← Product Code [OQG](/productcode/OQG) · K140130

# X-ALT HIGHLY CROSS LINKED ACETABULAR LINERS WITH VITAMIN E, NEUTRAL, 10 DEGREE HOODED, 20 DEGREE HOODED (K140130)

_Encore Medical L.P. · OQG · Nov 6, 2014 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K140130

## Device Facts

- **Applicant:** Encore Medical L.P.
- **Product Code:** [OQG](/productcode/OQG.md)
- **Decision Date:** Nov 6, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3358
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Joint replacement is indicated for patients suffering from disability due to: - noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis of the natural femoral head; - rheumatoid arthritis; - correction of functional deformity; - femoral fracture. This device may also be indicated in the salvage of previously failed surgical attempts. This device is intended for cementless use only

## Device Story

Acetabular liner component for hip joint replacement; manufactured from highly cross-linked polyethylene infused with pharmaceutical-grade alpha-tocopherol (Vitamin E). Device functions as bearing surface in hip arthroplasty; intended for cementless fixation. Used by orthopedic surgeons in clinical/surgical settings. Provides stable articulation for femoral head; Vitamin E infusion aims to improve oxidative stability and wear resistance of polyethylene. Output is physical structural support for hip joint; assists in restoring joint function and mobility. Device is an implantable orthopedic component.

## Clinical Evidence

No clinical data provided. Substantial equivalence supported by non-clinical mechanical testing demonstrating that the 32 mm liner configuration does not create a new worst-case for mechanical performance.

## Technological Characteristics

Highly cross-linked polyethylene infused with alpha-tocopherol (Vitamin E). Acetabular liner form factor; available in 32 mm ID with neutral, 10°, and 20° hooded configurations. Designed for cementless use. Sterilization and packaging methods are identical to predicate devices.

## Regulatory Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

## Predicate Devices

- DJO Surgical FMP Acetabular Shells with Standard Poly Liner ([K973119](/device/K973119.md), [K974093](/device/K974093.md), [K974095](/device/K974095.md))
- DJO Surgical X-alt Highly Cross Linked Acetabular Liners with Vitamin E ([K130365](/device/K130365.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of a human figure, with three profiles overlapping to create a sense of depth and connection.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 6. 2014

Encore Medical, L.P. Ms. Teffany Hutto Manager, Regulatory Affairs 9800 Metric Boulevard Austin, Texas 78758

Re: K140130

Trade/Device Name: X-alt™ Highly Cross Linked Acetabular Liner with Vitamin E Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: OQG, LPH Dated: September 12, 204 Received: September 15, 2014

Dear Ms. Hutto:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

Page 2 – Ms. Teffany Hutto

the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Lori A. Wiggins -S

for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known): _ K140130

Device Name: X-alt™ Highly Cross Linked Acetabular Liner with Vitamin E

Indications for Use:

### X-alt™ Highly Cross Linked Acetabular Liner with Vitamin E Indications for Use

Joint replacement is indicated for patients suffering from disability due to:

- noninflammatory degenerative joint disease including osteoarthritis and avascular ● necrosis of the natural femoral head;
- rheumatoid arthritis; ●
- correction of functional deformity; ●
- femoral fracture. ●

This device may also be indicated in the salvage of previously failed surgical attempts.

This device is intended for cementless use only

Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{3}------------------------------------------------

## 510(k) Summary

Date: November 4, 2014

Manufacturer: Encore Medical, L.P. 9800 Metric Blvd Austin, TX 78758

Contact Person: Teffany Hutto Manager, Regulatory Affairs Phone: (512) 834-6255 Fax: (512) 834-6313 Email: teffany.hutto@djoglobal.com

| Product                                                       | Classification | Product Code                                                                                                                                                                                                                                    |
|---------------------------------------------------------------|----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| X-alt™ Highly Cross Linked<br>Acetabular Liner with Vitamin E | Class II       | LPH - Hip joint metal/polymer/metal semi-constrained<br>porous-coated uncemented prosthesis per 21 CFR<br>888.3358<br>OQG - Hip prosthesis, semi-constrained, cemented,<br>metal/polymer, + additive, porous, uncemented per 21<br>CFR 888.3358 |

Description: A highly cross linked polyethylene acetabular liner infused with pure liquid pharmaceutical grade alpha-tocopheral. This application is for the addition of size 32 ID, with neutral, 10° and 20° hooded configurations available.

#### Indications for Use:

Joint replacement is indicated for patients suffering from disability due to:

- noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis of the natural femoral head;
- rheumatoid arthritis;
- correction of functional deformity; ●
- femoral fracture.

This device may also be indicated in the salvage of previously failed surgical attempts. This device is intended for cementless use only

#### Predicate Device:

- K973119, K974093, K974095 DJO Surgical FMP Acetabular Shells with Standard Poly Liner ● (Spiked, Hemispherical, Flared Shells)
- K130365 DJO Surgical X-alt Highly Cross Linked Acetabular Liners with Vitamin E ●

Comparable Features to Predicate Device(s): Features comparable to predicate devices include the same design features, materials, indications, sterilization, packaging, and intended use.

Non-Clinical Testing The 32 mm liner does not create a new worst-case for mechanical testing compared to the predicate devices.

#### Clinical Testing: None provided.

---

**Source:** [https://fda.innolitics.com/device/K140130](https://fda.innolitics.com/device/K140130)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
