ELECTRODE FOR URINARY INCONTINENCE

K140265 · Guangzhou Finecure Medical Equipment Co.,Ltd · KPI · Jan 23, 2015 · Gastroenterology, Urology

Device Facts

Record IDK140265
Device NameELECTRODE FOR URINARY INCONTINENCE
ApplicantGuangzhou Finecure Medical Equipment Co.,Ltd
Product CodeKPI · Gastroenterology, Urology
Decision DateJan 23, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5320
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Urinary Incontinence Probe, model PR-51 Vaginal Probe & PR-52 Anal Probe are intended to provide electromyographic feedback from pelvic musculature or electrical stimulation to pelvic musculature for the purpose of rehabilitation of weak pelvic floor muscles and restoration of neuromuscular control during the treatment of urinary incontinence.

Device Story

Urinary Incontinence Probe (models PR-51 Vaginal, PR-52 Anal) consists of lightweight cylinder with two independent conductive rings; used for pelvic floor muscle rehabilitation. Device provides electromyographic feedback or electrical stimulation to pelvic musculature. Operated in home or clinic settings by single user for up to one year. Probe connects to external stimulation/feedback devices via lead wires (2mm pigtail, 3.5mm female, or 2.35mm female terminals). Device is watertight for cleaning with soap and water; does not require sterilization. Output facilitates muscle strengthening and neuromuscular control, aiding treatment of urinary incontinence.

Clinical Evidence

No clinical data. Bench testing only, including biocompatibility testing per ISO 10993-1, ISO 10993-5, and ISO 10993-10.

Technological Characteristics

Cylindrical probe with two independent conductive rings; watertight design. Connectivity via 2mm pigtail, 3.5mm female, or 2.35mm female terminals. Materials evaluated per ISO 10993-1, 10993-5, and 10993-10. Non-sterile, reusable by single user. Passive electrode component; no internal software or active circuitry.

Indications for Use

Indicated for patients with urinary incontinence requiring pelvic floor muscle rehabilitation or neuromuscular control restoration via electromyographic feedback or electrical stimulation.

Regulatory Classification

Identification

A nonimplanted electrical continence device is a device that consists of a pair of electrodes on a plug or a pessary that are connected by an electrical cable to a battery-powered pulse source. The plug or pessary is inserted into the rectum or into the vagina and used to stimulate the muscles of the pelvic floor to maintain urinary or fecal continence. When necessary, the plug or pessary may be removed by the user. This device excludes an AC-powered nonimplanted electrical continence device and the powered vaginal muscle stimulator for therapeutic use (§ 884.5940).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three heads facing to the right. The eagle is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 23, 2015 Guangzhou Finecure Medical Equipment Co., Ltd % Field Fu Consultant Shenzhen Joyantech Consulting Co., Ltd. 4th Floor, Jinhui Building, Nanhai Blvd, Nanshan District, Shenzhen, 518000 China Re: K140265 Trade/Device Name: Electrode for Urinary Incontinence Regulation Number: 21 CFR 876.5320 Regulation Name: Non-Implanted electrical continence device Regulatory Class: II Product Code: KPI, HIR Dated: December 10, 2014 Received: December 14, 2014 Dear Field Fu, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA mav publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR {1}------------------------------------------------ Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Benjamin R. Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K140265 #### Device Name Electrode for Urinary Incontinence #### Indications for Use (Describe) The Urinary Incontinence Probe, model PR-51 Vaginal Probe are intended to provide electromyographic feedback from pelvic musculature or electrical stimulation for the purpose of rehabilitation of weak pelvic floor muscles and restoration of neuromuscular control during the treatment of urinary incontinence. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) {3}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Product: Electrode for Urinary Incontinence # SECTION 05 510(k) Summary This summary of 510(K) safety and effectiveness information is submitted As Required by requirements of SMDA and 21 CFR §807.92. ## 1 Administrative Information | Date of Summary<br>prepared | Jan, 19, 2014 | |--------------------------------------|---------------------------------------------------------------------| | Manufacturer<br>information | Company title:<br>Guangzhou Finecure Medical Equipment Co.,<br>Ltd. | | | Company address: | | | F19, No.1 Kesheng Road, Baiyun Torch | | | Building, No.1633 Beitai Road, Baiyun District | | | Guangzhou, CHINA | | | Phone: +86(020) 2802 6079 | | | Fax: +86(020) 2802 6059 | | | Contact Person: Chiping Ma | | | E-mail: Finecure_export@163.com | | Submission<br>Correspondent | Shenzhen Joyantech Consulting Co., Ltd. | | | 4th Floor, Jinhui Building, Nanhai BLVD, | | | Nanshan District, Shenzhen, Guangdong, | | | China. | | | Contact person: Mr. Field.Fu | | Image: Joyantech logo | E-Mail: cefda13485@163.com | | | QQ: 670312758 | | | Website: www.cefda.com | | Establishment registration<br>number | No. | ## 2 Device Information | Type of 510(k) submission: | Traditional | |----------------------------|------------------------------------| | Trade Name: | Electrode for Urinary Incontinence | | Model: | PR-51, PR-52 | {5}------------------------------------------------ Guangzhou Finecure Medical Equipment Co., LTD SEC_005:001_510K Summary Version:A/0 Product: Electrode for Urinary Incontinence | Classification name: | Stimulator, electrical, non-implantable, for<br>incontinence;<br>Perineometer. | |----------------------|--------------------------------------------------------------------------------| | Review Panel: | Gastroenterology/Urology;<br>Obstetrics/Gynecology. | | Product Code: | KPI; HIR | | Device Class: | II | | Regulation Number: | 876.5320;884.1425 | ### 3 Predicate Device Information | Sponsor: | VERYWAY MEDICAL INSTRUMENT CO., LTD | |----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device: | Everyway Incontinence Stimulation Electrode<br>Model: PR-02/02A, PR-03103A, PR-04104A, PR-10A,<br>PR-h1A, PR-14A for Life-Care VaginalProbe & PR-<br>06/06A, PR-1 2A, PR-1 3/13A for Life-Care Anal Probe. | | 510(K) Number: | K122194 | #### 4 Device Description The Urinary Incontinence Probe, model PR-51 Vaginal Probe & PR-52 Anal Probe are the light weight cylinder consisting of two independent conductive rings that are paired and isolated, physically and electrically. The cylinder is shaped with a handle for comfort positioning in vaginal/anal canal for incontinent treatment as above mentioned and easy for removing after treatment. It is watertight to allow for washing with soap and water between uses. The probe is designed for repeated intermittent use in home or clinic for up to one year by a single user. It does not require sterilization, but does required washing for reuse according to the validated cleaning method as recommended in user manual. To fit with different type of device connection, the lead wire of device was provided with three different type of wire connection terminals, 2mm pigtail, 3.5mm femelle and 2.35 mm femelle. Either of the PR-51 Vaginal Probe & PR-52 Anal Probe can be connected to any one of three connection terminals. ### 5 Intended Use The Urinary Incontinence Probe, model PR-51 Vaginal Probe & PR-52 Anal Probe are intended to provide electromyographic feedback from pelvic musculature or electrical stimulation to pelvic musculature for the purpose of rehabilitation of weak pelvic floor muscles and restoration of neuromuscular control during the treatment of urinary incontinence. {6}------------------------------------------------ Version:A/0 Product: Electrode for Urinary Incontinence ## 6 Indication for Use The Urinary Incontinence Probe. model PR-51 Vaginal Probe & PR-52 Anal Probe are intended to provide electromyographic feedback from pelvic musculature or electrical stimulation to pelvic musculature for the purpose of rehabilitation of weak pelvic floor muscles and restoration of neuromuscular control during the treatment of urinary incontinence. # 7 Technological characteristics of the proposed device compared to the predicate device The proposed device and the Predicate device have the same intended use, design principle, and similar material composition. The differences do not exert adverse effect on the proposed device. The proposed device is substantially equivalent to the predicate devices. ## 8 Brief discussion of the nonclinical tests Electrode for Urinary Incontinence conforms to the following standards: - & ISO 10993-1:2009 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process; - ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for � In Vitro cytotoxicity; - ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests � for irritation and skin sensitization. # 9 Brief discussion of clinical tests Not applicable. # 10 Other information (such as required by FDA guidance) No other information. # 11 Conclusions The subject device---Electrode for Urinary Incontinence PR-51, PR-52 are substantially equivalent to Everyway Incontinence Stimulation Electrode whose 510(k) number is K122194.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...