The Intra.Ox™ Handheld Tissue Oximeter is intended to non-invasively estimate the percent oxygen saturation (StO2) in a volume of tissue. The Intra.Ox™ Handheld Tissue Oximeter is indicated for use in monitoring patients during circulatory or perfusion examinations.
Device Story
Intra.Ox is a sterile, cordless, battery-powered, single-use handheld tissue oximeter. It uses spatially-resolved spectrophotometric oximetry via four LED wavelengths (760, 810, 850, 900 nm) to estimate StO2. The device is applied via direct physical contact with patient tissue; the StO2 estimate is displayed on an integrated screen. It is designed for use in clinical settings to monitor circulatory or perfusion status. By providing real-time StO2 estimates, the device assists clinicians in assessing tissue oxygenation, potentially identifying compromised circulation or perfusion. The device is fully disposable, eliminating the need for reusable control units or external power sources.
Clinical Evidence
Clinical study conducted on 11 healthy human volunteers (diverse age, gender, skin tone). Subjects underwent transient ischemic events to mimic compromised tissue. Comparison between Intra.Ox and T.Ox showed excellent agreement in the shape of ischemic events and similar StO2 ranges for healthy and compromised tissue. Mean baseline StO2 was 68% with a 30% desaturation dynamic range, consistent with literature. Bench testing included liquid phantom studies (n=3 devices) showing correlation >0.9 and blood desaturation studies showing limits of agreement of +8.49/-7.50 percentage points (acceptance criteria <±10).
Technological Characteristics
Biocompatible, single-use disposable construction. Spectrophotometric oximetry using four LED light sources (760, 810, 850, 900 nm). Battery-powered (4 Lithium AA, 6V total). Measurement range 1-99% StO2. Integrated sensor and control unit. Standalone device with built-in display.
Indications for Use
Indicated for monitoring patients during circulatory or perfusion examinations; intended for non-invasive estimation of tissue oxygen saturation (StO2).
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around a central graphic. The graphic depicts a stylized image of three human profiles facing right, with flowing lines suggesting movement or connection.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
September 12, 2014
Vioptix Inc. c/o Mr. Greg Holland Regulatory Specialist 3722 Ave. Sausalito Irvine, CA 92606 US
Re: K133983
> Trade/Device Name: Intra. Ox Handheld Tissue Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter, Tissue Saturation Regulatory Class: Class II Product Code: MUD Dated: August 5, 2014 Received: August 7, 2014
Dear Mr. Greg Holland:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{1}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, Misbranding by reference to premarket notification (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part
803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htmfor the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mitchell Stein
forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
K133983 Page 1 of 1
Indications for Use Statement
## Indications for Use
510(k) Number (if known): ______________________
Device Name: Intra.Ox™ Handheld Tissue Oximeter
Indications for Use:
The Intra.Ox™ Handheld Tissue Oximeter is intended to non-invasively estimate the percent oxygen saturation (StO2) in a volume of tissue.
The Intra.Ox™ Handheld Tissue Oximeter is indicated for use in monitoring patients during circulatory or perfusion examinations.
Prescription Use _ X (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) ==============================================================================================================================================================================
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
{3}------------------------------------------------
510(k) Summary
Image /page/3/Picture/2 description: The image shows the logo for ViOptix. The logo consists of the word "ViOptix" in blue font, with the "V" and "i" in a larger, bolder font than the rest of the word. To the right of the word is a green checkmark symbol, with a green circle inside the top part of the checkmark.
47224 Mission Falls Court · Fremont, CA 94539 Office: (510)226-5860 · Fax: (510)226-5864
| Submitter's Name | ViOptix, Inc. |
|--------------------------------------------------|--------------------------------------------|
| Address | 47224 Mission Falls Ct., Fremont, CA 94539 |
| Contact at ViOptix | Mark Lonsinger |
| Telephone | 510-360-7506 |
| Fax | 510-226-5864 |
| Email | info@vioptix.com |
| Date the Summary was prepared: December 20, 2013 | |
Information regarding Application Correspondent:
| Official Correspondent | Greg Holland |
|------------------------|-----------------------------------------|
| Address | 3722 Ave. Sausalito<br>Irvine, CA 92606 |
| Telephone | 949-262-0411 |
| Email | greg@regulatoryspecialists.com |
## Information regarding the device classification:
| Trade Name: | Intra.Ox™ Handheld Tissue Oximeter |
|---------------------------------|------------------------------------|
| Common Name: | Tissue Oximeter |
| Classification regulation: | 21 CFR 870.2700 |
| Classification regulation name: | Oximeter |
| Product code: | MUD |
| Device Class: | II |
#### Information regarding the legally marketed device to which we are claiming equivalence [807.92(a)(3)]:
510(k) Reference # Device Name 510(k) Holder
K042657 ODISsey Tissue Oximeter ViOptix
# Description of the Device:
The ViOptix Intra.Ox™ Handheld Tissue Oximeter is a sterile, cordless, batterypowered device that non-invasively estimates the percent oxygen saturation (StO2) in a volume of tissue. The device uses spatially-resolved optical
{4}------------------------------------------------
measurements at four wavelengths. The device performs measurements on the م patient by direct physical contact to the patient's tissue and displays the StO2 estimate on the built-in screen. The ViOptix Intra.Ox™ Handheld Tissue Oximeter is a single-use disposable constructed from biocompatible materials that can tolerate bodily fluids and other liquids such as disinfectants and marking materials.
### Indications for Use:
The Intra.Ox™ Handheld Tissue Oximeter is intended to non-invasively estimate the percent oxygen saturation (StO2) in a volume of tissue.
The Intra.Ox™ Handheld Tissue Oximeter is indicated for use in monitoring patients during circulatory or perfusion examinations.
### Intended Use:
The Intra.Ox™ Handheld Tissue Oximeter has the same intended use as the predicate, the T.Ox.
{5}------------------------------------------------
# Technological Characteristics:
| Parameter | Subject Device:<br>Intra.OxTM | Predicate Device:<br>T.Ox (formerly known as<br>ODISsey) |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510K number | This submission | K042657 |
| Manufacturer | ViOptix, Inc. | ViOptix, Inc. |
| Intended Use | | |
| Indications for Use | The Intra.OxTM Handheld<br>Tissue Oximeter is<br>intended to non-invasively<br>estimate the percent<br>oxygen saturation (StO2)<br>in a volume of tissue.<br><br>The Intra.OxTM Handheld<br>Tissue Oximeter is<br>indicated for use in<br>monitoring patients during<br>circulatory or perfusion<br>examinations. | The ViOptix ODISsey<br>Tissue Oximeter is<br>intended to non-invasively<br>estimate the percent<br>oxygen saturation (StO2) in<br>a volume of tissue. This is<br>performed in medical<br>environments including<br>physician offices, hospitals,<br>ambulatory care and<br>Emergency Medical<br>Services.<br><br>The ODISsey Tissue<br>Oximeter is indicated for<br>use in monitoring patients<br>during circulatory or<br>perfusion examinations of<br>skeletal muscle or when<br>there is a suspicion of<br>compromised circulation. |
| Measured<br>Parameters | Tissue oxygen saturation<br>(% StO2) | Tissue oxygen saturation<br>(% StO2) and trend graph |
| Operating Principle | Spectrophotometric<br>oximetry | Spectrophotometric<br>oximetry |
| Energy Delivered | Near-infrared light<br>Source: LED chips<br>Wavelengths: 760, 810,<br>850, 900 nm | Near-infrared light<br>Source: laser diodes<br>Wavelengths: 690, 830 nm |
| Single Patient Use? | Yes, integrated sensor<br>and control unit is single<br>patient use disposable. | Sensor is single patient<br>use disposable. Control<br>unit is reusable. |
| Power Source | Battery powered<br>Battery type: 4 Lithium AA<br>Battery voltage: 6 V total | Mains powered with<br>battery backup<br>Battery type: 3-cell Lithium<br>ion |
| Parameter | Subject Device:<br>Intra.Ox™ | Predicate Device:<br>T.Ox (formerly known as<br>ODISsey) |
| Measurement Range | 1-99% StO₂ | 1-99% StO₂ |
{6}------------------------------------------------
# Performance Testing
## Bench Tests
The Intra.Ox™ devices were found to measure absorption coefficients with a high degree of correlation to actual absorption coefficients in liquid phantoms prepared with Intralipid and swine whole blood. The correlation coefficient was greater than 0.9 for each of the four wavelengths used in the devices.
The Intra.Ox™ devices were shown to agree well with the predicate device in StO2% measurements. Over three full-scale (complete oxygenation to complete deoxygenation) blood desaturations, three different Intra.Ox™ devices as compared to two T.Ox devices showed combined limits of agreement of +8.49 and -7.50 percentage points. The acceptance criterion required the limits of agreement to be less than ±10 percentage points. Therefore, the Intra.Ox™ is demonstrated to be substantially equivalent to the T.Ox in estimating StO2%.
## Clinical Study
Performance was determined by measuring tissue oxygen saturation (StO2) with both devices during transient ischemic events on healthy human volunteers that temporarily mimics compromised tissue. Data from 11 subjects, who were nearevenly distributed over age, gender, and skin color as determined by the Fitzpatrick skin type, were analyzed.
There was excellent agreement in shape of the ischemic events between the Intra. Ox™ and T.Ox devices. A direct comparison with paired data showed good agreement considering the physiological variances inherent between measurement sites.
The mean baseline value of 68% and mean desaturation dynamic range of 30 percentage points agrees well with literature-reported values of skin and muscle transient ischemia.
Importantly, the Intra.Ox™ and T.Ox measure similar ranges of StO2 values for both healthy and compromised tissue, thus validating substantial equivalence.
{7}------------------------------------------------
K133983 Page 5 of 5
Image /page/7/Figure/1 description: The image is a boxplot comparing Intra.Ox and T.Ox in healthy and compromised conditions. The y-axis represents StO2% ranging from 0 to 100. The boxplot shows the median, 25%-75% range, and 9%-91% range for each condition. The plot indicates that both Intra.Ox and T.Ox have higher StO2% values in healthy conditions compared to compromised conditions.
### Conclusion
The Intra.Ox™ Handheld Tissue Oximeter has the predicate device identified above, has the same intended use as the predicate, has similar technology that does not raise new types of questions of safety or effectiveness, and performance data shows that this device provides reasonable assurance of safety and effectiveness to demonstrate substantial equivalence.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.