavelength range 810-1470mm, a power range 5-15W, using an SMA905 connector. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ## 1. Applicant Information - 1) Company : Diotech Co. - 2) Address : 301, Nakdong-daero, Saha-gu, Busan, Korea - 3) Device : Diotech Laser Fibers - 4) Phone Number : +82-51-292-6237 - 5) Fax Number : +82-51-292-6258 - 6) Homepage: http://www.diotech21.com ## 2. Device Information - 1) Trade Name : Diotech Laser Fibers - 2) Common Name : Diode Laser - 3) Regulation Name : Laser surgical instrument for use in general and plastic surgery and in dermatology - 4) Product code : GEX - 5) Regulation number : 878.4810 - 6) Class of device : Class II ### 3. The legally marketed device to which we are claiming equivalence - 1) 510(K) Number : K071959 - 2) Trade Name : AngioDynamics, Inc. NeverTouch 600 mm Fiber - 3) Product Code : GEX ### 4. Description of device Diotech Laser Fiber is free- beam delivery device that transits laser energy into intended direction. Fiber length is 2.9mters. And catheter tube length is 0.85meters. Catheter tube marked on per 10mm. This device can be connected with SMA905 connector and deliver to 810~1470m. This device can be connected with cleared surgical use laser diode. This device is single use only and sterilized by EO gas. #### 5. Intended Use Diotech Laser Fiber is indicated for endovascular cagulation of the great saphenous vein in patients with superficial vein reflux, for the treatment of varicose veins and varicosities associated with superficial reflux of the great saphenous vein, and for the treatment of incompetence and reflux of superficial veins of the lower extremity. The Diotech Laser Fiber used with CW laser with wavelength range 810~1470nm, a power range 5~15W, using an SMA905 connector. #### 6. Technological Characteristics Diotech laser fiber consists of two parts that are optic fiber and catheter. The optic fiber also consists of connector, optical fiber, and the tip part. Fiber cable is made of the optical fiber that delivers the laser beam, and tip part is made of glass that can deliver the laser beam diagonally. And the fiber may be used any laser wavelength between 810~1470 m that have been cleared for surgical use diode laser. Outer diameter of fiber cable is 0.6mm and length is 2900 mm. Catheter has a function that let the fiber cable be inserted into blood vessel as vein through its inner path. The catheter tube also consist of three parts as hub, tube and outer Inlet. Hub is used to connect with syringe. Tube is used for way that fiber cable may go through inner hole of tube. Outer Inlet is entrance hole for fiber to go through into tube. Catheter tube length is 0.85meters(850mm). It was marked on per 10mm to check depth to insert. The fiber core and cladding for the subject device are made from silica which is the same material used in the {5}------------------------------------------------ predicate device. This device can be connected with SMA905 connector. This device is single use only and sterilized by EO gas. ## 7. Performance Bench tests were performed. Bench testing included biocompatibility. The tests demonstrated that the device is as safe, as effective and performs in a substantially equivalent manner to the predicate device. #### 8. Predicate device comparison table | Trade Name | | | Diotech laser fiber | AngioDynamics, Inc. NeverTouch 600 um Fiber | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|-------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended use | | | This device is indicated for endovascular<br>coagulation of the great saphenous vein in<br>patients with superficial vein reflux, for the<br>treatment of varicose veins varicosities and<br>associated with superficial reflux of the great<br>saphenous vein, and for the treatment of<br>incompetence and reflux of superficial veins of<br>the lower extremity. | The AngioDynamics, Inc. NeverTouch 600gm<br>Fiber is indicated for endovascular coagulation of<br>the great saphenous vein in patients with superficial<br>vein reflux, for the treatment of varicose veins<br>varicosities and associated with superficial reflux<br>of the great saphenous vein, and for the treatment<br>of incompetence and reflux of superficial veins of<br>the lower extremity. | | 510(k) Number | | | K140458 | K071959 | | rediy | Design | | .1mm<br>0.6mm<br>10mm | | | | Diameter of Optical<br>Fiber | | 0.6 mm(DRF-A600-1, DBF-A600-1)<br>0.4 mm(DBF-A400-1) | 0.6 mm | | | Diameter of Glass Tube | | 1.1 mm | 0.905 mm | | | Length of Glass Tube | | 10 mm | 5.8 mm | | ទេវាទប្រពុជ | Design | | | | | | Total Length | | 85 cm | 80 cm & 90 cm | | | Operation Length | | 82 cm | 83 cm | | | Hub | | 5 cm | Similar | | | Outer Inlet | | 1.5 cm | Similar | | | Fiber | Optic Fiber | Silica | Silica | | sinə aid bitki növü. İstinadlar Rüspitaliya Şamaları Şahil Şamaları Şamaları Şamaları Şamaları Şamaları Şamaları Şamaları Şamaları Şamaları Şamaları Şamaları | | Connector | Stainless steel | Stainless steel & Plastic | | | Catheter | Hub | PP | Unknown | | | | Outer inlet | Polycarbonate | Unknown | | | | Tube | PE | PE | | Fiber transmission angle | | | 70~80° | 30~50° | | Wavelength range | | | 810~1470nm | 810~1470nm | {6}------------------------------------------------ | Maximum output power | 15W | Unknown | |----------------------|-------------------|-------------------| | Single use | Single use | Single use | | Biocompatibility | ISO10993 | ISO10993 | | Sterility | EO gas sterilized | EO gas sterilized | ## 9. Conclusion The Device is investigated for function and effectiveness to compare the operation of function between Diotech laser fiber and predicate devices . Comparison results demonstrate that the specifications and performance of the device are similar as functional and effective as the legally marketed predicate device. Therefore, it is concluded that Diotech laser fiber is substantially equivalent to the legally marketed predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...