K140455 · Trinity Biotech USA · PCL · May 22, 2014 · Microbiology
Device Facts
Record ID
K140455
Device Name
CAPTIA MEASLES IGM
Applicant
Trinity Biotech USA
Product Code
PCL · Microbiology
Decision Date
May 22, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3520
Device Class
Class 1
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K140455 · May 22, 2014
CAPTIA MEASLES IGM
Trinity Biotech USA
Public health department clinical serum samples (Texas Department of Health, North Carolina State Laboratory of Public Health, Kansas Department of Health and Environment)
Retrospective clinical samples from patients suspected of measles infection were used to evaluate the clinical performance (sensitivity/specificity) of the device against established comparator algorithms used in routine clinical/public health practice.
Retrospective clinical samples; Public health laboratory data; Clinical performance evaluation
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Texas and North Carolina Public Health Outbreak Samples; Retrospective clinical performance evaluation
Individuals suspected of measles infection in outbreak settings; Sample Size: 188; Number of Sites: 2
Comparator algorithm (including other Measles IgM ELISA, IFA, CF, HI, and IgG tests)
Clinical sensitivity and specificity
Kansas Department of Health and Environment Laboratory Study; Retrospective clinical performance evaluation
Patients suspected of measles infection; Sample Size: 8; Number of Sites: 1
Comparator algorithm (including PCR on urine, NPS, and/or throat swabs)
Clinical sensitivity and specificity
Indications for Use
The Trinity Biotech Captia™ Measles IgM Enzyme-Linked Immunosorbent Assay (ELISA) is intended for the qualitative detection of Measles IgM antibodies in human serum of patients suspected of measles (rubeola) infection. This assay is intended for use as an aid in the diagnosis of a current or recent measles (rubeola) infection in conjunction with other clinical information and laboratory findings. Performance characteristics have not been evaluated in immunocompromised or immunosuppressed individuals. This test is not intended for use in neonatal screening or for use at a point of care. This test is not intended for use in screening blood and plasma donors.
Device Story
ELISA-based diagnostic test for qualitative detection of Measles IgM antibodies in human serum. Device uses microassay plates coated with Edmonston strain measles antigen. Patient serum added; antigen-specific IgM binds to solid-phase antigen. After washing, goat anti-human IgM conjugated with horseradish peroxidase (HRP) is added, binding to antigen-antibody complexes. TMB substrate added; enzymatic reaction produces blue color, turning yellow upon addition of 1N H2SO4 stop solution. Color intensity measured via spectrophotometer/ELISA plate reader. Used in clinical laboratories by trained personnel. Output is Immune Status Ratio (ISR) calculated from optical density relative to cut-off calibrator. Results aid clinicians in diagnosing current/recent measles infection alongside other clinical/laboratory data. Benefits include rapid identification of acute infection to support clinical management and public health response.
Clinical Evidence
Clinical performance evaluated via three studies (n=196 total). Studies 1 and 2 (n=188) compared Captia™ Measles IgM against a composite comparator algorithm (including other ELISA, IFA, CF, HI, and IgG tests). Study 3 (n=8) compared against PCR-based algorithms. Seroconversion study (n=15 samples over 55 days) demonstrated detection of IgM onset. Normal population study (n=200) established expected values. Analytical performance included precision/reproducibility (inter-assay, inter-lot, intra-run), cross-reactivity (CMV, HSV, etc.), and interference testing (hemoglobin, bilirubin).
Indicated for qualitative detection of Measles IgM antibodies in human serum of patients suspected of measles (rubeola) infection. Aid in diagnosis of current/recent infection. Not for immunocompromised/immunosuppressed individuals, neonatal screening, point-of-care, or blood/plasma donor screening.
Regulatory Classification
Identification
Rubeola (measles) virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to rubeola virus in serum. The identification aids in the diagnosis of measles and provides epidemiological information on the disease. Measles is an acute, highly infectious disease of the respiratory and reticuloendothelial tissues, particularly in children, characterized by a confluent and blotchy rash.
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K140455
B. Purpose for Submission:
Clearance of New Device
C. Measurand:
Measles specific IgM antibodies
D. Type of Test:
Enzyme-Linked Immunosorbent Assay (ELISA)
E. Applicant:
Trinity Biotech
F. Proprietary and Established Names:
Captia™ Measles IgM
G. Regulatory Information:
1. Regulation section:
21 CFR 866.3520 Rubeola (measles) virus serological reagents
2. Classification:
Class I
3. Product code:
PCL
4. Panel:
83 (Micobiology)
H. Intended Use:
1. Intended use(s):
The Trinity Biotech Captia™ Measles IgM Enzyme-Linked Immunosorbent Assay (ELISA) is intended for the qualitative detection of Measles IgM antibodies in human serum of patients suspected of measles (rubeola) infection. This assay is intended for use as an aid in the diagnosis of a current or recent measles (rubeola) infection in conjunction with other clinical information and laboratory findings.
{2}
Performance characteristics have not been evaluated in immunocompromised or immunosuppressed individuals. This test is not intended for use in neonatal screening or for use at a point of care. This test is not intended for use in screening blood and plasma donors.
2. Indication(s) for use:
Same as Intended Use
3. Special conditions for use statement(s):
None
4. Special instrument requirements:
None
I. Device Description:
The Trinity Biotech Captia™ Measles IgM test is an Enzyme-Linked Immunosorbent Assays (ELISA). When measles antigen (Edmonston strain) is bound to the solid phase and brought into contact with a patient's serum, antigen specific antibody, if present, will bind to the antigen on the solid phase forming antigen-antibody complexes. Excess antibody is removed by washing. This is followed by the addition of goat anti-human IgM globulin conjugated with horseradish peroxidase which will bind to the antibody-antigen complexes. The excess conjugate is removed by washing, followed by the addition of Chromogen/Substrate tetramethylbenzidine (TMB). If specific antibody to the antigen is present in the patient's serum, a blue color develops. When the enzymatic reaction is stopped with 1N H₂SO₄, The contents of the wells turn yellow. The color, which is indicative of the concentration of antibody in the serum, can be read on a suitable spectrophotometer or ELISA microwell plate reader. Each kit contains the following components in sufficient quantities to perform the number of tests indicated on the package label.
1. Purified measles antigen coated microassay plate: 96 wells, configured in twelve 1x8 strips stored in a foil pouch with desiccant. (96T: one plate)
2. Calibrator: Human serum or defibrinated plasma. Sodium azide (< 0.1%) and pen/strep (0.01%) added as preservatives, with kit specific factor printed on vial label. The Calibrator is used to calibrate the assay to account for day-to-day fluctuations in temperature and other testing conditions. (96T: one vial, 0.4 mL)
3. Positive Control: Human serum or defibrinated plasma. Sodium azide (< 0.1%) and pen/strep (0.01%) added as preservatives, with established range printed on vial label. The Positive Control is utilized to control the positive range of the assay.(96T: one vial, 0.4 mL)
4. Negative Control: Human serum or defibrinated plasma. Sodium azide (< 0.1%) and pen/strep (0.01%) added as preservatives, with established range printed on vial label. The Negative Control is utilized to control the negative range of the assay (96T: one vial, 0.4 mL)
2
{3}
5. Horseradish-peroxidase (HRP) Conjugate: Ready to use. Goat anti-human IgM, containing ProClin® (0.1%) and gentamicin as preservatives. (96T: one bottle, 16 mL)
6. Serum Diluent Plus: Ready to use, Contains goat/sheep anti-human IgG for serum absorption to remove competing IgG, with protein stabilizers and ProClin® (0.1%) as a preservative. (96T: two bottles, 45 mL each)
7. Wash Buffer Type I (20X concentrate): dilute 1 part concentrate + 19 parts deionized or distilled water. Contains TBS, Tween-20 and ProClin® (0.1%) as a preservative. (96T: one bottle, 50 mL)
8. Chromogen/Substrate Solution Type I: Tetramethylbenzidine (TMB), ready to use. The reagent should remain closed when not in use. If allowed to evaporate, a precipitate may form in the reagent wells. (96T: one bottle, 15 mL)
9. Stop Solution: Ready to use, contains a 1N H₂SO₄ solution. (96T: one bottle, 15 mL)
10. Package Insert
J. Substantial Equivalence Information:
1. Predicate device name(s):
Gull Rubeola IgM ELISA kit
2. Predicate 510(k) number(s):
K904083
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Indications for use | The Trinity Biotech
Captia™ Measles IgM
Enzyme-Linked
Immunosorbent Assay (ELISA) is intended for the qualitative detection of Measles IgM antibodies in human serum of patients suspected of measles (rubeola) infection. This assay is intended for use as an aid in the diagnosis of a current or recent measles (rubeola) infection in conjunction with other clinical information and laboratory findings. | The Gull Rubeola IgM ELISA test is intended for the qualitative detection of IgM antibody to rubeola (measles) virus in human serum. When performed according to the instructions, the Rubeola IgM ELISA test is of value in the determination of recent exposure to and infection with rubeola virus and in the diagnosis of rubeola virus-associated disease. |
{4}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | Performance characteristics have not been evaluated in immunocompromised or immunosuppressed individuals. This test is not intended for use in neonatal screening or for use at a point of care. This test is not intended for use in screening blood and plasma donors. | |
| Technology | Enzyme-Linked Immunosorbent Assay (ELISA) | Same |
| Specimen Types | Human Serum | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| None | Not Applicable | Not Applicable |
K. Standard/Guidance Document Referenced (if applicable):
CLSI EP7-A2: Guidance for Interference Testing in Clinical Chemistry (Vol. 25, No.27 Second Ed) (Nov 2005)
CLSI H18-A4: Procedures for Handling & Processing Blood Specimens (2010)
CLSI EP5-A2: Guidance for Evaluation of Precision Performance of Quantitative Measurement Methods (Vol. 24, No. 25 Second Ed.) (Aug 2004).
CLSI EP12-A: Guidance for User Protocol for Evaluation of Qualitative Test Performance (Vol. 22, No. 14) (Sept 2002).
L. Test Principle:
The Trinity Biotech Captia™ Measles IgM test is an Enzyme-Linked Immunosorbent Assay (ELISA). When measles antigen (Edmonston strain) is bound to the solid phase and brought into contact with a patient's serum, antigen specific antibody, if present, will bind to the antigen on the solid phase forming antigen-antibody complexes. Excess antibody is removed by washing. This is followed by the addition of goat anti-human IgM globulin conjugated with horseradish peroxidase which will bind to the antibody-antigen complexes. The excess conjugate is removed by washing, followed by the addition of Chromogen/Substrate tetramethylbenzidine (TMB). If specific antibody to the antigen is present in the patient's
{5}
serum, a blue color develops. When the enzymatic reaction is stopped with $1\mathrm{N}$ $\mathrm{H}_2\mathrm{SO}_4$ , The contents of the wells turn yellow. The color, which is indicative of the concentration of antibody in the serum, can be read on a suitable spectrophotometer or ELISA microwell plate reader.
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
The Trinity Biotech Captia™ Measles IgM ELISA Test Kit was evaluated for inter-assay precision. This study consisted of six (6) blinded proficiency panel members, varying in reactivity: Four (4) Low to Moderate Positive samples and two (2) Negative samples were run in triplicate at three (3) separate facilities. Testing at each site was done over 10 days by two (2) operators at each site resulting in a total number of 180 data points. The results are summarized below:
Precision between all sites
| Sample ID | ISR Site 1 N=60 | ISR Site 2 N=60 | ISR Site 3 N=60 | MEAN | STD DEV | % CV |
| --- | --- | --- | --- | --- | --- | --- |
| 1 | 1.30 | 1.34 | 1.23 | 1.29 | 0.055678 | 4.3% |
| 2 | 0.57 | 0.68 | 0.67 | 0.64 | 0.060828 | 9.5% |
| 3 | 1.16 | 1.26 | 1.07 | 1.16 | 0.095044 | 8.2% |
| 4 | 1.92 | 1.84 | 1.76 | 1.84 | 0.08000 | 4.3% |
| 5 | 0.67 | 0.82 | 0.84 | 0.78 | 0.092916 | 12.0% |
| 6 | 1.28 | 1.18 | 1.01 | 1.16 | 0.136504 | 11.8% |
The Trinity Biotech Captia™ Measles IgM ELISA Test Kit was evaluated for lot to lot precision. This study consisted of six (6) blinded proficiency panel members, varying in reactivity: four (4) Low to Moderate Positive samples and two (2) Negative samples run in triplicate on three (3) separately manufactured kit lots. The results are summarized below:
Inter-lot Reproducibility
| Sample ID | ISR Lot 1 | ISR Lot 2 | ISR Lot 3 | Mean | Std Dev | % CV |
| --- | --- | --- | --- | --- | --- | --- |
| 1 | 1.22 | 1.43 | 1.36 | 1.34 | 0.120773 | 9.02 |
| 2 | 0.57 | 0.42 | 0.46 | 0.48 | 0.108972 | 22.55 |
| 3 | 1.19 | 1.31 | 1.24 | 1.25 | 0.103253 | 8.27 |
| 4 | 2.13 | 1.96 | 1.56 | 1.88 | 0.27924 | 14.83 |
| 5 | 0.79 | 0.75 | 0.56 | 0.7 | 0.109545 | 15.72 |
| 6 | 1.27 | 1.32 | 1.3 | 1.29 | 0.041265 | 3.19 |
{6}
A separate precision/reproducibility study was performed internally (Trinity Biotech, Jamestown) using one (1) lot of the Captia™ Measles IgM ELISA Test Kit. The study compared the consistency between two (2) Operators. This study consisted of six (6) blinded proficiency panel members, varying in reactivity: Four (4) Low to Moderate Positive samples and two (2) Negative samples were run by each of the operators over 10 days. The results are summarized below:
Intra-run Precision/Reproducibility Between users:
Total Precision at Site 1 with Two Operators Over Ten Days
| Sample ID | ISR Operator 1
N=30 | ISR Operator 2
N=30 | MEAN ISR | STD DEV | % CV |
| --- | --- | --- | --- | --- | --- |
| 1 | 1.26 | 1.33 | 1.3 | 0.0495 | 3.80% |
| 2 | 0.58 | 0.55 | 0.57 | 0.02121 | 3.80% |
| 3 | 1.14 | 1.17 | 1.16 | 0.02121 | 1.80% |
| 4 | 1.92 | 1.91 | 1.92 | 0.00707 | 0.40% |
| 5 | 0.69 | 0.65 | 0.67 | 0.02828 | 4.20% |
| 6 | 1.32 | 1.23 | 1.28 | 0.06364 | 5.00% |
b. Linearity/assay reportable range:
Not Applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability – The Controls and Calibrators are traceable against frozen internal Quality Control standard panels. The panels consist of patient samples whose values span the assay range.
Stability – Real time stability studies support the shelf life of the Reagents, Controls and Calibrators at $2 - 8^{\circ}\mathrm{C}$ up to the stated expiration date on the labeling.
Expected Values – The ISR (Immune Status Ratio) Values for the Positive and Negative Controls should be in their respective ranges printed on the vial labels. If the Control values are not within their respective ranges, the test should be considered invalid and should be repeated.
d. Detection limit:
Not Applicable
{7}
# e. Analytical specificity:
# Cross Reactivity:
A total of 47 samples were used for establishing the potential cross reactivity of the Trinity Biotech Captia™ Measles IgM ELISA Test Kit. The samples were selected as confirmed positive by Trinity Biotech Captia™ CMV IgM, HSV1 IgM, HSV2 IgM, Rubella IgM, VZV IgM and Mumps IgM test kits, for each of the following potentially cross-reacting agents: CMV, HSV-1, HSV-2, Rubella, VZV, Mumps and Parvo-B19. Fourteen (14) of the suspected cross reactive samples tested as measles IgM antibody positive or equivocal (12 positive and 2 equivocal) with the Trinity Biotech Captia™ Measles IgM ELISA test kit. These 14 samples were then run on a comparator Measles IgM IFA test kit and a comparator Measles IgM ELISA test kit. A sample was determined as measles IgM antibody positive if the results on two (2) measles IgM tests were positive. A sample was determined as measles IgM antibody negative if the results on two (2) measles IgM tests were negative. The results are presented in the table below:
| Cross Reacting Agents | Number of Samples | Trinity Biotech Captia™ Measles IgM Result | | | | Consensus Comparator Result* | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Neg | Pos | Equiv | | Neg | Pos |
| Mumps IgM | 4 | 4 | 0 | 0 | | NT | NT |
| VZV IgM | 4 | 4 | 0 | 0 | | NT | NT |
| Rubella IgM | 4 | 4 | 0 | 0 | | NT | NT |
| CMV IgM | 7 | 3 | 3 | 1 | | 3 | 1 |
| HSV1 IgM | 16 | 9 | 7 | 0 | | 5 | 2 |
| HSV 2 IgM | 10 | 7 | 2 | 1 | | 2 | 1 |
NT = Not tested
* Tested only if Captia™ Measles IgM result was positive.
Potential cross reactivity was observed with CMV IgM, HSV1 IgM and HSV2 IgM. Potential cross-reactivity with Parvovirus, Respiratory Syncytial Virus (RSV) and parainfluenza has not been ruled out, as either specimens were not tested or a limited number were tested.
# Interfering Substances:
Interfering substance testing was conducted using three (3) characterized samples for measles IgM, spiked with recommended interfering substances according to the guidance in the CLSI document EP7-A2 (Interference Testing in Clinical Chemistry; Approved Guideline 2nd Ed). Results from the three (3) samples tested as spiked with the interfering substance and unspiked (as a control) were favourable. No adverse results or significant change in results occurred with these samples tested with the following:
{8}
| Potential Interferent | Concentration |
| --- | --- |
| Hemoglobin | ≥500 mg/dL |
| Bilirubin (unconjugated) | ≥20 mg/dL |
Interferents with substances such as red cells, cholesterol, gamma globulin, triglycerides, beta carotene, protein, ascorbic acid and anticoagulants have not been tested.
## High Dose Hook Effect:
When antigen is evaluated for purchase, it is titrated to ensure that there is no dose hook effect (prozone) of antigen on the plate. If prozoning is apparent then the antigen is titrated out further past the prozone point.
Due to the nature of obtaining IgM positive serum in general, positive serum purchased for the Captia Measles IgM generally contains less antibody than would be present to demonstrate dose hook effect (prozone), although the positive control in each kit lot is titrated to negative at each kit QC test stage to ensure the positive control is not prozoning.
## IgM/IgG Competition:
Specific IgG may compete with the IgM for sites and may result in a false negative. Conversely, rheumatoid factor in the presence of specific IgG may result in a false positive reaction. The Serum Diluent Plus Solution diminishes competing virus-specific IgG and minimizes rheumatoid factor interference in samples. Studies indicate that the maximum amount of IgG which can be removed by the kit Serum Diluent Plus is in excess of the expected high end of the normal range for IgG > 1380 mg/dL. The highest titer of RF+ tested (1: 1280; 500 IU/mL) did not adversely affect the performance of the assay
## f. Assay cut-off:
The objective of this study was to demonstrate the performance of the Trinity Biotech Captia™ Measles IgM ELISA Test Kit; specifically to generate data to demonstrate determination of the cut-off. The results are summarized below:
Measles IgM negative sera (228) were assayed on the Trinity Biotech Captia™ Measles IgM ELISA Test Kit. Four (4) samples were disqualified as they were in the equivocal range on Trinity IgM ELISA or indeterminate on a comparator Measles IFA test kit (due to non-specific staining most likely due to the presence of ANA antibodies). The mean and standard deviation of the optical density readings for the sera were 0.105 and 0.070 respectively. The positive threshold for the assay was determined by adding the mean and 4 standard deviations (0.105 + 4(0.070) = 0.385). Positive sera were titrated to give a constant ratio of the threshold value to obtain a Cut-off Calibrator serum. On all subsequent assays, this serum was run and the assay was calibrated by multiplying the O.D. value for the calibrator by the ratio to the cut-
{9}
off to obtain the cut-off calibrator value. This value was then divided into the O.D. for the patient sera to obtain an Immune Status Ratio (ISR). By definition, the Cut-off ISR is equal to 1.00. To account for inherent variation in the immunoassay, values of 0.91 to 1.09 were considered equivocal. Therefore values $\leq 0.90$ are considered negative and values $\geq 1.10$ are considered positive.
2. Comparison studies:
a. Method comparison with predicate device:
Not Applicable
b. Matrix comparison:
Not Applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable
b. Clinical specificity:
Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable):
The clinical performance was evaluated using the results of three (3) studies that were conducted to evaluate the performance of the Trinity Biotech Captia™ Measles IgM ELISA test kit.
Studies 1 and 2:
A total of 188 samples were collected from individuals for whom a measles IgM test was ordered (suspected of measles infection in outbreak settings by the Texas Department of Health, Austin, Texas (66 samples) and the State Laboratory of Public Health in Raleigh, North Carolina (122 samples)). All serum samples were tested on the Trinity Biotech Captia™ Measles IgM ELISA test kit. All but one (1) sample was tested by a comparator testing algorithm used at each institution to determine the presence of current or recent measles infection, referred to as "comparator algorithm" in the performance tables. The algorithms used consisted of the results of other laboratory confirmation methods including comparator Measles IgM ELISA test kit, a comparator Measles IgM IFA test kit, Complement Fixation (CF), Hemagglutination Inhibition (HI) and the Trinity Biotech Captia™ Measles ELISA IgG test kit. See the table below for results.
9
{10}
10
# Comparator Algorithm
| | + | Indeterminate | - |
| --- | --- | --- | --- |
| + | 88 | 1 | 6 |
| Equivocal | 3 | 3 | 10 |
| - | 5 | 1 | 70 |
# Percent of Captia™ Measles IgM Results that are
Equivocal = 8.6% (16 of 187)
Performance with the indeterminate results from the comparator algorithm calculated against the performance of the Captia™ Measles IgM:
# Comparator Algorithm
| | + | Indeterminate | - |
| --- | --- | --- | --- |
| + | 88 | 1 | 6 |
| Equivocal | 3 | 3 | 10 |
| - | 5 | 1 | 70 |
# Study 3:
Eight (8) samples were submitted for testing at the Kansas Department of Health and Environment Laboratory (KDHE), Topeka, Kansas. All samples were tested on the Captia™ Measles IgM and Captia™ Measles IgG ELISA test kits as well as compared with the method used by the testing institution to determine the presence of current or recent measles infection, referred to as “comparator algorithm” in the performance table. The comparator algorithm included RNA polymerase chain reaction (PCR) on urine, nasopharyngeal swab (NPS) and/or throat swab (TS) samples. The final % agreement is presented in the table below:
{11}
11
| Trinity Biotech Captia™
Measles IgM | Comparator Algorithm | | | |
| --- | --- | --- | --- | --- |
| | | + | Indeterminate | - |
| | + | 3 | 0 | 0 |
| | Equivocal | 0 | 0 | 0 |
| | - | 0 | 1* | 4 |
| | # Agree | Total | Result | 95% CI |
| % Positive Agreement | 3 | 3 | 100.00% | 29.2-100.0% |
| % Negative Agreement | 4 | 4 | 100.00% | 39.8-100.0% |
* This sample tested NPS negative at two sites, urine positive at one site and urine indeterminate at one site (all testing done by PCR).
Performance with the indeterminate results from the comparator algorithm calculated against the performance of the Captia™ Measles IgM:
| Trinity Biotech Captia™
Measles IgM | Comparator Algorithm | | | |
| --- | --- | --- | --- | --- |
| | | + | Indeterminate | - |
| | + | 3 | 0 | 0 |
| | Equivocal | 0 | 0 | 0 |
| | - | 0 | 1* | 4 |
| | # Agree | Total | Result | 95% CI |
| % Positive Agreement | 3 | 4 | 75.00% | 30.1-95.4% |
| % Negative Agreement | 4 | 4 | 100.00% | 39.8-100.0% |
# Seroconversion study:
The BIOMEX Rubeola IgG and IgM Antibody Seroconversion Panel (Cat No. SCP-MEA-001) contained 15 samples depicting the onset and decline of IgG and IgM antibodies to Rubeloa (Measles) from one individual over a period of 55 days. The IgM response was tested on four (4) different lots of the Trinity Biotech Captia™ Measles IgM ELISA Test Kit. The seroconversion panel consisted of one patient with 15 draws during an approximately two month period from May 31st to July 25th, 1994. Seroconversion was demonstrated at Day 20 with the Trinity Biotech Captia™ Measles IgM ELISA Test Kit, and remained positive until Day 22. The first comparator assay (comparator 1) demonstrated seroconversion at Day 22 and was positive only at that bleed date. The second comparator assay (comparator 2) seroconversion happened earlier than the Trinity kit by 2 days and remained positive for the remainder of the timeframe. The results are summarized below:
{12}
| | Trinity Biotech Captia™ Measles IgM ELISA | | | | | | | | Comparator 1 Rubeola IgM | Comparator 2 Rubeola IgM | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Lot 027 | | Lot 28 | | Lot 29 | | Lot 030 | | | | | |
| Bleed Day | ISR | Result | ISR | Result | ISR | Result | ISR | Result | S/CO | Result | S/CO | Result |
| Day1 | 0.22 | Neg | 0.29 | Neg | 0.2 | Neg | 0.12 | Neg | NT | NT | 0.429 | Neg |
| Day 6 | 0.29 | Neg | 0.33 | Neg | 0.24 | Neg | 0.15 | Neg | 0.02 | Neg | 0.449 | Neg |
| Day 8 | 0.25 | Neg | 0.31 | Neg | 0.21 | Neg | 0.14 | Neg | -0.04 | Neg | 0.389 | Neg |
| Day 13 | 0.26 | Neg | 0.34 | Neg | 0.27 | Neg | 0.21 | Neg | 0.28 | Neg | 1.101 | Pos |
| Day 15 | 0.44 | Neg | 0.51 | Neg | 0.54 | Neg | 0.43 | Neg | 0.46 | Neg | 3.045 | Pos |
| Day 20 | 1.53 | Pos | 1.71 | Pos | 1.77 | Pos | 1.28 | Pos | 0.92 | Neg | 7.332 | Pos |
| Day 22 | 1.36 | Pos | 1.4 | Pos | 1.49 | Pos | 1.24 | Pos | 1.09 | Pos | 5.919 | Pos |
| Day 27 | 0.84 | Neg | 0.92 | equiv | 0.96 | Equiv | 0.71 | Neg | 0.5 | Neg | 3.785 | Pos |
| Day 29 | 0.59 | Neg | 0.8 | Neg | 0.89 | Neg | 0.54 | Neg | 0.33 | Neg | 3.296 | Pos |
| Day 35 | 0.54 | Neg | 0.63 | Neg | 0.62 | Neg | 0.47 | Neg | 0.61 | Neg | 2.449 | Pos |
| Day 37 | 0.44 | Neg | 0.53 | Neg | 0.53 | Neg | 0.79 | Neg | 0.24 | Neg | 2.113 | Pos |
| Day 43 | 0.44 | Neg | 0.6 | Neg | 0.54 | Neg | 0.38 | Neg | 0.17 | Neg | 2.04 | Pos |
| Day 45 | 0.37 | Neg | 0.5 | Neg | 0.5 | Neg | 0.31 | Neg | 0.15 | Neg | 1.802 | Pos |
| Day 49 | 0.34 | Neg | 0.59 | Neg | 0.39 | Neg | 0.28 | Neg | 0.04 | Neg | 1.471 | Pos |
| Day 54 | 0.38 | Neg | 0.5 | Neg | 0.42 | Neg | 0.29 | Neg | 0.27 | Neg | 1.47 | Pos |
Specificity in a Normal Population:
A total of 200 random serum samples were tested to establish the expected values in a population of individuals between the ages 18-65 with no known clinically apparent Measles infection. The table below summarizes the distribution of Trinity Biotech Measles IgM assay ISR Values observed for the population.
Distribution of Trinity Biotech Captia™ Measles IgM Assay ISR Values from a Normal Population
| Measles IgM ISR Range | Number of Specimens | Percent of Total |
| --- | --- | --- |
| 0.00-0.20 | 65 | 32.5% |
| 0.21-0.40 | 91 | 45.5% |
| 0.41-0.60 | 29 | 14.5% |
| 0.61-0.80 | 8 | 4.0% |
| 0.81-0.90 | 4 | 2.0% |
| 0.91-1.10 | 1 | 0.5% |
| 1.11-1.20 | 0 | 0.0% |
| 1.21-1.40 | 1 | 0.5% |
| >2.00 | 1 | 0.5% |
{13}

4. Clinical cut-off:
Not Applicable
5. Expected values/Reference range:
IgM antibodies appear in the first week of infection and usually peak within a month. Although uncommon, low levels of IgM may persist for one (1) year or longer. The annual incidence rates are reported to vary in different geographical areas. Recent reports (2004-2011) from the Centers for Disease Control and Prevention (CDC) indicate that annual measles incidence is $< 1$ reported case per 1 million population. A large proportion of cases $(40\%)$ are international imported or associated with importation. Unknown source cases are not linked to any endemic chain $(12\%)$ . Most imported cases do not lead to spread in the United States. This is attributed to high levels of vaccination coverage (CDC Report: Documentation and Verification of Measles, Rubella and Congenital Rubella Syndrome Elimination in the Region of the Americas, United States National Report, CDC, March 28, 2012).
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.