← Product Code [FLN](/productcode/FLN) · K133870

# FLUID LEVEL MONITOR (K133870)

_Coleman Laboratories · FLN · Apr 24, 2014 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K133870

## Device Facts

- **Applicant:** Coleman Laboratories
- **Product Code:** [FLN](/productcode/FLN.md)
- **Decision Date:** Apr 24, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.2420
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The Fluid Level Monitor is a passive sensing device that alarms when the infusion fluid level is low. The device provides both visual and audible alarms when preset condition is met. The Fluid Level Monitor is intended for low-fluid-level alarm use in any healthcare setting where gravity flow infusion is utilized. The use of the device improves efficiency within a busy medical setting by providing healthcare personnel of timely notification of the low-fluid-level status. The available models are: LM771: 1-channel, DC battery power LM781 1-channel, DC plug-in power LM785 - 5-channel, DC plug-in power

## Device Story

Passive sensing device; monitors infusion fluid levels in gravity flow systems. External sensor clips onto tubing; connects to main control unit via shielded cable. Control unit processes sensor input; triggers visual (LEDs) and audible (buzzer) alarms when fluid reaches preset low level. Operates on DC power (battery or plug-in). Used in clinical settings by healthcare personnel to improve monitoring efficiency. Provides timely notification of low-fluid status; assists in clinical decision-making regarding infusion management; benefits patient by preventing infusion interruption.

## Clinical Evidence

Bench testing only. Performance testing confirmed compliance with design specifications and verified that all functions operate as designed. Testing demonstrated consistent power consumption across battery/plug-in power and 1-channel/5-channel configurations.

## Technological Characteristics

Passive electronic monitor for gravity flow infusion. Components: main control unit, external sensor, shielded signal cable. Power: DC battery or DC plug-in. Monitoring capacity: 1-channel (LM771, LM781) or 5-channel (LM785). Output: Visual (LED) and audible (buzzer) alarms. Processor-based measurement.

## Regulatory Identification

An electronic monitor for gravity flow infusion systems is a device used to monitor the amount of fluid being infused into a patient. The device consists of an electronic transducer and equipment for signal amplification, conditioning, and display.

## Predicate Devices

- Drip Alert ([K030136](/device/K030136.md))
- Levelert ([K903193](/device/K903193.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 5. 510(k) Summary

The following information is provided as required by 21 CFR § 807.87 for Coleman Laboratories' Fluid Level Monitor 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the safety and effectiveness information upon which the substantial equivalence determination is based.

### 1) Submitter Information

| Company Name:          | Coleman Laboratories                                           |
|------------------------|----------------------------------------------------------------|
| Company Address:       | 1150 First Avenue, Suite 501<br>King of Prussia, PA 19406, USA |
| Contact:               | Le-Jun Yin, Vice President, Product Management                 |
| Email:                 | lejun.yin@colemanlabs.com                                      |
| Phone:                 | (484) 727-8812                                                 |
| Fax:                   | (484) 727-8828                                                 |
| Date Summary Prepared: | December 18th, 2013                                            |

## 2) Device Identification

| Generic Device Name:    | Electronic I.V. Monitor                              |
|-------------------------|------------------------------------------------------|
| Trade/Proprietary Name: | Fluid Level Monitor Model                            |
| Numbers:                | LM771, LM781, LM785                                  |
| Classification:         | Class II                                             |
| Panel:                  | General Hospital (80)                                |
| Regulation Number:      | 21 CFR 880.2420                                      |
| Regulation Name:        | Electronic monitor for gravity flow infusion systems |
| Product Code:           | FLN                                                  |
| 510(k) Number:          | K133870                                              |

## 3) Predicate Device

| 510(k) Number: | K030136, K903193     |
|----------------|----------------------|
| Trade Name:    | Drip Alert, Levelert |
| Product Code:  | FLN                  |

## 4) Description of Device

Fluid Level Monitor is a passive sensing device that alarms when the infusion fluid level is low. Fluid Level Monitor provides both visual and audible alarms when preset condition is met. Fluid Level Monitor operates on DC power source.

Fluid Level Monitor includes both main control unit and external sensor. There are status LEDs on the main control unit to display run, alarm and battery-low status. There is a buzzer on the main control unit to provide audible alarm.

The external sensor is connected to the main control unit via shielded signal cable. The sensor can be clipped on the tubing to monitor fluid level.

### 5) Intended Use

The Fluid Level Monitor is a passive sensing device that alarms when the infysion fluid level is low. The device provides both visual and audible alarms when preset condition is met.

The Fluid Level Monitor is intended for low-fluid-level alarm use in any healthcare setting where gravity flow infusion is utilized. The use of the device improves efficiency within a busy medical setting by providing healthcare personnel of timely notification of the low-fluid-level status. The available models are:

LM771: 1-channel, DC battery power LM781 1-channel, DC plug-in power

{1}------------------------------------------------

LM785 - 5-channel, DC plug-in power

Coleman Laboratories' Fluid Level Monitors (3 Models) are substantially equivalent to the predicate devices:

K030136 - Drip Alert, Drip Alert, Inc. K903193 - Smith & Nephew, Inc.

They have the same design principles, material, indications for use, and intended use.

| Device Name                                               | Fluid Level Monitor |              |              | Drip Alert    | Levelert     |
|-----------------------------------------------------------|---------------------|--------------|--------------|---------------|--------------|
| Device Model                                              | LM771               | LM781        | LM785        | -             | -            |
| 510(K) Number                                             |                     | K133870      |              | K030136       | K903193      |
| Passive device                                            | Yes                 | Yes          | Yes          | Yes           | Yes          |
| Uses a processor to<br>perform<br>measurements            | Yes                 | Yes          | Yes          | Yes           | Yes          |
| Use sensor to detect<br>fluid-level-low status            | Yes                 | Yes          | Yes          | Yes           | Yes          |
| Visual and audible<br>alarm on fluid-level-<br>low status | Yes                 | Yes          | Yes          | Yes           | Yes          |
| Patient has no contact<br>with device                     | Yes                 | Yes          | Yes          | Yes           | Yes          |
| Class II device                                           | Yes                 | Yes          | Yes          | Yes           | Yes          |
| Power source                                              | DC<br>battery       | DC<br>plugin | DC<br>plugin | DC<br>battery | DC<br>plugin |
| Monitoring Point                                          | 1                   | 1            | 5            | 1             | 1            |

# 7) Summary Performance Data

Based on the risk analysis, the verification and validation tests that were performed and the acceptance criteria applied for each are listed in Section 18. Performance testing was conducted to confirm compliance to design specifications; all functions were verified to operate as designed.

{2}------------------------------------------------

The performance test data demonstrated that the system power consumption remains same when using:

- a) DC battery or DC plug-in power
- b) 1-channel or 5-channel sensing

We concluded that the design difference on number of monitoring points does not raise safety or effectiveness concerns.

### 8) Conclusion

Based on the analysis on nonclinical test data, and comparison on design principle, operation procedure and indications for use, we demonstrated that the device is as safe, as effective, and performs as well as the predicate devices. The design difference does not raise questions on safety or effectiveness.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, a symbol often associated with medicine and healthcare.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 24, 2014

Coleman Laboratories Le-Jun Yin Vice President, Product Management 1150 First Avenue, Suite 501 King of Prussia, PA 19406, USA

Re: K133870

Trade/Device Name: Fluid Level Monitor, Models LM771, LM781, and LM785 Regulation Number: 21 CFR 880.2420 Regulation Name: Electronic monitor for gravity flow infusion systems Regulatory Class: II Dated: March 13, 2014 Received: March 28, 2014

Dear Mr. Yin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Yin

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Susan R. DODDS, MA

Erin I. Keith, M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.

#### 510(k) Number (if known) K133870

Device Name

Fluid Level Monitor, Models LM771, LM781, and LM785

#### Indications for Use (Describe)

The Fluid Level Monitor is a passive sensing device that alarms when the infusion fluid is low. The device provides both visual and audible alarms when preset condition is met.

The Fluid Level Monitor is intended for low-fluid-level alarm use in any healthcare setting where gravity flow infirsed. The use of the device improves efficiency within a busy medical setting by providing healthcare personnel of the low-fluid-level status.

LM771: 1-channel, DC battery power LM781: 1-channel, DC plug-in power LM785: 5-channel DC plug-in power

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Digitally signed by Sajjad H. Syed -S DN: c=US, o=U.S. Government, ou=HHS, ou=FDA, ou=People, cn=Sajjad H. Syed -S, 0.9.2342.19200300.100.1.1=2000601742 Date: 2014.04.24 12:11:28 -04'00' PSC Publishing Services (30) ) 443-6740 Page 1 of 2 FORM FDA 3881 (1/14)

---

**Source:** [https://fda.innolitics.com/device/K133870](https://fda.innolitics.com/device/K133870)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
