VC150 VITAL SIGNS MONITOR

K133810 · Innokas Yhtyma · MWI · Oct 30, 2014 · Cardiovascular

Device Facts

Record IDK133810
Device NameVC150 VITAL SIGNS MONITOR
ApplicantInnokas Yhtyma
Product CodeMWI · Cardiovascular
Decision DateOct 30, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2300
Device ClassClass 2
AttributesPediatric

Indications for Use

The VC150 is intended to monitor a single patient's vital signs at the site of care or during intra-hospital transport. The noninvasive oscillometric blood pressure parameter is intended for measurement of systolic, and mean arterial blood pressure, as well as pulse rate, for adult, pediatric and neonatal patients. The optional GE TruSignal pulse oximetry and accessories are indicated for continuous nonitoring of functional oxygen saturation (SpO2) and pulse rate, including monitoring conditions of clinical patient motion or low perfusion, with adult, pediatric and neonatal patients. The optional Masimo Rainbow SET® Pulse CO-oximetry and accessories are indicated for the continuous noninvasive monitoring of: · functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate, during both no motion and motion conditions, and for patients who are well or poorly perfusion) for adult, pediatric and neonatal patients, · carboxyhemoglobin saturation (SpCO) for adult and pediatric patients, - · methemoglobin saturation (SpMet) for adult, pediatric and neonatal patients, · total hemoglobin concentration (SpHb) for adult and pediatric patients, and/or · respiratory rate (RRa) for adult and pediatric patients. The optional Nellcor™ oximetry and accessories are indicated for the continuous, noninyasive monitoring of arterial oxygen saturation (SpO2) and pulse rate of adult, pediatric, and neonatal patients during both motion and non-motion conditions, and for patients who are well or poorly perfused. The optional Oximax SPD™ Alert (SPD) feature is intended only for facility-use care of adults to desaturation indicative of repetitive reductions in airflow through the upper airway and into the lungs. The optional Nellcor™ Respiration Rate parameter is intended for the continuous noninvasive monitoring of respiration rate in adult patients who are well perfused during non-motion conditions. The optional Welch Allyn SureTemp Plus electronic thermometer is intended to measure oral, axillary, and rectal temperature of adult and pediatric pational Exergen TemporalScanner thermometer is intended for the intermittent measurement of human body temperature of patients of all ages. A wireless network connection is provided to transmit clinical data into various hospital information systems. An optional remote alarm cable connection is intended to complement visual and audible alarms and not replace the need for the presence of a caregiver. The portable device is designed for use in hospital-type facilities. The VC150 can also be used in satellite areas or alternate care settings. "Portable" refers to the ability of the VC150 to be easily moved by the caregiver, such as on a roll stand. The VC150 is not intended to be used for continuous monitoring during patient transport.

Device Story

VC150 is a portable, battery-operated vital signs monitor for hospital/clinical use. Inputs include NIBP (oscillometric), SpO2/CO-oximetry (via GE, Masimo, or Nellcor modules), respiration rate, and temperature (SureTemp or Exergen). Device processes signals to display real-time vitals on an 8.4-inch touchscreen. Features include wireless data transmission to hospital systems, optional nurse-call alarm integration, and PDF record export via cable. Operated by clinicians to monitor single patients; provides visual/audible alarms. Benefits include continuous, non-invasive patient assessment to support clinical decision-making.

Clinical Evidence

No clinical studies were required to support substantial equivalence. Evidence based on bench testing, design verification, system integration, and usability testing.

Technological Characteristics

Portable monitor with 8.4-inch LCD touchscreen. Parameters: NIBP (GE SuperSTAT/Classic), SpO2/CO-oximetry (GE TruSignal, Nellcor OxiMax, Masimo Rainbow SET), Temperature (Welch Allyn SureTemp, Exergen TemporalScanner). Connectivity: Wireless network, optional remote alarm cable, USB/cable for PDF export. Power: Battery-operated. Software: Configurable measurement modules.

Indications for Use

Indicated for adult, pediatric, and neonatal patients requiring monitoring of vital signs (BP, pulse, SpO2, SpCO, SpMet, SpHb, respiration rate, temperature) in hospital or clinical settings. Oximax SPD Alert indicated for adults to detect airflow reduction patterns. Exergen thermometer indicated for all ages.

Regulatory Classification

Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's symbol, which is a stylized representation of a human figure. The symbol is composed of three overlapping profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 October 30, 2014 lnnokas Yhtyma Oy Ms. Tiina Kotipalo Director of Quality and Regulatory Affairs Vihikari 10 90440 Kempele Finland Re: K133810 Trade/Device Name: Vital Signs Monitor VC150 Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (including cardiotachometer and rate alarm) Regulatory Class: Class II Product Code: MWI Dated: October 23, 2014 Received: October 27, 2014 Dear Ms. Kotipalo: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Page 2 - Ms. Tiina Kotipalo Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. # Melissa A. Torres -S Bram D. Zuckerman, M.D. For Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K133810 Device Name VC150 #### Indications for Use (Describe) The VC150 is intended to monitor a single patient's vital signs at the site of care or during intra-hospital transport. The noninvasive oscillometric blood pressure parameter is intended for measurement of systolic, and mean arterial blood pressure, as well as pulse rate, for adult, pediatric and neonatal patients. The optional GE TruSignal pulse oximetry and accessories are indicated for continuous nonitoring of functional oxygen saturation (SpO2) and pulse rate, including monitoring conditions of clinical patient motion or low perfusion, with adult, pediatric and neonatal patients. The optional Masimo Rainbow SET® Pulse CO-oximetry and accessories are indicated for the continuous noninvasive monitoring of: · functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate, during both no motion and motion conditions, and for patients who are well or poorly perfusion) for adult, pediatric and neonatal patients, · carboxyhemoglobin saturation (SpCO) for adult and pediatric patients, - · methemoglobin saturation (SpMet) for adult, pediatric and neonatal patients, · total hemoglobin concentration (SpHb) for adult and pediatric patients, and/or · respiratory rate (RRa) for adult and pediatric patients. The optional Nellcor™ oximetry and accessories are indicated for the continuous, noninyasive monitoring of arterial oxygen saturation (SpO2) and pulse rate of adult, pediatric, and neonatal patients during both motion and non-motion conditions, and for patients who are well or poorly perfused. The optional Oximax SPD™ Alert (SPD) feature is intended only for facility-use care of adults to desaturation indicative of repetitive reductions in airflow through the upper airway and into the lungs. The optional Nellcor™ Respiration Rate parameter is intended for the continuous noninvasive monitoring of respiration rate in adult patients who are well perfused during non-motion conditions. The optional Welch Allyn SureTemp Plus electronic thermometer is intended to measure oral, axillary, and rectal temperature of adult and pediatric pational Exergen TemporalScanner thermometer is intended for the intermittent measurement of human body temperature of patients of all ages. A wireless network connection is provided to transmit clinical data into various hospital information systems. An optional remote alarm cable connection is intended to complement visual and audible alarms and not replace the need for the presence of a caregiver. The portable device is designed for use in hospital-type facilities. The VC150 can also be used in satellite areas or alternate care settings. "Portable" refers to the ability of the VC150 to be easily moved by the caregiver, such as on a roll stand. The VC150 is not intended to be used for continuous monitoring during patient transport. {3}------------------------------------------------ X Prescription Use (Part 21 CFR 801 Subpart D) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Innokas Medical. The logo consists of a circular icon with the letter "m" inside, and the words "Innokas Medical" to the right of the icon. The word "Innokas" is in a darker green color, while the word "Medical" is in a lighter gray color. ## 510(k) Summary 510(k) number: K133810 Device Trade Name: VC150 In accordance with 21 CFR 807.92 the following summary of information is provided: | Date: | Oct 30th, 2014 | |------------------------------|--------------------------------------------------------------| | Submitter: | Innokas Yhtymä Oy | | | Vihikari 10 | | | 90440 Kempele | | | Finland | | Primary Contact<br>Person: | Tiina Kotipalo<br>Director of Quality and Regulatory Affairs | | | Innokas Yhtymä Oy | | | Vihikari 10 | | | 90440 Kempele, Finland | | | Phone: +358 44 5623 131 | | | Fax: +358 8 562 3151 | | | Email: tiina.kotipalo@innokasmedical.fi | | Secondary Contact<br>Person: | Risto Salmio<br>Program Manager | | | Innokas Medical Engineering Oy | | | Elimäenkatu 20 A | | | 00510 Helsinki, Finland | | | Phone: +358 40 5520 878 | | | Fax: +358 9 730 535 | | | Email: risto.salmio@innokasmedical.fi | | Device Trade<br>Name: | VC150 | | Common/Usual<br>Name: | Physiological or Vital Signs Monitor; Patient Monitor | | Classification<br>Names: | 21 CFR 870.2300 | | Product Code: | MWI | | Predicate<br>Device(s): | K102426 CARESCAPE V100 Vital Signs Monitor | | | K101680 Welch Allyn Spot Ultra Vital Signs and Welch Allyn | | | Spot Vital Signs LXi | {5}------------------------------------------------ Image /page/5/Picture/0 description: The image contains the logo for Innokas Medical. The logo consists of a circular icon with a stylized letter "m" inside, accompanied by the text "Innokas" in a dark green, sans-serif font, and "Medical" in a lighter gray, sans-serif font positioned below and slightly to the right of "Innokas". The overall design is clean and professional, suggesting a company in the healthcare or medical technology sector. K110028 Masimo Rainbow SET® Radical 7 Pulse CO-Oximeter and Accessories K121806 Bedside Respiratory Patient Monitoring System Device Description: The VC150 is a portable vital signs monitor for use in hospitals and hospital-type facilities. It can be used also in satellite areas or alternate care settings. The monitor is used for adult, pediatric, and neonatal patients, one at a time. The battery-operated monitor offers non-invasive determination of blood pressure (systolic, diastolic, and mean arterial blood pressure), pulse rate, temperature, oxygen saturation, total hemoglobin, carboxyhemoglobin, methemoglobin, and respiration rate. The VC150 monitor includes features and measurement modules that are optional or configurable. The monitors can be equipped with the following parameters and technologies: - NIBP and pulse rate: GE SuperSTAT or Classic (auscultatory) algorithms - SpO2 and pulse rate: GE TruSignal, Nellcor OxiMax, or ● Masimo Rainbow SET - Respiration rate: Nellcor OxiMax or Masimo Rainbow SET - Hemoglobin (SpHb), Methemoglobin (SpMet), Oxygen Content (SpOC), and Carboxyhemoglobin (SpCO): Masimo Rainbow SET - . Temperature: Welch Allyn SureTemp Thermometer or Exergen Temporal Arterial Thermometer The VC150 monitors include a wireless network connection. An integrated printer is optional. The VC150 has an 8.4-inch LCD panel with a touch screen and an alarm light integrated in the front bezel on top of the display. An optional cable connection is available to interface with a traditional nurse-call system for triggering remote alarms. A computer can be connected to the VC150 via a cable to retrieve electronic patient records stored on the device in PDF format. The VC150 is intended to monitor a single patient's vital signs at Intended use: the site of care or during intra-hospital transport. > The noninvasive oscillometric blood pressure parameter is intended for measurement of systolic, diastolic, and mean arterial blood pressure, as well as pulse rate, for adult, pediatric and {6}------------------------------------------------ Image /page/6/Picture/0 description: The image contains the logo for Innokas Medical. The logo consists of a circular icon with a stylized letter "m" inside, accompanied by the text "Innokas" in a dark green, sans-serif font, with the word "Medical" underneath in a lighter gray color. The icon is positioned to the left of the text, creating a balanced and professional visual identity. neonatal patients. The optional GE TruSignal pulse oximetry and accessories are indicated for continuous noninvasive monitoring of functional oxygen saturation (SpO₂) and pulse rate, including monitoring during conditions of clinical patient motion or low perfusion, with adult, pediatric and neonatal patients. The optional Masimo Rainbow SET® Pulse CO-oximetry and accessories are indicated for the continuous noninvasive monitoring of: - . functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate, during both no motion and motion conditions, and for patients who are well or poorly perfused (low perfusion) for adult, pediatric and neonatal patients, - . carboxyhemoglobin saturation (SpCO) for adult and pediatric patients, - methemoglobin saturation (SpMet) for adult, pediatric and neonatal patients, - total hemoglobin concentration (SpHb) for adult and ● pediatric patients, and/or - . respiratory rate (RRa) for adult and pediatric patients. The optional Nellcor™ oximetry and accessories are indicated for the continuous, noninvasive monitoring of arterial oxygen saturation (SpO2) and pulse rate of adult, pediatric, and neonatal patients during both motion and non-motion conditions, and for patients who are well or poorly perfused. The optional Oximax SPD™ Alert (SPD) feature is intended only for facility-use care of adults to detect patterns of desaturation indicative of repetitive reductions in airflow through the upper airway and into the lungs. The optional Nellcor™ Respiration Rate parameter is intended for the continuous, noninvasive monitoring of respiration rate in adult patients who are well perfused during non-motion conditions. The optional Welch Allyn SureTemp Plus electronic thermometer {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for Innokas Medical. The logo consists of a circular icon with the letter "m" inside, and the words "Innokas Medical" to the right of the icon. The word "Innokas" is in a darker green color, while the word "Medical" is in a lighter gray color. is intended to measure oral, axillary, and rectal temperature of adult and pediatric patients. The optional Exergen TemporalScanner thermometer is intended for the intermittent measurement of human body temperature of patients of all ages. A wireless network connection is provided to transmit clinical data into various hospital information systems. An optional remote alarm cable connection is intended to complement visual and audible alarms and not replace the need for the presence of a caregiver. The portable device is designed for use in hospitals and hospitaltype facilities. The VC150 can also be used in satellite areas or alternate care settings. "Portable" refers to the ability of the VC150 to be easily moved by the caregiver, such as on a roll stand. The VC150 is not intended to be used for continuous monitoring during patient transport. - The VC150 employs the same fundamental scientific Technology: technology as its predicate devices. Conclusion from the substantial equivalence comparison submitted with this 510(k) demonstrate that VC150 monitor is as safe and effective and performs substantially equivalent as the following predicate devices: - CARESCAPE V100 Vital Signs Monitor (K102426) ● - Welch Allyn Spot Ultra Vital Signs and Welch Allyn ● Spot Vital Signs LXi (K101680) - Masimo Rainbow SET® Radical 7 Pulse CO-Oximeter ● and Accessories (K110028) - Bedside Respiratory Patient Monitoring System (K121806) Determination of Summary of Non-Clinical Tests: > Substantial Equivalence: The VC150 and its applications comply with voluntary standards as detailed in this premarket submission. The following quality assurance measures have been applied to the development of the system: {8}------------------------------------------------ Image /page/8/Picture/0 description: The image contains the logo for Innokas Medical. The logo consists of a circular icon with a stylized letter "m" inside, accompanied by the text "Innokas" in a darker, bolder font above the word "Medical" in a lighter font. The overall design is clean and professional, suggesting a company in the healthcare or medical field. - Establishment of requirements - Risk management activities - Design reviews . - Design verification testing - System integration testing - System verification and validation testing ● - Performance testing - Safety testing - Usability testing . Summary of Clinical Tests: The subject of this premarket submission, VC150, did not require clinical studies to support substantial equivalence. - The results of these activities demonstrate that the VC150 Conclusion: monitor is as safe, as effective, and performs as well as or better than the predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...