NORDICBRAINEX

K133910 · Nordicneurolab · LLZ · Apr 4, 2014 · Radiology

Device Facts

Record IDK133910
Device NameNORDICBRAINEX
ApplicantNordicneurolab
Product CodeLLZ · Radiology
Decision DateApr 4, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

nordicBrainEx is an advanced visualization and processing software, with specific focus on providing algorithms designed to analyze functional MR data of the brain. The software runs on a standard "offthe-shelf" PC workstation and can be used with data and images acquired through DICOM compliant imaging devices and modalities. The software is intended to be used by medical personnel, such as radiologists or medical technicians, trained in the methods provided by the application. In order to best accommodate this group of users, it is specifically designed to have an easy to use and streamlined workflow, as well as an intuitive graphical user interface.

Device Story

nordicBrainEx is a post-processing software application for dynamic MRI brain data; operates on standard Windows PC workstations. Inputs: DICOM-compliant MR images. Processing: algorithms analyze signal changes over time to derive functional parameters, including BOLD (blood-flow/neuronal activity), DTI (water diffusion/white matter structure), and DSC (perfusion/blood vessel characteristics). Outputs: visualization of functional maps, 2D/3D images, and reports. Used in clinical environments by radiologists or medical technicians to optimize workflow efficiency. Healthcare providers use outputs to assess brain function and structure, aiding clinical decision-making regarding neurological conditions.

Clinical Evidence

Bench testing only. Verification and validation activities were performed to ensure the software meets design and performance specifications and user needs.

Technological Characteristics

Software-based post-processing application; runs on standard off-the-shelf Windows PC workstations. Connectivity: DICOM-compliant. Development framework: Embarcadero C++ Builder XE2. Programming language: C++. Functionality: 2D MPR, 3D visualization, and dynamic analysis (BOLD, DTI, DSC).

Indications for Use

Indicated for analysis and visualization of dynamic brain MRI data to present derived properties and parameters in a clinical context. Intended for use by trained medical personnel (radiologists or technicians).

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ # APR 0 4 2014 # 510(k) Summary NordicNeuroLab AS nordicBrainEx Software | Submitter: | NordicNeuroLab AS | |------------|------------------------| | | Møllendalsveien 65C | | | N-5009 Bergen | | | Norway | | | Phone: +47 55 70 70 95 | | Fax: | +47 55 70 70 96 | Primary Contact: Chandana Gurung Bhandari (chandana@nordicneurolab.com) | Proprietary Name: | nordicBrainEx Software | |----------------------------|--------------------------------------------| | Device Common Name: | PACS | | Device: | System, image processing, radiological | | Classification Name: | Picture archiving and communication system | | Classification Regulation: | 892.2050 | | Class: | II | | Panel: | Radiology | | Product Code: | LLZ | Predicate device name: Nordic Image Control and Evaluation (nordiclCE) Software K090546 # Device Description The nordicBrainEx Software is a post-processing application for dynamic MRI data developed with focus on ease of use and high performance on a standard Windows workstation. The software provides comprehensive functionality for dynamic image analysis and visualization of MRI data, where signal changes over time are analyzed to determine various modality dependent functional parameters. The following algorithms provide the main functional analyses of the application. - BOLD: BOLD fMRI analysis is used to highlight small magnetic susceptibility changes in the ● human brain in areas with altered blood-flow resulting from neuronal activity. - DTI: Diffusion analysis is used to visualize local water diffusion properties from the analysis of � diffusion-weighted MRI data. Fiber tracking utilizes the directional dependency of the diffusion to display the white matter structure in the brain. - DSC: Calculations of perfusion related parameters that provide information about the blood . vessel structure and characteristics. Examples of such maps are blood volume, blood flow, time to peak, mean transit time and leakage. NordicNeuroLab AS- Traditional 510(k) nordicBrainEx Software 510 (k) Summary Page 5-2 {2}------------------------------------------------ #### 510(k)- K133910 In addition to these specific functional analyses, the application also provides general visualization tools, a database for data handling, and a reporting feature. This is to ensure that the workflow of the application is optimized to ensure efficiency and high throughput in a clinical environment. ### Intended Use nordicBrainEx is an advanced visualization and processing software, with specific focus on providing algorithms designed to analyze functional MR data of the brain. The software runs on a standard "offthe-shelf" PC workstation and can be used with data and images acquired through DICOM compliant imaging devices and modalities. The software is intended to be used by medical personnel, such as radiologists or medical technicians, trained in the methods provided by the application. In order to best accommodate this group of users, it is specifically designed to have an easy to use and streamlined workflow, as well as an intuitive graphical user interface. ## Indications For Use nordicBrainEx provides analysis and visualization capabilities of dynamic MRI data of the brain, presenting the derived properties and parameters in a clinically useful context. The indications for use of the predicate device, nordic!CE, is formulated in a more general way with focus on delivering specific functionality to a clinical environment. However, since those functionalities are the same as those in nordicBrainEx the more specific indications of nordicBrainEx are considered to be substantially equivalent to those of nordicICE. See SE discussion for more details. ## Technological Characteristics and Substantial Equivalence The nordicBrainEx Software is substantially equivalent to the nordicICE Software (K090546) in technological characteristics and operational characteristics. | | nordicBrainEx | nordicICE | |-----------------------|-----------------------------------------------------------------|-----------------------------------------------------------------| | Development framework | Embarcadero C++ Builder XE2 | Embarcadero C++ Builder 2010 | | Programming language | C++ | C++ | | Operating environment | "Off-the-shelf" windows PC<br>workstation | "Off-the-shelf" windows PC<br>workstation | | Input data | DICOM compliant MR data | DICOM compliant MR data<br>RAW<br>Analyze<br>Nift! | | General functionality | 2D MPR visualization<br>3D Visualization<br>Volumes of interest | 2D MPR visualization<br>3D Visualization<br>Regions of interest | NordicNeuroLab AS- Traditional 510(k) nordicBrainEx Software 510 (k) Summary Page 5-3 {3}------------------------------------------------ 510(k)- K133910 K133910 Page 3 of 3 | | Measurement tools<br>DICOM compliant node<br>Reporting tool | Measurement tools<br>DICOM compliant node | |------------------|-------------------------------------------------------------|-------------------------------------------| | Dynamic analyses | BOLD<br>DTI<br>DSC | BOLD<br>DWI<br>DTI<br>DSC<br>DCE | More details can be found in the substantial equivalence discussion. The rationale for determining the substantial equivalence between nordicBrainEx and nordiclCE is based on the defined intended use, indications for use, and the technical and operational characteristics of the two applications. To verify that nordicBrainEx fulfils the defined characteristics and requirements, it has been subject to extensive in-house testing (see section below). The successful completion of said tests verifies the claimed characteristics of nordicBrainEx, and thus supports the determination of substantial equivalence. ## Performance Testing Prospectively defined verification and validation activities for the nordicBrainEx Software assure that the nordicBrainEx Software meets design and performance specifications as well as user needs when operated according to the operating instructions. #### NordicNeuroLab AS- Traditional 510(k) nordicBrainEx Software 510 (k) Summary Page 5-4 {4}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with outstretched wings, rendered in a simple, bold line drawing. Encircling the bird figure is text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The text is in all capital letters and appears to be in a sans-serif font. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 4, 2014 NordicNeuroLab AS % Ms. Chandana Gurung Bhandari VP Quality Mollendalsveien 65C N-5009 Bergen NORWAY Re: K133910 Trade/Device Name: nordicBrainEx v 2.0 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: March 21, 2014 Received: March 26, 2014 Dear Ms. Bhandari: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (2) CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {5}------------------------------------------------ #### Page 2 - Ms. Bhandari If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Michael D. O'Hara Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indications for Use Statement 510(k) Number (if known): K133910 Device Name: nordicBrainEx Software Indications for Use: nordicBrainEx provides analysis and visualization capabilities of dynamic MRI data of the brain, presenting the derived properties and parameters in a clinically useful context. Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Michael D. O'Hara ______________________________________________________________________________________________________________________________________________________________________________ (Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health K133910 510(k) _______________________________________________________________________________________________________________________________________________________________________ NordicNeuroLab AS - Traditional 510(k) nordicBrainEx Software Indication for Use Statement Page 4-2
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...