ELIA PR3S IMMUNOASSAY, ELIA MPOS IMMUNOASSAY, ELIA GBM IMMUNOASSAY, ELIA ANCA/GBM POSITIVE CONTROL 100, ELIA ANCA/GBM PO
K140225 · Phadia US, Inc. · MOB · Oct 27, 2014 · Immunology
Device Facts
Record ID
K140225
Device Name
ELIA PR3S IMMUNOASSAY, ELIA MPOS IMMUNOASSAY, ELIA GBM IMMUNOASSAY, ELIA ANCA/GBM POSITIVE CONTROL 100, ELIA ANCA/GBM PO
Applicant
Phadia US, Inc.
Product Code
MOB · Immunology
Decision Date
Oct 27, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5660
Device Class
Class 2
Indications for Use
EliA PR3s is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to proteinase 3 (PR3) in human serum and plasma (heparin, EDTA, citrate) to aid in the clinical diagnosis of Granulomatosis with Polyangiitis (GPA; formerly known as Wegener's granulomatosis) in conjunction with other laboratory and clinical findings. EliA PR3 uses the EliA IgG method on the instruments Phadia 100 and Phadia 250. EliA MPO is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to myeloperoxidase (MPO) in human serum and plasma (heparin, EDTA, citrate) as an aid in the clinical diagnosis of microscopic polyangiitis (MPA) in conjunction with other laboratory and clinical findings. EliA MPO uses the EliA IgG method on the instruments Phadia 100 and Phadia 250. EliA GBM is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to a.3 chain of collagen IV in human serum and plasma (heparin, EDTA, citrate) as an aid in the clinical diagnosis of Goodpasture syndrome in conjunction with other laboratory and clinical findings. EliA GBM uses the EliA IgG method on the instruments Phadia 100 and Phadia 250. EliA ANCA/GBM Positive Control 100 is intended for laboratory use in monitoring the performance of in vitro measurement of ANCA/GBM antibodies with Phadia 100 using the EliA IgG method. EliA ANCA/GBM Positive Control 250 is intended for laboratory use in monitoring the performance of in vitro measurement of ANCA/GBM antibodies with Phadia 250 using the EliA IgG method.
Device Story
EliA immunoassays utilize antigen-coated wells to detect specific IgG antibodies in human serum/plasma samples. Input: patient specimen; process: non-competitive solid-phase ELISA; patient antibodies bind to coated antigens (PR3, MPO, or collagen IV alpha3 chain); enzyme-labeled anti-human IgG conjugate added; fluorescence measured after development solution reaction. Output: semi-quantitative IgG concentration. Used in clinical laboratories on Phadia 100/250 automated analyzers. Results interpreted by clinicians alongside other findings to aid diagnosis of autoimmune vasculitis or Goodpasture syndrome. Benefits: standardized, automated diagnostic support for specific autoimmune conditions.
Clinical Evidence
Clinical validation included 455 samples for PR3, 425 for MPO, and 460 for GBM. Sensitivity/specificity for GPA (PR3): 83.0%/95.8%. Sensitivity/specificity for MPA (MPO): 58.8%/92.2%. Sensitivity/specificity for Goodpasture syndrome (GBM): 98.6%/99.5%. Method comparison against predicate ELISAs showed high agreement. Analytical performance included precision (CVs generally <15%), linearity, and lack of interference from bilirubin, hemoglobin, lipids, or rheumatoid factor.
Technological Characteristics
Polystyrene wells coated with human PR3, MPO, or recombinant alpha3 chain of collagen IV. Detection via beta-galactosidase labeled mouse monoclonal anti-IgG conjugate and MUG substrate. Fluorescence signal measured on Phadia 100/250 instruments. Automated, 37°C controlled reaction. Calibrated against WHO-standardized IgG calibrators. Sample matrix: serum, Li-heparin, EDTA, or citrate plasma.
Indications for Use
Indicated for semi-quantitative measurement of IgG antibodies to PR3, MPO, or alpha3 chain of collagen IV in human serum/plasma to aid diagnosis of Granulomatosis with Polyangiitis, microscopic polyangiitis, or Goodpasture syndrome, respectively. For prescription use only.
Regulatory Classification
Identification
A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.