← Product Code [BZD](/productcode/BZD) · K140124

# S9 ELOUERA (K140124)

_Resmed Corp · BZD · May 12, 2014 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K140124

## Device Facts

- **Applicant:** Resmed Corp
- **Product Code:** [BZD](/productcode/BZD.md)
- **Decision Date:** May 12, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5905
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The S9 Elouera self-adjusting device is indicated for the treatment of Obstructive Sleep Apnea (OSA) in patients (female patients with mild to moderate OSA when using AfH treatment mode) weighing more than 66 lb (30 kg). It is intended for home and hospital use.

## Device Story

S9 Elouera is a microprocessor-controlled flow generator providing Continuous Positive Airway Pressure (CPAP) to treat OSA. It functions as a non-continuous ventilator, delivering pressure (4-20 cmH2O) to maintain an 'air splint' in the airway. Used in home or hospital settings by patients under clinician guidance. Inputs include real-time patient breathing patterns (inspiratory flow limitation, snore, apnea) detected via the device. The system features three therapy modes: fixed CPAP, AutoSet (automatic pressure adjustment), and AutoSet for Her (AfH), which provides modified responses to flow limitations with slower pressure rise/decay tailored for female patients. It includes Cheyne-Stokes Respiration (CSR) detection and reporting. Output is therapeutic air pressure delivered via mask and tubing. Clinicians use reported data to monitor treatment efficacy. Optional pulse oximeter accessory confirms apnea events. Benefits include automated, personalized pressure delivery to improve OSA management.

## Clinical Evidence

Single-blind, randomized, cross-over non-inferiority clinical trial compared the new AfH algorithm to the standard AutoSet algorithm (predicate S8 Aspen). Primary endpoints included Apnea-Hypopnea Index (AHI) and Oxygen Desaturation Index (ODI). Results demonstrated the AfH algorithm effectively treated female OSA patients. No adverse events or complications were reported. Bench testing verified CSR detection against human scoring, meeting predefined sensitivity and specificity criteria.

## Technological Characteristics

Microprocessor-controlled blower system; 4-20 cmH2O pressure range. Components: flow generator, integrated humidifier, mask, tubing. Materials in gas pathway classified as permanent external communicating devices, tested per ISO 10993-3, -5, -6, -10. Connectivity: optional pulse oximeter. Standards: IEC 60601-1, IEC 60601-1-2, IEC 62366. Software: rule-based/algorithmic pressure adjustment.

## Regulatory Identification

A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.

## Predicate Devices

- S8 Aspen ([K091947](/device/K091947.md))
- ApneaLink Plus ([K083575](/device/K083575.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

|                              | 510(k) Summary - S9 Elouera                                                                                                                                                                   | K140124<br>MAY 1 2 2014 |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| Required                     | By section 807.92 (c)                                                                                                                                                                         |                         |
| Date Prepared<br>Owners Name | 12th May, 2014<br>ResMed Ltd<br>1 Elizabeth Macarthur Drive<br>Bella Vista, NSW 2153, Australia                                                                                               |                         |
| Submitter                    | Greg Dockar<br>+ 61 2 8884 2157 (Phone)<br>+ 61 2 8884 2004 (FAX)<br>GregD@resmed.com.au                                                                                                      |                         |
| Official Contact             | Mr Jim Cassi - Vice President - Quality Assurance Americas<br>ResMed Corp.<br>9001 Spectrum Center Boulevard,<br>San Diego, CA 92123<br>+ 1 (858) 836 6081 (Phone)<br>+1 (858) 836 5519 (FAX) |                         |
| Product Codes                | 73 BZD<br>73 MNR                                                                                                                                                                              |                         |
| Class                        | II                                                                                                                                                                                            |                         |
| Classification Reference     | (21 CFR 868.5905 Product code 73 BZD)<br>(21 CFR 868.2375 Product code 73 MNR)                                                                                                                |                         |
| Common/Usual Name            | Non continuous ventilator (IPPB)<br>Breathing Frequency Monitor                                                                                                                               |                         |
| Proprietary Name             | S9 Elouera                                                                                                                                                                                    |                         |
| Predicate Device(s)          | S8 Aspen (K091947)<br>ApneaLink Plus (K083575)                                                                                                                                                |                         |
| Reason for submission        | New Device                                                                                                                                                                                    |                         |

{1}------------------------------------------------

### Indication for Use

The S9 Elouera self-adjusting device is indicated for the treatment of Obstructive Sleep Apnea (OSA) in patients (female patients with mild to moderate OSA when using AfH treatment mode) weighing more than 66 lb (30 kg).

It is intended for home and hospital use.

### Substantial Equivalence

The new device has the following similarities to the previously cleared predicate devices.

- > Similar intended use Same
- > operating principle Similar
- > technologies
- > Same manufacturing process

The S9 Elouera retains all the same operating/technologies/manufacturing characteristics cleared in the S8 Aspen (K091947). The S9 Elouera includes an additional treatment mode called AutoSet for Her (AH). AutoSet for Her mode is based on key aspects of ResMed's AutoSet algorithm and delivers therapeutic responses tailored to the characteristics of female OSA patients. These include modified responses to Flow Limitations resulting in a slower pressure rise and decay when compared to standard AutoSet algorithm.

These changes have been fully tested within the clinical trial as shown in the "Clinical Testing" section of this 510(k) summary.

The S9 Elouera includes Cheyne-Stokes Respiration (CSR) breathing pattern recognition and reporting, this feature is equivalent to ResMed's ApneaLink Plus (K083575).

Design and Verification activities were performed on the S9 Elouera device as a result of the risk analysis and design requirements. All tests confirmed the product met the predetermined acceptance criteria. ResMed has determined that the new device has not altered the safety and effectiveness of CPAP treatment. The S9 Elouera complies with the applicable requirements referenced in the FDA guidance documents:

- > FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005)
- > FDA Draft Guidance for Industry and FDA Staff Design Considerations for Devices Intended for Home Use- Document Issued on: December 12, 2012
- > FDA Draft Guidance for Industry and FDA Staff Radio Frequency Wireless Technology in Medical Devices - Document Issued on: August 13, 2013
- > Reviewer Guidance for Premarket Notification Submissions, ARDB, CDRH, FDA, November 1993.

### Clinical Testing:

A clinical trial demonstrated that the modified AutoSet algorithm performed as expected in treating female patients in a single-blind, randomised, cross-over non-inferiority study comparing the efficacy of the new AfH algorithm when compared to the standard AutoSet algorithm cleared in the S8 Aspen (K091947). The trial showed that the AfH algorithm effectively treated female OSA patients as reflected by the apneahypopnea index (AHI) and oxygen desaturation index (ODI) outcomes.

No adverse event or complications occurred as a result of the trial.

{2}------------------------------------------------

# Non-Clinical Testing:

Side-by-Side bench testing was performed to verify that the S9 Elouera met the requirements of the S9 Elouera System Specification when compared to the predicate devices (S8 Aspen (K091947) and Apneal.ink Plus (K083575)).

This bench testing included testing the performance of each therapy mode which included:

- > Pressure stability
- A Response to apneas
- Response to flow limitations and snore. >
- Reporting of Closed Airway Detection (CAD) A

A breathing machine simulates patient breathing patterns, which results in the Flow Generator responding in a manner consistent with maintaining the CPAP treatment pressure (CPAP mode) or adjusting the CPAP pressure based upon the patient's condition in real-time (self-adjusting mode). The clinical Pass/Fail requirements are traced to the S9 Elouera System Specification and to the predicate device's performance.

The S9 Elouera also includes an optional pulse oximeter accessory which is used to confirm apneas when linked with the apnea events.

Bench testing for the S9 Elouera includes Chevnes Stokes Respiration (CSR) breathing pattern recognition and reporting, this feature is equivalent to ResMed's ApneaLink Plus (K083575). Side-by-side testing using the same digitised breathing patterns was used. This test executes patient script files. Each patient script file is treated as an individual scenario, which the S9 Elouera reports either:

- > No CSR CSR > CSR + OSA > OSA
The S9 Elouera CSR detector results are then compared to human scoring where sensitivity and specificity are assessed. The S9 Elouera met the predefined Clinical Pass/Fail criteria.

In this test, a patient is reported as having CSR once the CSR duration on the FG reaches a value of 15 mins. Once the CSR duration of 15 mins is met, the scenario is complete and the next scenario in the feature file can begin. Should the CSR duration fail to reach 15 mins, the patient script file will run to completion, before the scenario is completed and moves on to the next one. The S9 Elouera met the predefined Clinical Pass/Fail criteria.

Materials used in the construction of components that contact the heated humidified gas pathway are classified as permanent "external communicating devices" (with tissue/bone/dentin). The appropriate biological tests conducted and passed for these components, in accordance with FDA guidance #G95-1 were:

- ISO 10993-3 Genotoxicity, .
- ISO 10993-5 Cytotoxicity, .
- ISO 10993-6 Implantation and ●
- ISO 10993-10 Sensitization & Irritation. .

Testing for particulate matter and volatiles demonstrated compliance to EPA requirements.

{3}------------------------------------------------

The S9 Elouera has been tested to appropriate standards and other applicable requirements. The S9 Elouera with integrated heated humidifier was designed and tested according to:

- > IEC 60601-1:2005. Medical electrical equipment Part 1: General requirements for safety Medical electrical equipment - General requirements for basic safety and essential performance
- > IEC 60601-1-2:2007, Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests

### Device Description

The S9 Elouera retains all the same hardware and performance features of the predicate device(s). Key features include in-line power supply, fully integrated humidifier, tubing, colour LCD and simplified controls which provides improved usability. The S9 Elouera contains a Micro-processor controlled blower system that generates Continuous Positive Airway Pressure (CPAP) from 4-20 cmHzO as required to maintain an "air splint" for effective treatment of OSA. The system comprises the flow generator, patient tubing, mask (patient interface) and humidifier.

The S9 Elouera flow generator includes three therapy modes. These include:

- CPAP mode the device delivers a continuous positive airway pressure throughout the entire . therapy session
- . AutoSet mode - the device automatically adjusts pressure in response to inspiratory flow limitation, snore and apnea.
- t AutoSet for Her mode - the device automatically adjusts pressure in response to female-specific OSA characteristics.

The functional characteristics of the S9 Elouera system includes all the clinician and user friendly features of the predicate device which have been verified during usability studies in accordance with IEC 62366 Medical devices - Application of usability engineering to medical devices.

# Characteristics between predicate and new device

| Characteristic            | Predicate Device S8 Aspen (K091947)                                                                                                                                                                                                | New device S9 Elouera                                                                                                                                                                                                                                                         | Comments                                                                                    |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| Indication for use        | The S8 Aspen self-adjusting system is indicated for the treatment of obstructive sleep apnea (OSA) in patients weighing more than 66 lb (>30 kg).<br><br>The S8 Aspen self-adjusting system is intended for home and hospital use. | The S9 Elouera self-adjusting device is indicated for the treatment of Obstructive Sleep Apnea (OSA) in patients (female patients with mild to moderate OSA when using AfH treatment mode) weighing more than 66 lb (30 kg).<br><br>It is intended for home and hospital use. | Equivalent<br>Name change and reduced patient population from predicate regarding AfH mode. |
| Location of use (primary) | Hospital/Home                                                                                                                                                                                                                      | Hospital/Home                                                                                                                                                                                                                                                                 | Equivalent                                                                                  |
| Performance               |                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                               |                                                                                             |
| Pressure range            | 4-20 cm H2O (all modes) EPR +3 cm H2O (all modes)                                                                                                                                                                                  | 4-20 cm H2O (all modes) EPR +3 cm H2O (all modes)                                                                                                                                                                                                                             | Equivalent                                                                                  |

{4}------------------------------------------------

# ResMed

# S9 Elouera Traditional 510(k) Submission May 2014

| Ramp<br>features      | User selected as "Off" to 45 minutes in 5 minute increments Max Ramp time set at clinician's discretion                         | User selected as "Off" to 45 minutes in 5 minute increments Max Ramp time set at clinician's discretion                                                                          | Equivalent                                                                                                                                                |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Modes of<br>operation | CPAP Mode (Fixed-pressure) AutoSet Mode (maximum to 20cm H2O with CAD) CPAP and AutoSet with EPR (maximum to 20cm H2O with CAD) | CPAP Mode (Fixed-pressure) AutoSet Mode (maximum to 20cm H2O with CAD) CPAP, AutoSet with EPR(maximum to 20cm H2O with CAD) AutoSet for Her (AfH) (maximum to 20cm H2O with CAD) | Equivalent:<br>Addition of AfH treatment mode.<br>Clinical testing demonstrated that<br>this new mode is substantially<br>equivalent to the AutoSet mode. |

| Characteristic       | Predicate Device S8 Aspen<br>(K091947)                                                            | New device S9 Elouera                                           | Comments   |
|----------------------|---------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|------------|
| System<br>Components | Flow generator Integrated humidifier<br>(HumidAire 4i Plus) Mask, air tubing and heated<br>tubing | Flow generator Humidifier Mask, air tubing and heated<br>tubing | Equivalent |

| Characteristic                    | Predicate Device ApneaLink Plus<br>(K083575)                                                             | New device S9 Elouera                                                                                       | Comments   |
|-----------------------------------|----------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|------------|
| CSR<br>Detection and<br>Reporting | Scoring of apnea events such as<br>Obstructive Sleep Apnea (OSA) and<br>Cheyne-Stokes Respiration (CSR). | Scoring of apnea events such as<br>Obstructive Sleep Apnea<br>(OSA) and Cheyne-Stokes<br>Respiration (CSR). | Equivalent |

### Conclusion

The S9 Elouera is substantially equivalent to the predicate devices (S8 Aspen K091947) and ApneaLink Plus (K083575) and is as safe and as effective as the predicate devices.

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.

May 12, 2014

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Resmed Corporation Mr. Jim Cassi Vice President, Quality Assurance Americas 9001 Spectrum Center Blvd. San Diego. CA 92123

Re: K140124

Trade/Device Name: S9 Elouera Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator Regulatory Class: II Product Code: BZD, MNR Dated: April 7, 2014 Received: April 9, 2014

Dear Mr. Cassi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Page 2 - Mr. Cassi

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/Reportally/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

ManFDAiner-S

Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{7}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

#### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.

#### 510(k) Number (if known) K140124

Device Name S9 Eloucra

#### Indications for Use (Describe)

The S9 Elovera self-adjusting device is indicated for the treatment of Obstructive Sleep Apnea (OSA) in patients with mild to moderate OSA when using AfH treatment mode) weighing more than 66 lb (30 kg).

It is intended for home and hospital use.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

の「 「 」 「 」 「 」 「 FOR FDALUSE ONLY」 ( 2 ) ( 」 ( ) ( ) ) ( ) ) ( ) ) ( ) ) ) ( ) ) ) ) ( ) ) ) ) ( ) ) ) ) ( ) ) ) ( ) ) ) ( ) ) ) ( ) ) ) ( ) ) ) ( ) ) ) ( ) ) ) ( ) ) ) ( ) Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Image /page/7/Picture/14 description: The image contains a logo on the left and text on the right. The text reads "Anya C. Harry -S", "2014.05.12", and "04:24:11 -04'00'". The logo is a stylized design, possibly representing a company or organization.

FORM FDA 3881 (1/14)

{8}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.'

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

---

**Source:** [https://fda.innolitics.com/device/K140124](https://fda.innolitics.com/device/K140124)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
