← Product Code [BSO](/productcode/BSO) · K140110

# FLEXTIP PLUS EPIDURAL CATHETER;FLEXBLOCK CONTINOUS PERIPHERAL NERVE BLOCK CATHETER KIT/SET (K140110)

_Teleflex, Inc. · BSO · Mar 27, 2015 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K140110

## Device Facts

- **Applicant:** Teleflex, Inc.
- **Product Code:** [BSO](/productcode/BSO.md)
- **Decision Date:** Mar 27, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5120
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The Arrow Epidural Catheter permits access to the epidural space for the administration of epidural anesthetic. The epidural catheter is intended for use up to 72 hours.

## Device Story

The ARROW FlexTip Plus Epidural Catheter is a single-lumen, wire-reinforced polyurethane catheter used to provide access to the epidural space for anesthetic administration. It features centimeter markings to assist clinicians with accurate placement. The device is intended for short-term use (up to 72 hours). This submission updates the MR Conditional labeling to align with current best practices based on magnetic resonance testing performed by Shellock R&D Services. The device remains identical to the predicate in design, materials, and technological characteristics.

## Clinical Evidence

No clinical data. Bench testing only: evaluation of magnetic field interactions, heating, and artifacts performed by Shellock R&D Services to support updated MR Conditional labeling.

## Technological Characteristics

Single-lumen catheter; wire-reinforced polyurethane body; includes centimeter markings for placement. MR Conditional. No changes to materials or design from predicate.

## Regulatory Identification

An anesthesia conduction catheter is a flexible tubular device used to inject local anesthetics into a patient and to provide continuous regional anesthesia.

## Predicate Devices

- FlexTip Plus Epidural Catheter ([K103658](/device/K103658.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are rendered in a dark color, contrasting with the white background of the seal.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 27, 2015

Teleflex, Inc. ARROW International, Inc. Chet Jones Regulatory Affairs Specialist 2400 Bernville Rd. Reading, PA 19605

Re: K140110

Trade/Device Name: ARROW FlexTip Plus Epidural Catheter Regulation Number: 21 CFR 868.5120 Regulation Name: Anesthesia Conduction Catheter Regulatory Class: Class II Product Code: BSO, CAZ Dated: February 23, 2015 Received: February 24, 2015

Dear Mr. Jones:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{1}------------------------------------------------

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Erin I. Keith -S

Erin I. Keith, M.S. Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known) K140110

#### Device Name

ARROW® FlexTip Plus® Epidural Catheter

#### Indications for Use (Describe)

The Arrow Epidural Catheter permits access to the administration of epidural anesthetic. The epidural catheter is intended for use up to 72 hours.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image features the logo for Arrow International. The word "ARROW" is written in large, white, sans-serif font, with a horizontal line extending from the left side of the "A" to the left side of the "R". Below "ARROW", the word "INTERNATIONAL" is written in a smaller, white, sans-serif font. The background is a solid blue color.

#### Submitter Information

| Name           | ARROW International, Inc. (subsidiary of Teleflex Inc.) |
|----------------|---------------------------------------------------------|
| Address        | 4024 Stirrup Creek Dr., Suite 720                       |
|                | PO Box 12600                                            |
|                | Durham, NC 27703                                        |
| Phone number   | 919-433-2579                                            |
| Fax number     | 919-433-4996                                            |
| Contact Person | Chet Jones                                              |
| Date prepared  | June 30, 2014 (Revised March 26, 2015)                  |

## Trade Name

1. ARROW® FlexTip Plus® Epidural Catheters

#### Common/Usual Name

1. Epidural Catheter

## Classification Name

Catheter:

BSO - Anesthesia Conduction Catheter 21 CFR 868.5120 Class 2

Kit:

CAZ – Anesthesia Conduction Kit 21 CFR 868.5140 Class 2

## Predicate Device

1. FlexTip Plus Epidural Catheter - K103658

#### Modification and Changes to Predicate

The purpose of this traditional 510(k) is to update the MR Conditional statements in the Instructions for Use (IFU). The update is based on magnetic resonance testing performed by Shellock R&D Services. The purpose of the update is to make a correction and to align the MR Conditional statements to the current best practices.

The update to the MR Conditional statements in the IFU is the only modification. The devices that are the subject of this submission are identical to their respective predicates in all other respects, including their technological characteristics, materials, indications for use, and FDA classification.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Arrow International. The word "ARROW" is in large, bold, white letters on the top line. Below that, the word "INTERNATIONAL" is in smaller, white letters. The background is a solid blue color.

### Device Description

ARROW® FlexTip Plus® Epidural Catheters are single lumen catheters featuring a wire re-enforced polyurethane body with centimeter markings to facilitate placement.

#### Indications for Use

The Arrow Epidural Catheter permits access to the epidural space for the administration of epidural anesthetic. The epidural catheter is intended for use up to 72 hours.

#### Substantial Equivalence Comparison to Predicates

ARROW FlexTip Plus Epidural Catheters were cleared under K103658. No changes have been made to the catheters or components since receiving clearance. Substantial equivalence between the device subject of this submission and the predicate device is based upon the fact that they are identical in regards to technological characteristics, design, materials, indications for use, and FDA classification.

The only change for these devices is an update to the MR Conditional labeling based on magnetic resonance testing performed by Shellock R&D Services. The update are provided in Section 014 Proposed Labeling. The test report is provided in Attachment C.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for Arrow International. The word "ARROW" is in large, white, sans-serif font on the top line. Below that, the word "INTERNATIONAL" is in a smaller, white, sans-serif font. The background is a solid blue color.

| Characteristic                | Comparison to Predicate                                                                                                                                                                                          |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Classification        | No change                                                                                                                                                                                                        |
| Intended Use                  | No change                                                                                                                                                                                                        |
| Technological Characteristics | No change                                                                                                                                                                                                        |
| Materials                     | No change                                                                                                                                                                                                        |
| MR Conditional Labeling       | See Table 1 in Section 014 Proposed Labeling.                                                                                                                                                                    |
| Performance Testing           | Evaluation of Magnetic Field Interactions, Heating and<br>Artifacts for FlexTip Plus Epidural Catheter, 19-gauge x 90-<br>cm, Nonmagnetic SS Version by Shellock R&D Services,<br>12/09/11, Revised 7/15/2013. * |

### Table 1 Substantial Equivalence Comparison to Predicates

* The same testing was submitted in support of the MR Conditional statements cleared in the predicate 510(k) K103658. However, the test was found to contain an error and was revised, producing the 7/15/2013 version of the report. This performance testing is presented in support of the MR Conditional statements and is not intended to support substantial equivalence. Substantial equivalence is based upon the devices being identical to their predicates in all respects.

#### Substantial Equivalence Conclusion

The proposed update to the MR Conditional labeling establishes parameters for the safe use of these catheters in a magnetic resonance environment. It does not impact the safety and effectiveness of ARROW FlexTip Plus Epidural Catheters, and does not raise new questions of safety or efficacy.

---

**Source:** [https://fda.innolitics.com/device/K140110](https://fda.innolitics.com/device/K140110)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
