The Cryotop®SC is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos.
Device Story
CryotopSC is a single-use, sterile cryopreservation storage device used in assisted reproduction laboratories. It consists of an ABS handle shaft, a PET fine tip with a flat film area for loading embryos, and a weighted straw enclosure. During vitrification, embryos are loaded onto the film tip; the device is then inserted into the straw, which is heat-sealed by the user. The device is used with a separately sold Aluminum Block to prevent liquid nitrogen ingress and stabilize manipulation. The device maintains embryo integrity during freezing and thawing. It is operated by laboratory personnel. The output is a cryopreserved embryo, which benefits patients by enabling storage for future use, potentially reducing the need for multiple extraction procedures.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical bench testing: cooling rate (3,000°C/min), warming rate (44,000°C/min), tensile strength (5N), endotoxin levels (≤0.5 EU/device), sterility, and Mouse Embryo Assay (≥80% development at 96 hours).
Technological Characteristics
Materials: PET, ABS, Polypropylene. Form factor: Two-piece assembly (handle shaft with fine tip and weighted straw). Sterilization: Radiation. Connectivity: None. Principle: Vitrification storage. Cooling rate: 3,000°C/min. Warming rate: 44,000°C/min.
Indications for Use
Indicated for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos.
Regulatory Classification
Identification
Assisted reproduction labware consists of laboratory equipment or supplies intended to prepare, store, manipulate, or transfer human gametes or embryos for in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures. These include syringes, IVF tissue culture dishes, IVF tissue culture plates, pipette tips, dishes, plates, and other vessels that come into physical contact with gametes, embryos or tissue culture media.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when it is a dish or plate intended for general assisted reproduction technology procedures, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
## 510(k) Summary
for
## CRYOTOP®SC
#### 1. Submission Sponsor
KITAZATO BioPharma Co., Ltd. 81 Nakajima, Fuji Shizuoka 416-0907 JAPAN Phone: +(81) 545-66-2202 Fax: +(81) 545-60-5772 Contact: Futoshi INOUE, President
#### 2. Submission Correspondent
Emergo Group 816 Congress Avenue, Suite 1400 Austin, TX 78701 Cell Phone: (508) 838.9139 Office Phone: (512) 327.9997 Fax: (512) 327.9998 Contact: Richard VINCINS, Vice President, QA Email: project.management@emergogroup.com
#### 3. Date Prepared
1 April 2014
#### 4. Device Identification
Cryotop®SC Trade/Proprietary Name: Cryopreservation storage device Common/Usual Name: Assisted Reproduction Labware Classification Name: 884.6160 Classification Regulation: MQK Product Code: Device Class: Class II Classification Panel: Obstetrics/Gynecology
## 5. Legally Marketed Predicate Device(s)
Cryo Bio System, HSV Straw, K092398
..............................................................................................................................................................................
# APR 0 3 2014
{1}------------------------------------------------
#### 6. Device Description
CryotopSC is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos. Extracting embryos from the body can be costly and burdensome process for the patient to go through multiple extraction procedures. As part of the vitrification procedure, the cells to be stored are loaded on the tip of the CryotopSC device for subsequent storage.
The CryotopSC device is composed of an ABS handle shaft with a PET fine tip and a straw enclosure. The fine tip has a flat film area for loading embryos. The CryotopSC device has a "straw" enclosure system for when the unit is placed in the liquid nitrogen. The straw cap is designed to be heat sealed by the user. The straw cap has a weight at the distal end to place the straw cap and the shaft inside of the cap in a correct position in the liquid nitrogen. The CryotopSC device is provided sterile and is for single use only. The CryotopSC device has been designed to maintain the integrity of the human embryos through the freezing and thawing process.
The Aluminum Block is required for use with the CryotopSC device for the vitrification and thawing process and is sold separately. The Aluminum Block is to be used with the CryotopSC device to prevent liquid nitrogen from entering the straw and to assist in stabilizing the manipulation of the straw.
The CryotopSC device conforms to product quality test specifications of our company: appearance, dimension, durability, tensile strength, colorfastness, endotoxin and Mouse Embryo Assay. The sterilization dose of CryotopSC is validated by sterilization to maintain the sterility of the device through anticipated storage and handling.
#### 7. Indication for Use Statement
The Cryotop®SC is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos.
#### Substantial Equivalence Discussion 8.
The following table compares the CryotopSC to the predicate device with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.
| Manufacturer<br>Trade Name | KITAZATO BioPharma Co., Ltd.<br>Cryotop®SC | Cryo Bio System<br>HSV Straw |
|----------------------------|--------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | Not assigned | K092398 |
| Product Code | MQK | MQK |
| Regulation Number | 884.6160 | 884.6160 |
| Regulation Name | Assisted Reproduction Labware | Assisted Reproduction Labware |
| Indications for Use: | The CryotopSC is a<br>cryopreservation storage device<br>that is intended for use in | The HSV Straw is a cryopreservation<br>storage device that is intended for<br>use in vitrification procedures to |
| | Table 5A - Comparison of Characteristics |
|--|------------------------------------------|
| | |
{2}------------------------------------------------
| Manufacturer | KITAZATO BioPharma Co., Ltd. | Cryo Bio System |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Cryotop ® SC | HSV Straw |
| | vitrification procedures to contain<br>and maintain human 4-8 cell and<br>blastocyst stage embryos. | contain and maintain human 4-8 cell<br>and blastocyst stage embryos. |
| Overall Design | The CryotopSC is designed to<br>contain, freeze and maintain<br>embryos. The device consists of a<br>two piece assembly comprised of<br>the main part containing the fine<br>tip film area and the "straw." The<br>straw is designed to be sealed. The<br>straw is weighted to allow proper<br>alignment in the storage container.<br>The CryotopSC device is packaged<br>in a single barrier sterilization<br>pouch. | The HSV Straw is designed to<br>contain, vitrify and maintain<br>embryos.<br>The device consists of 3 parts: a resin<br>straw, a capillary tube with a pre-<br>formed gutter attached to a colored<br>handling rod, and a blue plastic<br>insertion device.<br>The straw is designed to be sealed.<br>The HSV Straw is packaged in a peel<br>off blister pack. |
| Method of Action | Vitrification Method | Vitrification Method |
| Sterile | Radiation | Radiation |
| Cooling Rate | 3,000°C/min | 2,900°C/min |
| Rewarming Rate | 44,000°C/min | 25,000°C/min |
| Material Composition | PET, ABS, Polypropylene | Medical Grade Styrene-Butadiene<br>Copolymer, Ionomeric resin |
| Contact with Warming<br>Medium | The tip (film) and the shaft of<br>CryotopSC are taken out from the<br>straw. Directly immerse the tip<br>(film) into thawing solution. | The curved spatula, containing the<br>cryopreserved embryo, is immersed<br>in thawing solution where thawing<br>and dilution in the thawing solution<br>occur simultaneously. |
| Performance Testing of<br>Device | Passed | Passed |
| Mouse Embryo Test<br>Passed | Yes | Yes |
| Sterility Validation<br>Passed | Yes | Yes |
#### Non-Clinical Performance Data 9.
The CryotopSC device has been evaluated for the cooling/warming rate, mechanically tested, sterility tested, and mouse embryo assay supporting that all the specifications have met the acceptance criteria for the device. The following testing has been performed to support substantial equivalence:
Performance Testing
- Cooling Rate Testing: Cooling rate of 3,000 °C/min passed .
- Warming Rate Testing: Warming rate of 44,000 °C/min passed .
- Dimensional Testing: Passes outer diameter and length according to specifications .
- Mechanical Tensile Testing: Tensile strength to withstand 5N .
- Endotoxin Testing: Endotoxin values conform to the value ≤0.5 EU/device .
- Sterility Testing: No microbial growth from sterility testing .
- Mouse Embryo Assay: ≥80% of 1-cell control embryos develop at 96 hours .
{3}------------------------------------------------
Note: The performance testing, Mouse Embryo Assay (MEA), and sterility test are all performed on samples from routine manufactured lots; a Certificate of Analysis is provided with each lot of CryotopSC device.
The CryotopSC device passed all testing and supports the claims of substantial equivalence. The CryotopSC meets all the requirements for overall design, sterilization, functional, and mouse embryo assay testing confirms that the output meets the design inputs and specifications. The CryotopSC passed all testing stated above as shown by the acceptable results obtained.
The CryotopSC device complies with the applicable voluntary standards for sterilization. The device passed all the testing in accordance with national and international standards.
#### 10. Clinical Performance Data
There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. The substantial equivalence of the device is supported by the non-clinical testing. The validation testing of sterility testing and mouse embryo testing was found to acceptable and supports the claims of substantial equivalence.
#### 11. Statement of Substantial Equivalence
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.
KITAZATO BioPharma's Cryotop®SC, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol resembling a caduceus, featuring a staff with a snake winding around it.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 3, 2014
KITAZATO BioPharma Co., Ltd. % Richard A. VINCINS, CQA, CBA, RAC (US, EU) Vice President, QA Emergo Group 816 Congress Avenue, Suite 1400 Austin, TX 78701
K140072 Re:
> Trade/Device Name: Cryotop®SC Regulation Number: 21 CFR§ 884.6160 Regulation Name: Assisted reproduction labware Regulatory Class: II Product Code: MQK Dated: January 10, 2014 Received: January 13, 2014
Dear Richard A. VINCINS,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I loast of has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{5}------------------------------------------------
#### Page 2 - Richard A. VINCINS, CQA, CBA, RAC (US, EU)
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Image /page/5/Picture/6 description: The image shows the name "Benjamin Fisher -S" in a bold, sans-serif font. The letters "FDA" are stylized and overlaid on top of the name. The letters "FDA" are outlined with a thick border and have a decorative pattern inside.
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
#### 510(k) Number (if known)
#### K140072
Device Name Cryotop@SC
#### Indications for Use (Describe)
The Cryotop®SC is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos.
Type of Use (Select one or both, as applicable)
[x] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
1 PRESERT A FIFOR FDA USE ONLY CENTER FOR FREE S 3 1. 2. 1. 2. 1. 2. 1. 2. 1. 2. 1. 2. 1. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2 : 1
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{7}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.