← Product Code [HIG](/productcode/HIG) · K133799

# THERMEDX FLUID MANAGMENT SYSTEM (K133799)

_Thermedx, LLC · HIG · Jun 20, 2014 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K133799

## Device Facts

- **Applicant:** Thermedx, LLC
- **Product Code:** [HIG](/productcode/HIG.md)
- **Decision Date:** Jun 20, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.1700
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology
- **Attributes:** Therapeutic

## Indications for Use

The Thermedx Fluid Management System is intended for irrigation and fluid warming in laparoscopic procedures, and distention, fluid warming, and volume/deficit measurements in endoscopic procedures within gynecology, urology, and orthopedic disciplines.

## Device Story

Thermedx Fluid Management System is a stand-mounted surgical device for fluid management. It utilizes infrared fluid heating technology and a peristaltic fluid pump with closed-loop controls to regulate irrigation and distention fluids. Optional monopolar electrocautery probes are available for cutting and coagulation. The system is operated by clinical staff in surgical settings. It provides real-time fluid warming and volume/deficit measurements to assist surgeons during endoscopic and laparoscopic procedures. The device helps maintain patient body temperature and provides precise fluid control, which is critical for patient safety during minimally invasive surgeries.

## Clinical Evidence

No clinical data provided. Substantial equivalence is supported by non-clinical bench testing, including ISO 10993-1:2009 biological testing (cytotoxicity, sensitization, irritation) and software verification testing of setting limits and valve crack pressures.

## Technological Characteristics

Stand-mounted unit with infrared fluid heating and peristaltic pump. Closed-loop control system. Includes disposable tubing sets. Software-controlled settings for fluid management. Sterilization methods and materials remain unchanged from the predicate.

## Regulatory Identification

A hysteroscopic insufflator is a device designed to distend the uterus by filling the uterine cavity with a liquid or gas to facilitate viewing with a hysteroscope.

## Predicate Devices

- Thermedx 37-5 ([K091939](/device/K091939.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

JUN 2 0 2014

#### 510(k) Summary

## Thermedx Fluid Management System Thermedx, LLC

510(k) Submitter Thermedx, LLC 31200 Solon Road, Unit #1 Solon. Ohio 44139 Douglas L. Carr Contact Person Phone: (440) 542-0883 (440) 542-0920 Fax: E-mail: Dcarn@thermedx.com 30 April 2014 Date Prepared Proprietary Name: Thermedx Fluid Management System Device Name Common Name: Hysteroscopic Insuffiator Requlation Number: 21 CFR & 884 1700 Requiatory Class: II Classification Name: HIG, LGZ, HRX The Thermedx Fluid Management System is intended for irrigation and fluid Intended Use warming in laparoscopic procedures, and distention, fluid warming, and volume/deficit measurements in endoscopic procedures within gynecology. urology, and orthopedic disciplines. The Thermedx device is a fluid management system intended for surgical Device Description applications. The device consists of a main unit and disposable tubing sets. A rolling stand-mounted device houses infrared fluid heating tecnnology and a peristaltic fluid pump with closed-loop controls. The Thermedx monopolar electrocautery probes are an optional accessory for electrosurgical cutting and coaqulation, Thermedx Fluid Management System is identical to the previous version. with Technological no technological differences. Packaging and sterilization means remain Comparison unchanged as well. Changes relate solely to the individual procedure settings and the addition of a high pressure disposable tubing set HLL003 Non clinical testing performed including the following. Non Clinical Tests Biological testing of tube-sets based on 10993-1:2009 including: 1. a. Cytotoxicity using the MEM Elution Assay method (10993-5.2009) b. Sensitization using the Kligman or Buehler methods (10993-10 2010) Irritation using the Intraculaneous Reactivity/Irritation Test (10993-ﻥ 10.2010) 2. Software verification testing: Verify individual setting limits within table and regression testing via running range of simulated use cases. Verify crack pressures of valve for HLL003 is within 240-360mmHg range. 3. Clinical In support to the changes to the procedures and settings please refer to the Information numerous journal articles, practical guidelines, hospital guidelines. Substantial The Thermedx Fluid Management System is substantially equivalent in terms Equivalence of safety and effectiveness to the device as originally cleared. Thermedx 37-5 (K091939).

Page 3 of 26

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 20, 2014

Thermedx LLC Jeff B. Williams Engineering Manager 31200 Solon Rd., Unit 1 Solon, OH 44139

Re: K133799

> Trade/Device Name: Thermedx Fluid Management System Regulation Number: 21 CFR 884.1700 Regulation Name: Hysteroscopic Insufflator Regulatory Class: Class II Product Code: HIG, LGZ, HRX Dated: May 23, 2014 Received: May 27, 2014

Dear Jeff B. Williams,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (sec above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act`s requirements, including, but not limited to: registration and listing (21CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{2}------------------------------------------------

Page 2 - Jeff B. Williams

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (2)CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301)796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Image /page/2/Picture/8 description: The image shows the name "Herbert P. Lerner - S" in a bold, sans-serif font. The text is black against a white background, making it highly legible. The name appears to be a signature or title, with the "- S" possibly indicating a designation or affiliation.

for Benjamin R. Fisher, Ph.D.

Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

#### 510(k) Number

K133799

#### Device Name

Thermedx Fluid Management System

#### Indications for Use

The Thermedx Fluid Management System is intended for irrigation and fluid warming in laparoscopic procedures, and distention, fluid warming, and volume/deficit messurements in endoscopic procedures within gynecology, urology, and orthopedic disciplines.

Type of Use

8 Prescription Use (Part 21 CFR 801 Subpart D)

J Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH)

# Herbert P. Lerner 2014.06.20 12:40:56 -04'00'

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of Information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

---

**Source:** [https://fda.innolitics.com/device/K133799](https://fda.innolitics.com/device/K133799)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
