← Product Code [NJD](/productcode/NJD) · K133811

# RESTORIS POROUS PARTIAL KNEE SYSTEM (K133811)

_Mako Surgical Corp. · NJD · Jul 8, 2014 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K133811

## Device Facts

- **Applicant:** Mako Surgical Corp.
- **Product Code:** [NJD](/productcode/NJD.md)
- **Decision Date:** Jul 8, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3535
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Restoris® Porous Partial Knee System components are intended for unicompartmental knee arthroplasty to treat one or more of the following conditions: . Moderately disabling joint disease of the knee resulting from painful osteo- or post traumatic arthritis. . Revision of previous unsuccessful surgical procedures, either involving, or not involving, previous use of a unicompartmental knee prosthesis. . As an alternative to tibial osteotomy in patients with unicompartmental osteoarthritis. The Restoris® Porous Femoral Component and PST Tibial Baseplate are intended for cementless or cemented fixation. The Tibial Baseplate may be used in conjunction with ancillary screw fixation. The porous surfaces of both the femoral and tibial tray components provide biological fixation when used in a cementless application.

## Device Story

Restoris Porous Partial Knee System is a unicompartmental knee arthroplasty implant. Components include cast CoCr femoral components (sizes 1-8) and Ti6Al4V porous-coated tibial baseplates (sizes 1-8). Designed for cementless or cemented fixation; porous surfaces facilitate biological fixation in cementless applications. Femoral condyle features extended posterior radius for flexion up to 155 degrees. Used by orthopedic surgeons in clinical settings. Compatible with predicate MAKO Restoris MCK Tibial Onlay Inserts. Provides mechanical stability and joint articulation for patients with degenerative knee conditions.

## Clinical Evidence

Bench testing only. Evaluations included femoral component fatigue, tibial baseplate fatigue, and porous surface characterization. Testing previously submitted for predicates (K090763, K133039) supported locking mechanism strength, femorotibial congruency, contact area, and range of motion. No clinical data presented.

## Technological Characteristics

Materials: Cast CoCr (femoral), Ti6Al4V (tibial baseplate). Design: Asymmetric components, sizes 1-8. Fixation: Cemented or cementless (porous coating). Features: Extended posterior radius for 155-degree flexion. Ancillary screw fixation optional for tibial baseplate. No software or electronic components.

## Regulatory Identification

A knee joint femorotibial (uni-compartmental) metal/polymer porous-coated uncemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surface. It has no linkage across-the-joint. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement. This identification includes fixed-bearing knee prostheses where the ultra-high molecular weight polyethylene tibial bearing is rigidly secured to the metal tibial baseplate.

## Special Controls

*Classification.* Class II (special controls). The special control is FDA's guidance: “Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA.” See § 888.1 for the availability of this guidance.

## Predicate Devices

- Restoris MCK Knee System ([K090763](/device/K090763.md))
- MMCK Uni Knee System ([K082172](/device/K082172.md))
- Natural-Knee® II Unicompartmental Knee System ([K033810](/device/K033810.md))
- MAKO Restoris® Hip System ([K112802](/device/K112802.md))

## Reference Devices

- MAKO Restoris MCK Tibial Onlay Inserts ([K133039](/device/K133039.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# Restoris® Porous Partial Knee System 510(k) Summary

JUL 0882014

| Device Proprietary Name:   | Restoris Porous Partial Knee System                                                                                                                                                                             |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name:               | Artificial Partial Knee System                                                                                                                                                                                  |
| Classification regulation: | 888.3530 (Knee joint femorotibial metal/polymer<br>semi-constrained cemented prosthesis) and<br>888.3535 (Knee joint femorotibial (uni-<br>compartmental) metal/polymer porous-coated<br>uncemented prosthesis) |
| Device Class:              | Class II                                                                                                                                                                                                        |
| Product Codes:             | HRY and NJD                                                                                                                                                                                                     |
| Submitter's Name:          | MAKO Surgical Corp.                                                                                                                                                                                             |
| Address:                   | 2555 Davie Road<br>Fort Lauderdale, FL 33317                                                                                                                                                                    |
| Contact Person:            | Robert C. Cohen                                                                                                                                                                                                 |
| Telephone Number:          | (973) 267-8800                                                                                                                                                                                                  |
| Fax Number:                | (973) 267-8810                                                                                                                                                                                                  |
| Date Summary Prepared:     | December 13, 2013                                                                                                                                                                                               |

### Device Description:

The Restoris® Porous Partial Knee System is a unicompartmental knee system that includes porous coated, cast CoCr, asymmetric femoral components in sizes 1 – 8; and Ti6Al4V, asymmetric, porous tibial baseplate components in sizes 1 - 8. The subject femoral and tibial baseplate components are compatible and are intended for use with predicate MAKO Restoris MCK Tibial Onlay Inserts in standard UHMWPE or highly crosslinked vitamin E UHMWPE (K090763, K133039). The sagittal articular surface of the femoral condyle has an extended posterior radius that accommodates flexion up to 155%.

#### Intended Use

Restoris® Porous Partial Knee System components are The intended for unicompartmental knee arthroplasty to treat one or more of the following conditions:

- . Moderately disabling joint disease of the knee resulting from painful osteo- or post traumatic arthritis.
- . Revision of previous unsuccessful surgical procedures, either involving, or not involving, previous use of an unicompartmental knee prosthesis.

{1}------------------------------------------------

# Restoris® Porous Partial Knee System 510(k) Summary

- As an alternative to tibial osteotomy in patients with unicompartmental . osteoarthritis.
The Restoris® Porous Femoral Component and PST Tibial Baseplate are intended for cementless or cemented fixation. The Tibial Baseplate may be used in conjunction with ancillary screw fixation. The porous surfaces of both the femoral and tibial tray components provide biological fixation when used in a cementless application.

### Predicate Devices:

Restoris® Porous Partial Knee System is similar to the following predicates:

| Trade/Proprietary Name                                   | Manufacturer                                       | 510(K) # | Clearance Date |
|----------------------------------------------------------|----------------------------------------------------|----------|----------------|
| Restoris MCK Knee System                                 | MAKO Surgical                                      | K090763  | 6/17/2009      |
| MMCK Uni Knee System                                     | MAKO Surgical                                      | K082172  | 11/28/2008     |
| Natural-Knee® II Unicompartmental<br>Knee System         | CenterPulse<br>Orthopedics/Zimmer                  | K033810  | 3/5/2004       |
| MAKO Restoris® Hip System<br>(Pipeline Total Hip System) | MAKO (Pipeline<br>Orthopedics acquired<br>by MAKO) | K112802  | 03/12/2012     |

## Purpose of Submission:

The 510(k) submission is for new unicompartmental knee system femoral and tibial tray . components.

## Technological Characteristics:

The Restoris® Porous Partial Knee System has the same indications for use, design features, and materials as one or more of the identified predicate knee systems. The porous surface of the subject Restoris® Porous Partial Knee System tibial baseplates is the same as the porous surface of the predicate MAKO Restoris® Total Hip System (K112802) acetabular shells.

### Performance Testing:

The Restoris® Porous Partial Knee System has been evaluated, either by new testing and analysis submitted in this 510(k) for femoral component fatigue, tibial baseplate fatigue, and porous surface characterization, or by testing previously submitted in predicate 510(k) #K090763 or #K133039 for the tibial insert/baseplate locking mechanism strength, femorotibial conguency, femorotibial contact area, and range of motion. The testing demonstrates that the Restoris® Porous Partial Knee System when used as intended with the predicate MAKO Restoris MCK Tibial Onlay Inserts, is capable of withstanding expected in vivo loading and is substantially equivalent to predicate knee systems.

{2}------------------------------------------------

# Restoris® Porous Partial Knee System 510(k) Summary

## Conclusions:

The Restoris® Porous Partial Knee System has the same indications for use, is manufactured from the same materials, and has similar design features as compared with one or more of the predicate knee systems. Characterization of the device porous surfaces has shown them equivalent to predicate device porous coatings. Engineering analyses and mechanical testing demonstrate that the performance characteristics of the Restoris® Porous Partial Knee System are equivalent to those of other legally marketed knee systems. Therefore, the Substantial Equivalence of the subject device for the proposed indications for use is demonstrated.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows a logo with the words "HUMAN SERVICES - USA" arranged in a circular pattern around a symbol. The symbol is a stylized depiction of an eagle or bird in flight, with its wings spread and feathers suggested by curved lines. The logo appears to be a government or organizational emblem, possibly related to health or social services.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 8, 2014

MAKO Surgical Corp. % Ms. Terry Sheridan Powell Consultant M Squared Associates, Inc. 575 Eighth Avenue Suite 1212 New York, New York 10018

Re: K133811

Trade/Device Name: Restoris Porous Partial Knee System Regulation Number: 21 CFR 888.3535 Regulation Name: Knee joint femorotibial (uni-compartmental) metal/polymer porouscoated uncemented prosthesis

Class: Class II Product Code: NJD, HRY Dated: May 23, 2014 Received: May 27, 2014

Dear Ms. Powell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing

{4}------------------------------------------------

Page 2 - Ms. Terry Sheridan Powell

(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Lori A. Wiaains

for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use Statement

K133811 510(k) Number: To be assigned -

Device Name: Restoris® Porous Partial Knee System

Indications for Use:

The Restoris® Porous Partial Knee System components are intended for unicompartmental knee arthroplasty to treat one or more of the following conditions:

- . Moderately disabling joint disease of the knee resulting from painful osteo- or post traumatic arthritis.
- . Revision of previous unsuccessful surgical procedures, either involving, or not involving, previous use of a unicompartmental knee prosthesis.
- . As an alternative to tibial osteotomy in patients with unicompartmental osteoarthritis.

The Restoris® Porous Femoral Component and PST Tibial Baseplate are intended for cementless or cemented fixation. The Tibial Baseplate may be used in conjunction with ancillary screw fixation. The porous surfaces of both the femoral and tibial tray components provide biological fixation when used in a cementless application.

AND/OR -Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)

Over-The-Counter Use _ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Casey L. Hanley, Ph.D.

Division of Orthopedic Devices

Page 1 of ____________________________________________________________________________________________________________________________________________________________________

---

**Source:** [https://fda.innolitics.com/device/K133811](https://fda.innolitics.com/device/K133811)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
