← Product Code [QYW](/productcode/QYW) · K133973

# TEMPUS PRO WITH ENHANCED FEATURES (K133973)

_Remote Diagnostic Technologies , Ltd. · QYW · Aug 14, 2014 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K133973

## Device Facts

- **Applicant:** Remote Diagnostic Technologies , Ltd.
- **Product Code:** [QYW](/productcode/QYW.md)
- **Decision Date:** Aug 14, 2014
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 870.1025
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The Tempus Pro is a portable vital signs monitor intended to be used by clinicians and medically qualified personnel for the attended or unattended monitoring of single or multiple vital signs in clinical and pre-hospital care applications. The device is indicated for: 3 & 5 Lead ECG monitoring; 12 Lead ECG recording; real-time arrhythmia detection / alarming; QT measurement / alarming and ST measurements / alarming: impedance pneumography: noninvasive blood pressure (NIBP); end-tidal CO2 (ETCO2) and respiration rate; pulse oximetry (SpO2); contact temperature; and invasive pressure. The monitor is intended to be used as a stand-alone monitor or as a telemedicine system (transmitting patient data to other medical professionals located elsewhere). The device is indicated for adults, paediatrics and neonates.

## Device Story

Portable multi-parameter vital signs monitor for pre-hospital and clinical use by healthcare professionals. Inputs: ECG leads, NIBP, SpO2, ETCO2, temperature, invasive pressure sensors; integrated camera, GPS, and voice headset. Operation: processes physiological signals for real-time arrhythmia detection, ST-segment, and QT interval measurements; transmits data via wired/wireless connections to remote i2i software system. Output: visual display of vital signs, alarms, video, and location data. Clinical utility: enables remote monitoring and telemedicine; supports clinical decision-making through real-time alarming and data transmission to response centers.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance of arrhythmia, ST, and QT algorithms was validated against standard ECG databases per FDA-recognized consensus standard EC57. Bench testing confirmed reliable data transmission and operation across specified environmental ranges (temperature, altitude, humidity, vibration, shock).

## Technological Characteristics

Portable vital signs monitor. Connectivity: wired/wireless data transmission to remote software. Standards: IEC 60601-1 (safety), IEC 60601-1-8 (alarms), IEC 60601-1-2 (EMC), AAMI EC11/EC13, IEC 60601-2-25/27 (ECG), RTCA DO-160, MIL810, EN 1789, EN 13718-1, EN 60068 (environmental), IEC 60529 (ingress). Software: IEC 62304 compliant.

## Regulatory Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

## Predicate Devices

- Tempus Pro ([K130773](/device/K130773.md))
- Mindray iPM Patient Monitor ([K123074](/device/K123074.md))
- Schiller Welch Allyn 1500 Patient Monitor ([K101619](/device/K101619.md))
- Mortara surveyor central system ([K131929](/device/K131929.md))
- Philips Medical IntelliVue Patient Monitor MP2 ([K120366](/device/K120366.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

November 15, 2023

Remote Diagnostic Technologies Ltd Leigh Cornock Director Of Systems Engineering And Test The Old Coach House, The Avenue, Farleigh Wallop Basingstoke, RG252HT United Kingdom

Re: K133973

Trade/Device Name: Tempus Pro With Enhanced Features Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including St-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: QYW, MHX

Dear Leigh Cornock:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated August 14, 2014. Specifically, FDA is updating this SE Letter because FDA has created a new product code to better categorize your device technology.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Jennifer Kozen, OHT2: Office of Cardiovascular Devices, 301-796-5813, Jennifer.Kozen@fda.hhs.gov.

Sincerely.

Jennifer W.Shih-S

Jennifer Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three stylized human profiles facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 14, 2014

REMOTE DIAGNOSTIC TECHNOLOGIES LTD Leigh Cornock Director of System Engineering and Tests The Old Coach House, The Avenue, Farleigh Wallop Basingstoke, RG252HT GB

Re: K133973 Trade/Device Name: TEMPUS PRO WITH ENHANCED FEATURES Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including St-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MHX Dated: December 23, 2013 Received: July 14, 2014

Dear Leigh Cornock,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{2}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

AthelBStei

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use Statement

510(k) Number (if known): K133973

Device Name: Tempus Pro

Indications for Use:

The Tempus Pro is a portable vital signs monitor intended to be used by clinicians and medically qualified personnel for the attended or unattended monitoring of single or multiple vital signs in clinical and pre-hospital care applications. The device is indicated for: 3 & 5 Lead ECG monitoring; 12 Lead ECG recording; real-time arrhythmia detection / alarming; QT measurement / alarming and ST measurements / alarming: impedance pneumography: noninvasive blood pressure (NIBP); end-tidal CO2 (ETCO2) and respiration rate; pulse oximetry (SpO2); contact temperature; and invasive pressure.

The monitor is intended to be used as a stand-alone monitor or as a telemedicine system (transmitting patient data to other medical professionals located elsewhere).

The device is indicated for adults, paediatrics and neonates.

Prescription Use _ V_ AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{4}------------------------------------------------

Remote Diagnostic Technologies Ltd - Tempus Pro™ Patient Monitor With Enhanced Features 510(k)

# 510(k) Summary of Safety and Effectiveness

### 5.1 Submitter's Information

The submitter of this abbreviated pre-market notification is:

| Name:                  | Remote Diagnostic Technologies Limited                                         |
|------------------------|--------------------------------------------------------------------------------|
| Address:               | The Old Coach House, Farleigh Wallop, Basingstoke,<br>RG25 2HT, United Kingdom |
| Company Phone No:      | +44 (0) 1256 362 400                                                           |
| Company Fax No:        | +44 (0) 1256 362 415                                                           |
| Contact Person:        | Dr Leigh Cornock (Director of Systems Eng. & Test)                             |
| Date summary prepared: | 9th July, 2014                                                                 |

### 5.2 Device Identification

| Device Trade Name:    | Tempus Pro                                            |
|-----------------------|-------------------------------------------------------|
| Common Name:          | Patient Monitor (with arrhythmia detection or alarms) |
| Class:                | II                                                    |
| Classification Panel: | 74                                                    |
| Product Code:         | MHX                                                   |
| Regulation Number:    | 870.1025                                              |

### 5.3 Device Description

The Tempus Pro is a multi-parameter vital signs monitor designed for use in pre-hospital care and remote clinical locations by trained healthcare professionals. It provides 3&5 lead ECG monitoring, 12 lead ECG recording, real-time arrhythmia detection / alarming, QT measurement / alarming and ST measurements / alarming pulse oximetry, non-invasive blood pressure, sidestream capnometry, 2 channels of contact temperature, impedance respiration, up to 4 channels of invasive pressure and user configurable alarms.

In addition, it provides the ability to transmit all vital signs data via wired or wireless connections to a software system (called i2i) expected to be based in a facility far from the user e.g. a response centre facility. In addition to sending all vital signs, the system can also send pictures or video via an integrated camera, geographic position by an integrated GPS receiver and voice via a wired or wireless headset.

#### 5.4 Indications for Use

The following indications for use for the Tempus Pro remain unchanged from the previously cleared Tempus Pro apart from the addition of the arrhythmia detection, ST and QT measurement / alarming features,

"The Tempus Pro is a portable vital signs monitor intended to be used by clinicians and medically qualified personnel for the attended or unattended monitoring of single or multiple vital signs in clinical and pre-hospital care applications. The device is indicated for: 3 & 5 Lead ECG monitoring; 12 Lead ECG recording; realtime arrhythmia detection / alarming; QT measurement / alarming and ST measurements / alarming; impedance pneumography; non-invasive blood pressure (NIBP); end-tidal CO2 (ETCO2) and respiration rate; pulse oximetry (SpO2); contact temperature; and invasive pressure.

The monitor is intended to be used as a stand-alone monitor or as a telemedicine system (transmitting patient data to other medical professionals located elsewhere).

The device is indicated for adults, paediatrics and neonates."

{5}------------------------------------------------

# Remote Diagnostic Technologies Ltd - Tempus Pro™ Patient Monitor With Enhanced Features 510(k)

### 5.5 Comparison with Cleared Device

The intended use and indications for use, plus the fundamental technology used in the Tempus Pro device, remain essentially unchanged by the addition of optional enhanced software feature to the ECG monitoring capability, which is specifically:

- . Arrhythmia detection and alarms
- ST Elevation and depression measurement and alarms 트
- I QT Interval measurement and alarm

### 5.6 Substantial Equivalence

In adding these optional features, substantial equivalence to the previously cleared Tempus Pro (K130773) has been claimed as well as substantial equivalence to the following predicate devices:

- . Arrhythmia detection - Mindray iPM Patient Monitor (K123074)
	- Schiller Welch Allyn 1500 Patient Monitor (K101619)
	- Mortara surveyor central system (K131929)
- I QT and ST measurement - Philips Medical IntelliVue Patient Monitor MP2 (K120366)

With respect to arrhythmia analysis and alarms, ST elevation and QT interval measurements, the implementation of algorithms utilized in previously 510(k) cleared devices, plus the additional alarm feature, were evaluated in accordance with FDA guidance on arrhythmia detection and alarm.

This guidance calls for use of the FDA recognized consensus standard EC57 (FDA Rec No: 3-118), which defines methods of presenting results when testing against specified ECG databases. A summary report demonstrating compliance with the guidance and associated standard is presented as part of this 510(k) submission.

## 5.7 Summary of Non-Clinical Testing

The above modifications are software only and have been made under well-established design control procedures, which ensure that appropriate risk management processes have been carried out to determine their impact and ensure that appropriate verification/ validation testing is performed.

The non-clinical testing carried out in relation to addition of the enhanced features described above is summarized in the following table:

| Area                                             | Testing Performed                                                                                                                                                                                                                                                                            |
|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Safety                                           | This is a minor software only change so no impact on the safety of the<br>device. The device has previously been tested to IEC 60601-1.                                                                                                                                                      |
| Defibrillation and<br>electrosurgical protection | This is a minor software only change so no impact on the<br>performance of the device. The device has previously been tested for<br>operation with a defibrillator and operation with an electro-surgical unit<br>according to IEC 60601-1 (and relevant particular standards).              |
| Environmental                                    | This is a minor software only change so no impact on the<br>performance of the device. The device has previously been tested to<br>a range of environmental (temperature, altitude, humidity, vibration,<br>shock) tests according to RTCA DO-160, MIL810, EN 1789, EN<br>13718-1, EN 60068. |

{6}------------------------------------------------

# Remote Diagnostic Technologies Ltd - Tempus Pro™ Patient Monitor With Enhanced Features 510(k)

| Area                                    | Testing Performed                                                                                                                                                                                                                                                                                          |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Ingress Protection                      | This is a minor software only change so no impact on the<br>performance of the device. The device has previously been tested to<br>IEC 60529 for solid and water ingress.                                                                                                                                  |
| ECG monitoring                          | This is a minor software only change so no impact on the<br>performance of the device. No change is made to the ECG board,<br>applied parts, or software relating to communication with the ECG<br>board Testing to AAMI EC11 & EC13 and IEC 60601-2-25 & IEC<br>60601-1-27 has previously been performed. |
| EMC                                     | This is a minor software only change so no impact on the<br>performance of the device. The device has previously been tested to<br>IEC 60601-1-2 for emissions and immunity including 20 V/m radiated<br>immunity.                                                                                         |
| Alarms                                  | This is a minor software only change so no impact on the<br>performance of the device. The alarm physical functions of the<br>product have previously been tested to IEC 60601-1-8.                                                                                                                        |
| Comparative testing<br>to<br>predicates | Comparative testing has been performed to demonstrate that the<br>performance of the device is equivalent to the predicates.                                                                                                                                                                               |
| Software                                | The requirements of the FDA document Guidance for the Content of<br>Premarket Submissions for Software in Pre-Market Submissions has<br>been applied. In addition, the requirements of IEC 62304 have been<br>addressed.                                                                                   |
| Bench testing                           | All parameters of the device have been tested to confirm they operate<br>to specification across their stated performance range and across<br>their stated temperature range.                                                                                                                              |
| Bench testing                           | The product has been bench tested to confirm that all data is<br>transmitted reliably and accurately.                                                                                                                                                                                                      |

In each case the results of this testing confirmed that acceptance criteria defined by the relevant standard, or other appropriate reference document had been met.

With respect to usability, no additional user validation was considered necessary, as the Tempus Pro is almost identical to the predicate Tempus Pro in terms of physical and interface features, including size and weight, user interface data layouts, button styles, menus, and layout of connectors, resulting the physical interaction with the device being unchanged.

## 5.8 Conclusion

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this Premarket Notification, Remote Diagnostic Technologies Limited conclude that the Tempus Pro Patient Monitor with the enhancement of arrhythmia detection and alarms is safe and effective, and substantially equivalent to the previously cleared version of this device and other cleared devices used as predicates.

---

**Source:** [https://fda.innolitics.com/device/K133973](https://fda.innolitics.com/device/K133973)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
