← Product Code [JCX](/productcode/JCX) · K140031

# SUCTION PUMP (K140031)

_Ding Hwa Co., Ltd. · JCX · Apr 28, 2014 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K140031

## Device Facts

- **Applicant:** Ding Hwa Co., Ltd.
- **Product Code:** [JCX](/productcode/JCX.md)
- **Decision Date:** Apr 28, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4780
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

AS-XXXX Suction Pump is intended to be used to remove body fluids from a patient's airway or respiratory system.

## Device Story

AS-XXXX Suction Pump is a portable, AC-powered device designed to remove bodily fluids from a patient's airway or respiratory system. The system generates negative pressure (vacuum) to draw fluids through a disposable suction catheter and tubing into a collection container for disposal. Components include a pump unit, on/off switch, power cord, collection container, relief valve, pressure adjustment knob, pressure gauge, and microbial filter. The device is operated by a physician or a trained caregiver under physician order. The vacuum level is manually adjusted by the user via the pressure knob and monitored via the pressure gauge. The device facilitates airway clearance, potentially benefiting patients with respiratory secretions. It is intended for use in clinical or home settings, provided the caregiver has received documented training.

## Clinical Evidence

Bench testing only. The device underwent non-clinical performance testing to verify compliance with IEC 60601-1 (General safety), IEC 60601-1-2 (EMC), and ISO 10079-1 (Safety requirements for electrically powered suction equipment). All results met established acceptance criteria.

## Technological Characteristics

Portable AC-powered suction pump. Components: pump unit, collection container, relief valve, pressure adjustment knob, pressure gauge, microbial filter, and tubing. Standards: IEC 60601-1 (3rd Ed.), IEC 60601-1-2, and ISO 10079-1 (2009). Manual pressure regulation. Standalone operation.

## Regulatory Identification

A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.

## Predicate Devices

- EMG Suction Unit, Model SUA01-AXX Series ([K112421](/device/K112421.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# CO., LTD. IING H

#### 510(k) Summary – Ding Hwa AS-XXXX (Family Series) Suction Pump Sec. 5:

| Date Summary was<br>Prepared                    | December 25, 2013                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Submitter                                | Ding Hwa Co. Ltd.<br>#121 Sec. 3 Zhongshan Road<br>Dacun, Chang-Hwa, Taiwan 51542                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Primary Contact for this<br>510(k) Submission   | Robert O. Dean<br>Compliance Systems International, LLc.<br>1083 Delaware Rd.<br>Buffalo, NY 14223<br>716.440.7362<br>compliancesystems@yahoo.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Device Common Name                              | Powered suction pump                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Trade Name                                      | AS-XXXX Suction Pump (family series)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Device Product Codes and<br>Classification Name | JCX, 21CFR878.4780, Powered suction pump                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Predicate Device                                | EMG Suction Unit, Model SUA01-AXX Series 510(k)112421                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Device Description                              | The AS-XXXX (FAMILY SERIES) suction pump is a portable AC-powered suction pump.<br>The AS-XXXX (FAMILY SERIES) suction pump creates a negative pressure (vacuum)<br>that draws fluids through a disposable suction catheter that is connected by suction<br>tubing to a collection container. The suctioned fluids are then trapped in the collection<br>container for proper disposal. The device must only be used on the order of a physician.<br>If practiced outside of the hospital setting, the care giver must be trained by a certified<br>healthcare professional and the training must be recorded and documented. The AS-<br>XXXX (FAMILY SERIES) suction pump consists of an on/off switch, a pump unit, a power<br>cord, a collection container, relief valve, pressure adjustment knob, pressure gauge,<br>microbial filter, long intermediate tubing, short intermediate tubing, and suction tubing.<br>Disposable suction catheters are not packaged with this device and must be purchased<br>separately. |
| Intended Use                                    | AS-XXXX Suction Pump is intended to be used to remove body fluids from a patient's<br>airway or respiratory system.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Technological<br>Characteristics                | Ding Hwa AS-XXXX Suction Pump is of the same design, material and performance<br>characteristics and intended use as the predicate device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Summary of Testing                              | Ding Hwa AS-XXXX Suction Pump is substantially equivalent and meets the same<br>acceptance criteria as the predicate device, EMG Suction Unit SUA01-AXX<br>510(k)112421. Non-clinical performance testing includes IEC 60601-1, 3rd Ed.<br>MEDICAL ELECTRICAL EQUIPMENT - PART 1: GENERAL REQUIREMENTS FOR<br>BASIC SAFETY AND ESSENTIAL PERFORMANCE,<br>IEC 60601-1-2: Medical electrical equipment - Part 1-2: General requirements for basic<br>safety and essential performance - Collateral standard: Electromagnetic compatibility<br>and ISO 10079-1 (2009): Medical suction equipment -- Part 1: Electrically powered<br>suction equipment -- Safety requirements.<br>All results of the testing met acceptance criteria.                                                                                                                                                                                                                                                                                           |
| Substantial Equivalence                         | The AS-XXXX Suction Pump described in this 510(k) submission are substantially<br>equivalent in all specifications and performance compared to the predicate device<br>identified in K112421 and do not raise any new types of safety and effectiveness<br>issues when compared to the predicate device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |

EXPORT DEPT:Rm. 4B-02/03, Taipei World Trade Center, 4/Fl.,#5, Hsin-Yi Road, Section 5, Taipei,Taiwan 11011 TEL:+886- 2- 2345- 1868 FAX : +886- 2- 2345- 3162 FACTORY : # 121, Sec. 3, Zhongshan Road, Dacun, Chang-Hwa, Taiwan 51542 TEL : +886- 4- 888- 1486 FAX : +886- 4- 888- 1499

- FACTORY : # 43, Shin- Kung 1st Road, Pei- Tou Industrial Park, Pei- Tou, Chang- Hwa, Taiwan 52149
	- TEL : +886- 4- 888- 1486 FAX : +886- 4- 888- 1499

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 28, 2014

Ding Hwa Company, Ltd. Mr. Robert O. Dean Compliance Systems International, LLC. 1083 Delaware Road Buffalo, New York 14223

Re: K140031

Trade/Device Name: Ding Hwa Co., Ltd - AS-XXXX Suction Pump (Family Series) Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: JCX Dated: March 26, 2014 Received: April 1, 2014

Dear Mr. Dean:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

Image /page/1/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a caduceus, composed of three curved lines.

{2}------------------------------------------------

Page 2 - Mr. Robert O. Dean

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.goy/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

## David Krause -S

for

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices · Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

510(k) Number (if known) 140031

Device Name

Ding Hwa Co., Ltd - AS-XXXX Suction Pump (Family Series)

Indications for Use (Describe)

Intended Use: AS-XXX Suction Pump is intended to be used to remove bodily fluids from a patient's airvay or respiratory system.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

# (

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (1/14)

Page 1 of 1

 ध PSC Particles Samices (10) | 43-4211)

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

g-et

---

**Source:** [https://fda.innolitics.com/device/K140031](https://fda.innolitics.com/device/K140031)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
