K133966 · Halifax Biomedical, Inc. · LLZ · Mar 6, 2014 · Radiology
Device Facts
Record ID
K133966
Device Name
MODEL-BASED RSA SOFTWARE
Applicant
Halifax Biomedical, Inc.
Product Code
LLZ · Radiology
Decision Date
Mar 6, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Orthopaedic specialists and/or Halifax Biomedical Inc image processing labs use the Model-based RSA Software as standalone analytical software package for the evaluation of orthopaedic implant fixation , bone segment motion. Model-based RSA software measures the in-vivo 3D position and/or relative motion of metal implants, marker beads, and/or bone segments. When interpreted by trained physicians these measurements may be useful to derive conclusions for patient treatment. NOT FOR MAMMOGRAPHY.
Device Story
Software-only product for Windows workstations; performs Roentgen Stereophotogrammetric Analysis (RSA). Inputs: stereo roentgen images (DICOM CR/DX) or scanned films (BMP). Operation: measures 3D relative position/motion of orthopaedic implants, tantalum marker beads, or bone segments. Used by orthopaedic specialists or image processing labs for pre-operative planning and post-operative assessment of implant wear, fixation, or bone instability (e.g., spine). Output: 3D motion/position measurements; interpreted by physicians to guide patient treatment decisions. Benefits: provides objective data on implant stability and bone segment motion.
Clinical Evidence
Five clinical studies evaluated the accuracy and precision of migration calculations. Results demonstrated that migration calculations are unbiased with high accuracy and precision. Bench testing included phantom experiments using a micromanipulator to validate calculated migrations against known translations and rotations.
Technological Characteristics
Windows-based software; standalone analytical package. Processes DICOM (CR/DX) and bitmap (BMP) images. Employs RSA (Roentgen Stereophotogrammetric Analysis) for 3D measurement of implants/markers/bones. No hardware components. Software validation performed per FDA guidance (May 11, 2005).
Indications for Use
Indicated for orthopaedic specialists and image processing labs to evaluate orthopaedic implant fixation and bone segment motion by measuring in-vivo 3D position and relative motion of metal implants, marker beads, or bone segments. Not for mammography.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
510(k) Summary K133966 Halifax Biomedical Inc. 11493 Route 19, Mabou Nova Scotia, Canada, BOE 1X0 +1 902 945 2762 Phone Fax +1 902 945 2324 Contact: Chad Munro, CEO President Date Prepared: February 20, 2014
MAR - 6 2014
- 1. Identification of the Device: Proprietary-Trade Name: Product Code Device: Regulation Description: Regulation Medical Specialty Regulation Number Device Class
Model-based RSA Software LLZ System, Image Processing, Radiological Picture archiving and communications system. Radiology 892.2050 2
- Equivalent legally marketed devices:. The Model-based RSA Software employs the software 2. cleared in K042383, the RSA-CMS.
- 3. Indications for Use (intended use): Orthopaedic specialists and/or Halifax Biomedical Inc image processing labs use the Model - based RSA Software as standalone analytical software package for the evaluation of orthopaedic implant fixation, bone segment motion. Model - based RSA software measures the in - vivo 3D position and/or relative motion of metal implants, marker beads, and/or bone segments. When interpreted by trained physicians these measurements may be useful to derive conclusions for patient treatment. NOT FOR MAMMOGRAPHY.
- 4. Description of the Device: This is a Windows based software only product. It represents an advancement and extension of the same software cleared in K042383. Model-based RSA (Roentgen Stereophotogrammetric Analysis) is a stand-alone analytical software package for RSA digital image post-processing that runs on standard workstations running a Microsoft Windows operating system. This software is used to analyze roentgen images. It accepts digital images in the specific formats from all the major roentgen manufacturers (DICOM - CR and DX modality) as well as scanned roentgen films in bitmap (BMP) format. Generally, a pair of stereo roentgen images is taken of a patient's joint pre-operatively or postoperatively at one or more time points. Model-based RSA software is then used to measure the three dimensional (3D) relative position and/or relative motion of 3D models in the RSA images. Models may generally represent orthopaedic implants, a group of implanted markers (small tantalum beads), or bones. The 3D relative position and/or relative motion measures may provide information regarding of implants, wear of implants, and excessive or reduced motion between bones such as in spine instability and spine fusion
- 5. Safety and Effectiveness, comparison to predicate device. The results of software validation, clinical evaluation, and risk analysis indicates that the new device is as safe and effective as the predicate device.
- 6. Substantial Equivalence Chart, Model-based RSA Software: Please see the chart below.
{1}------------------------------------------------
K133966
Page 2 of 2
| Characteristic | K042383, the RSA-CMS. | Model-based RSA Software |
|---------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: | RSA-CMS has been developed for the<br>objective and reproducible analysis on<br>digital roentgen images (DICOM CR<br>or DX) or digitised images in a PACS<br>environment. Orthopedic specialist<br>and core labs use the RSA-CMS<br>standalone analytical software<br>package In image post-processing for<br>the evaluation of new implant<br>designs, coatings and new<br>cementation techniques in clinical<br>trials. When interpreted by trained<br>physicians these parameters may be<br>useful to derive conclusions from<br>these clinical trials. | Orthopaedic specialists and/or Halifax<br>Biomedical Inc. image processing labs use<br>the Model-based RSA Software as<br>standalone analytical software package for<br>the evaluation of orthopaedic implant<br>fixation , bone segment motion. Model-<br>based RSA software measures the in-vivo<br>3D position and/or relative motion of metal<br>implants, marker beads, and/or bone<br>segments. When interpreted by trained<br>physicians these measurements may be<br>useful to derive conclusions for patient<br>treatment. NOT FOR MAMMOGRAPHY. |
| Target Users | Trained physicians | SAME |
| Specialty | Orthopaedic | SAME |
| Image Types | DICOM, BMP, TIFF | Same |
| PC Workstation | Yes | Yes |
| Receive digital images<br>from various sources | Yes | Yes |
| Transmit data to remote<br>viewing PACS stations<br>over a medical imaging<br>network | Yes | No |
| Digital image post<br>processing | Yes | Yes |
| Scaling of image facility | Yes | Yes |
| Measurement<br>capabilities | Yes | Yes |
| Preoperative Planning | No | Yes |
| Assessment of 2D<br>migration of prostheses | Yes | Yes |
| Assessment of 3D<br>migration of prostheses | Yes | Yes |
- Description of non-clinical (bench) testing: Software documentation was executed according to the 7. recommendations of the FDA Guidance Document: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document issued on: May 11, 2005. This included validation and risk analysis. Validation of Model-based RSA software is done by comparing the calculated migrations with accurately applied translations and rotations using a micromanipulator in phantom experiments.
- Description of clinical testing: Clinical experiments studied the accuracy and precision of migration 8. calculations using Model-based RSA. Five studies in all showed that migration calculations by the Modelbased RSA software are not biased and have high accuracy and precision.
- 9. Conclusion: After analyzing software validation, bench tests, clinical data, and risk analysis, it is the conclusion of Halifax Biomedical that the Model-Based RSA Software is as safe and effective as the predicate devices, has no significant technological differences, and has no new indications for use, (In fact USING the predicate devices) thus rendering it substantially equivalent to the predicate devices.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its wings and body. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter of the circle. The text is written in all capital letters.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 6, 2014
Halifax Biomedical, Inc. % Mr. Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Court NAPLES FL 34114
Re: K133966
Trade/Device Name: Model-Based RSA Software Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: December 20, 2013 Received: December 24, 2013
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2-Mr. Kamm
. . .
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Smh.7)
ਿੱ
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): K133966
Device Name: Model-Based RSA Software
Indications For Use:
Orthopaedic specialists and/or Halifax Biomedical Inc image processing labs use the Model-based RSA Software as standalone analytical software package for the evaluation of orthopaedic implant fixation , bone segment motion. Model-based RSA software measures the in-vivo 3D position and/or relative motion of metal implants, marker beads, and/or bone segments. When interpreted by trained physicians these measurements may be useful to derive conclusions for patient treatment. NOT FOR MAMMOGRAPHY.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of In Vitro Diagnostics and Radiological Health (OIR)
Smh.7)
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health
> K133966 510(k)
> > Page I of I
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.