LUXER SHADED ZIRCONIA

K140070 · Continental Dental Ceramics, Inc. · EIH · May 7, 2014 · Dental

Device Facts

Record IDK140070
Device NameLUXER SHADED ZIRCONIA
ApplicantContinental Dental Ceramics, Inc.
Product CodeEIH · Dental
Decision DateMay 7, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.6660
Device ClassClass 2

Indications for Use

Luxer Shaded Zirconia blanks/discs are intended for use with CAD/CAM technology to produce all-ceramic dental restorations (full contour crowns and bridges) as prescribed by a dentist.

Device Story

Luxer Shaded Zirconia blanks consist of Yttria-stabilized zirconia blocks (12-20 mm thickness) pigmented with TOSOH TZ-Series components (white, yellow, pink, gray) to match VITA shades A-D. Used in dental laboratories with CAD/CAM milling machines; milled blocks are sintered to create high-strength (>1100 Mpa), water-insoluble, biocompatible full contour crowns and bridges. Final restorations are placed by dentists to restore dental function and esthetics. Device relies on established material properties and CAD/CAM fabrication workflows.

Clinical Evidence

Bench testing only. Device validated against ISO 6872:2008 standards for physical and chemical properties, including flexural strength and solubility. Biocompatibility was established via equivalence to predicate materials.

Technological Characteristics

Yttria-stabilized zirconia; TOSOH TZ-Series pigments. Form factor: 12-20 mm blanks/discs. Material properties: >1100 Mpa flexural strength, water insoluble, biocompatible. Conforms to ISO 6872:2008. Processing: CAD/CAM milling followed by sintering.

Indications for Use

Indicated for patients requiring all-ceramic dental restorations, specifically full contour crowns and bridges, produced via CAD/CAM technology as prescribed by a dentist.

Regulatory Classification

Identification

Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ ### 510(k) Summary # MAY C 7 2014 ## Luxer Shaded Zirconia K140070 - 1. Date of Summary May 6, 2014 Preparation: 2. Submitting Firm: Continental Dental Ceramics, Inc. 3. Contact Person: Jerry Doviack, CDT President Continental Dental Ceramics, Inc. 1873 Western Way Torrance, CA 90501 USA jdoviack@continentaldental.com T: (310) 618-8821 F: (310) 618-1238 4. Name of Medical Device Proprietary Name: Luxer Shaded Zirconia Porcelain Powder for Clinical Use Regulation Name: Regulation Number: 872.6660 Product Code: EIH Powder, Porcelain Classification Name: Classification: Class II Panel: Dental - K112710 ZENO Zr ട്. Predicate Device: Wieland Dental + Technik GmbH & Co. KG #### 6. Description of Medical Device: Luxer Shaded Zirconia blanks are pressed and sintered blocks of Yttria stabilized Zirconia for use in CAD/CAM milling machines. It comes in different thicknesses (12 mm - 20 mm) and classic VITA shades A-D. In order to achieve pigmentation, four basic color components are used from the TOSOH TZ-Series of zirconia {white, yellow, pink, and gray} that yield corresponding VITA shades (A-D). After the Zirconia block is milled, it is sintered into a high strength all ceramic material suitable for full contour crowns and bridges. The material is biocompatible, has a high flexural strength (>1100 Mpa), and is water insoluble. These characteristics meet the ISO standard 6872 for an esthetic dental ceramic material. 510(k) Summary {2}------------------------------------------------ #### 7. Intended Use Luxer Shaded Zirconia blanks/discs are intended for use with CAD/CAM technology to produce all-ceramic dental restorations (full contour crowns and bridges) as prescribed by a dentist. #### Substantial Equivalence 8. Continental Dental believes that Luxer Shaded Zirconia blanks are substantially equivalent to other legally marketed devices in the United States, such as K112710 ZENO Zr, Wieland Dental + Technik Gmbh & Co. KG. Both Luxer Shaded blanks and the predicate device are made from porcelain powder (product code EIH) and exhibit similar physical and chemical properties. In comparison to the predicate, using ISO standard 6872 the data from that testing concludes that both materials are fundamentally the same. They both have the same fundamental technology, indications for use, physical properties (flexural strength, biocompatibility, solubility) and other technical similarities. Based on the comparative data, Luxer Shaded Zirconia is essentially the same as currently marketed devices for the same indication, with similar physical and chemical properties, and supports our claim for substantial equivalence. #### ರು Non-Clinical Testing Non-clinical testing was performed in order to validate the design against the company's specified design requirements for physical and chemical properties, as well as flexural strength, and to assure conformance with the voluntary design standard ISO 6872:2008. Please see attached Standard Data Report (Form 3654). #### 10. Biocompatibility Biocompatibility testing was not performed since identical materials are used in the predicate device with the same type and duration of patient contact. Our device is comprised of identical materials and manufacturing methods as the predicate device and other legally marketed devices and does not introduce any new issues of safety or effectiveness. #### 11. Safety & Effectiveness The successful prior use of the components of Luxer Shaded Zirconia product in legally marketed devices, the similarity of the formulations used in this device and earlier devices, and the substantially equivalence of Luxer Shaded Zirconia blanks to prior cleared devices support the safety and effectiveness of this product for the intended use. {3}------------------------------------------------ It has been shown in this 510(k) submission that the difference between Luxer Shaded Zirconia discs and the predicate device do not raise any questions regarding its safety and effectiveness. ## END OF SECTION ### 510(k) Summary : . : : {4}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three abstract human figures connected in a flowing, dynamic design. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MI) 20993-0002 May 7, 2014 Continental Dental Ceramics, Incorporated Mr. Jerry Doviack, CDT President 1873 Western Way Torrance, CA 90501 ・:・ . . . . . . Re: K140070 Trade/Device Name: Porcelain Powder for Clinical Use Regulation Number: 21 CFR 872.6660 Regulation Name: Powder, Porcelain Regulatory Class: II Product Code: EIH Dated: April 30, 2014 Received: May 1, 2014 Dear Mr. Doviack: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, MarySRunner-S Erin 1. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indications for Use 510(k) Number (if known): __K140070 Device Name: Luxer Shaded Zirconia Range of Indications: Luxer Shaded Zirconia blanks are intended for use with CAD/CAM technology to produce all ceramic dental restorations (full contour crowns and bridges) as prescribed by a dentist. Prescription Use x (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) Concurrence of CDRH, Office of Device Evaluation (ODE) She 2014.05.07 14 Indications for Use
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...