BD PROBETEC NEISSERIA GONORRHOEAE (GC) QX AMPLIFIED DNA ASSAY
K140448 · Becton, Dickinson & CO · LSL · May 20, 2014 · Microbiology
Device Facts
Record ID
K140448
Device Name
BD PROBETEC NEISSERIA GONORRHOEAE (GC) QX AMPLIFIED DNA ASSAY
Applicant
Becton, Dickinson & CO
Product Code
LSL · Microbiology
Decision Date
May 20, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3390
Device Class
Class 2
Indications for Use
The BD ProbeTec Neisseria gonorrhoeae (GC) Qx Amplified DNA Assay, when tested with either the BD Viper™ System in Extracted Mode or the BD Viper™ LT System, uses Strand Displacement Amplification technology for the direct, qualitative detection of Neisseria gonorrhoeae DNA in clinician-collected female endocervical and male urethral swab specimens, patient-collected vaginal swab specimens (in a clinical setting), and male and female urine specimens (both UPT and Neat). The assay is also intended for use with gynecological specimens collected in BD SurePath™ Preservative Fluid or PreservCyt™ Solution using an aliquot that is removed prior to processing for either the BD SurePath or ThinPrep™ Pap test. The assay is indicated for use with asymptomatic and symptomatic individuals to aid in the diagnosis of gonococcal urogenital disease.
Device Story
The BD ProbeTec GCQ Assay is a molecular diagnostic test for Neisseria gonorrhoeae. It utilizes Strand Displacement Amplification (SDA) technology to amplify and detect target DNA sequences. The assay uses dried reagents in disposable microwells: a Priming Microwell (primers, detector probe, nucleotides) and an Amplification Microwell (DNA polymerase, restriction endonuclease). The device is designed for use on the BD Viper LT System, a bench-top laboratory instrument that automates nucleic acid extraction, amplification, and detection. The system processes various specimen types, including swabs, urine, and liquid-based cytology (LBC) media. The instrument provides qualitative results to laboratory personnel, which are used by clinicians to aid in the diagnosis of gonococcal urogenital infections. The assay benefits patients by providing an automated, standardized method for detecting N. gonorrhoeae, facilitating timely clinical decision-making.
Clinical Evidence
Clinical performance was evaluated using 653 female and 170 male subjects across four clinical sites. Comparison panels of randomized positive and negative specimens were tested on the BD Viper LT System and compared to reference results from the BD Viper System (extracted mode). Total Positive Percent Agreement (PPA) was 98.4% (440/447) and Negative Percent Agreement (NPA) was 99.0% (1152/1164). Reproducibility was assessed across three sites over eight days, showing high concordance across specimen types.
Technological Characteristics
Uses Strand Displacement Amplification (SDA) with Bst DNA polymerase and BsoBI restriction enzyme. Reagents are provided in dried Priming and Amplification microwells. Detects GC pilin gene inverting protein homologue using a fluorescein-labeled detector probe with a Dabcyl quencher. Operates on the BD Viper LT System, a bench-top automated instrument. Qualitative NAAT assay.
Indications for Use
Indicated for the qualitative detection of Neisseria gonorrhoeae DNA in symptomatic and asymptomatic individuals to aid in the diagnosis of gonococcal urogenital disease. Applicable to males and females using clinician-collected endocervical/urethral swabs, patient-collected vaginal swabs, urine (neat/UPT), and gynecological specimens in SurePath or PreservCyt media.
Regulatory Classification
Identification
Neisseria spp. direct serological test reagents are devices that consist of antigens and antisera used in serological tests to identify Neisseria spp. from cultured isolates. Additionally, some of these reagents consist of Neisseria spp. antisera conjugated with a fluorescent dye (immunofluorescent reagents) which may be used to detect the presence of Neisseria spp. directly from clinical specimens. The identification aids in the diagnosis of disease caused by bacteria belonging to the genus Neisseria, such as epidemic cerebrospinal meningitis, meningococcal disease, and gonorrhea, and also provides epidemiological information on diseases caused by these microorganisms. The device does not include products for the detection of gonorrhea in humans by indirect methods, such as detection of antibodies or of oxidase produced by gonococcal organisms.
Predicate Devices
BD ProbeTec Neisseria gonorrhoeae (GC) Qx Amplified DNA Assay (K081825)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K140448
B. Purpose for Submission:
The purpose of this submission is to migrate the BD ProbeTec *Neisseria gonorrhoeae* (GC) Q^x Amplified DNA Assay from the previously cleared (K081825) BD Viper System operating in extracted mode to the new BD Viper LT System.
C. Measurand:
*Neisseria gonorrhoeae* DNA
D. Type of Test:
Strand displacement nucleic acid amplification (SDA) assay
E. Applicant:
Becton, Dickinson and Company
F. Proprietary and Established Names:
BD ProbeTec™ *Neisseria gonorrhoeae* (GC) Q^x Amplified DNA Assay
G. Regulatory Information:
1. Regulation section:
21 CFR 866.3390, *Neisseria* spp. direct serological test reagents
2. Classification:
Class II
3. Product code:
LSL
4. Panel:
Microbiology (83)
H. Intended Use:
1. Intended use(s):
The BD ProbeTec *Neisseria gonorrhoeae* (GC) Q^x Amplified DNA Assay, when tested with either the BD Viper™ System in Extracted Mode or the BD Viper™ LT System, uses Strand
{1}
Displacement Amplification technology for the direct, qualitative detection of Neisseria gonorrhoeae DNA in clinician-collected female endocervical and male urethral swab specimens, patient-collected vaginal swab specimens (in a clinical setting), and male and female urine specimens (both UPT and Neat). The assay is also intended for use with gynecological specimens collected in BD SurePath™ Preservative Fluid or PreservCyt™ Solution using an aliquot that is removed prior to processing for either the BD SurePath or ThinPrep™ Pap test. The assay is indicated for use with asymptomatic and symptomatic individuals to aid in the diagnosis of gonococcal urogenital disease.
2. Indication(s) for use:
Same as intended use
3. Special conditions for use statement(s):
Prescription only
4. Special instrument requirements:
The BD ProbeTec GC Q® Amplified DNA Assay uses SDA technology on either the BD Viper System or the BD Viper™ LT System.
I. Device Description:
The BD ProbeTec GC Q® Amplified DNA Assay uses SDA technology for the detection of GC DNA. The BD Viper LT System is a table-top instrument that is designed to be fully contained on a standard laboratory bench-top. The system performs automated extraction of nucleic acids from multiple specimen types in addition to amplification and detection of target nucleic acid sequences when utilized with legally marketed in vitro diagnostic assays.
J. Substantial Equivalence Information:
1. Predicate device name(s):
BD ProbeTec™ Neisseria gonorrhoeae (GC) Q® Amplified DNA Assay on the BD Viper System
2. Predicate 510(k) number(s):
K081825
3. Comparison with predicate:
| Item | Predicate Device: GCQ Assay on the BD Viper System – K081825 | Subject Device: GCQ Assay on the BD Viper LT System – K140448 |
| --- | --- | --- |
| Intended Use | The BD ProbeTec GC Q® | The BD ProbeTec Neisseria |
{2}
| Item | Predicate Device: GCQ Assay on the BD Viper System - K081825 | Subject Device: GCQ Assay on the BD Viper LT System - K140448 |
| --- | --- | --- |
| | Amplified DNA Assay, when tested with either the BD ViperTM System in Extracted Mode, uses Strand Displacement Amplification technology for the direct, qualitative detection of Neisseria gonorrhoeae DNA in clinician-collected female endocervical and male urethral swab specimens, patient-collected vaginal swab specimens (in a clinical setting), and male and female urine specimens. The assay is indicated for use with asymptomatic and symptomatic individuals to aid in the diagnosis of gonococcal urogenital disease. | gonorrhoeae (GC) QxAmplified DNA Assay, when tested with either the BD ViperTM System in Extracted Mode or the BD ViperTM LT System, uses Strand Displacement Amplification technology for the direct, qualitative detection of Neisseria gonorrhoeae DNA in clinician-collected female endocervical and male urethral swab specimens, patient-collected vaginal swab specimens (in a clinical setting), and male and female urine specimens. The assay is also intended for use with gynecological specimens collected in BD SurePathTM Preservative Fluid or PreservCytTM Solution using an aliquot that is removed prior to processing for either the BD SurePath or ThinPrepTM Pap test. The assay is indicated for use with asymptomatic and symptomatic individuals to aid in the diagnosis of gonococcal urogenital disease. |
| Assay Results | Qualitative | Qualitative |
| Technology | Strand Displacement Amplification (SDA) | Strand Displacement Amplification (SDA) |
| Specimen Types | Female specimens: Endocervical swab Patient-collected vaginal swab Neat urine UPT urine | Female specimens: Endocervical swab Patient-collected vaginal swab Neat urine UPT urine LBC specimens collected in |
{3}
| Item | Predicate Device: GCQ Assay on the BD Viper System – K081825 | Subject Device: GCQ Assay on the BD Viper LT System – K140448 |
| --- | --- | --- |
| | Male Specimens: Urethral swab Neat urine UPT urine | SurePath preservative fluid LBC specimens collected in PreservCyt Solution Male Specimens: Urethral swab Neat urine UPT urine |
| Priming Microwell | | |
| Primers | Target a sequence within the GC pilin gene inverting protein homologue | Same as predicate |
| Detector | Linear Detector Flourescein (fluorophore) Dabcyl (quencher) | Same as predicate |
| Nucleotides | 4 of 4 nucleotides required for SDA | Same as predicate |
| Non-specific reagents and cofactors | Buffering components, magnesium ions, salt, and stabilizing reagents | Same as predicate |
| Amplification Microwell | | |
| Restriction enzyme | BsoBI restriction enzyme | Same as predicate |
| Polymerase | Bst DNA polymerase | Same as predicate |
| Nucleotides | 0 of 4 nucleotides required for SDA | Same as predicate |
| Non-specific reagents and cofactors | Buffering components, magnesium ions, salt, and stabilizing reagents | Same as predicate |
# K. Standard/Guidance Document Referenced (if applicable):
1. CLSI. Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline - Second Edition. Sections 5 and 7 - Protocols for Evaluation of the Limit of Blank & Limit of Detection and Verification of Detection Capability Claims. CLSI document EP 17-A2. Wayne, PA: Clinical and Laboratory Standards Institute; 2012.
2. CLSI. Collection, Transport, Preparation, and Storage of Specimens for Molecular Methods; Approved Guideline. Sections 5, 6, and 8 - Specimen Collection, Specimen Transport and Storage, & Sample Storage. CLSI document MM13-A. Wayne, PA:
{4}
Clinical and Laboratory Standards Institute; 2005.
3. CLSI. Molecular Diagnostic Methods for Infectious Diseases; Approved Guideline – Second Edition. Sections 10 and 13 – Establishment and Evaluation of Performance Characteristics of Molecular Diagnostic Tests & Controlling False-Positive Nucleic Acid Target Amplification Reactions. CLSI document MM3-A2. Wayne, PA: Clinical and Laboratory Standards Institute; 2006.
4. FDA. Assay Migration Studies for in vitro Diagnostic Devices. Guidance for Industry and FDA Staff; 2013.
L. Test Principle:
The BD ProbeTec GC Q® Amplified DNA Assay is based on the simultaneous amplification and detection of target DNA using amplification primers and a fluorescently-labeled detector probe. The reagents for SDA are dried in two separate disposable microwells: the Priming Microwell contains the amplification primers, fluorescently-labeled detector probe, nucleotides, and other reagents necessary for amplification, while the Amplification Microwell contains the two enzymes (a DNA polymerase and a restriction endonuclease) that are required for SDA. The BD Viper LT System pipettes a portion of the purified DNA solution from each Extraction Tube into a Priming Microwell to rehydrate the contents. After a brief incubation, the reaction mixture is transferred to a corresponding, pre-warmed Amplification Microwell which is sealed to prevent contamination and then incubated in one of the two thermally-controlled fluorescent readers. The presence or absence of Neisseria gonorrhoeae DNA is determined by calculating the peak fluorescence (Maximum Relative Fluorescence Units (Max RFU)) over the course of the amplification process and by comparing this measurement to a predetermined threshold value.
In addition to the fluorescent probe used to detect amplified N. gonorrhoeae target DNA, a second fluorescently-labeled oligonucleotide is incorporated into each reaction. The Extraction Control (EC) oligonucleotide is labeled with a different dye than that used for detection of the N. gonorrhoeae-specific target and is used to confirm the validity of the extraction process. The EC is dried in the Extraction Tubes and is re-hydrated upon addition of the specimen and extraction reagents. At the end of the extraction process, the EC fluorescence is monitored by the BD Viper LT instrument and an automated algorithm is applied to both the EC and N. gonorrhoeae-specific signals to report specimen results as positive, negative, or EC failure.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
{5}
# Reproducibility
To create the panel for the reproducibility study, CT (serovar H) and GC (ATCC strain 19424) were spiked into female UPT urine, PreservCyt LBC fluid or vaginal matrix in $\mathbf{Q}^{\mathrm{x}}$ diluent at four levels: negative, high negative $(\mathrm{C}_{20} - \mathrm{C}_{80})$ , low positive (1.5x LoD), and moderate positive (3x LoD). The reproducibility panel was comprised of four levels of analyte for each of the three matrix types (4 levels x 3 matrices = 12 panel members). Each panel member was tested in duplicate on three BD Viper LT instruments over eight days with two operators testing a panel per day (2 replicates x 3 instruments x 8 days x 2 operators = 96 tests per panel member). Testing was conducted at two external sites (each had one BD Viper LT) and one internal site (one BD Viper LT and three BD Viper Instruments). For the BD Viper System, two operators tested the same number of panel members on three instruments over a period of four days. Results were tabulated by day, by operator, by replicate, across days within a site, and across sites for each sample level. Positive and negative percent agreement was calculated along with $95\%$ confidence intervals for each assay by specimen type and level over all sites and days.
Table 1: Reproducibility of the BD ProbeTec GCQ Assay on the BD Viper LT
| | | | | | Within Run | | Between Run within Day | | Between Day within Site | | Between Site | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Specimen Type | Panel | % Expected Results** | 95% CI | Mean of Max RFU | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| PreservCyt LBC | Negative | 100.0% (96/96) | (96.2 - 100.0%) | 3.3 | 9.2 | 280.1 | 0.0 | 0.0 | 0.0 | 0.0 | 2.2 | 65.4 | 9.5 | 287.6 |
| | High Negative | 20.8% (20/96) | (13.9 - 30.0%) | 560.2 | 425.0 | 75.9 | 49.0 | 8.7 | 0.0 | 0.0 | 0.0 | 0.0 | 427.8 | 76.4 |
| | Low Positive | 100.0% (96/96) | (96.2 - 100.0%) | 1415.9 | 231.4 | 16.3 | 172.0 | 12.1 | 0.0 | 0.0 | 28.1 | 2.0 | 289.7 | 20.5 |
| | Moderate Positive | 100.0%(94/94*) | (96.1 - 100.0%) | 1631.9 | 169.7 | 10.4 | 93.7 | 5.7 | 70.9 | 4.3 | 0.0 | 0.0 | 206.4 | 12.6 |
| Vaginal Swab | Negative | 99.0% (95/96) | (94.3 - 99.8%) | 41.6 | 180.1 | 432.6 | 13.2 | 31.6 | 0.0 | 0.0 | 0.0 | 0.0 | 180.6 | 433.8 |
| | High Negative | 13.5% (13/96) | (8.1 - 21.8%) | 871.5 | 562.4 | 64.5 | 0.0 | 0.0 | 0.0 | 0.0 | 88.2 | 10.1 | 569.2 | 65.3 |
| | Low Positive | 100.0%(95/95*) | (96.1 - 100.0%) | 1687.5 | 297.7 | 17.6 | 0.0 | 0.0 | 0.0 | 0.0 | 34.7 | 2.1 | 299.7 | 17.8 |
| | Moderate Positive | 100.0% (96/96) | (96.2 - 100.0%) | 1819.2 | 163.3 | 9.0 | 48.2 | 2.7 | 43.3 | 2.4 | 73.3 | 4.0 | 190.3 | 10.5 |
| Female UPT | Negative | 100.0% (96/96) | (96.2 - 100.0%) | 3.6 | 8.0 | 221.8 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 8.0 | 221.8 |
| | High Negative | 18.8% (18/96) | (12.2 - 27.7%) | 766.6 | 502.1 | 65.5 | 0.0 | 0.0 | 75.8 | 9.9 | 15.8 | 2.1 | 508.0 | 66.3 |
| | Low Positive | 100.0% (96/96) | (96.2 - 100.0%) | 1593.6 | 224.9 | 14.1 | 86.6 | 5.4 | 36.7 | 2.3 | 0.0 | 0.0 | 243.8 | 15.3 |
| | Moderate Positive | 100.0% (96/96) | (96.2 - 100.0%) | 1741.5 | 126.1 | 7.2 | 86.2 | 5.0 | 35.1 | 2.0 | 21.5 | 1.2 | 158.2 | 9.1 |
* Two invalid LBC results and one invalid swab result occurred due to extraction transfer errors.
** The results for Negative panel members calculated according to an expected result of 'negative for GC'. All other panel members calculated according to an expected result of 'positive for GC'.
{6}
Table 2: Reproducibility of the BD ProbeTec GCQ Assay on the BD Viper
| | | | | | Within Run | | Between Run within Day | | Between Day within Site | | Between Site | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Specimen Type | Panel | % Expected Results* | 95% CI | Mean of Max RFU | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| PreservCyt LBC | Negative | 100.0% (96/96) | (96.2 - 100.0%) | 0.1 | 0.5 | 438.2 | 0.3 | 245.6 | 0.0 | 0.0 | 0.1 | 128.3 | 0.5 | 518.5 |
| | High Negative | 18.8% (18/96) | (12.2 - 27.7%) | 1047.1 | 672.7 | 64.2 | 205.0 | 19.6 | 0.0 | 0.0 | 0.0 | 0.0 | 703.2 | 67.2 |
| | Low Positive | 99.0% (95/96) | (94.3 - 99.8%) | 1590.9 | 356.3 | 22.4 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 356.3 | 22.4 |
| | Moderate Positive | 100.0%(96/96) | (96.2 - 100.0%) | 1943.2 | 110.8 | 5.7 | 12.2 | 0.6 | 0.0 | 0.0 | 0.0 | 0.0 | 111.5 | 5.7 |
| Vaginal Swab | Negative | 100.0% (96/96) | (96.2 - 100.0%) | 1.1 | 4.1 | 379.5 | 0.7 | 61.5 | 0.6 | 55.3 | 0.7 | 68.3 | 4.2 | 394.4 |
| | High Negative | 35.4% (34/96) | (26.6 - 45.4%) | 594.5 | 622.7 | 104.7 | 81.1 | 13.6 | 0.0 | 0.0 | 108.7 | 18.3 | 637.4 | 107.2 |
| | Low Positive | 99.0%(95/96) | (94.3 - 99.8%) | 1589.4 | 433.1 | 27.2 | 0.0 | 0.0 | 137.5 | 8.7 | 52.9 | 3.3 | 457.5 | 28.8 |
| | Moderate Positive | 100.0% (96/96) | (96.2 - 100.0%) | 1666.6 | 337.3 | 20.2 | 76.6 | 4.6 | 0.0 | 0.0 | 58.5 | 3.5 | 350.8 | 21.1 |
| Female UPT | Negative | 100.0% (96/96) | (96.2 - 100.0%) | 0.3 | 1.5 | 468.2 | 0.0 | 0.0 | 0.0 | 0.0 | 0.4 | 133.1 | 1.6 | 486.8 |
| | High Negative | 8.3% (8/96) | (4.3 - 15.6%) | 1127.9 | 673.0 | 59.7 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 673.0 | 59.7 |
| | Low Positive | 100.0% (96/96) | (96.2 - 100.0%) | 1806.2 | 256.7 | 14.2 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 256.7 | 14.2 |
| | Moderate Positive | 100.0% (96/96) | (96.2 - 100.0%) | 1937.9 | 143.3 | 7.4 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 143.3 | 7.4 |
* The results for Negative panel members calculated according to an expected result of 'negative for GC'. All other panel members calculated according to an expected result of 'positive for GC'.
The data demonstrates good reproducibility as it shows $\geq 95\%$ detection of GC at low positive/close to the detection limit and moderate positive levels on both instruments. Detection was $\leq 21\%$ for the high negative samples on the BD Viper LT, which is acceptable and typical for SDA assays. The total percent variance for each of the specimen types across both the old and the new system are approximately equivalent.
b. Linearity/assay reportable range: N/A
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
# Post Pre-Warm Specimen Stability
The existing workflow previously cleared for the BD ProbeTec GC Q $^{\text{x}}$ Amplified DNA Assay on the BD Viper includes a pre-warm step for swab and urine samples. The workflow proposed in K140448 for the BD Viper LT includes a pre-warm step for LBC specimens in SurePath or PreservCyt media so all sample types (urine, swabs, and LBC specimens) can be tested on the same instrument run. For this reason, a post pre-warm specimen stability study was conducted to support the storage and transport stability claims for LBC specimens in SurePath or PreservCyt LBC media.
{7}
Mini-pools were created by diluting CT and NG negative LBC specimens in either SurePath or PreservCyt LBC media at a ratio of $1:3.4\mathrm{mL}$ (12 mini-pools were created for each LBC media). Each mini-pool was split equally. Half were spiked with CT serovar H and NG strain ATCC 19424 (at 90 EB/mL and 300 cells/mL, respectively) and half were not spiked (negative samples). From these split mini-pools, $2.2\mathrm{mL}$ aliquots were removed and dispensed into BD LBC Specimen Dilution Tubes (15 samples per group; 30 samples total/mini-pool). Fourteen of these tubes were pre-warmed in the BD Viper LT heat block at $114^{\circ}\mathrm{C}$ for $15\mathrm{min}$ ; one tube was read at "baseline" (no warming).
After the pre-warm step, the seven tubes were stored at the following conditions: two tubes at $2 - 8^{\circ}\mathrm{C}$ for 7 days, two tubes at $30^{\circ}\mathrm{C}$ for 7 days, and two tubes at $-20^{\circ}\mathrm{C}$ for 90 days (one tube was available for testing and one was available for retesting, if required). At each storage condition/time point, one negative and one positive tube from each of the 12 mini-pools was removed and tested in a dual dispensing mode, which generated two replicate test results (1 tube x 2 replicates x 12 mini-pools = 24 measurements/storage condition/time point for each LBC media type). Four lots of extraction reagents, 2-3 lots of amplification reagents, three lots of the control set for the BD ProbeTec CT/GC Q $^{\text{X}}$ Amplified DNA Assays, and three BD Viper LT Instruments were included in this study. Results are summarized below.
Table 3: PreservCyt Specimens Post Pre-Warm Stability Results
| Temperature | Timepoint | Positives | Negatives |
| --- | --- | --- | --- |
| N/A | Baseline | 24/24 (Pass) | 0/24 (Pass) |
| 2-8 °C | Day 3 | 24/24 (Pass) | 0/24 (Pass) |
| 30 °C | Day 3 | 24/24 (Pass) | 0/24 (Pass) |
| 2-8 °C | Day 7 | 24/24 (Pass) | 0/24 (Pass) |
| 30 °C | Day 7 | 24/24 (Pass) | 0/24 (Pass) |
| -20 °C | Day 90 | 24/24 (Pass) | 0/24 (Pass) |
| -20 °C | Day 90 | 24/24 (Pass) | 0/24 (Pass)* |
*Data generated by Out of Specification (OOS) testing after contaminated environment was deemed clean.
Table 4: SurePath Specimens Post Pre-Warm Stability Results
| Temperature | Timepoint | Positives | Negatives |
| --- | --- | --- | --- |
| N/A | Baseline | 24/24 (Pass) | 0/24 (Pass) |
| 2-8 °C | Day 3 | 24/24 (Pass) | 0/24 (Pass) |
| 30 °C | Day 3 | 24/24 (Pass) | 0/24 (Pass) |
| 2-8 °C | Day 7 | 24/24 (Pass) | 0/24 (Pass) |
| 30 °C | Day 7 | 24/24 (Pass) | 0/24 (Pass) |
{8}
| -20 °C | Day 90 | 24/24 (Pass) | 0/24 (Pass) |
| --- | --- | --- | --- |
| -20 °C | Day 90 | 24/24 (Pass) | 0/24 (Pass) |
The data indicates that, for all conditions and time points tested, the percent positive rate for pre-warmed GC positive and percent negative rate for GC negative LBC samples was 100% for samples stored in either PreservCyt or SurePath LBC media.
## Run Control Failure Rate
This analysis was conducted to determine the failure rate for positive and negative controls across four different studies. One positive and negative control result was recorded for each run for each of the following:
Clinical agreement study = 56 runs
Carry-over study (in K140448, referred to as "contamination validation study") = 63 runs
LoD/LoB validation studies = 71 runs
Reproducibility study = 51 runs
A total of 241 results were collected for each control type. There were no false negatives found in the positive control samples assessed (0/241 x 100 = 0% failure rate). There were 2 false positives found in the negative control samples assessed (2/241 x 100 = 0.83% failure rate).
## Environmental Study
The environmental study was conducted to assess the performance of the BD ProbeTec GC Q® Amplified DNA Assay on the BD Viper LT when the instrument is exposed to variations in environmental temperature and humidity. One BD Viper LT instrument was placed in an environmental chamber and equilibrated to either 1) 18°C with 20% humidity (low temperature, low humidity) or 2) 27°C with 85% humidity (high temperature, high humidity). Three lots of positive and negative controls from the control set for the BD ProbeTec CT/GC Q® Amplified DNA Assays were reconstituted with Viper QC Wash Buffer and run as samples. For the negative controls, 30 "samples" were prepared from each lot with one run per lot. An extraction control (EC) was included in each of the negative controls to show that the negative result was valid (30 negative samples/ECs x 3 lots = 90 measurements). For the positive controls, 90 "samples" were prepared from each lot with 3 runs/lot (90 positive samples/lot x 3 lots = 270 measurements).
Each rack was loaded with all positive or all negative controls (30/rack) that were logged in as "samples", and one pair of positive and negative controls that were run normally
{9}
(e.g., reconstituted, extracted, and amplified by the BD Viper LT). This study required 12 runs total with one lot of extraction and amplification reagents for testing with the assay. The results are summarized in the table below.
Table 5 - Environmental System Testing Summary
| Conditions/Parameter | GCQ Percent Correct (N) | | | Pass/Fail ≥97% |
| --- | --- | --- | --- | --- |
| | EC | Negative | Positive | |
| 1: 18°C / 20% RH | 100% (90/90) | 100% (90/90) | 100% (270/270) | Pass |
| 2: 27°C / 85% RH | 100% (90/90) | 100% (90/90) | 100% (270/270) | Pass |
There were no false negatives or false positives detected in this study, indicating that the performance of the BD ProbeTec GC Q $^x$ Assay is not affected by the temperature and humidity tested in this study.
# d. Detection limit:
# Limit of Detection (LoD)
Sample mini-pools were prepared from pre-screened CT/GC negative vaginal swabs in $\mathbf{Q}^{\mathrm{x}}$ diluent, urine in UPT, endocervical swabs in SurePath LBC diluent, and endocervical swabs in PreservCyt LBC diluent. Four mini-pools were prepared for each matrix to be tested.
Four mini-pools of matrix were used to create panels with six analyte levels per minipool and six replicates per analyte level (4 minipools x 6 levels x 6 replicates = 144 total tubes). The panels were tested across six instruments. The table below demonstrates the setup of a rack on one instrument:
Table 6 - LoD Testing Rack Layout
| | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | | | | |
| | | Level 5 | Level 4 | Level 3 | Level 2 | Level 1 | null |
| | | MP 1 | MP 1 | MP 1 | MP 1 | MP 1 | MP 1 |
| | | Level 5 | Level 4 | Level 3 | Level 2 | Level 1 | null |
| | | MP 2 | MP 2 | MP 2 | MP 2 | MP 2 | MP 2 |
| | | Level 5 | Level 4 | Level 3 | Level 2 | Level 1 | null |
| | | MP 3 | MP 3 | MP 3 | MP 3 | MP 3 | MP 3 |
| Pos Cont | Neg Cont | Level 5 | Level 4 | Level 3 | Level 2 | Level 1 | null |
| | | MP 4 | MP 4 | MP 4 | MP 4 | MP 4 | MP 4 |
(Level $=$ analyte level; $\mathrm{MP} =$ mini-pool)
Each rack was run twice on the same instrument (4 replicates x 2 runs x 3 instruments = 24 measurements).
{10}
Spiking levels for the sets consisted of the following concentrations of GC: 0, 5, 20, 50, 100, or 200 cells/mL. The LoD concentrations for each instrument were determined for two strains of GC by identifying the lowest concentration of GC cells/mL which provided a detection rate of at least 95%. The table below summarizes the LoD for the two strains of GC in each matrix type as detected by the two instruments:
Table 7: LoD comparison between BD Viper LT and BD Viper
| | | BD Viper LT | BD Viper |
| --- | --- | --- | --- |
| Matrix | GC Strain | LoD (GC cells/mL) | LoD (GC cells/mL) |
| Vaginal Matrix | 19424 | 50 | 50 |
| | 49226 | 100 | 200 |
| UPT Urine | 19424 | 50 | 50 |
| | 49226 | 50 | 50 |
| Surepath matrix in LBC Diluent | 19424 | 50 | 20 |
| | 49226 | 50 | 50 |
| PreservCyt matrix in LBC Diluent | 19424 | 50 | 20 |
| | 49226 | 50 | 20 |
The difference in LoD as determined on the BD Viper LT compared to the BD Viper System ranged from 0.5x to 2.5x GC cells/mL. The observed ratio is acceptable as 'approximately equivalent'.
# Limit of Blank (LoB)
Data from GC negative specimens (Level 0) from the analytical LoD studies described above were combined to calculate an overall LoB for the BD ProbeTec GC Q $^{\text{x}}$ Assay on the BD Viper LT. Data from all specimen matrices were compared and the worst (or highest) LoB estimate among all the specimen matrices was selected as the overall LoB (70 MaxRFU). Because this value is below the assay threshold of 125 MaxRFU, this demonstrates that the assay has an acceptable separation between background fluorescence obtained from GC negative specimens and the assay cut-off for a positive result.
# e. Analytical specificity:
{11}
12
# Carry-Over/System Contamination
The carry-over study was conducted to look for the presence of contamination in negative specimens due to carry-over of DNA during nucleic acid extraction and PCR amplification in the BD Viper LT. Each run consisted of 15 high positive and 15 negative samples arranged in alternating positive or negative rows. The CT Qx Assay was selected as the representative analyte test due to better analytical sensitivity of CT. For information on the CT Qx assay, see K140446. To create high positive samples, 1 x 105 EB/mL of CT was spiked into either one of two diluents (listed below). These diluents were selected because they represent two most challenging settings that can cause sample contamination.
1. 3 mL of Qx Swab Diluent: This volume was selected to represent a maximum fill volume, thereby maximizing the amount of fluid adhering to the outside of the pipette tip.
2. PreservCyt LBC media diluted in BD LBC Specimen Dilution tube diluent (1.7 mL LBC diluent + 0.5 mL PreservCyt = 2.2 mL; “LBC Specimen Matrix”): This LBC media was selected because it has a high alcohol content and is therefore more likely to drip from the pipette tips during sample transfer due to low surface tension.
Each run consisted of 30 samples. Seven runs were conducted per instrument with 3 BD Viper LT Systems for a total of (30 samples/run x 7 runs x 3 instruments) 630 measurements for each diluent. Runs were conducted by three operators over a period of three days. Two lots of extraction reagents, one lot of amplification reagents, and two lots of the BD CT/GC Qx Control Set were used in this study.
Table 8 – Carry Over on Viper LT
| BD Viper LT System | Qx Sample Diluent | | | LBC Specimen Matrix | | |
| --- | --- | --- | --- | --- | --- | --- |
| | n | Positive Results | Percent Positive | n | Positive Results | Percent Positive |
| 1 | 210 | 0 | 0.00% | 210 | 0 | 0.00% |
| 2 | 210 | 1 | 0.48% | 210 | 0 | 0.00% |
| 3 | 210 | 1 | 0.48% | 210 | 0 | 0.00% |
| Overall | 630 | 2 | 0.32% | 630 | 0 | 0.00% |
The overall rate of carry-over was 0.32% (2/630) for CT spiked into the Qx Swab Diluent and 0.0% (0/630) for CT spiked into PreservCyt media diluted in BD LBC Specimen Dilution tube diluent. These results are acceptable as they are comparable to the levels of carry-over cleared for the BD ProbeTec GCQ Assay on the BD Viper in K081825.
f. Assay cut-off:
Assay cut-off was established in K081825
{12}
2. Comparison studies:
a. Method comparison with predicate device:
## Clinical Comparison Study
A clinical comparison study was performed to compare results obtained with the BD ProbeTec GC Q® Amplified DNA Assay tested on the BD Viper LT system and the BD Viper system run in extracted mode. Clinical samples were collected between October 2012 and April 2013 from four geographically diverse clinical sites in North America (OB/GYN, sexually transmitted disease and family planning clinics). For each female subject, specimens were collected in the following order: (1) a first void urine specimen, (2) five patient-collected vaginal swab specimens, and (3) a SurePath liquid based cytology (LBC) specimen and a PreservCyt LBC specimen. The LBC specimen collection order was randomized throughout the study. For each male subject, a first void urine specimen was collected. Each urine specimen (male and female) was aliquoted into five UPTs (BD urine Preservative Transport for the Q® Amplified DNA Assays).
A total of 823 subjects were enrolled, 170 males and 653 females. Specimens were collected from subjects presenting with symptoms of chlamydia and/or gonorrhoeae (symptomatic) and shipped to BD on cold packs for screening, aliquoting, and panel assembly. Following exclusion of non-compliant subjects and non-evaluable specimens, a total of 617 compliant female subjects and 167 compliant male subjects were available for the clinical comparison panel assembly.
To create the clinical comparison panel, for each subject, a UPT specimen was tested on the BD Viper System operating in extracted mode (BD Viper). These results were used to confirm that the target number of CT and NG positive panel members had been met. The clinical comparison panel consisted of randomly selected positive and negative specimens with four aliquots created from each specimen type. The number of GC positive and negative specimens chosen for the clinical comparison panel was based on minimums suggested by FDA in previous correspondence (minimum of 25 GC positive and 40 negative). The positive and negative specimens were randomized within the panel and the testing sites were blinded to the results for each panel member. Panels were identical across all testing sites. Three external sites tested the panels on the BD Viper LT instrument and compared the results (positive and negative percent agreement) to the results obtained on the BD Viper. The BD Viper panel testing was conducted internally at BD.
13
{13}
Table 9 - Clinical Comparison Study Results - Female Specimens
| | | Positive Percent Agreement | | Negative Percent Agreement | |
| --- | --- | --- | --- | --- | --- |
| Specimen Type | Site | Percent | 95% CI | Percent | 95% CI |
| Vaginal Swab | A | 100.0% (27/27) | (87.5%, 100.0%) | 94.9% (75/79) | (87.7%, 98.0%) |
| | B | 96.3% (26/27) | (81.7%, 99.3%) | 96.2% (76/79) | (89.4%, 98.7%) |
| | C | 96.3% (26/27) | (81.7%, 99.3%) | 96.2% (76/79) | (89.4%, 98.7%) |
| | Total | 97.5% (79/81) | (92.6%, 100.0%) | 95.8% (227/237) | (92.0%, 98.7%) |
| UPT | A | 96.3% (26/27) | (81.7%, 99.3%) | 100.0% (79/79) | (95.4%, 100.0%) |
| | B | 100.0% (27/27) | (87.5%, 100.0%) | 100.0% (79/79) | (95.4%, 100.0%) |
| | C | 96.3% (26/27) | (81.7%, 99.3%) | 100.0% (79/79) | (95.4%, 100.0%) |
| | Total | 97.5% (79/81) | (92.6%, 100.0%) | 100.0% (237/237) | NA |
| SurePath | A | 96.4% (27/28) | (82.3%, 99.4%) | 100.0% (78/78) | (95.3%, 100.0%) |
| | B | 96.4% (27/28) | (82.3%, 99.4%) | 100.0% (78/78) | (95.3%, 100.0%) |
| | C | 96.4% (27/28) | (82.3%, 99.4%) | 98.7% (77/78) | (93.1%, 99.8%) |
| | Total | 96.4% (81/84) | (89.3%, 100.0%) | 99.6% (233/234) | (98.7%, 100.0%) |
| PreservCyt | A | 100.0% (27/27) | (87.5%, 100.0%) | 100.0% (79/79) | (95.4%, 100.0%) |
| | B | 100.0% (27/27) | (87.5%, 100.0%) | 100.0% (79/79) | (95.4%, 100.0%) |
| | C | 100.0% (27/27) | (87.5%, 100.0%) | 100.0% (79/79) | (95.4%, 100.0%) |
| | Total | 100.0% (81/81) | NA | 100.0% (237/237) | NA |
{14}
Table 10 – Clinical Comparison Study Results – Male Specimens
| | | Positive Percent Agreement | | Negative Percent Agreement | |
| --- | --- | --- | --- | --- | --- |
| Specimen Type | Site | Percent | 95% CI | Percent | 95% CI |
| UPT | A | 100.0% (40/40) | (91.2%, 100.0%) | 100.0% (73/73) | (95.0%, 100.0%) |
| | B | 100.0% (40/40) | (91.2%, 100.0%) | 100.0% (73/73) | (95.0%, 100.0%) |
| | C | 100.0% (40/40) | (91.2%, 100.0%) | 98.6% (72/73) | (92.6%, 99.8%) |
| | Total | 100.0% (120/120) | NA | 99.5% (218/219) | (98.6%, 100.0%) |
b. Matrix comparison:
N/A
3. Clinical studies:
a. Clinical Sensitivity and Specificity
N/A
4. Clinical cut-off:
N/A
5. Expected values/Reference range:
N/A
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.