K133890 · Aesculap, Inc. · GAW · Jul 15, 2014 · General, Plastic Surgery
Device Facts
Record ID
K133890
Device Name
AESCULAP OPTILENE NONABSORBABLE SUTURE
Applicant
Aesculap, Inc.
Product Code
GAW · General, Plastic Surgery
Decision Date
Jul 15, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Optilene Nonabsorbable Polypropylene Surgical sutures are indicated for use in all types of general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic, microsurgery and neural tissue.
Device Story
Optilene is a sterile, nonabsorbable monofilament surgical suture composed of 95% polypropylene and 5% polyethylene. Available in USP sizes 10-0 through 0, with or without attached needles and optional PTFE pledgets. Used by surgeons in clinical settings for soft tissue approximation and ligation. Device provides mechanical wound closure; performance verified via tensile strength, diameter, and needle attachment testing per USP standards. Biocompatibility confirmed via ISO 10993-1 testing. Sterilized via ethylene oxide.
Clinical Evidence
No clinical data. Bench testing only. Mechanical testing (diameter, tensile strength, needle attachment) performed per USP 35. Biocompatibility testing performed per ISO 10993-1, including cytotoxicity, sensitization, intracutaneous irritation, hemolysis, genotoxicity (Mouse Lymphoma, Mouse Peripheral Blood Micronucleus, Bacterial Reverse Mutation), and 12-week muscle implantation. Chemical characterization and biological risk assessment supported material equivalency.
Technological Characteristics
Synthetic nonabsorbable monofilament suture; 95% polypropylene/5% polyethylene. Dyed with [Phthalacyaninato(2-)] copper (21 CFR 74.3045). Sizes 10-0 to 0. Optional PTFE pledgets. Sterilized via ethylene oxide. Mechanical specs meet USP 35. Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for general soft tissue approximation and/or ligation in patients requiring cardiovascular, ophthalmic, microsurgical, or neural procedures.
Regulatory Classification
Identification
Nonabsorbable polypropylene surgical suture is a monofilament, nonabsorbable, sterile, flexible thread prepared from long-chain polyolefin polymer known as polypropylene and is indicated for use in soft tissue approximation. The polypropylene surgical suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be undyed or dyed with an FDA approved color additive; and the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
PremiCron Nonabsorbable PET Surgical Suture (K012201)
Submission Summary (Full Text)
{0}------------------------------------------------
510(k) Premarket Notification
Optilene Nonabsorbable Surgical Suture
# JUL 15 2014
Page 1 of 4
#### B. 510(k) SUMMARY (as required by 21 CFR 807.92)
Aesculap Optilene Nonabsorbable Surgical Suture July 3. 2014
| COMPANY: | Aesculap®, Inc.<br>3773 Corporate Parkway<br>Center Valley, PA 18034<br>Establishment Registration Number: 2916714 |
|----------------------|--------------------------------------------------------------------------------------------------------------------|
| CONTACT: | Kathy A. Racosky<br>610-984-9291 (phone)<br>610-791-6882 (fax)<br>kathy.racosky@aesculap.com |
| TRADE NAME: | Aesculap Optilene Nonabsorbable Suture |
| COMMON NAME: | Nonabsorbable polypropylene/polyethylene surgical suture |
| CLASSIFICATION: | Class II |
| CLASSIFICATION NAME: | Suture, Nonabsorbable, Synthetic, Polypropylene |
| REGULATION NUMBER: | 878.5010 |
| PRODUCT CODE: | GAW |
# SUBSTANTIAL EQUIVALENCE
Aesculap®, Inc. believes that the Optilene Nonabsorbable Polypropylene/Polyethylene Surgical Suture is substantially equivalent to:
- . Premilene Nonabsorbable Polypropylene Surgical Suture, Aesculap Inc. (K980703)
- · Prolene Polypropylene Nonabsorbable Surgical Suture, Ethicon Inc. (N16374)
- · USS Polypropylene Suture, United States Surgical (K050947/K010909/K954808)
The Optilene Nonabsorbable Polypropylene/Polyethylene Surgical Suture pledget is substantially equivalent to:
- · PremiCron Nonabsorable PET Surgical Suture, Aesculap Inc. (K012201)
### DEVICE DESCRIPTION
:
Optilene is a nonabsorbable monofilament surgical suture which is supplied sterile. Optilene is composed of 95% polypropylene and 5% polyethylene. The Optilene suture will be offered dyed with the FDA approved colorant [Phthalacyaninato(2-)] copper in accordance with Title 21 CFR, \$74.3045. The Optilene suture will be offered in diameters ranging from USP size 10-0 through 0 and with be available in a variety of cut lengths with or without needles attached. The Optilene sutures will also be offered with or without pledgets composed of 100% polytetrafluorothylene (PTFE). The pledgets are also available separately packaged six per pouch.
{1}------------------------------------------------
Page 2 of 4
### INDICATIONS FOR USE
Optilene Nonabsorbable Polypropylene/Polyethylene Surgical sutures are indicated for use in all types of general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic, microsurgery and neural tissue.
### TECHNOLOGICAL CHARACTERISTICS(compared to Predicate(s)}
As established in this submission, the Aesculap Optilene suture is a synthetic nonabsorbable monofilament surgical suture offered dyed in the same range of diameters and cut lengths that are substantially equivalent to other predicate devices cleared by FDA. The subject device is shown to be substantially equivalent and has the same technological characteristics to its predicate devices through comparison in design, intended use, material composition, function and range of sizes.
Please see attached Substantial Equivalence table comparing the Optilene Nonabsorbable Polypropylene/Polyethylene Suture to the predicate devices.
## PERFORMANCE DATA
As recommended by the FDA's Class II Special Control Guidance Document for Surgical Sutures, including mechanical testing in accordance to USP 35 and biocompatibility testing in accordance to ISO 10993-1 has been performed to demonstrate that the Optilene Nonabsorbable Suture is substantially equivalent to other predicate devices.
Tests were conducted for diameter, tensile strength, and needle attachment. All specifications were met. Biocompatibility testing within this submission includes the following: Cytotoxicity, Sensitization, Intracutaneous Irritation, Hemolysis, Genotoxicity - Mouse Lymphoma, Mouse Peripheral Blood Micronucleus, Bacterial Reverse Mutation and Muscle Implantation (12-week).
Testing demonstrated that the device is as safe, as effective, and performs as well as the predicate devices.
Additional chemical characterization testing and a biological risk assessment were conducted in order to evaluate the chemical equivalency of the polypropylene/polyethylene raw material. The results of this testing demonstrate that the Optilene Nonabsorbable Suture is substantially equivalent to the predicate devices.
The Optilene Nonabsorbable Suture is blister packed and sterilized by ethylene oxide. Accelerated aging data for the Optilene Nonabsorbable Suture has been generated to support this submission.
{2}------------------------------------------------
10(k) Premarket Notification
Optilene Nonabsorbable Surgical Suture
Page 3 of 4
| K# | Indications | Material | Dyed, Un-dyed<br>Colorant | Structure | Size | Physical:<br>- Diameter<br>- Length<br>- Needle<br>Attachment<br>- Tensile Strength<br>Needles | Pledgets | Packaging | Sterilization |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|--------------------------------------------------------|--------------|-------------------------------------|------------------------------------------------------------------------------------------------|-------------------------|-------------------------------------------------------------------------------------------------------------------------------|---------------------|
| | Aesculap, Inc.<br>Optilene Nonabsorbable<br>Suture | | | | | | | | |
| Pending | Indicated for use in all types of<br>general soft tissue<br>approximation and/or ligation,<br>including use in cardiovascular,<br>ophthalmic, microsurgery and<br>neural tissue. | 95% polypropylene and 5%<br>Polyethylene | Dyed, Un-dyed<br>Dyed<br>[Phthalacyaninato(2-)] copper | Monofilament | 10-0 through 0 (various<br>lengths) | with or w/out needles attached<br>with or w/out pledgets attached | | Oval HDPE support with an<br>oval card fold card lid in a<br>tyvek peel pouch and polyester-<br>polyethylenerephatalate film. | Ethylene Oxide (EO) |
| K980703 | Aesculap, Inc.<br>Premilene Nonabsorbable<br>Suture<br>Indicated for use in all types of<br>general soft tissue<br>approximation and ligation,<br>including use in cardiovascular,<br>ophthalmic, microsurgery and<br>neural tissue. | Polypropylene | Un-dyed and Dyed<br>[Phthalacyaninato(2-)] copper | Monofilament | 11-0 through 2 (various<br>lengths) | with or w/out needles attached | N/A | Oval HDPE support with an<br>oval card fold card lid in a<br>tyvek peel pouch and polyester-<br>polyethylenerephatalate film. | Ethylene Oxide (EO) |
| K050947/K010909/K954808 | USS Polypropylene<br>Nonabsorbable Suture<br>Indicated for use in general soft<br>tissue approximation and/or<br>ligation, including use in<br>cardiovascular, ophthalmic,<br>and neurological surgery. | Polypropylene and<br>Polyethylene | Un-dyed and Dyed<br>Copper phthalocyanine, CI<br>74160 | Monofilament | 10-0 through 2 (various<br>lengths) | with or w/out needles attached<br>with or w/out pledgets attached | polytetrafluoroethylene | Foil packaging in a second<br>outer peel-pack with paper and<br>plastic film | Ethylene Oxide (EO) |
| N16374 | Prolene Polypropylene<br>Nonabsorbable<br>Indicated for use in general soft<br>tissue approximation and/or<br>ligation, including use in<br>cardiovascular, ophthalmic,<br>and neurological surgery. | Polypropylene | Un-dyed and Dyed<br>Copper phthalocyanine, Cl<br>74160 | Monofilament | 10-0 through 1 (various<br>lengths) | with or w/out needles attached<br>with or w/out pledgets attached | polytetrafluoroethylene | Oval HDPE support with an<br>oval card fold card lid in a<br>tyvek peel pouch and polyester-<br>polyethylenerephatalate film. | Ethylene Oxide (EO) |
| | | | | | | All characteristics meet USP<br>Requirements. | | | |
| | | | | | | All characteristics meet USP<br>Requirements | | | |
| | | | | | | All characteristics meet USP<br>Requirements, except for<br>diameter. | | | |
| | | | | | | All characteristics meet USP<br>Requirements | | | |
| | | | | | | 300-series stainless steel | | | |
| | | | | | | 300 or 455 series stainless steel | | | |
| | | | | | | 300-series stainless steel and<br>tungsten-rhenium | | | |
·
.
.
{3}------------------------------------------------
510(k) Premarket Notificatio
・
......
Page 3 of 4
| K# | | |
|----------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Aesculap, Inc.<br>Optilene Nonabsorbable Suture | Aesculap, Inc.<br>PremiCron Nonabsorbable Suture |
| Indications | Pending<br>Indicated for use in all types of general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic, microsurgery and neural tissue. | K980703<br>Indicated for use in all types of general soft tissue approximation and ligation, including use in cardiovascular, ophthalmic, microsurgery and neural tissue. |
| Material | 95% polypropylene and 5% Polyethylene | poly(ethylene terephthalate) |
| Dyed, Un-dyed | Dyed | Un-dyed and Dyed |
| Colorant | [Phthalacyaninato(2-)] copper | D&C Green no. 6 |
| Structure | Monofilament | Multifilament |
| Size | 10-0 through 0 (various lengths)<br>with or w/out needles attached<br>with or w/out pledgets attached | 8-0 through 5 (various lengths)<br>with or w/out needles attached<br>with or w/out pledgets attached |
| Physical:<br>- Diameter<br>- Length<br>- Needle Attachment<br>- Tensile Strength | All characteristics meet USP Requirements. | All characteristics meet USP Requirements |
| Needles | 300-series stainless steel | 300 or 400 series stainless steel |
| | polytetrafluoroethylene | polytetrafluoroethylene |
| Pledgets | | |
| Packaging | Oval HDPE support with an oval card fold card lid in a tyvek peel pouch and polyester-polyethylenerephatalate film. | Oval HDPE support with an oval card fold card lid in a tyvek peel pouch and polyester-polyethylenerephatalate film. |
| Sterilization | Ethylene Oxide (EO) | Ethylene Oxide (EO) |
1000
:
.
.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 15, 2014
Aesculap® Incorporated Ms. Kathy A. Racovsky Senior Regulatory Affairs Specialist 3773 Corporate Parkway Center Valley, Pennsylvania 18034
Re: K133890
Trade/Device Name: Aesculap Optilene Nonabsorbable Surgical Suture Regulation Number: 21 CFR 878.5010 Regulation Name: Nonabsorbable polypropylene surgical suture Regulatory Class: Class II Product Code: GAW Dated: June 11, 2014 Received: June 12, 2014
Dear Mr. Racovsky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Ms. Kathy A. Racovsky
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fdagov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K133890
#### Device Name
Aesculup Optilene Nonahsorbable Surgical Suture
#### Indications for Use (Describe)
Optilene Nonabsorbable Polypropylene Surgical sutures are indicated for use in all types of general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic, microsurgery and neural tissue.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Peter L. Hudson -S
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the rime to review instructions, search exisling data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaffi@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.