← Product Code [OQO](/productcode/OQO) · K140029

# AMIPLIVUE HSV 1&2 ASSAY (K140029)

_Quidel Corporation · OQO · Mar 26, 2014 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K140029

## Device Facts

- **Applicant:** Quidel Corporation
- **Product Code:** [OQO](/productcode/OQO.md)
- **Decision Date:** Mar 26, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.3305
- **Device Class:** Class 2
- **Review Panel:** Microbiology
- **Attributes:** Pediatric

## Indications for Use

The AmpliVue® HSV 1+2 Assay is an in vitro diagnostic test for the direct, qualitative detection and differentiation of Herpes Simplex Virus 1 (HSV-1) and Herpes Simplex Virus 2 (HSV-2) DNA in cutaneous or mucocutaneous lesion specimens from symptomatic patients. The test is intended for use as an aid in diagnosis of HSV infection in symptomatic patients. Warning: The AmpliVue® HSV 1+2 Assay is not FDA cleared for use with cerebrospinal fluid (CSF). The assay is not intended for prenatal screening.

## Device Story

In vitro diagnostic test for qualitative detection and differentiation of HSV-1 and HSV-2 DNA in cutaneous/mucocutaneous lesion specimens. Process: specimen dilution in transport medium; isothermal Helicase-Dependent Amplification (HDA) at 64°C for 45 minutes; target-specific hybridization probes; colorimetric detection via lateral-flow strip in a self-contained disposable cassette. Includes internal control to monitor inhibition/failure. Results visually read by clinician as colored lines. Aids diagnosis of HSV infection in symptomatic patients. Benefits include rapid, point-of-care-compatible molecular testing without complex instrumentation.

## Clinical Evidence

Clinical performance evaluated at five US sites with 1,355 specimens from symptomatic patients. Compared against ELVIS viral culture. For HSV-1: sensitivity 100% (cutaneous), 94.9% (mucocutaneous); specificity 97.1% (cutaneous), 96.5% (mucocutaneous). For HSV-2: sensitivity 98.3% (cutaneous), 97.4% (mucocutaneous); specificity 95.6% (cutaneous), 95.8% (mucocutaneous). Analytical studies included LoD (1.1x10^5 TCID50/mL for HSV-1; 1.1x10^4 TCID50/mL for HSV-2), cross-reactivity (89 microorganisms), and interference testing.

## Technological Characteristics

Isothermal HDA (Helicase-Dependent Amplification) at 64°C. Detection via lateral-flow strip with anti-DNP, anti-DIG, and anti-FAM antibodies. Kit includes dilution tubes, reaction tubes, and disposable detection cassettes. Qualitative visual readout. Internal control (IC) included for inhibition/failure monitoring. No external instrumentation required.

## Regulatory Identification

Herpes simplex virus serological assays are devices that consist of antigens and antisera used in various serological tests to identify antibodies to herpes simplex virus in serum. Additionally, some of the assays consist of herpes simplex virus antisera conjugated with a fluorescent dye (immunofluorescent assays) used to identify herpes simplex virus directly from clinical specimens or tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by herpes simplex viruses and provides epidemiological information on these diseases. Herpes simplex viral infections range from common and mild lesions of the skin and mucous membranes to a severe form of encephalitis (inflammation of the brain). Neonatal herpes virus infections range from a mild infection to a severe generalized disease with a fatal outcome.

## Special Controls

*Classification.* Class II (special controls). The device is classified as class II (special controls). The special control for the device is FDA's revised guidance document entitled “Class II Special Controls Guidance Document: Herpes Simplex Virus Types 1 and 2 Serological Assays.” For availability of the guidance revised document, see § 866.1(e).

## Predicate Devices

- IsoAmp® HSV Assay ([K111951](/device/K111951.md))

## Reference Devices

- ELVIS® HSV ID/Typing Test System ([K971662](/device/K971662.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:
K140029

B. Purpose for Submission:
Clearance of New Device

C. Measurand:
Target DNA sequences from Herpes Simplex Virus type 1 (HSV-1) and Herpes Simplex Virus type 2 (HSV-2)

D. Type of Test:
An *in vitro* molecular diagnostic test for the direct, qualitative detection and differentiation of HSV-1 and HSV-2 DNA in cutaneous or mucocutaneous lesion specimens

E. Applicant:
Quidel Corporation

F. Proprietary and Established Names:
AmpliVue® HSV 1+2 Assay

G. Regulatory Information:
1. Regulation section: 21 CFR 866.3305
2. Classification: Class II
3. Product code: OQO
4. Panel: Microbiology (83)

H. Intended Use:
1. Intended use(s):
The AmpliVue® HSV 1+2 Assay is an *in vitro* diagnostic test for the direct, qualitative detection and differentiation of Herpes Simplex Virus 1 (HSV-1) and Herpes Simplex

{1}

Virus 2 (HSV-2) DNA in cutaneous or mucocutaneous lesion specimens from symptomatic patients. The test is intended for use as an aid in diagnosis of HSV infection in symptomatic patients.

**Warning**: The AmpliVue® HSV 1+2 Assay is not FDA cleared for use with cerebrospinal fluid (CSF). The assay is not intended for prenatal screening.

2. **Indication(s) for use**:

Same as Intended Use

3. **Special conditions for use statement(s)**:

For prescription use only

4. **Special instrument requirements**:

N/A

I. **Device Description**:

The AmpliVue® HSV 1+2 Assay is an *in vitro* diagnostic test for the qualitative detection and differentiation of HSV-1 and HSV-2 DNA isolated from cutaneous or mucocutaneous lesion specimens collected with swabs from symptomatic patients. The assay utilizes helicase-dependent amplification (HDA) for the amplification of highly conserved HSV-1 and HSV-2 gene sequences and a self-contained disposable amplification detection device that allows for visual evaluation of assay results without analytical instrumentation.

The AmpliVue® HSV 1+2 Assay consists of three major steps: 1) specimen preparation, 2) amplification of target gene sequences specific to HSV-1 and HSV-2 by HDA, and 3) detection of the amplified DNA by target-specific hybridization probes by a colorimetric reaction on a lateral-flow strip which is embedded in a self-contained disposable cassette to prevent amplicon contamination.

The AmpliVue® HSV 1+2 Assay kit contains 16 tests per kit. The components are provided in a single box and are: 16 each of Dilution Tubes, Reaction Tubes, Detection Cassettes, and Amplicon Cartridges.

J. **Substantial Equivalence Information**:

1. **Predicate device name(s)**:

IsoAmp® HSV Assay (Biohelix)

{2}

Reference Method:

ELVIS® HSV ID/Typing Test System (Diagnostic Hybrid, Inc.) for clinical evaluation (K971662)

2. Predicate 510(k) number(s):

K111951

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Features | Quidel Corporation AmpliVue® HSV 1+2 Assay (K140029) | BioHelix Corporation IsoAmp® HSV Assay (K111951)  |
|  Intended Use | The AmpliVue® HSV 1+2 Assay is an in vitro diagnostic test for the direct, qualitative detection and differentiation of Herpes Simplex Virus 1 (HSV-1) and Herpes Simplex Virus 2 (HSV-2) DNA in cutaneous or mucocutaneous lesion specimens from symptomatic patients. The test is intended for use as an aid in diagnosis of HSV infection in symptomatic patients. | The IsoAmp® HSV Assay is an in vitro diagnostic test for the direct, qualitative detection of the Herpes Simplex Virus (HSV-1 & HSV-2) DNA in male and female genital and oral lesions. The test is intended for use as an aid in diagnosis of HSV infection in symptomatic patients.  |
|  Assay Results | Qualitative/ Visual colored band | Qualitative/ Visual colored band  |
|  Detection of HSV-1 & HSV-2 | Yes | Yes  |
|  Methodology | HDA (Helicase-Dependent Amplification) | HDA (Helicase-Dependent Amplification)  |

{3}

|  Differences  |   |   |
| --- | --- | --- |
|  Features | Quidel Corporation AmpliVue® HSV 1+2 Assay (K140029) | BioHelix Corporation IsoAmp® HSV Assay (K111951)  |
|  Typing of HSV-1 & HSV-2 | Yes | No  |
|  Packaging | The product is supplied as a kit in one box; 16 tests per kit 1. Amplification-related Kit Components (ARKC) 2. Non-amplification related Kit Components (NKC) | The product is supplied as two separate labeled boxes; 50 tests per/kit 1. Amplification-related Kit Components (ARKC) 2. Non-amplification related Kit Components (NKC)  |
|  Kit Reagent Storage Conditions | ARKC: 2°C to 8°C NKC: 2°C to 30°C | ARKC: <-15°C NKC: 15-30°C  |

# K. Standard/Guidance Document Referenced (if applicable): N/A

# L. Test Principle:

The AmpliVue® HSV 1+2 Assay consists of three major steps: 1) specimen preparation, 2) amplification of target gene sequences specific to HSV-1 and HSV-2 by HDA, and 3) detection of the amplified DNA by target-specific hybridization probes via a colorimetric reaction on a lateral-flow strip which is embedded in a self-contained disposable cassette to prevent amplicon contamination.

Specimen preparation involves one dilution step in which specimens in viral transport medium are diluted 80-fold in Dilution Tubes. The diluted samples are transferred into a 0.2 mL Amplification Tube containing lyophilized HDA reagents. Incubation at  $64^{\circ}\mathrm{C}$  for 45 minutes results in the release of the HSV DNA and subsequent isothermal amplification of the target sequence. The amplified DNA is detected by a set of specific detection probes included in the Amplification Tube: HSV-1 target hybridizes to two specific probes, one labeled with Biotin (BioTEG) and the other with Digoxigenin (DIG), while HSV-2 target hybridizes to two specific probes labeled with Biotin (BioTEG) and 6-carboxyfluorescein (6-FAM). A competitive internal control (IC) is included in the Amplification Tube to monitor for inhibition, reagent failure or device failure. The IC target is amplified by HSV-2 specific primers and hybridizes to the biotin-labeled HSV-2 probe and an IC specific probe labeled with 2, 4-dinitrophenyl (DNP-TEG).

Detection of the amplified DNA with specific probes is achieved by AmpliVue cassettes. The self-contained AmpliVue cassettes carry lateral-flow DNA detection strips coated with anti-DNP antibodies (C line), anti-DIG antibodies (T1 line) and anti-FAM antibodies (T2 line). HSV-1 amplicon complexed with BioTEG and DIG-labeled probes is captured by anti-DIG antibodies at the T1 line and HSV-2 amplicon complexed with BioTEG and FAM-labeled probes is captured by anti-FAM antibodies at the T2 line, while the IC amplicon complexed

{4}

with BioTEG and DNP-labeled probes is captured by anti-DNP antibodies at the C line. The biotin in the amplicon-probe complexes captures the streptavidin-conjugated color particles for visualization and the test result is shown as colored lines that are visually read.

## Interpretation of the assay results:

- Positive: Always read the Test lines (T1 and T2) first.
- When T1 line is visible, report the assay result as “HSV-1 DNA detected”.
- When T2 line is visible, report the assay result as “HSV-2 DNA detected”.
- When T1 and T2 are visible, report the assay result as “HSV-1 and HSV-2 Positive: HSV-1 and HSV-2 DNA detected”

- Negative: When no visible T line is present (T1-/T2-), a visible C line indicates that the Internal Control DNA has been amplified and detected, eliminating the possibility of a false negative due to failure of amplification or device, and thus the assay result should be reported as negative - “no HSV DNA detected”. The C line intensity may vary with each test. Any pink to red colored visible line in the control signifies a valid test.

- Invalid: If T and C lines are not present (T1- /T2-/C-), then the assay is invalid and the test needs to be repeated.

## Visual Interpretation of Assay Results

![img-0.jpeg](img-0.jpeg)

{5}

The interpretation of the assay results is carried out according to the following criteria:

|  Test line 1 (T1) Reading | Test line 2 (T2) Reading | Control line (C) Reading | Interpretation of result  |
| --- | --- | --- | --- |
|  T1+ | T2- | C+ or C- | HSV-1 Positive: HSV-1 DNA detected  |
|  T1- | T2+ | C+ or C- | HSV-2 Positive: HSV-2 DNA detected  |
|  T1+ | T2+ | C+ or C- | HSV-1 and HSV-2 Positive: HSV-1 and HSV-2 DNA detected  |
|  T1- | T2- | C+ | Negative: No HSV-1 or HSV-2 DNA detected  |
|  T1- | T2- | C- | Invalid: Failure due to inhibitory specimen, reagent failure, or device failure. Repeat test with original specimen.  |
|  Note: The absence of a C line (control) in conjunction with a positive test line (T1, T2 or T1 and T2) means that target material was successfully amplified. This occurs because of the over abundance of amplicons that generates competition with the internal control target.  |   |   |   |

# M. Performance Characteristics (if/when applicable):

# 1. Analytical performance:

a. Reproducibility: The reproducibility of the AmpliVue® HSV 1+2 Assay was evaluated at three test sites using a panel consisting of four panel members: HSV 1+2 High Negative; HSV-1 Low Positive; HSV 1+2 Moderate Positive; and HSV-2 Low Positive members. The HSV-1 Low Positive member served as a HSV-2 Negative member and the HSV-2 Low Positive member served as a HSV-1 Negative member. The panel members were prepared in HSV Negative Matrix that consisted of a pool of HSV negative cheek swabs in M4 medium. HSV Negative Matrix was spiked with quantified HSV-1 and HSV-2 stocks at pre-determined  $\mathrm{TCID}_{50}$  concentrations. The HSV-1 and HSV-2 stocks were diluted in the HSV Negative Matrix to three (3) different concentration levels, defined as High Negative (0.3 x LoD), Low Positive (1 x LoD) and Moderate Positive (3 x LoD level).

Each run tested the four member panel in triplicate and also included three each of HSV-1 positive control, HSV-2 positive control and negative control. The external positive controls used were HSV-1 and HSV-2 plasmid DNA. Two (2) operators per test site each carried out one run of the four member panel plus controls per test day for five (5) days using one lot of the AmpliVue® HSV 1+2 Assay.

Results of the reproducibility study for the AmpliVue® HSV 1+2 Assay performed at three sites are presented in the tables below.

{6}

|  Reproducibility Study Summary for HSV-1  |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Category | Site |   |   |   |   |   | Rate of Detection | Overall Percent Agreement | 95% Confidence Interval  |
|   |  Site #1 |   | Site #2 |   | Site #3  |   |   |   |   |
|   |  Rate of Detection | Percent Agreement | Rate of Detection | Percent Agreement | Rate of Detection | Percent Agreement  |   |   |   |
|  HSV 1+2 High Negative | 16/30 | 47% | 9/30 | 70% | 20/30 | 33% | 45/90 | 50% | 40% - 60%  |
|  HSV-1 Low Positive | 30/30 | 100% | 29/30 | 97% | 30/30 | 100% | 89/90 | 99% | 94% - 100%  |
|  HSV 1+2 Moderate Positive | 30/30 | 100% | 30/30 | 100% | 30/30 | 100% | 90/90 | 100% | 96% - 100%  |
|  HSV-2 Low Positive | 0/30 | 100% | 0/30 | 100% | 0/30 | 100% | 0/90 | 100% | 96% - 100%  |
|  HSV-1 Positive Control | 30/30 | 100% | 30/30 | 100% | 30/30 | 100% | 90/90 | 100% | 96% - 100%  |
|  Negative Control | 0/30 | 100% | 0/30 | 100% | 0/30 | 100% | 0/90 | 100% | 96% - 100%  |
|  Reproducibility Study Summary for HSV-2  |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Category | Site |   |   |   |   |   | Rate of Detection | Overall Percent Agreement | 95% Confidence Interval  |
|   |  Site #1 |   | Site #2 |   | Site #3  |   |   |   |   |
|   |  Rate of Detection | Percent Agreement | Rate of Detection | Percent Agreement | Rate of Detection | Percent Agreement  |   |   |   |
|  HSV 1+2 High Negative | 20/30 | 33% | 17/30 | 43% | 13/30 | 57% | 50/90 | 44% | 35% - 55%  |
|  HSV-2 Low Positive | 30/30 | 100% | 30/30 | 100% | 30/30 | 100% | 90/90 | 100% | 96% - 100%  |
|  HSV 1+2 Moderate Positive | 30/30 | 100% | 30/30 | 100% | 30/30 | 100% | 90/90 | 100% | 96% - 100%  |
|  HSV-1 Low Positive | 0/30 | 100% | 0/30 | 100% | 0/30 | 100% | 0/90 | 100% | 96% - 100%  |
|  HSV-2 Positive Control | 30/30 | 100% | 30/30 | 100% | 30/30 | 100% | 90/90 | 100% | 96% - 100%  |
|  Negative Control | 0/30 | 100% | 0/30 | 100% | 0/30 | 100% | 0/90 | 100% | 96% - 100%  |

{7}

# Precision:

The repeatability study used the same four member panel as used for the reproducibility study. As for the reproducibility study, the HSV-1 Low Positive member served as a HSV-2 Negative member and the HSV-2 Low Positive member served as a HSV-1 Negative member. The HSV-1 and HSV-2 stocks were diluted in the HSV Negative Matrix to three (3) different concentration levels, defined as High Negative  $(0.3 \times \mathrm{LoD})$ , Low Positive  $(1 \times \mathrm{LoD})$ , and Moderate Positive  $(3 \times \mathrm{LoD}$  level).

Each run tested the panel of four members in triplicate by two (2) operators, twice a day (2X) for twelve (12) days on all three instruments (triplicate testing x 2 operators x 12 days = 72 results per level for each virus). Positive and negative controls were run in triplicate with each test run. The external positive controls used were HSV-1 and HSV-2 plasmid DNA. Results of the Repeatability study for the AmpliVue® HSV 1+2 Assay performed at one sites are presented in the tables below.

|  Repeatability Study Summary for HSV-1  |   |   |   |
| --- | --- | --- | --- |
|  Category | Rate of Detection | Overall Percent Agreement | 95% Confidence Interval  |
|  HSV 1+2 High Negative | 35/72 | 51% | 40 – 63%  |
|  HSV-1 Low Positive | 72/72 | 100% | 95 – 100%  |
|  HSV 1+2 Moderate Positive | 72/72 | 100% | 95 – 100%  |
|  HSV-2 Low Positive | 0/72 | 100% | 95 – 100%  |
|  HSV-1 Positive Control | 72/72 | 100% | 95 – 100%  |
|  Negative Control | 0/72 | 100% | 95 – 100%  |
|  Repeatability Study Summary for HSV-2  |   |   |   |
| --- | --- | --- | --- |
|  Category | Rate of Detection | Overall Percent Agreement | 95% Confidence Interval  |
|  HSV 1+2 High Negative | 43/72 | 40% | 30 – 52%  |
|  HSV-2 Low Positive | 72/72 | 100% | 95 – 100%  |
|  HSV 1+2 Moderate Positive | 72/72 | 100% | 95 – 100%  |
|  HSV-1 Low Positive | 0/72 | 100% | 95 – 100%  |
|  HSV-2 Positive Control | 72/72 | 100% | 95 – 100%  |
|  Negative Control | 0/72 | 100% | 95 – 100%  |

b. Linearity/assay reportable range: N/A

{8}

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

## Internal Control

The competitive internal control (IC) consists of plasmid DNA. The IC target sequence is amplified using the same primer set that amplifies the HSV-2 target sequence. The internal sequence of the IC target is different from the HSV target sequences and is detected by an IC-specific probe. After amplification, the IC amplicon-probe complexes are detected as a visible Control line by anti-DNP antibodies. The IC DNA and probes are pre-mixed in the Amplification Reagent.

## External Controls

The assay Positive Controls are commercially available external positive HSV-1 and HSV-2 from patient specimens. The controls are intended to monitor reagent and cassette failure. Previously characterized positive HSV-1 and HSV-2 specimens may be used in lieu of commercial HSV-1 and HSV-2 controls. The assay Negative Control consists of blank viral transport medium or previously characterized negative specimen and is used to detect reagent or environmental contamination (or carry-over) by either HSV DNA or amplicons.

**Note**: For the analytical studies, HSV-1 and HSV-2 plasmid DNA were used as the positive external controls. The plasmid DNA contains HSV-1 or HSV-2 target sequences. For the clinical studies, positive HSV-1 and HSV-2 patient specimens were used as the external controls.

d. Detection limit:

A Limit of Detection (LoD) study was performed to evaluate the analytical sensitivity of the AmpliVue® HSV 1+2 Assay using two representative viral strains of HSV-1 (McIntyre &amp; HF) and two representative strains of HSV-2 (G &amp; MS). Quantified (TICD₅₀/mL) cultures of the HSV-1 and HSV-2 strains were serially diluted to five (5) concentrations in HSV-negative matrix pools and tested in replicates of ten (10) with three reagent lots. The observed LoD of a HSV strain was determined as the lowest concentration level that had a positivity rate of ≥95%. The observed LoDs for HSV-1 and HSV-2 were determined to be 1.1 × 10⁵ TCID₅₀/mL and 1.1 × 10⁴ TCID₅₀/mL, respectively.

9

{9}

Results from HSV LoD Panels

|  Strain | Concentration TCID50/mL | Positive/Total | Positivity rate | 95% CI  |   |
| --- | --- | --- | --- | --- | --- |
|  HSV-1 McIntyre | 1.00 x106 | 30/30 | 100.0% | 88.65% | 100.00%  |
|   |  3.33 x106 | 30/30 | 100.0% | 88.65% | 100.00%  |
|   |  1.1 x105 | 30/30 | 100.0% | 88.65% | 100.00%  |
|   |  3.70 x104 | 24/30 | 80.0% | 62.69% | 90.50%  |
|   |  1.23 x104 | 8/30 | 26.7% | 14.18% | 44.45%  |
|  HSV-1 HF | 1.00 x106 | 30/30 | 100.0% | 88.65% | 100.00%  |
|   |  3.33 x105 | 30/30 | 100.0% | 88.65% | 100.00%  |
|   |  1.11 x105 | 30/30 | 100.0% | 88.65% | 100.00%  |
|   |  3.70 x104 | 23/30 | 76.7% | 59.07% | 88.21%  |
|   |  1.23 x104 | 9/10 | 30.0% | 16.66% | 47.88%  |
|  HSV-2 G | 1.00 x105 | 30/30 | 100.0% | 88.65% | 100.00%  |
|   |  3.33 x104 | 30/30 | 100.0% | 88.65% | 100.00%  |
|   |  1.11 x104 | 29/30 | 100.0% | 88.30% | 100.00%  |
|   |  3.70 x103 | 28/30 | 93.3% | 78.68% | 98.15%  |
|   |  1.23 x103 | 24/30 | 73.3% | 55.55% | 85.82%  |
|  HSV-2 MS | 1.00 x105 | 30/30 | 100.0% | 88.65% | 100.00%  |
|   |  3.33 x104 | 30/30 | 100.0% | 88.65% | 100.00%  |
|   |  1.11 x104 | 30/30 | 100.0% | 88.65% | 100.00%  |
|   |  3.70 x103 | 30/30 | 100.0% | 88.65% | 100.00%  |
|   |  1.23 x103 | 27/30 | 90.0% | 74.38% | 96.54%  |

The LoDs determined in the initial study were confirmed with the same two HSV-1 and two HSV-2 reference strains diluted to the observed LODs (1.1 x  $10^{5}$  TCID $_{50}$ /mL for HSV-1 and 1.1 x  $10^{4}$  TCID $_{50}$ /mL for HSV-2) and tested with twenty (20) replicates using three (3) lots of the AmpliVue HSV 1+2 Assay. Since all HSV-1 and HSV-2 strains showed positivity rates of  $\geq 95\%$  with all three (3) validation lots, the observed LoDs were confirmed for both HSV-1 and HSV-2.

Clinical Isolates Testing: In addition, twenty (20) HSV-1 and twenty (20) HSV-2 clinical isolates were cultured and quantified in  $\mathrm{TCID}_{50} / \mathrm{mL}$ . Each isolate was diluted to the corresponding LoD in HSV-negative matrix and tested in triplicate. The AmpliVue HSV 1+2 Assay was able to detect all 20 HSV-1 and 20 HSV-2 clinical isolates at their corresponding LoD.

Assay LoD: The final assay LoD claim is  $1.1 \times 10^{5} \mathrm{TCID}_{50} / \mathrm{mL}$  for HSV-1 and  $1.1 \times 10^{4} \mathrm{TCID}_{50} / \mathrm{mL}$  for HSV-2.

{10}

# e. Analytical specificity:

A study was performed to evaluate the performance of the AmpliVue® HSV 1+2 Assay in the presence of eighty-nine (89) microorganisms that might be found in cutaneous or mucocutaneous lesion specimens. The panel members were obtained from suppliers as purified genomic DNA (GD) or quantified cultures (QC), or prepared in house (IHC) by growing each organism and quantifying the culture. Each potentially interfering or cross-reactive microorganism was tested in three (3) replicates in negative matrix. Clinically relevant levels of viruses and bacteria are typically  $10^{6}\mathrm{cfu / ml}$  or higher for bacteria and  $10^{5}\mathrm{pfu / ml}$  or higher for viruses. Purified and quantified DNA was used for nine (9) of the microorganisms. For these microorganisms,  $10^{6}$  copies per ml (cp/ml) or higher was used. None of the eighty-nine (89) microorganisms showed cross-reactivity with the assay.

Cross Reactivity Panel

|  Microorganism | Member Type (GD, QC, IHC) | Test concentration  |
| --- | --- | --- |
|  Bacteria (N=52)  |   |   |
|  Acholeplasma laidlawi PG8 | QC | 7.1 x 10^6 cfu/mL  |
|  Acinetobacter calcoaceticus | QC | 9.80 x 10^6 cfu/mL  |
|  Acinetobacter lwoffii | IHC | 4.55 x 10^6 cfu/mL  |
|  Bacteroides fragilis Z029 | QC | 8.8 x 10^6 cfu/mL  |
|  Bordetella bronchiseptica | QC | 1.17 x 10^6 cfu/mL  |
|  Bordetella pertussis E431 | QC | 1.73 x 10^6 cfu/mL  |
|  Chlamydia trachomatis VR-347 | QC | 3.00 x 10^6 cfu/mL  |
|  Chlamydia trachomatis D-UW3 | QC | 7.83 x 10^7 IFU/mL  |
|  Chlamydia trachomatis LGV-II 434 DNA | GD | 1.5 x 10^7 cp/mL  |
|  Chlamydophila pneumoniae DNA | GD | 1.6 x 10^6 cp/mL  |
|  Clostridium difficile NAP1 | QC | 6.77 x 10^6 cfu/mL  |
|  Clostridium perfringens Type A | QC | 1.06 x 10^6 cfu/mL  |
|  Corynebacterium diphtheriae | QC | 3.44 x 10^6 cfu/mL  |
|  Enterobacter cloacae Z101 | QC | 5.70 x 10^6 cfu/mL  |
|  Enterococcus faecalis VSE | QC | 8.60 x 10^6 cfu/mL  |
|  Escherichia coli ATCC 43895 | QC | 1.13 x 10^6 cfu/mL  |
|  Fusobacterium nucleatum | IHC | 8.05 x 10^6 cfu/mL  |
|  Gardnerella vaginalis | QC | 1.20 x 10^6 cfu/mL  |
|  Haemophilus influenzae type A | QC | 4.00 x 10^6 cfu/mL  |
|  Haemophilus ducreyi Class I DNA | GD | 2.97 x 10^6 cp/mL  |
|  Haemophilus influenzae type B | GD | 2.97 x 10^6 cp/mL  |
|  Haemophilus influenzae type C | GD | 2.97 x 10^6 cp/mL  |
|  Haemophilus influenzae type D | GD | 2.97 x 10^6 cp/mL  |
|  Haemophilus influenzae type E | GD | 2.97 x 10^6 cp/mL  |
|  Haemophilus influenzae type F | GD | 2.97 x 10^6 cp/mL  |

{11}

|  Microorganism | Member Type (GD, QC, IHC) | Test concentration  |
| --- | --- | --- |
|  Klebsiella pneumoniae | QC | 9.75 x 10^6 cfu/mL  |
|  Lactobacillus acidophilus Z048 | QC | 2.00 x 10^6 cfu/mL  |
|  Legionella pneumophila | QC | 1.42 x 10^6 cfu/mL  |
|  Mobiluncus curtisii V125 [DSM 2711] ATCC 43063 | QC | 3.2 x 10^6 cfu/mL  |
|  Mobiluncus mulieris ATCC 35240 | QC | 1.76 x 10^6 cfu/mL  |
|  Moraxella cartarrhalis Ne11 | QC | 9.90 x 10^6 cfu/mL  |
|  Mycoplasma hominis LBD-4 | QC | 1.30 x 10^6 cfu/mL  |
|  Mycoplasma hyorhinis BTS-7 | QC | 6.6 x 10^6 cfu/mL  |
|  Mycoplasma orale CH 19299 | QC | 3.08 x 10^6 cfu/mL  |
|  Mycoplasma pneumoniae M129 | QC | 3.16 x 10^6 CCU^1/mL ('Color Changing Units)  |
|  Mycoplasma salivarium H110 | QC | 1.67 x 10^6 cfu/mL  |
|  Neisseria gonorrhoeae Z017 | QC | 5.73 x 10^6 cfu/mL  |
|  Neisseria meningitidis SerogroupA | QC | 7.07 x 10^6 cfu/mL  |
|  Prevotella melaninogenica ATCC 25845 | QC | 7.3 x 10^6 cfu/mL  |
|  Proteus mirabilis | QC | 1.19 x 10^6 cfu/mL  |
|  Pseudomonas aeruginosa | QC | 1.32 x 10^6 cfu/mL  |
|  Salmonella enteriditis | QC | 5.40 x 10^6 cfu/mL  |
|  Salmonella typhimurium | QC | 4.60 x 10^6 cfu/mL  |
|  Staphylococcus aureus MRSA | IHC | 7.52 x 10^6 cfu/mL  |
|  Staphylococcus aureus MSSA | IHC | 7.02 x 10^6 cfu/mL  |
|  Staphylococcus epidermidis MRSE | IHC | 1.75 x 10^6 cfu/mL  |
|  Staphylococcus saprophyticus | QC | 3.00 x 10^6 cfu/mL  |
|  Streptococcus agalactiae | QC | 2.20 x 10^6 cfu/mL  |
|  Streptococcus mitis | QC | 2.43 x 10^6 cfu/mL  |
|  Streptococcus mutans Z072 | QC | 1.17 x 10^6 cfu/mL  |
|  Streptococcus pneumonia | QC | 2.8 x 10^6 cfu/mL  |
|  Streptococcus pyogenes ATCC 9898 | QC | 6.38 x 10^6 cfu/mL  |
|  Streptococcus salivarius | IHC | 2.75 x 10^6 cfu/mL  |
|  Toxoplasma gondii | QC | 6.6 x 10^6 tachyzoites/mL  |
|  Toxoplasma hyopneumoniae | QC | 1.0 x 10^6 cfu/mL  |
|  Toxoplasma hyorhinis | QC | 1.0 x 10^6 cfu/mL  |
|  Toxoplasma pneumoniae | QC | 1.0 x 10^6 cfu/mL  |
|  T. aerogenes ATCC 10001 | QC | 1.0 x 10^6 cfu/mL  |
|  T. aerogenes ATCC 10002 | QC | 1.0 x 10^6 cfu/mL  |
|  T. aerogenes ATCC 10003 | QC | 1.0 x 10^6 cfu/mL  |
|  T. aerogenes ATCC 10004 | QC | 1.0 x 10^6 cfu/mL  |
|  T. aerogenes ATCC 10005 | QC | 1.0 x 10^6 cfu/mL  |
|  T. aerogenes ATCC 10006 | QC | 1.0 x 10^6 cfu/mL  |
|  T. aerogenes ATCC 10007 | T. aerogenes | 1.0 x 10^6 cfu/mL  |
|  T. aerogenes ATCC 10008 | T. aerogenes | 1.0 x 10^6 cfu/mL  |
|  T. aerogenes ATCC 10009 | T. aerogenes | 1.0 x 10^6 cfu/mL  |
|  T. aerogenes ATCC 10010 | T. aerogenes | 1.0 x 10^6 cfu/mL  |
|  T. aerogenes ATCC 10011 | T. aerogenes | 1.0 x 10^6 cfu/mL  |
|  T. aerogenes ATCC 10012 | T. aerogenes | 1.0 x 10^6 cfu/mL  |

{12}

|  Microorganism | Member Type (GD, QC, IHC) | Test concentration  |
| --- | --- | --- |
|  Treponema pallidum Nichols | QC | 2.0 x 10^{6}
Treponema pallidum/mL  |
|  Trichomonas vaginalis Z070 | QC | 1.65 x 10^{6}
trophzoites/mL  |
|  Ureaplasma uralyticum NCTC 10177 DNA | GD | 1.23 x 10^{6} cp/mL  |
|  **Yeast (N=7)**  |   |   |
|  Candida albicans | QC | 2.00 x 10^{6} cfu/mL  |
|  Candida glabrata Z007 | QC | 9.73 x 10^{6} cfu/mL  |
|  Candida guilliermondii Z008 | QC | 9.96 x 10^{6} cfu/mL  |
|  Candida krusei Z009 | QC | 5.33 x 10^{6} cfu/mL  |
|  Candida lusitaniae Z010 | QC | 6.56 x 10^{6} cfu/mL  |
|  Candida parapsilosis Z011 | QC | 1.24 x 10^{6} cfu/mL  |
|  Candida tropicalis Z012 | QC | 1.0 x 10^{6} cfu/mL  |
|  **Virus (N=30)**  |   |   |
|  Influenza A/Mexico/4108/2009 H1N1 | QC | 4.08 x 10^{6} TCID_{50}/mL  |
|  Adenovirus 2 | QC | 1.02 x 10^{5} TCID_{50}/mL  |
|  Adenovirus 7 VR-7 | QC | 1.58 x 10^{5} TCID_{50}/mL  |
|  Coronavirus OC43 VR-1558 | QC | 2.42 x 10^{5} TCID_{50}/mL  |
|  Coxsackievirus B4 | QC | 1.08 x 10^{5} TCID_{50}/mL  |
|  Cytomegalovirus AD-169 | QC | 9.55 x 10^{5} TCID_{50}/mL  |
|  Echovirus 11 ODH-37285 | QC | 2.14 x 10^{5} TCID_{50}/mL  |
|  Enterovirus Type 71 | QC | 1.00 x 10^{5} TCID_{50}/mL  |
|  Epstein-Barr Virus B95-8 | GD | 2.22 x 10^{5} cp/mL  |
|  Influenza B Hong Kong VR-791 | QC | 9.53 x 10^{6} TCID_{50}/mL  |
|  Hepatitis B Virus | QC | 3.44 x 10^{5} IU/mL  |
|  Hepatitis C Virus | QC | 7.58 x 10^{5} IU/mL  |
|  HHV-8 | QC | 1.26 x 10^{5} TCID_{50}/mL  |
|  HIV-1 Subtype B RNA | GD | 1.14 x 10^{5} cp/mL  |
|  hMPV (Italy) A1 | QC | 3.66 x 10^{5} TCID_{50}/mL  |
|  Human Herpes 6 virus Z29 strain | QC | 1.95 x 10^{5} TCID_{50}/mL  |
|  Human Herpes 7 virus SB strain | QC | 1.15 x 10^{5} TCID_{50}/mL  |
|  Human papilloma virus 16 DNA | GD | 4.3 x 10^{5} cp/mL  |
|  Human papilloma virus 18 DNA | GD | 1.8 – 3.6 x 10^{5} cp/mL  |

{13}

|  Microorganism | Member Type (GD, QC, IHC) | Test concentration  |
| --- | --- | --- |
|  Measles virus | QC | 1.95 x 105TCID50/mL  |
|  Mumps virus | QC | 5.89 x 105TCID50/mL  |
|  Parainfluenza Type 1 #2 | QC | 3.97 x 105TCID50/mL  |
|  Parainfluenza Type 2 | QC | 3.15 x 105TCID50/mL  |
|  Parainfluenza Type 3 NY14 | QC | 2.36 x 105TCID50/mL  |
|  Parainfluenza Type 4B VR-1377 | QC | 1.37 x 105TCID50/mL  |
|  RSV A Long VR-26 | QC | 4.36 x 104TCID50/mL  |
|  RSV B Washington VR-1401 | QC | 3.43 x 105TCID50/mL  |
|  Rubella virus | QC | 4.17 x 105TCID50/mL  |
|  Simian Virus type 40 Pa-57 ATCC strain VR-239 | QC | 3.16 x 105TCID50/mL  |
|  VZV DNA | GD | 1.5 x 105cp/mL  |

# f. Interference Studies

This study was performed to evaluate potential interference with the AmpliVue® HSV 1+2 Assay with a panel of thirty-three (33) substances, five different viral transport media, and the 87 microorganisms from the cross reactivity panel that may be present in clinical specimens. The study was carried out in the presence of HSV-1 and HSV-2 at 3 times the limit of detection (3 x LoD) to evaluate potential interference with the detection of the HSV targets. The study was also carried out in the absence of HSV to evaluate potential interference with the detection of the internal control of the AmpliVue® HSV 1+2 Assay. Each potential interfering substance was tested in triplicate.

Interfering Substances: The analytical performance of the AmpliVue® HSV 1+2 Assay was characterized in the presence of interfering substances at the potentially highest ("the worst case") concentrations to evaluate the susceptibility of the HSV assay to interference. Each interfering substance was introduced into the assay by directly wetting a clean, dry Remel M4 kit swab with the substance and placing the swab directly in transport medium. Calculated concentrations are based on an estimated volume of  $200\mu \mathrm{L}$  of substance introduced by the swab. Each panel member was tested in triplicate spiked with HSV-1 HF and HSV-2 G strains separately at  $3\times$  LoD. The panel was also tested in triplicate in the absence of HSV transport medium to see if the potentially interfering substances interfere with the detection of the internal control. No interference was observed in the presence of the potential interfering substances tested.

{14}

15

Interfering Substance Panel

|  Substance | Test conc.  |
| --- | --- |
|  Seminal Fluid | 7%  |
|  Cornstarch | 1.25 mg/mL  |
|  Acetamidophenol | 5 mg/mL  |
|  Feces | 7%  |
|  Acetylsalicylic Acid | 10 mg/mL  |
|  Chlorpheniramine | 5 mg/mL  |
|  Dextromethorphan | 10 mg/mL  |
|  Whole blood with EDTA | 7%  |
|  Female Urine | 7%  |
|  Male Urine | 7%  |
|  Acyclovir (Acycloguanosine) | 7 mg/mL  |
|  Albumin | 3.3 mg/mL  |
|  Casein | 7 mg/mL  |
|  K-Y Brand Jelly | 7%  |
|  Douche | 7%  |
|  Monistat 1 | 7%  |
|  Monistat 3 | 7%  |
|  Tioconazole 1 | 7%  |
|  Preparation H | 7%  |
|  Lanacane | 7%  |
|  Listerine | 7%  |
|  Abreva | 7%  |
|  Carmex Cold Sore Lip Balm | 7%  |
|  Releev cold sore treatment | 7%  |
|  Crest | 7%  |
|  Mucin (Bovine Submaxillary Gland, type I-S) | 60 μg/mL  |
|  Buffy coat | 7%  |
|  YeastGard | 7%  |
|  Vagisil Crème | 7%  |
|  Lip clear Lysine+ | 7%  |
|  Clotrimazole 3 Vaginal Cream | 7%  |
|  Balneol Hygienic Cleansing Lotion | 7%  |
|  Ortho Options Gynol II Extra Strength Vaginal Contraceptive Jelly | 7%  |

{15}

Viral Transport Media: The performance of the AmpliVue® HSV 1+2 Assay was assessed with the following viral transport media: Remel M4RT, Remel M5, Bartels VTM, and BD Universal Viral Transport (UVT)/Copan UTM. Remel M4 had previously been assessed and found to not interfere with the assay. Each transport medium was tested after spiking with HSV-1 HF and HSV-2 G strain to a final concentration of approximately 3 x LOD to determine if the viral transport medium interferes with the detection of HSV targets in positive samples. The media were tested in the absence of HSV-1 and HSV-2 (transport medium only) to see if the viral transport media interfere with the detection of the internal control in negative samples.

There was no interference observed with the Remel M4RT, Remel M5, Bartels VTM, and BD UVT/Copan UTM media for the detection of HSV-1 and HSV-2 target or the internal control. Remel M4RT, Remel M5, Bartels VTM, and BD UVT/Copan UTM did not interfere with the detection of HSV-1 and HSV-2 target or the internal control.

Cross-Reactivity Panel Members: The performance of the AmpliVue® HSV 1+2 Assay was characterized by testing the eighty-nine (89) microorganisms that were evaluated for cross reactivity in the presence of HSV-1 HF and HSV-2 G at 3 x LoD separately to see if the presence of any of these 89 organisms interferes with the detection of HSV targets. Each panel member was tested in triplicate. None of the cross reactivity panel members interfered with the detection of HSV-1 and HSV-2 target.

g. Specimen Stability

A study was performed to assess the stability of HSV-1 and HSV-2 in viral transport media in accordance with recommended storage and handling specifications of each medium tested. Based on the supporting data, HSV-1 and HSV-2 were stable in all five (5) media for 7 days at 2 - 8°C, and for 34 days at -70°C.

h. Competitive Inhibition

The performance of the AmpliVue® HSV 1+2 Assay was assessed for competitive interference using simulated samples in two studies mimicking co-infections. The first study used simulated samples with one target at a concentration near the LoD (3 x LOD) and the other target at higher concentrations (30 x LOD to 3000 x LOD). The second study used simulated samples that had equal concentrations of HSV-1 virus and HSV-2 virus (3 x LOD to 3000 x LOD).

In the first study, competitive inhibition was not observed with simulated samples containing one target at a concentration near the LOD (3 x LOD) and the other target at up to 300 x LOD. However, competitive inhibition was observed for both HSV-1 and HSV-2 with simulated samples containing one target at a concentration near the LOD (3 x LOD) and the other target at 3000 x LOD.

16

{16}

In the second study, competitive inhibition was not observed with simulated samples containing equal concentrations of HSV-1 virus and HSV-2 virus, from 3 x LOD to 3000 x LOD.

i. Carry-over/Cross Contamination

Test results confirm that carry-over/cross contamination does not occur with the AmpliVue® HSV 1+2 Assay. High HSV-1 (HSV-2) positive samples were tested in series alternating with negative samples. In order to challenge the device, cultured and quantified viral stock served as high positive sample. The HSV-1 HF (7.96 x 10⁸ TCID₅₀/mL) and HSV-2 G (2.27 x 10⁷ TCID₅₀/mL) stocks were used directly without dilution, for the highest concentration available. Remel M4 viral transport medium was used as the negative sample. Ten (10) replicates of negative sample together with assay controls were run by two (2) operators to confirm that negative samples (Remel M4 viral transport medium) generate a negative result 100% of the time. Five (5) replicates of high-concentration positive and negative samples were tested in a series with alternating sample types. All HSV-1 and HSV-2 high positive samples gave positive results and all the negative samples gave HSV negative results.

2. Comparison Studies:

a. Method comparison with predicate device:

The method comparison/clinical performance evaluation was done against a gold standard/reference method using the FDA cleared, ELVIS® HSV ID/Typing Test System (Diagnostic Hybrid, Inc.), i.e., cell culture using an enzyme linked virus inducible system with HSV typing by fluorescently labeled antibodies. The testing description and data are listed in the Clinical Studies Section below (Section 3).

b. Matrix comparison: N/A

3. Clinical Studies:

a. Clinical Sensitivity: N/A
b. Clinical Specificity: N/A
c. Other clinical supportive data (when a. and b. are not applicable):

The performance of the AmpliVue® HSV 1+2 Assay was compared with the ELVIS® HSV ID/Typing Test System (Diagnostic Hybrid, Inc.) which is a gold standard reference method i.e., Cell Culture using an enzyme linked virus inducible system with HSV typing by fluorescently labeled antibodies.

Clinical Performance

The performance of the AmpliVue® HSV 1+2 Assay was evaluated at five geographically diverse locations within the United States. A total of one thousand

17

{17}

three hundred and fifty-five (1,355) specimens from symptomatic male and female patients was tested. Three (3) cultures (0.2%) were unable to be read due to gross microbial contamination (there was insufficient quantity of each sample to repeat the culture). Four (4) specimens (0.3%) were positive for HSV in the ELVIS culture system, but the virus could not be identified as HSV-1 or HSV-2. Ten (10) specimens remained invalid upon repeat testing by the AmpliVue® HSV 1+2 Assay, and two (2) specimens were unavailable for re-testing. These nineteen (19) specimens were removed from performance analysis. The ages of the patient population ranged from 5 years to 60 years. The swab specimens were categorized as cutaneous (e.g., skin lesion, genital - penis), or mucocutaneous (e.g., anorectal, genital - vaginal/cervical, nares, ocular, oral lesion and urethral).

The reference ELVIS viral culture used in this study is unable to detect co-infected specimens and cannot identify HSV-1 if HSV-2 is identified first. Consequently, if a specimen was positive for HSV-2, it was removed from the calculation of the HSV-1 clinical performance. Two hundred and eleven (211) specimens identified as HSV-2 positive by ELVIS were removed from the initial 1,336 specimens. The tables below show 1,125 HSV-1 specimens from cutaneous and mucocutaneous lesions.

|  Combined Sites – HSV-1 Cutaneous Lesions (N=340)  |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Reference Method |   |   |   |   |   |  | 95% CI  |   |
|  AmpliVue HSV 1&2 Assay |  | POS | NEG | Total | Sensitivity | 100% | 88.6% | 100%  |
|   |  POS | 30 | 9 | 39 | Specificity | 97.1% | 94.6% | 98.5%  |
|   |  NEG | 0 | 301 | 301 |   |   |   |   |
|   |  Total | 30 | 310 | 340  |   |   |   |   |
|  Combined Sites – HSV-1 Mucocutaneous Lesions (N=785)  |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Reference Method |   |   |   |   |   |  | 95% CI  |   |
|  AmpliVue HSV 1&2 Assay |  | POS | NEG | Total | Sensitivity | 94.9% | 90.3% | 97.4%  |
|   |  POS | 149 | 22 | 171 | Specificity | 96.5% | 94.8% | 97.7%  |
|   |  NEG | 8 | 606 | 614 |   |   |   |   |
|   |  Total | 157 | 628 | 785  |   |   |   |   |

The tables below show the HSV-2 results for 1,336 specimens from cutaneous and mucocutaneous lesions.

{18}

|  Combined Sites – HSV-2 Cutaneous Lesions (N=399)  |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Reference Method |   |   |   |   |   |  | 95% CI  |   |
|  AmpliVue HSV 1&2 Assay |  | POS | NEG | Total | Sensitivity | 98.3% | 91.0% | 99.7%  |
|   |  POS | 58 | 15 | 73 | Specificity | 95.6% | 92.8% | 97.3%  |
|   |  NEG | 1 | 325 | 326 |   |   |   |   |
|   |  Total | 59 | 340 | 399  |   |   |   |   |
|  Combined Sites – HSV-2 Mucocutaneous Lesions (N=937)  |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Reference Method |   |   |   |   |   |  | 95% CI  |   |
|  AmpliVue HSV 1&2 Assay |  | POS | NEG | Total | Sensitivity | 97.4% | 93.4% | 99.0%  |
|   |  POS | 148 | 33 | 181 | Specificity | 95.8% | 94.2% | 97.0%  |
|   |  NEG | 4 | 752 | 756 |   |   |   |   |
|   |  Total | 152 | 785 | 937  |   |   |   |   |

4. Clinical cut-off: N/A

5. Expected values/Reference range:

The prevalence of HSV-1 and HSV-2 with the AmpliVue® HSV 1+2 Assay in cutaneous (e.g., skin lesion, genital - penis) or mucocutaneous (e.g., anorectal, genital - vaginal/cervical, nares, ocular, oral lesion and urethral) specimens was estimated for the 1,343 specimens with valid AmpliVue® HSV 1+2 Assay results. Seven of the 1,343 specimens that had valid AmpliVue® HSV 1+2 Assay results were not included in the performance analysis due to contamination or invalid results by the reference method; these seven samples were included in the calculations of prevalence

The prevalence of HSV-1 and HSV-2 with the AmpliVue® HSV 1+2 Assay was calculated for the combined sites based on the age of the patient and the specific source of specimen and the results are presented below.

{19}

|  Combined Study – Cutaneous Prevalence by Age  |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|   | HSV- 1 |   |   | HSV-2  |   |   |
|  Age | Total # | Total Positive | Prevalence | Total # | Total Positive | Prevalence  |
|  ≤ 5 years | 37 | 2 | 5.4% | 37 | 1 | 2.7%  |
|  6 to 21 years | 68 | 13 | 19.1% | 68 | 6 | 8.8%  |
|  22 to 59 years | 225 | 20 | 8.9% | 225 | 49 | 21.8%  |
|  ≥ 60 years | 70 | 5 | 7.1% | 70 | 18 | 25.7%  |
|   | % | 95% CI |   | % | 95% CI  |   |
|  Positive Predictive Value | 76.9% | 88.6% to 100% |   | 79.5% | 68.8% to 87.1%  |   |
|  Negative Predictive Value | 100% | 94.6% to 98.5% |   | 99.7% | 98.3% to 99.9%  |   |
|  Combined Study – Cutaneous Prevalence by Specific Source  |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|   | HSV- 1 |   |   | HSV-2  |   |   |
|  Source | Total # | Total Positive | Prevalence | Total # | Total Positive | Prevalence  |
|  Skin lesion | 271 | 27 | 10.0% | 271 | 48 | 17.7%  |
|  Genital - penis | 129 | 13 | 19.1% | 129 | 26 | 20.2%  |

{20}

|  Combined Study – Mucocutaneous Prevalence by Age  |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|   | HSV- 1 |   |   | HSV-2  |   |   |
|  Age | Total # | Total Positive | Prevalence | Total # | Total Positive | Prevalence  |
|  ≤ 5 years | 39 | 10 | 25.6% | 39 | 1 | 2.6%  |
|  6 to 21 years | 190 | 42 | 22.1% | 190 | 34 | 17.9%  |
|  22 - 59 years | 606 | 104 | 17.2% | 606 | 132 | 21.8%  |
|  ≥ 60 years | 107 | 16 | 15.0% | 107 | 17 | 15.9%  |
|  Not provided | 1 | 1 | 100% | 1 | 0 | 0%  |
|   | % | 95% CI |   | % | 95% CI  |   |
|  Positive Predictive Value | 87.1% | 81.3% to 91.3% |   | 81.8% | 75.5% to 86.7%  |   |
|  Negative Predictive Value | 98.7% | 97.5% to 99.3% |   | 99.5% | 98.6% to 99.8%  |   |
|  Combined Study – Mucocutaneous Prevalence by Specific Source  |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|   | HSV- 1 |   |   | HSV-2  |   |   |
|  Source | Total # | Total Positive | Prevalence | Total # | Total Positive | Prevalence  |
|  Anorectal | 35 | 2 | 5.7% | 35 | 8 | 22.9%  |
|  Genital – vaginal/cervical | 691 | 109 | 15.9% | 691 | 168 | 24.3%  |
|  Nasal | 16 | 5 | 31.3% | 16 | 2 | 12.5%  |
|  Ocular | 18 | 3 | 16.7% | 18 | 1 | 5.6%  |
|  Oral lesion | 165 | 54 | 32.7% | 165 | 2 | 1.2%  |
|  Urethral | 18 | 0 | N/A | 18 | 3 | 16.7%  |

## N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

## O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/device/K140029](https://fda.innolitics.com/device/K140029)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
