← Product Code [INK](/productcode/INK) · K140095

# PRIME ZOOM STRETCHER (K140095)

_Stryker Corporation · INK · May 20, 2014 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K140095

## Device Facts

- **Applicant:** Stryker Corporation
- **Product Code:** [INK](/productcode/INK.md)
- **Decision Date:** May 20, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.3690
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

The Prime Series Stretcher with Zoom Motorized Drive is an electromechanical stretcher that provides a method of transporting patients within healthcare facilities. The stretcher may be used for minor procedures and short-term stay, typical of existing stretcher applications such as short-term outpation, treatment, minor procedure, and as a short-term outpatient recovery platform. The drive-assist Zoom® feature provides a healthcare professional and/or trained representative greater maneuverability in steering and moving the stretcher with significantly less force. The Prime Series® Stretcher with Zoom® Motorized Drive is intended for use in all establishments and may include use in, but not limited to, the Emergency Department (ED), including the Trauma area, and Postanesthesia Care Unit (PACU). The Prime Series® Stretcher with Zoom® Motorized Drive has a safe working load up to 700 pounds (318 kg) and is intended to support and transport all patients, including those mildly to crime Series® Stretcher with Zoom Motorized Drive may also be used to transport deceased patients within an enclosed healthcare facility. The Prime X option is intended to provide an articulating radiographic patient support surface and a platform below the patient support surface for x-ray cassette placement to assist in the capture of clinical x-rays when used in conjunction with a medical x-ray system.

## Device Story

Electromechanical stretcher for patient transport within healthcare facilities; operated by healthcare professionals or trained representatives. Features Zoom Motorized Drive system to assist maneuverability and reduce physical force required during transport. Manual push capability available during power loss. Includes hydraulic lift for height adjustment, Trendelenburg positioning, and siderails. Prime X option provides articulating radiographic support surface with under-platform cassette placement for use with external medical x-ray systems. Device supports up to 700 lbs. Benefits include improved caregiver ergonomics and maneuverability during patient transport.

## Clinical Evidence

Bench testing only. No clinical data. Performance verified through mechanical, functional, and software testing against specified requirements and recognized standards (ISO 10993, IEC 60601 series).

## Technological Characteristics

DC-powered wheeled stretcher; 120V charging system; 2x12V 31Ah batteries. Materials: cold/hot rolled steel, HDPE, glass-filled polypropylene, ABS, rigid PU foam, aluminum, powdercoat paint. Features: 4-caster brake system, hydraulic lift, manual overrides. Prime X option: articulating radiographic surface. Connectivity: none. Sterilization: not intended to be sterilized.

## Regulatory Identification

A powered wheeled stretcher is a battery-powered table with wheels that is intended for medical purposes for use by patients who are unable to propel themselves independently and who must maintain a prone or supine position for prolonged periods because of skin ulcers or contractures (muscle contractions).

## Special Controls

(1) Appropriate analysis and nonclinical testing must demonstrate that the safety controls are adequate to ensure safe use of the device and prevent user falls from the device in the event of a device failure; 
(2) Appropriate analysis and nonclinical testing must demonstrate the ability of the device to withstand the rated user weight load with an appropriate factor of safety; 
(3) Appropriate analysis and nonclinical testing must demonstrate the longevity of the device to withstand external forces applied to the device and provide the user with an expected service life of the device; 
(4) Appropriate analysis and nonclinical testing must demonstrate proper environments of use and storage of the device to maximize the longevity of the device; 
(5) Appropriate analysis and nonclinical testing (such as outlined in appropriate FDA-recognized consensus standards) must validate electromagnetic compatibility and electrical safety; 
(6) Appropriate analysis and nonclinical testing (such as outlined in appropriate FDA-recognized consensus standards) must validate that the skin-contacting components of the device are biocompatible; 
(7) Appropriate analysis and nonclinical testing (such as outlined in appropriate FDA-recognized consensus standards) must validate the software life cycle and that all processes, activities, and tasks are implemented and documented; 
(8) Appropriate analysis and nonclinical testing must validate that the device components are found to be nonflammable; 
(9) Appropriate analysis and nonclinical testing (such as outlined in appropriate FDA-recognized consensus standards) must validate that the battery in the device performs as intended over the anticipated service life of the device; 
(10) Adequate labeling is provided to the user to document proper use and maintenance of the device to ensure safe use of the device in the intended use environment; and 
(11) Appropriate risk assessment including, but not limited to, evaluating the dimensional limits of the gaps in hospital beds, and mitigation strategy to reduce entrapment.

## Predicate Devices

- Stryker Powered Wheeled Stretcher ([K022309](/device/K022309.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 140095

#### SECTION 5 - 510(k) SUMMARY or 510(k) STATEMENT

# MAY 2 0 2014

#### 510(k) Summary

Stryker Corporation - Medical Division

Prime Series® Stretcher with Zoom® Motorized Drive

#### SUBMITTER/510(K) HOLDER

Date Prepared:

| Name:           | Stryker Corporation - Medical Division     |
|-----------------|--------------------------------------------|
| Address:        | 3800 East Centre Ave.<br>Portage, MI 49002 |
| Contact Person: | Brian L. Orwat                             |
| Telephone:      | 269 389 6817                               |

#### DEVICE NAME

| Proprietary Name:     | Prime Series® Stretcher with Zoom® Motorized Drive |
|-----------------------|----------------------------------------------------|
| Catalog Numbers:      | 1125                                               |
| Common/Usual Name:    | Patient Transport Stretcher                        |
| Classification Name:  | Powered Wheeled Stretcher                          |
| Classification:       | 21 CFR 890.3690                                    |
| Product Code:         | INK                                                |
| Classification Panel: | Physical Medicine                                  |

11 March 2014

#### PREDICATE DEVICES

Stryker Medical ("Stryker") claims substantial equivalence to: 1. Stryker Powered Wheeled Stretcher (K022309)

#### DEVICE DESCRIPTION

The Prime Series® Stretcher with Zoom® Motorized Drive, is a powered wheeled stretcher that consists of a platform mounted on a wheeled frame that is designed to transport patients in a substantially horizontal position within the interior of a healthcare facility by health professionals and/or trained representatives of the user facility. The electric-drive system, called the Zoom Drive System, assists the health professional and/or trained representative by assisting stretcher movement and maneuverability in various healthcare facilities. The device can be manually pushed by the user in the event of power loss to the Zoom Drive System. The device has siderails, supports for fluid infusion equipment, and various options and accessories that assist with the transport of the patient. The Prime X option is intended to provide an articulating radiographic patient support surface and a platform below the patient support surface for x-ray cassette placement to assist in the capture of clinical x-rays when used in conjunction with a medical x-ray system.

{1}------------------------------------------------

#### INTENDED USE

The Prime Series Stretcher with Zoom Motorized Drive is an electromechanical stretcher that provides a method of transporting patients within healthcare facilities. The stretcher may be used for minor procedures and short-term stay, typical of existing stretcher applications such as short-term outpation, treatment, minor procedure, and as a short-term outpatient recovery platform. The drive-assist Zoom® feature provides a healthcare professional and/or trained representative greater maneuverability in steering and moving the stretcher with significantly less force. The Prime Series® Stretcher with Zoom® Motorized Drive is intended for use in all establishments and may include use in, but not limited to, the Emergency Department (ED), including the Trauma area, and Postanesthesia Care Unit (PACU). The Prime Series® Stretcher with Zoom® Motorized Drive has a safe working load up to 700 pounds (318 kg) and is intended to support and transport all patients, including those mildly to crime Series® Stretcher with Zoom Motorized Drive may also be used to transport deceased patients within an enclosed healthcare facility.

The Prime X option is intended to provide an articulating radiographic patient support surface and a platform below the patient support surface for x-ray cassette placement to assist in the capture of clinical x-rays when used in conjunction with a medical x-ray system.

#### TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE

Substantial equivalence is based on one predicate device, Stryker Power Wheeled Stretcher (K022309). The referenced predicate device is a powered wheeled stretcher as defined in 21 CFR § 890.3690. The product code of the cleared predicate device is INK.

Like the predicate device, the Prime Series Stretcher with Zoom Motorized Drive is a DC powered wheeled stretcher that operates as an electromechanical stretcher which provides a method of transporting patients within healthcare facilities. The Prime Series® Stretcher with Zoom® Motorized Drive may be used for minor procedures and short-term stay, typical of existing stretcher applications. The drive-assist Zoom system professional and/or trained representative greater manewerability in steering and moving the stretcher with significantly less force. The Prime Series® Stretcher with Zoom® Motorized Drive is equivalent in operational characteristics to the predicate device in that the mobility, drive-assist Zoom system, and caregiver input for patient transport operate in the same manner. Both devices have a four (4) caster brake system, side rails, hydraulic lift system for height adjustment, support, and trendelenburg features of the patient surface. Both devices also feature similarities in construction materials and surface mattresses.

The Prime Series® Stretcher with Zoom® Motorized Drive includes an available option that allows for a full-length articulating radiographic patient support surface and platform. A dual-deck design of the patient platform allows for the positioning of x-ray cassettes at any point under the patient, from head to side to side. With an open architecture design, this design will accommodate the majority of sizes of commercially available x-ray cassette receptors. The predicate device featured an optional x-ray cassette holder behind the fowler portion of the patient platform. Verification and validation of design and performance for the Prime Series® Stretcher with Zoom® Motorized Drive demonstrates that these technology differences do not adverselv affectiveness of the device when used as labeled, as the device has been fully tested for use and performance to demonstrate its safe and effective use.

{2}------------------------------------------------

| Table 5-1 Side-by-Side Comparison of                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                |
|--------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Prime Series® Stretcher with Zoom® Motorized Drive with Predicate Device |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Category                                                                 | Subject Device: Prime Series® Stretcher with Zoom® Motorized Drive                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Predicate Device: Stryker Powered Wheeled Stretcher K022309                                                                                                                                                                                                                                                                                                                                                                    |
| Indications for Use                                                      | The Prime Series® Stretcher with Zoom® Motorized Drive is an electromechanical stretcher that provides a method of transporting patients within healthcare facilities. The stretcher may be used for minor procedures and short-term stay, typical of existing stretcher applications, such as short-term outpatient clinical evaluation, treatment, minor procedure, and as a short-term outpatient recovery platform. The drive-assist Zoom® feature provides a healthcare professional and/or trained representative greater maneuverability in steering and moving the stretcher with significantly less force. The Prime Series® Stretcher with Zoom® Motorized Drive is intended for use in all establishments and may include use in, but not limited to, the Emergency Department (ED), including the Trauma area, and Postanesthesia Care Unit (PACU). The Prime Series® Stretcher with Zoom® Motorized Drive has a safe working load up to 700 pounds (318 kg) and is intended to support and transport all patients, including those mildly to critically ill. The Prime Series® Stretcher with Zoom® Motorized Drive may also be used to transport deceased patients within an enclosed healthcare facility.<br><br>The Prime X option is intended to provide an articulating radiographic patient support surface and a platform below the patient support surface for x-ray cassette placement to assist in the capture of clinical x-rays when used in conjunction with a medical x-ray system. | The Stryker Powered Wheeled Stretcher is an electromechanical stretcher that provides a method of transporting patients within healthcare facilities. The stretcher may be used for minor procedures and short-term stay, typical of existing stretcher applications. The drive-assist Big Wheel provides the healthcare caregiver greater maneuverability in steering and moving the stretcher with significantly less force. |
| Product Weight                                                           | 416 – 483 lbs.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Less than 450 lbs.                                                                                                                                                                                                                                                                                                                                                                                                             |
| Patient Capacity                                                         | 700 lbs. max                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 500-700 lbs. max                                                                                                                                                                                                                                                                                                                                                                                                               |
| Overall Dimensions                                                       | Length: 86"<br>Width: 31" to 34"                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Length: 84"<br>Width: 31" to 34"                                                                                                                                                                                                                                                                                                                                                                                               |
| Patient Surface Dimensions                                               | Length: 75.25"<br>Width: 26" or 30"                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Length: 75.5"<br>Width: 26" to 29"                                                                                                                                                                                                                                                                                                                                                                                             |
| Height Range (to litter top)                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                |

.

Table 5-1 Side-by-Side Comparison of Predicate

{3}------------------------------------------------

| - High                                     | 37.25"                                                                                                                                                    | 35.5" - 36"                                                                                                                                                   |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| - Low                                      | 26.25"                                                                                                                                                    | 21" - 22"                                                                                                                                                     |
| Trendelenburg/<br>Reverse<br>Trendelenburg | +/- 16 degrees                                                                                                                                            | +/- 16 degrees                                                                                                                                                |
| Articulation:<br>- Fowler<br>(electric)    | Available option, not in conjunction<br>with Prime X option                                                                                               | None                                                                                                                                                          |
| - Fowler<br>(manual)                       | 0-90 degrees                                                                                                                                              | 0 - 90 degrees                                                                                                                                                |
| - Knee Gatch                               | 0-40 degrees, not in conjunction with<br>Prime X option                                                                                                   | 0 - 35 degrees                                                                                                                                                |
| Siderails<br>- Length                      | Glideaway collapsible siderails<br>54"                                                                                                                    | Glideaway collapsible siderails<br>54"                                                                                                                        |
| - Height                                   | 12" (Prime X Option); 14"                                                                                                                                 | 14"                                                                                                                                                           |
| - In-Rail Controls                         | Available option, not in conjunction<br>with Prime X option                                                                                               | None                                                                                                                                                          |
| - Patient Lock-out                         | Available option, not in conjunction<br>with Prime X option                                                                                               | None                                                                                                                                                          |
| Lift System                                | Manual Hydraulic pedestal jacks                                                                                                                           | Manual Hydraulic pedestal jacks                                                                                                                               |
| Brake and Steer                            | Located at head/foot end<br>4 caster brake system                                                                                                         | Located at head/foot end<br>4 caster brake system                                                                                                             |
| Mobility<br>- Zoom Drive                   | Drive-assisted "Big Wheel" –<br>controlled by the caregiver                                                                                               | Drive-assisted "Big Wheel" - controlled by the<br>caregiver                                                                                                   |
| - Push Control<br>Handles                  | Electrical push handles used to<br>maneuver loaded/unloaded stretcher                                                                                     | Electrical push handles used to maneuver<br>loaded/unloaded stretcher                                                                                         |
| - Performance on<br>Ramps (Grades)         | Drive-assisted "Big Wheel" will aid the<br>healthcare user                                                                                                | Drive-assisted "Big Wheel" will aid the healthcare<br>user                                                                                                    |
| Zoom Drive<br>Speed                        | Healthcare user sets the speed of the<br>stretcher using the push control handles                                                                         | Healthcare user sets the speed of the stretcher using<br>the push control handles                                                                             |
| Casters<br>- Diameter                      | 8"                                                                                                                                                        | 8"                                                                                                                                                            |
| - Casters                                  | Covered                                                                                                                                                   | Covered/Uncovered available                                                                                                                                   |
| Mattress                                   | 3" h (standard), 4", & 5" h optional (5"<br>not available in conjunction with Prime<br>X option)                                                          | 3" h (standard), 4" & 5" h (optional)                                                                                                                         |
| Energy Source -<br>Charging System         | 120 V, 60Hz, 4 A                                                                                                                                          | 120 V, 60Hz, 4 A                                                                                                                                              |
| Energy Source -<br>Battery                 | 2 x 12 V, 31Ah Battery                                                                                                                                    | 2 x 12 V, 31Ah Battery                                                                                                                                        |
| Materials<br>- Stretcher                   | Cold Rolled Steel, Hot Rolled Steel,<br>HDPE Plastic, Glass Filled<br>Polypropylene Plastic, ABS Plastic,<br>Rigid PU Foam, Aluminum,<br>Powdercoat paint | Low Carbon Steel, Fiber-Resin or Low Carbon Steel<br>Litter Panels, Siderails are chrome-plated low<br>carbon steel with plastic or low carbon steel toprail. |
| - Mattress                                 | Polyurethane foam with vinyl or                                                                                                                           | Polyurethane foam with vinyl or polyurethane-                                                                                                                 |

:

.

(

{4}------------------------------------------------

|                                     | polyurethane-coated nylon covering.                               | coated nylon covering.                                               |
|-------------------------------------|-------------------------------------------------------------------|----------------------------------------------------------------------|
| Manual<br>Overrides                 | All manual overrides                                              | All manual overrides                                                 |
| Standard<br>Operating<br>Conditions |                                                                   |                                                                      |
| - Storage                           | Temp: -4°F to +140°F, Rel Humidity –<br>Up to 95%                 | Temp: -32°F to +140°F, Rel Humidity – Up to<br>100% (non-condensing) |
| - Operation                         | Temp: 50°F to 100°F, Rel Humidity –<br>30% to 75% RH              | Temp: 60°F to 100°F, Rel Humidity – 0% to 60%<br>RH (non-condensing) |
| Sterility                           | Device is not intended to be sterilized                           | Device is not intended to be sterilized                              |
| X-Ray Surface                       | Articulating radiographic patient<br>support surface and platform | Optional radiolucent surface via X-Ray Cassette<br>Holder (Fowler)   |

#### SUMMARY

Stryker Medical has verified and validated that the Prime Series® Stretcher with Zoom® Motorized Drive meets its functional, performance, safety, and efficacy specifications and requirements. Physical and mechanical testing has been performed on individual components and on the system, including Software testing that has been completed. The Prime Series Stretcher with Zoom® Motorized Drive has successfully passed bench and software testing, Test results demonstrate that both the individual units and system meet specified performance requirements.

The Prime Series Stretcher with Zoom Motorized Drive has been designed and evaluated according to the following FDA Recognized and international Standards:

- AAMI/ANSI/ISO 10993-1: 2009; Biological Evaluation of Medical Devices -- Part 1: Evaluation and . Testing within a Risk Management Process
- ISO 10993-2: 2006; Biological Evaluation of Medical Devices -- Part 2: Animal Welfare Requirements .
- AAMI/ANSI/ISO 10993-5: 2009; Biological Evaluation Of Medical Devices -- Part 5: Tests For In Vitro . Cytotoxicity
- AAMI/ANSI/ISO 10993-10: 2010; Biological Evaluation Of Medical Devices Part 10: Tests For Irritation . And Skin Sensitization
- ISO 10993-11: 2006; Biological Evaluation Of Medical Devices Part 11: Tests For Systemic Toxicity .
- AAMI/ANSI ES 60601-1: 2005/(R)2012 and C1:2009(R)2012 and A2:2010/(R)2012 ; Medical Electrical . Equipment - Part 1: General Requirements For Basic Safety And Essential Performance
- . IEC 60601-1-2 Edition 3:2007-03; Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility -Requirements And Tests
- . IEC 60601-1-3 Edition 2.0 2008-01, Medical Electrical Equipment - Part 1-3: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Radiation Protection In Diagnostic X-Ray Equipment

{5}------------------------------------------------

- IEC 60601-2-52 Edition 1.0 2009-12, Medical Electrical Equipment Part 2-52: Particular Requirements . For Basic Safety And Essential Performance Of Medical Beds
- . IEC 60601-2-54 Edition 1.0 2009-06, Medical Electrical Equipment - Part 2-54: Particular Requirements For The Basic Safety And Essential Performance Of X-Ray Equipment For Radiography And Radioscopy

Based on the Prime Series® Stretcher with Zoom Motorized Drive and predicate design, operational and technical characteristics and completed bench testing, the Prime Series® Stretcher with Zoom® Motorized Drive is substantially equivalent to and as safe and effective as that of the predicate device. The Prime Series with Zoom Motorized Drive's intended uses are substantially supported by the previously cleared predicate device. Any differences described between the Prime Series® Stretcher with Zoom® Motorized Drive and the predicate device does not raise any new issues of safety or effectiveness. The Prime Series® Stretcher with Zoom Motorized Drive's indication for use statement includes all of the same indications as the previously cleared predicate device.

#### CONCLUSION

In summary, Stryker Medical has demonstrated that the Prime Series® Stretcher with Zoom® Motorized Drive is as safe and effective as and is substantially equivalent to the predicate device.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the eagle.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 20, 2014

Stryker Corporation - Medical Division c/o Mark Job Regulatory Technology Services LLC 1394 25th Street NW Buffalo, MN 55313

Re: K140095

Trade/Device Name: Prime Series® Stretcher with Zoom® Motorized Drive Regulation Number: 21 CFR 890.3690 Regulation Name: Powered Wheeled Stretcher Regulatory Class: Class II Product Code: INK Dated: March 19, 2014 Received: March 28, 2014

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{7}------------------------------------------------

Page 2 - Mr. Mark Job

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance,

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Felipe Aquel -S

Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation

Center for Devices and Radiological Health

Enclosure

•

{8}------------------------------------------------

# Indications for Use

510(k) Number (if known) K140095

Device Name

Prime Series® Stretcher with Zoom® Motorized Drive

#### Indications for Use (Describe)

The Prime Series® Stretcher with Zoom® Motorized Drive is an electromechanical stretcher that provides a method of transporting patients within healthcare facilities. The stretcher may be used for minor procedures and short-term stay, typical of existing stretcher applications, such as short-term outpation, treatment, minor procedure, and as a short-term outpatient recovery platform. The drive-assist Zoom® feature provides a healthcare professional and/or trained representative greater maneuverability in steering and moving the stretcher with significantly less force. The Prime Series® Stretcher with Zoom® Motorized Drive is intended for use in all establishments and may include use in, but not limited to, the Emergency Department (ED), including the Trauma area, and Postanesthesia Care Unit (PACU). The Prime Series® Stretcher with Zoom® Motorized Drive has a safe working load up to 700 pounds (318 kg) and is intended to support and transport all patients, including those mildly to critically ill. The Prime Series® Stretcher with Zoom® Motorized Drive may also be used to transport deceased patients within an enclosed healthcare facility.

The Prime X option is intended to provide an articulating radiographic patient support surface and a platform below the patient support surface for x-ray cassette placement to assist in the capture of clinical x-rays when used in conjunction with a medical x-ray system.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

# ം Date: 2014.05.20 15:38:03
2014.05.2014.05.20 15:38:03
2014.05.2014.05.20 15:38:03
2014.05.2014.05.2014.05.2014.00

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the I i ne to review instructions, search existing data sources, gather and maintain the data needed and complete time to review the collection. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

---

**Source:** [https://fda.innolitics.com/device/K140095](https://fda.innolitics.com/device/K140095)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
