← Product Code [MQL](/productcode/MQL) · K140317

# EMBRYOGEN (K140317)

_Origio A/S · MQL · Jun 16, 2014 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K140317

## Device Facts

- **Applicant:** Origio A/S
- **Product Code:** [MQL](/productcode/MQL.md)
- **Decision Date:** Jun 16, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.6180
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

EmbryoGen® is for culture of human embryos until the 2-8 cell stage. EmbryoGen® can also be used for embryo transfer at day 2 or 3.

## Device Story

EmbryoGen® (1204) is a ready-to-use, aseptically filtered, non-viscous culture medium for human embryos. It contains recombinant human GM-CSF (2 ng/mL) to support embryo development. The device is used by professionals in assisted reproduction clinics. It serves as the environment for embryo culture until the 2-8 cell stage and facilitates embryo transfer to the uterus. The medium provides essential nutrients, including amino acids, physiological salts, and bicarbonate, while maintaining pH and osmolality. By providing a controlled environment, it supports embryo viability during the critical early stages of development prior to implantation. The device is supplied in 3 ml glass bottles.

## Clinical Evidence

Bench testing only. Performance data includes sterility, endotoxin levels, osmolality, pH, and GM-CSF potency (TF-1 cell assay). Mouse Embryo Assay (MEA) results were ≥80%. Biocompatibility testing per ISO 10993-1:2009 (cytotoxicity, sensitization, irritation) confirmed materials are biocompatible. Stability studies validated a 26-week shelf life.

## Technological Characteristics

Aseptically filtered, non-viscous aqueous solution. Contains physiological salts, bicarbonate, glucose, lactate, pyruvate, 21 amino acids, folic acid, calcium pantothenate, sodium hyaluronate (0.05 g/l), 5 mg/mL HSA, gentamicin sulphate (0.01 mg/mL), and 2 ng/mL recombinant human GM-CSF. pH 7.2-7.4; osmolality 272-288 mOsm/kg; endotoxin <0.15 EU/mL. Packaged in 3 ml glass bottles with polypropylene caps.

## Regulatory Identification

Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).

## Special Controls

*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.

## Predicate Devices

- EmbryoGen® (Cat. No. 1203) ([K120136](/device/K120136.md))

## Reference Devices

- EmbryoAssist® ([K080473](/device/K080473.md))
- ORIGIO Sequential Cleav™ ([K133912](/device/K133912.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows a stylized logo or wordmark that reads "origio". The letters are designed with a somewhat retro or vintage aesthetic, featuring thick, rounded strokes. The "g" is uniquely styled with a comma-like shape below it, and the "i" is formed by two vertically stacked circles, giving the wordmark a distinctive and memorable appearance.

JUN 1 6 2014

K140317- EmbryoGen®

### 510(K) SUMMARY

### Summary of Safety and Effectiveness Information Supporting a Substantially Equivalent Determination

| Submitted by:         | ORIGIO a/s<br>Knardrupvej 2,<br>2760 Måløv,<br>Denmark                    |
|-----------------------|---------------------------------------------------------------------------|
| Contact person:       | Tove Kjær<br>Director Corporate Regulatory Affairs<br>ORIGIO a/s          |
|                       | Phone: +45 4679 0220<br>Fax: +45 4679 0300                                |
|                       | Date Submitted: June 04, 2014                                             |
| Device Identification |                                                                           |
| Trade name:           | EmbryoGen® (Cat. No. 1204)                                                |
| Common name:          | EmbryoGen® (Cat. No. 1204)                                                |
| Classification name:  | Reproductive media and supplements (21 CFR 884.6180, Product<br>Code MQL) |

### Predicate devices:

EmbryoGen® (Cat. No. 1203) (K120136)

### Description

EmbryoGen® (1204) is a modification of EmbryoGen® (1203) (K120136), where the base medium has been changed from EmbryoAssist® (K080473) to the new ORIGIO Sequential Cleav™ (K133912) medium. EmbryoGen® (1204) contains GM-CSF (Granulocyte macrophage- colony stimulating factor), in the same concentration as EmbryoGen® (1203).

EmbryoGen® (1204) is an aseptically filtered, colorless, non viscous solution, which is ready to use by professionals within assisted reproduction. EmbryoGen® (1204) is contained in 3 ml transparent glass bottles with white polypropylene caps, available in a single piece card board box, individually labeled and with instruction for use provided as a package insert. )

### Indication for use

EmbryoGen® is for culture of human embryos until the 2-8 cell stage. EmbryoGen® can also be used for embryo transfer at day 2 or 3.

{1}------------------------------------------------

The indication for use for EmbryoGen® (1204) has been narrowed and does not include fertilization as EmbryoGen (1203). The indication for EmbryoGen® (1204) is considered comparable to the predicate since both media are for culture until 2-8 cell stage including human embryo transfer. Thus the reduction in indication does not represent a new intended use nor does it pose any safety or effectiveness issues.

### Technological Characteristics

Table 1 compares the technological characteristics of EmbryoGen® (1204) to the predicate EmbryoGen® (1203). Both similarities and differences are illustrated.

| Product                           | EmbryoGen® (1204)                                                                                                                      | EmbryoGen® (1203)                                                                                                                   |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Indication for use                | EmbryoGen® is for culture of human embryos until the 2-8 cell stage.<br>EmbryoGen® can also be used for embryo transfer at day 2 or 3. | EmbryoGen® is for fertilization and culture until 2-8 cell stage.<br>EmbryoGen® can also be used for embryo transfer at day 2 or 3. |
| Product specification             |                                                                                                                                        |                                                                                                                                     |
| pH                                | 7.2-7.4                                                                                                                                | 7.3-7.5                                                                                                                             |
| Osmolality (mOsm/kg)              | 272-288                                                                                                                                | 272-288                                                                                                                             |
| Endotoxin (EU/mL)                 | <0.15                                                                                                                                  | <0.15                                                                                                                               |
| Aseptically filtered              | X                                                                                                                                      | X                                                                                                                                   |
| 1-cell MEA                        | ≥80%                                                                                                                                   | ≥80%                                                                                                                                |
| GM-CSF ELISA test                 | >80%                                                                                                                                   | >80%                                                                                                                                |
| GM-CFS potency (TF-1 cell assay)  | 80-125%                                                                                                                                | 80-125%                                                                                                                             |
| Base Medium                       |                                                                                                                                        |                                                                                                                                     |
| Physiological salts               | X                                                                                                                                      | X                                                                                                                                   |
| Bicarbonate                       | X                                                                                                                                      | X                                                                                                                                   |
| HEPES                             | -                                                                                                                                      | X                                                                                                                                   |
| Glucose, lactate, pyruvate        | X                                                                                                                                      | X                                                                                                                                   |
| Amino Acids                       | 21 amino acids                                                                                                                         | 11 amino acids                                                                                                                      |
| EDTA                              | X                                                                                                                                      | X                                                                                                                                   |
| Vitamins                          | Ca-pantothenate, Folic acid                                                                                                            | Mixture of 10 vitamins                                                                                                              |
| Sodium Hyaluronate                | X                                                                                                                                      | -                                                                                                                                   |
| Synthetic Serum Replacement (SSR) | -                                                                                                                                      | X                                                                                                                                   |
| Protein source                    |                                                                                                                                        |                                                                                                                                     |
| HSA Protein Supplementation       | 5 mg/mL                                                                                                                                | 2 mg/mL                                                                                                                             |
| Drugs                             |                                                                                                                                        |                                                                                                                                     |
| Gentamicin sulphate               | 0.01 mg/mL                                                                                                                             | 0.01 mg/mL                                                                                                                          |
| Insulin¹                          | -                                                                                                                                      | 0.51 µg/mL                                                                                                                          |
| Product                           | EmbryoGen® (1204)                                                                                                                      | EmbryoGen® (1203)                                                                                                                   |
| Recombinant human GM-<br>CSF      | 2 ng/mL                                                                                                                                | 2 ng/mL                                                                                                                             |

Table 1. Comparison of EmbryoGen® (1204) and the predicate EmbryoGen® (1203).

{2}------------------------------------------------

Insulin is a component in SSR

The technological characteristics of EmbryoGen® (1204) are comparable to those of the predicate device. The differences are all in the base medium which is identical to the cleared ORIGIO® Sequential Cleav™ (K1333912) with the same intended use as EmbryoGen (1204). The main differences between EmbryoGen (1204) and EmbryoGen (1203) are:

- Synthetic Serum Replacement (SSR %: SSR has been removed in EmbryoGen (1204), . SSR is a metal ion buffer primarily containing a balanced mixture of trace elements and human recombinant insulin . The majority of the commercially available ART media do not contain SSR or similar components. Moreover, as EmbryoGen (1204) is identical in formulation to the cleared ORIGIO Sequential Cleav (K1333912) which also does not contain SSR, it does not represent a new technology.
- EDTA: The concentration of EDTA is decreased compared to EmbryoGen® (1203) but is . identical to the EDTA concentration in ORIGIO® Sequential Cleav™(K1333912). A decreased EDTA concentration in EmbryoGen® (1204) therefore do not represent a new technology
- Human Serum Albumin (HSA): EmbryoGen® (1204) contains 5 mg/mL HSA compared to 2 . mg/mL HSA in EmbryoGen® (1203). The majority of commercially available culture media with the same intended use including ORIGIO® Sequential Cleav™ (K1333912) contain 5 mg/mL HSA and thus, do not represent a new technology.
- Amino Acids: EmbryoGen® (1204) contains 21 amino acids (both non-essential and . essential amino acids), whereas EmbryoGen® (1203) contains 11 non-essential amino acids only. EmbryoGen® (1204) are equivalent to the cleared ORIGIO® Sequential Cleav™(K1333912) in regards of the different amino acids as well as concentration and thus, do not represent a new technology.
- Vitamins: EmbryoGen® (1204) contains the vitamins folic acid and calcium pantothenate in similar concentration range as predicate and other cleared ART media (ORIGIO® Sequential Cleav™ (K1333912)). Thus, folic acid and calcium pantothenate have a history of use in other ART media and do not represent a new technology
- Sodium hyaluronate: EmbryoGen® (1204) contains sodium hyaluronate whereas . EmbryoGen 91203) does not. The concentration of sodium hyaluronate is 0.05 g/l which is in the range found in other commercial IVF media with the same intended use( ORIGIO® Sequential Cleav™ (K1333912)).

The technological characteristics of EmbryoGen® (1204) are comparable to the predicate device. do not impact substantial equivalence, and do not raise new types of safety or effectiveness questions.

### Performance data

The product specifications for EmbryoGen® (1204) and EmbryoGen® (1203) are similar regarding sterility, endotoxin, osmolality, GM-CSF ELISA test, GM-CSF potency (TF-1 cell assay) and Mouse Embryo Assay (MEA) test.

Regarding the pH, the specification limit is a little lower for EmbryoGen® (1204) than EmbryGen® (1203). However, the specification for EmbryoGen® (1204) is identical to the pH specification for ORIGIO® Sequential Cleav™ (K1333912) with the identical indication for use for culture of human

{3}------------------------------------------------

embryos until 2-8 cell stage and embryo transfer. Thus, the pH specification for EmbryoGen® (1204) does not raise any safety concerns.

The shelf life of EmbryoGen® has been validated in stability studies to 26 weeks. The parameters which have been tested in the stability studies through shelf life includes pH, osmolality, endotoxin, · HSA concentration, MEA, GM-CSF ELISA, GM-CSF potency (TF-1 Cell assay) and sterility.

In general, EmbryoGen® (1204) medium is subject to the same control methods and, to a significant degree, contain the same components as the predicate device EmbryoGen® (1203). EmbryoGen® (1204) has similar handling procedures and storage conditions. Therefore, EmbryoGen® (1204) is considered substantially equivalent to the predicate device EmbryoGen® (1203) (K133912).

#### Biocompatibility

EmbryoGen® (1204) is categorized as a medium in direct contact with embryos until 2-8 cell stage. Since EmbryoGen® (1204) can also be used for human embryo transfer, it is also in direct contact with the uterus (patient). Biocompatibility testing was conducted on EmbryoGen® per ISO 10993-1:2009 and included assessments of cytotoxicity, sensitization, and irritation. Results of these studies demonstrated that EmbryoGen® materials were biocompatible. EmbryoGen® (1204) is therefore considered safe for culture of human embryos as well as transfer of embryos to the patient (uterus).

#### Conclusion

The conclusion from the performance and safety data, intended use comparison, product formulation comparison and test specification comparison, demonstrates that EmbryoGen® (1204) is suitable for its intended use, and that the technological characteristics are substantially equivalent to the predicate device EmbryoGen® (1203) in which substantial equivalence has been demonstrated.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it. The symbol is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The text is written in all capital letters.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 16, 2014

Origio A/S Tove Kjær Director Corporate Regulatory Affairs Knardrupvej 2 2760 Måløv Denmark

Re: K140317

> Trade/Device Name: EmbryoGen® Regulation Number: 21 CFR§ 884.6180 Regulation Name: Reproductive media and supplements Regulatory Class: II Product Code: MQL Dated: March 17, 2014 Received: March 20, 2014

Dear Tove Kjær,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{5}------------------------------------------------

Page 2 - Tove Kjær

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Image /page/5/Picture/6 description: The image shows the name "Herbert P. Lerner -S" in a bold, sans-serif font. The name is written in black ink on a white background. The letters are evenly spaced and the name is easy to read. The letters "P. L" are surrounded by a faint, dotted circle.

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known) K140317

Device Name EmbryoGen®

Indications for Use (Describe)

EmbryoGen® is for culture of human embryos until the 2-8 cell stage. EmbryoGen® can also be used for embryo transfer at day 2 or 3.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY : Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Herbert P. Lerner 2014.06.16 14:5

FORM FDA 3881 (9/13)

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

---

**Source:** [https://fda.innolitics.com/device/K140317](https://fda.innolitics.com/device/K140317)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
