BAXTER SIGMA SPECTRUM INFUSION PUMP WITH MASTER DRUG LIBRARY
K133801 · Baxter Healthcare Corporation · FRN · May 7, 2014 · General Hospital
Device Facts
Record ID
K133801
Device Name
BAXTER SIGMA SPECTRUM INFUSION PUMP WITH MASTER DRUG LIBRARY
Applicant
Baxter Healthcare Corporation
Product Code
FRN · General Hospital
Decision Date
May 7, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used for the controlled administration of fluids. These may include pharmaceutical drugs, blood, blood products and mixtures of required patient therapy. The intended routes of administration consist of the following clinically accepted routes; intravenous, arterial, subcutaneous, epidural or irrigation of fluid space. The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used in conjunction with legally marketed and compatible intravenous administration sets and medications provided by the user. The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is suitable for a variety of patient care environments such as, but not limited to hospitals and outpatient care areas. The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to reduce operator interaction through guided programming, thereby helping to reduce errors. The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used by trained healthcare professionals.
Device Story
Large volume infusion pump; linear peristaltic mechanism; delivers fluids via compatible administration sets. Controlled by primary/secondary processors; software-driven. Master Drug Library (MDL) Editor software (Windows-based) allows facility-specific drug profiles, limits, and advisories to reduce medication errors. Data loaded via wireless network or IrDA. Used in hospitals/outpatient areas; pole-mounted. Healthcare professionals program delivery parameters; pump provides smooth fluid dynamics and volumetric accuracy. Benefits include reduced operator interaction and error mitigation through preprogrammed drug libraries and safety limits.
Clinical Evidence
No clinical testing performed. Evidence consists of non-clinical bench testing (functional, performance, hardware, software, sensors, alarms, network interface) and human factors evaluation in a simulated environment to ensure minimal risk of use errors.
Technological Characteristics
Linear peristaltic infusion pump; electromechanical; software-controlled. Features dual-beam ultrasonic air-in-line detector, set-based anti-free flow clamp. Power: 120VAC/9VDC adapter or Lithium Ion battery (standard or wireless 802.11b/g). Connectivity: IrDA, wireless network, serial port. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-24, UL 60601-1, CAN/CSA C22.2 No 601.1-M90.
Indications for Use
Indicated for controlled administration of fluids (pharmaceuticals, blood, blood products, therapy mixtures) via IV, arterial, subcutaneous, epidural, or irrigation routes. For use by trained healthcare professionals in hospitals and outpatient settings.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
SIGMA Spectrum and Spectrum with Master Drug Library (K042121)
Submission Summary (Full Text)
{0}------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Traditional 5 I0(k) Premarket Notification
Baxter SIGMA Spectrum Infusion Pump with MDL
、
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Page 1 of 12 Section 5-510(k) Summary
| 510(K) SUMMARY | K133801 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | May 6, 2014 |
| Contact: | Baxter Healthcare Corporation<br>711 Park Avenue<br>Medina, New York 14103<br>Registration No: 1314492 |
| | Thomas Sampogna<br>Sr. Director, Regulatory Affairs<br>Phone: 224-270-4307<br>Fax: 847-786-5116<br>Email: thomas_sampogna@baxter.com |
| Trade Name: | Baxter SIGMA Spectrum Infusion Pump with Master Drug Library<br>Model 35700 |
| Common Name: | Infusion Pump |
| Classification Name: | Pump, Infusion |
| Classification: | Infusion Pump as defined in 21 CFR 880.5725, Class II |
| Product Code: | FRN, PHC |
| Predicate Device: | The Baxter SIGMA Spectrum Infusion Pump with Master Drug<br>Library is substantially equivalent to the following predicate<br>device: |
| | SIGMA Spectrum and Spectrum with Master Drug Library,<br>cleared Aug. 26, 2004 (K042121). |
| Device Description: | The Baxter SIGMA Spectrum Infusion Pump with Master Drug<br>Library is a large volume infusion pump system that provides for<br>safe and effective delivery of fluids into a patient in a controlled<br>manner, as identified in 21 CFR, 880.5725. The pump is a<br>software controlled, electromechanical device used for the infusion<br>of pharmaceutical drugs, blood, blood products and mixtures of<br>required patient therapy through administration sets at user<br>selectable rates and volumes. The feedback-controlled, motorized. |
{1}------------------------------------------------
pumping mechanism is of linear peristaltic design and uses inlet and outlet valves for flow control. The pump utilizes a primary and secondary processor to assure safe operation while providing infusion pump capabilities for a wide range of applications.
The pump is specifically manufactured and calibrated for the application of a manufacturer's brand of standard gravity administration sets, as indicated in the device labeling. For use, the administration set is loaded into the infusion pump. After acceptance of program parameters, the pump is started and fluid is propelled by the peristaltic action of the pumping mechanism against the outside surface of the administration set tubing. The pump is controlled to create smooth fluid dynamics, precision volumetric accuracy and uniformity of flow rate. None of the pump materials contact the administration set's fluid path.
The infusion pump is small in comparison to the traditional Large Volume Parenteral (LVP) infusion pumps currently on the market. It is designed to be used in a variety of patient care environments such as, but not limited to hospitals and outpatient care areas using an IV pole mounted configuration.
The Master Drug Library (MDL) Editor is a software application that allows the generation, configuration and management of a downloadable drug library into a SIGMA Spectrum infusion pump. The drug library can be loaded directly into the SIGMA Spectrum infusion pump through a wireless network host or through an Infrared Data Association (IrDA) device. The MDL Editor software operates on a Microsoft Windows® platform.
Using the MDL Editor software application, a facility can provide preprogrammed delivery profiles, advisories and limits for a corresponding drug that is intended for a specific use classification or clinical care area, thus reducing the risk of medication errors. The MDL Editor software application allows the ability to generate both standard or customized drug and fluid reports by clinical care area. The MDL Editor software application also provides a feature to restrict/limit the access of data to only appropriate personnel, providing additional security and rights to specific users.
{2}------------------------------------------------
... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ..
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
| Intended Use/<br>Indication for Use: | The Baxter SIGMA Spectrum Infusion Pump with Master Drug<br>Library is intended to be used for the controlled administration of<br>fluids. These may include pharmaceutical drugs, blood, blood<br>products and mixtures of required patient therapy. The intended<br>routes of administration consist of the following clinically accepted<br>routes; intravenous, arterial, subcutaneous, epidural or irrigation of<br>fluid space. The Baxter SIGMA Spectrum Infusion Pump with<br>Master Drug Library is intended to be used in conjunction with<br>legally marketed and compatible intravenous administration sets<br>and medications provided by the user. |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The Baxter SIGMA Spectrum Infusion Pump with Master Drug<br>Library is suitable for a variety of patient care environments such<br>as, but not limited to hospitals and outpatient care areas. |
| | The Baxter SIGMA Spectrum Infusion Pump with Master Drug<br>Library is intended to reduce operator interaction through guided<br>programming, thereby helping to reduce errors. The Baxter<br>SIGMA Spectrum Infusion Pump with Master Drug Library is<br>intended to be used by trained healthcare professionals. |
| Technological<br>Characteristics: | The Baxter SIGMA Spectrum Infusion Pump with Master Drug<br>Library is substantially equivalent to the predicate device with<br>regards to design, performance and intended use. The following<br>provides a comparison summary of the technical characteristics of |
| Characteristic | Proposed Device | Predicate Device (K042121 |
|---------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Baxter SIGMA Spectrum Infusion Pump | | |
| Pump Type | Linear peristaltic pump | Linear peristaltic pump |
| Routes of<br>Administration | • Intravenous<br>• Arterial<br>• Subcutaneous<br>• Epidural<br>• Irrigation of fluid space | • Intravenous<br>• Arterial<br>• Subcutaneous<br>• Epidural<br>• Irrigation of fluid space<br>• Intrathecal |
| Characteristic | Proposed Device | Predicate Device (K042121) |
| AC Power | Input: 120VAC, 60 Hz / 300 mA<br>Output (Power Adapter P/N<br>35727): 9VDC/1200 mA, | Input: 100VAC - 240VAC, 50-<br>60 Hz / 200 mA<br>Output (Power Adapter P/N<br>35714): 9VDC/800 mA |
| Alarm Volume | Variable at three levels: high,<br>medium and low | Variable at three levels: high,<br>medium and low |
| Air-In-Line | Air In Line: dual-beam ultrasonic<br>detector alarms for large bubbles but<br>allows smaller bubbles to pass.<br>Detects air bubbles >2.5 cm (>1<br>in) (approximately 140 µL in<br>Baxter sets)<br>Detects >1 mL of accumulated<br>air over 15 min., excluding <50<br>µL bubbles, at room temperature<br>Detects >1.5 mL of accumulated<br>air over 15 min., excluding <50<br>µL bubbles, at 15.5℃ (60°F) | Inline Air Detection - Air bubbles of<br>>5/8", alarm if >1 ml over 15<br>minutes, <50 µL bubbles are not-<br>summed |
| Anti-Free Flow System | Set-based, utilizing IV set slide<br>clamp | Set-based, utilizing IV set slide<br>clamp |
| Battery Power and<br>Capacity - Standard<br>Battery | Lithium Ion, minimum 1700<br>mA/h, 7.4 VDC nominal.<br>Capacity 8 hrs (at 125 mL/hr at<br>the highest backlight settings)<br>12 hr recharge time | Lithium Ion, 1800 mA/h, 7.2<br>VDC nominal.<br>Capacity < 11.5hrs (at 115 mL) at<br>10 mL/hr, 14 hrs (1750 mL) at<br>125 mL/hr, 12 hrs (3000mL) at<br>999 mL/hr with backlight off.<br>8 hr. recharge time |
| Battery Power and<br>Capacity - Wireless<br>Battery (802.11b/g) | Lithium Ion, minimum 1700<br>mA/h, 7.4 VDC nominal<br>Capacity 4 hrs (at 125 mL/hr at<br>the highest backlight settings<br>with wireless communication<br>on)<br>16 hr recharge time | N/A - No Wireless Battery Module |
| Characteristic | Proposed Device | Predicate Device (K042121) |
| Dose Modes | mL/hr, mL/kg/min, mL/kg/hr, mL, mL/kg g/hr, mg/hr, mg/kg/hr, mg/min, mg/kg/min, mg/kg/day, mcg/hr, mg/kg/hr, mcg/min, mcg/kg/min, mcg/kg/day, ng/min, ng/kg/min, grams, grams/kg, grams/m², mg, mg/kg, mg/m², mcg, mcg/kg, mcg/m² Units/hr, Units/kg/hr, Units/min, Units/kg/min, mUnits/min, mUnits/kg/hr, mUnits /kg/min, Units, Units/kg, Units/m² mEq/hr, mEq/kg/hr, mEq, mEq/kg mmol/hr, mmol/kg/hr, mmol, mmol/kg | mL/hr g/hr, mg/hr, mg/kg/hr, mg/min, mg/kg/min, mcg/hr, mg/kg/hr, mcg/min, mcg/kg/min, ng/min, ng/kg/min Units/hr, Units/kg/hr, Units/min, Units/kg/min, mUnits/min mEq/hr, mEq/kg/hr |
| External Interfaces/<br>Communications | Bi-directional IrDA Additional Asynchronous Serial Port expansion bus available at battery terminals (manufacturer use only) and wireless-enabled battery pack | Bi-directional IrDA Additional Asynchronous Serial Port expansion bus available at battery terminals (manufacturer use only) |
| Flow rate | 0.5 to 999 mL/hr with 0.1 mL/hr increments from 0.5 to 99.9 mL/hr 1.0 mL/hr increments from 100 to 999 mL/hr | 0.5 to 999 mL/hr with 0.1 mL/hr increments from 0.5 to 99.9 mL/hr 1 mL/hr increments thereafter |
| Infusion Modes | Continuous Primary and Secondary Multi-Step Cyclic TPN Rate Change Bolus Amount/Time (Primary/Secondary) | Continuous Primary and Secondary Titration Ramp/Taper Bolus |
| Characteristic | Proposed Device | Predicate Device (K042121) |
| Intended Use/<br>Indications for Use | The Baxter SIGMA Spectrum<br>Infusion Pump with Master Drug<br>Library is intended to be used for the<br>controlled administration of fluids.<br>These may include pharmaceutical<br>drugs, blood, blood products and<br>mixtures of required patient therapy.<br>The intended routes of<br>administration consist of the<br>following clinically accepted routes;<br>intravenous, arterial, subcutaneous,<br>epidural or irrigation of fluid space.<br>The Baxter SIGMA Spectrum<br>Infusion Pump with Master Drug<br>Library is intended to be used in<br>conjunction with legally marketed<br>and compatible intravenous<br>administration sets and medications<br>provided by the user.<br><br>The Baxter SIGMA Spectrum<br>Infusion Pump with Master Drug<br>Library is suitable for a variety of<br>patient care environments such as,<br>but not limited to hospitals and<br>outpatient care areas.<br><br>The Baxter SIGMA Spectrum<br>Infusion Pump with Master Drug<br>Library is intended to reduce<br>operator interaction through guided<br>programming, thereby helping to<br>reduce errors. The Baxter SIGMA<br>Spectrum Infusion Pump with<br>Master Drug Library is intended to<br>be used by trained healthcare | The Spectrum and Spectrum with<br>Master Drug Library is intended to<br>be used for the controlled<br>administration of intravenous fluids.<br>These may include pharmaceutical<br>drugs, blood, blood products and<br>mixtures of required patient therapy.<br>The intended routes of<br>administration consist of the<br>following clinically accepted routes;<br>intravenous, arterial, subcutaneous,<br>intrathecal, epidural or irrigation of<br>fluid space. The spectrum is<br>intended to be used in conjunction<br>with legally marketed intravenous<br>administration sets and medications<br>provided by the user.<br><br>The Spectrum and Spectrum with<br>Master Drug Library is suitable for<br>many user facility applications such<br>as but not limited to hospitals,<br>outpatient care areas, homecare and<br>ambulatory care services.<br><br>The Spectrum and Spectrum with<br>Master Drug Library is intended to<br>reduce operator interaction through<br>automated programming, thereby<br>helping to reduce errors associated<br>with complex device programming.<br><br>Parameter programming requires<br>trained healthcare professional<br>confirmation of limits and drug<br>therapy to physician's directive. |
| Characteristic | Proposed Device | Predicate Device (K042121) |
| Keep Vein Open<br>(KVO) | At the completion of a primary<br>infusion, the pump will infuse at the<br>KVO rate configured per drug in the<br>Drug Library or the current infusion<br>rate, whichever is lower.<br>The default KVO rate is set at 1<br>mL/hr, but may be configured to<br>between 0.5 - 50 mL/hr.<br>At the completion of a secondary<br>infusion program with secondary<br>callback enabled, the pump will run<br>at a fixed KVO rate of 1 mL/hr. | KVO rate of either 1 mL/hr or the<br>actual rate, whichever is lower at<br>infusion completion alarm. |
| Occlusion Pressure | Adjustable:<br>• High (19 ±9 PSI)<br>• Medium (13 ±6 PSI)<br>• Low (6 ±4 PSI) | Adjustable:<br>• High (19 ±9 PSI)<br>• Medium (13 ±6 PSI)<br>• Low (6 ±4 PSI) |
| Operational Conditions | With Standard Battery<br>• Operating temperature: 60 to<br>90°F (15.6 to 32.2° C), 20 to<br>90% relative humidity non-<br>condensing<br>With Wireless Battery Module<br>• Operating temperature: 60 to<br>80°F (15.6 to 26.7° C), 20 to<br>90% relative humidity non-<br>condensing | With Standard Battery<br>• Operating temperature: 60 to<br>90°F (15.6 to 32.2° C), 20 to<br>90% relative humidity non-<br>condensing<br>N/A - No Wireless Battery Module |
| Overall Size (Pump) | With Standard Battery<br>• Without IV pole clamp - 5.8" H<br>x 4.2" W x 2.5" D<br>• With standard IV pole clamp -<br>5.8" H x 6.4" W x 4.7" D<br>With Wireless Battery Module<br>• Without IV pole clamp - 6.3" H<br>x 4.2" W x 2.5" D<br>• With standard IV pole clamp -<br>6.3" H x 6.4" W x 4.7" D | With Standard Battery<br>• Without IV pole clamp - 5.8" H<br>x 4.2" W x 2.5" D<br>• With standard IV pole clamp -<br>5.8" H x 6.4" W x 4.7" D<br>N/A - No Wireless Battery Module |
| Characteristic | Proposed Device | Predicate Device (K042121) |
| Volumetric Accuracy<br>Volumetric accuracy at stated flow<br>rates: | | Volumetric accuracy using standard<br>Abbott Sets (now Hospira) |
| | 2.0-999 mL/hr ± 5% | 2-999 mL/hr ± 5%…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.