CHLORADERM ANTIMICROBIAL TRANSPARENT THIN FILM DRESSING, 2 3/8 X 2 3/4, 4 X 4 3/4

K140389 · Entrotech Life Sciences, Inc. · FRO · Nov 7, 2014 · SU

Device Facts

Record IDK140389
Device NameCHLORADERM ANTIMICROBIAL TRANSPARENT THIN FILM DRESSING, 2 3/8 X 2 3/4, 4 X 4 3/4
ApplicantEntrotech Life Sciences, Inc.
Product CodeFRO · SU
Decision DateNov 7, 2014
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Intended Use

The entrotech life sciences inc. ChloraDerm™ Antimicrobial Transparent Thin Film Dressings are intended to cover and protect a wound caused by percutaneous medical devices such as drains, chest tubes, orthopedic pins, fixtures and wires. ChloraDerm™ may also be used to cover and secure primary dressings. ChloraDerm™ inhibits microbial growth within the dressing and prevents external contamination.

Device Story

ChloraDerm is a single-use, transparent, thin polyurethane film dressing coated with an acrylic-based pressure-sensitive adhesive containing chlorhexidine. It is applied to the skin surface to cover and protect percutaneous medical device sites (e.g., drains, pins) or to secure primary dressings. The device acts as a physical barrier to prevent external contamination and utilizes the incorporated chlorhexidine to inhibit microbial growth within the dressing. It is intended for use for up to 7 days. The dressing features a framed delivery system for application. It is used in clinical settings where percutaneous devices are present. The device benefits patients by providing a protective, antimicrobial environment for wound sites, potentially reducing the risk of infection associated with percutaneous access.

Clinical Evidence

Bench testing only. Performance testing included physical characteristics (Water Vapor Transmission Rate, flammability, synthetic blood/viral penetration, tear/tensile strength, adhesion) and antimicrobial efficacy (Zone of Inhibition, spectrum, in vitro kill time). Results demonstrated ChloraDerm met all specifications and performed equivalently to predicate devices. Analytical testing confirmed no detectable p-chloroaniline (PCA) content. Biocompatibility, shelf life, packaging, and sterilization validation were also performed.

Technological Characteristics

Polyurethane film dressing with acrylic-based pressure-sensitive adhesive containing chlorhexidine. Framed delivery system. Dimensions vary by part number. Sterilized via gamma radiation. Single-use, 7-day wear duration. Transparent for site visibility.

Indications for Use

Indicated for covering and protecting wounds caused by percutaneous medical devices (drains, chest tubes, orthopedic pins, fixtures, wires) and securing primary dressings in patients requiring such wound management.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 7, 2014 Entrotech Life Sciences Incorporated % Ms. Valerie Defiesta-Ng Experien Group, LLC 755 North Mathilda Avenue, Suite 100 Sunnyvale, California 94085 Re: K140389 Trade/Device Name: ChloraDerm™ Antimicrobial Transparent Thin Film Dressing Regulatory Class: Unclassified Product Code: FRO Dated: October 8, 2014 Received: October 9, 2014 Dear Ms. Defiesta-Ng: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # David Krause -S for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration #### Indications for Use 510(k) Number (if known) K140389 Device Name ChloraDerm™ Antimicrobial Transparent Thin Film Dressings Indications for Use (Describe) The entrotech life sciences inc. ChloraDerm™ Antimicrobial Transparent Thin Film Dressings are intended to cover and protect a wound caused by percutaneous medical devices such as drains, chest tubes, orthopedic pins, fixtures and wires. ChloraDerm™ may also be used to cover and secure primary dressings. ChloraDerm™ inhibits microbial growth within the dressing and prevents external contamination. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 510(k) Notification K140389 #### GENERAL INFORMATION #### Applicant: entrotech life sciences inc. 409 Illinois Street San Francisco, CA 94158 U.S.A. Phone: 510-384-0657 Fax: 800-580-0161 #### Contact Person: Valerie Defiesta-Ng Regulatory Consultant Experien Group, LLC 755 N. Mathilda Avenue, Suite 100 Sunnyvale, CA 94085 U.S.A. Phone: 408-400-0856 Fax: 408-400-0865 Date Prepared: October 7th, 2014 #### DEVICE INFORMATION Trade Name: ChloraDerm™ Antimicrobial Transparent Thin Film Dressings #### Generic/Common Name: Wound Dressing, Antimicrobial Classification: Antimicrobial Dressing, Unclassified Product Code: FRO {4}------------------------------------------------ ### PREDICATE DEVICE(S) - Covalon SurgiClear™ Antimicrobial Clear Silicone Adhesive Dressing with o Chlorhexidine and Silver ("SurgiClear") (K121819) - 3M™ Tegaderm™ Transparent Dressings ("Tegaderm") (K973036) ● # INDICATIONS FOR USE The entrotech life sciences inc. ChloraDerm™ Antimicrobial Transparent Thin Film Dressings are intended to cover and protect a wound caused by percutaneous medical devices such as drains, chest tubes, orthopedic pins, fixtures and wires. ChloraDerm may also be used to cover and secure primary dressings. ChloraDerm inhibits microbial growth within the dressing and prevents external contamination. # PRODUCT DESCRIPTION The entrotech life sciences inc. ChloraDerm™ Antimicrobial Transparent Thin Film Dressings ("ChloraDerm") are intended to cover and protect a wound caused by percutaneous medical devices such as drains, chest tubes, orthopedic pins, fixtures and wires. The ChloraDerm may also be used to cover and secure primary dressings. The ChloraDerm inhibits microbial growth within the dressing and prevents external contamination. ChloraDerm consists of a transparent, thin plastic film coated with an acrylic-based pressure sensitive adhesive containing chlorhexidine as an antimicrobial agent. The ChloraDerm dressing is a single-use device, intended to be used for up to 7 davs. ### TECHNOLOGICAL CHARACTERISTICS The technological characteristics of the ChloraDerm are substantially equivalent to the predicate devices. Table 5.1 lists the technological characteristics of the ChloraDerm and the predicate device and provides the rationale to support a determination of substantial equivalence. Any differences in the technological characteristics of the device do not affect the safety and effectiveness of the device. {5}------------------------------------------------ # 510(k) SUMMARY (CONT. | Table | |-------------| | | | | | | | | | | | | | | | | | | | | | Equivalence | | | | | | | | | | | | | | Substantial | | | | | | | | | | Table 5.1: | | | | | | | | | | | | | | | | Table 5.1: Substantial Equivalence Table | | | | | |-----------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Feature | entrotech life sciences inc.<br>ChloraTMDerm Antimicrobial<br>Transparent Thin Film<br>Dressings | Covalon SurgiClearTM<br>Antimicrobial Clear Silicone<br>Adhesive Dressing with<br>Chlorhexidine and Silver | 3MTM TegadermTM Transparent<br>Dressing | Substantial Equivalence<br>Rationale | | 510(k) Number | K140389 | K121819 | K973036 | -- | | Instruction for Use | Single Use | Single Use | Single Use | N/A (Same) | | Duration of Use | Up to 7 days | Up to 7 days | Up to 7 days | N/A (Same) | | Anatomical Site | Surface of skin | Surface of skin | Surface of skin | N/A (Same) | | Physical Composition | Polyurethane film | Polyurethane film | Polyurethane film | N/A (Same) | | Antimicrobial Agent | Chlorhexidine in the adhesive | Chlorhexidine and Silver salts<br>in the adhesive | None | Antimicrobial Efficacy Testing-<br>All acceptance criteria were met<br>and the device performed as<br>intended establishing<br>antimicrobial effectiveness.<br>Physical Performance Testing-<br>All acceptance criteria were met<br>and the device performed as<br>intended. | | Adhesive | Acrylic-based pressure sensitive<br>adhesive | Silicone-based pressure sensitive<br>adhesive | Acrylic-based pressure sensitive<br>adhesive | | | Design Features | Transparent for site visibility,<br>Framed Delivery System | Transparent for site visibility,<br>Pull away release tabs for<br>delivery | Transparent for site visibility,<br>Framed Delivery System | N/A (Same) | | Dimensions | 1624CH: 2-3/8" x 2-3/4"<br>1626CH: 4" x 4-3/4" | TWBD1017: 1.6" x 2.8"<br>TWBD1018: 1.6" x 5"<br>TWBD1019: 4" x 4.8"<br>TWBD1021: 1.6" x 1.6"<br>TWBD1023: 2.5" x 8"<br>TWBD1024: 2.6" x 10"<br>TWBD1025: 2.5" x 12" | Multiple sizes/Multiple Part<br>Numbers | The dimensional differences raise<br>no new issues of safety or<br>efficacy. | | Feature | Sterility | Target Microbes | | Substantial Equivalence<br>Rationale | | entrotech life sciences inc.<br>ChloraDerm™ Antimicrobial<br>Transparent Thin Film<br>Dressings | Gamma | Gram-positive bacteria,<br>Gram-negative bacteria and<br>yeast including:<br>• Methicillin-Resistant<br>Staphylococcus aureus<br>(MRSA)<br>• Methicillin-Resistant<br>Staphylococcus epidermidis<br>(MRSE)<br>• Vancomycin-Resistant<br>Enterococcus faecalis (VRE)<br>• Multiple Drug-Resistant<br>Enterococcus faecium (MDR)<br>• Enterococcus faecium<br>• Pseudomonas aeruginosa<br>• Escherichia coli<br>• Serratia marcescens<br>• Candida albicans<br>• Candida tropicalis<br>• Candida parapsilosis | | | | Covalon SurgiClear™<br>Antimicrobial Clear Silicone<br>Adhesive Dressing with<br>Chlorhexidine and Silver | Ethylene Oxide | Gram-positive bacteria,<br>Gram-negative bacteria and yeast<br>including:<br>• Methicillin-Resistant<br>Staphylococcus aureus (MRSA)<br>• Methicillin-Resistant<br>Staphylococcus epidermidis<br>• Vancomycin-Resistant<br>Enterococcus faecalis (VRE)<br>• Klebsiella pneumoniae<br>• Pseudomonas aeruginosa<br>• Enterobacter cloacae<br>• Candida albicans<br>• Candida tropicalis | | Sterilization validation<br>performed, ensures a SAL of 10-6<br>as per device specifications. | | 3M™ Tegaderm™ Transparent<br>Dressing | Radiation | | | Antimicrobial Efficacy Testing-<br>All acceptance criteria were met<br>and the device performed as<br>intended establishing<br>antimicrobial effectiveness. | | | | | Not specified on IFU | | {6}------------------------------------------------ # 510(k) SUMMARY (CONT. {7}------------------------------------------------ ### SUBSTANTIAL EQUIVALENCE The indications for use of the ChloraDerm is identical to that of the Covalon SurgiClear predicate device. All three devices are intended to cover and protect wounds caused by percutaneous medical devices and are also to be used as a secondary dressing to cover and secure a primary dressing. The ChloraDerm is similar to the Tegaderm in its physical performance and to the SurgiClear in its antimicrobial efficacy. Available performance data support the determination of substantial equivalence in terms of both device physical performance and antimicrobial efficacy and confirmed that any differences in the technological characteristics between the ChloraDerm and the predicate devices do not raise any new issues of safety or effectiveness. Thus, the ChloraDerm is substantially equivalent to the predicate devices. # TESTING IN SUPPORT OF SUBSTANTIAL EQUIVALENCE DETERMINATION All necessary performance testing including head comparison testing with the predicate devices was conducted on the ChloraDerm to support a determination of substantial equivalence to the predicate devices. The following table lists the non-clinical testing performed and the results for each test. | Test Type | Test Description | Results | |---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Physical<br>Performance | Water Vapor Transmission Rate<br>Flammability<br>Synthetic Blood Penetration<br>Viral Penetration<br>Tear Resistance<br>Tensile Elongation<br>Recovery/Elongation to Break<br>Primary Release<br>Peel Adhesion/Shear Adhesion | The ChloraDerm passed all functional testing and<br>met all product specification requirements in<br>addition to demonstrating equivalent performance to<br>that of the predicate device. | | Antimicrobial<br>Efficacy | Zone of Inhibition/Activity<br>Spectrum<br>In vitro Kill Time | The ChloraDerm demonstrated that it had adequate<br>antimicrobial activity that was equivalent to that of<br>the predicate device. | | Analytical | p-Chloroaniline (PCA) Content | The test results from zero-time and accelerated aged<br>samples demonstrated that no detectable levels of<br>p-chloroaniline were observed in the ChloraDerm.<br>However, p-chloroaniline was detected in all of the<br>Covalon SurgiClear samples evaluated | In addition to the above performance testing, entrotech conducted Biocompatibility, Shelf life, Packaging and Sterilization Validation testing on the ChloraDerm. The collective results of the performance testing demonstrate that the materials chosen, manufacturing processes, and design of the ChloraDerm meet the established specifications necessary for consistent performance during its intended use. In addition, the collective bench testing demonstrates that the ChloraDerm does not raise new questions of safety or effectiveness when compared to the predicate devices. {8}------------------------------------------------ # CONCLUSION The entrotech life sciences inc. ChloraDerm is substantially equivalent to the Covalon SurgiClear and 3M Tegaderm dressings. The indications for use of the ChloraDerm is identical to that of the Covalon SurgiClear predicate device. The product performance testing demonstrated that ChloraDerm is as safe, as effective and performs as well as the predicate devices in terms of intended use, safety and technological characteristics, and patient populations. The key difference between the devices is the presence and/or choice of antimicrobial agent that is incorporated into the adhesive. The difference in antimicrobial agent between the subject device and the SurgiClear predicate device raises no issue of safety and effectiveness. Additionally, the difference in physical performance between the subject device and the 3M Tegaderm product raises no issues of safety and effectiveness. The information contained in this 510(k) premarket notification demonstrates the substantial equivalence of the ChloraDerm to the SurgiClear and to the Tegaderm.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%