← Product Code [MSD](/productcode/MSD) · K133796

# NEXSITE HD, HEMODIALYSIS CATHETER FOR LONG TERM USE (32CM), NEXSITE HD, HEMODIALYSIS CATHETER FOR LONG TERM USE (36CM), (K133796)

_Marvao Medical Devices, Ltd. · MSD · Jan 31, 2014 · Gastroenterology, Urology · SESK_

**Canonical URL:** https://fda.innolitics.com/device/K133796

## Device Facts

- **Applicant:** Marvao Medical Devices, Ltd.
- **Product Code:** [MSD](/productcode/MSD.md)
- **Decision Date:** Jan 31, 2014
- **Decision:** SESK
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.5540
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

Indicated for use in attaining long term vascular access for hemodialysis and apheresis. It may be inserted percutaneously and is primarily inserted in the internal jugular vein of an adult patient. Alternate insertion sites include the subclavian vein.

## Device Story

Polyurethane hemodialysis catheter for long-term vascular access; includes Dacron cuff and subcutaneous Polyurethane/Dacron Dermal Ingrowth Support Collar (DISC) for catheter direction. Provided in 32cm, 36cm, and 40cm lengths. Supplied with placement accessories (tunneler, guidewire, introducer/dilator, scalpel). Sterile, single-use device; ethylene oxide sterilized. Used by clinicians for hemodialysis and apheresis procedures. Provides stable vascular access; DISC assists in catheter management.

## Clinical Evidence

Bench testing only. Performance testing included visual and dimensional analysis and pressure vs. flow testing to confirm substantial equivalence to the predicate device. No clinical data provided.

## Technological Characteristics

Polyurethane catheter with Dacron cuff and subcutaneous Polyurethane/Dacron Dermal Ingrowth Support Collar (DISC). Available in 32cm, 36cm, 40cm lengths. Sterilized via ethylene oxide. Complies with ISO 10555-1 standards. Mechanical device; no software or electronic components.

## Regulatory Identification

A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device. (2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle). (3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).

## Special Controls

*Classification.* (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
*1* ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
*2* ) Proper care and maintenance of the device and device exit site;(
*3* ) Removal of the device;(
*4* ) Anticoagulation;(
*5* ) Management of obstruction and thrombus formation; and(
*6* ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

## Predicate Devices

- Marvao NexSite HD Hemodialysis Catheter for long term use ([K121933](/device/K121933.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# JAN 3 1 2014

#### 6. 510(k) Summary

### General Information

|                  | Date Compiled December 12, 2013                                                                                             |
|------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Classification   | Class III, 21 CFR § 876.5540, Blood Access Devices and Accessories, Product<br>code MSD (Catheter, Hemodialysis, Implanted) |
| Trade Name       | NexSite™ HD, Hemodialysis Catheter for long term use                                                                        |
| Model<br>Numbers | NEXHD1553201: NexSite HD, Hemodialysis Catheter for long term use (32 cm)                                                   |
|                  | NEXHD1553601: NexSite HD, Hemodialysis Catheter for long term use (36 cm)                                                   |
|                  | NEXHD1554001: NexSite HD, Hemodialysis Catheter for long term use (40 cm)                                                   |
| Submitter        | Marvao Medical Devices, Ltd.                                                                                                |
|                  | GMIT Innovation in Business Centre, Dublin Road                                                                             |
|                  | Galway, Ireland                                                                                                             |
| Contact          | Christine Nichols RAC                                                                                                       |
|                  | Boston Biomedical Associates                                                                                                |
|                  | 386 West Main Street, Suite 7                                                                                               |
|                  | Northborough, MA 01532                                                                                                      |
|                  | Phone: (508) 351-8632 ext 252                                                                                               |
|                  | Fax: (508) 351-8637                                                                                                         |
|                  |                                                                                                                             |

### Intended Use

Indicated for use in attaining long term vascular access for hemodialysis and apheresis. It may be inserted percutaneously and is primarily inserted in the internal jugular vein of an adult patient. Alternate insertion sites include the subclavian vèin.

# Predicate Devices

#### K121933

Marvao NexSite HD Hemodialysis Catheter for long term use (24cm and 28cm lengths) Manufactured by Marvao Medical Devices Ltd.

{1}------------------------------------------------

Special 510(k) Premarket Notification Submission: Marvao Modified NexSite™ HD Hemodialysis Catheter for Long Term Use

#### Device Description

The Modified NexSite HD Hemodialysis Catheter for long term use is intended for use in attaining long term vascular access for hemodialysis and apheresis. The polyurethane catheter has a Dacron cuff distal to the bifurcation hub. A Polyurethane/Dacron Dermal Ingrowth Support Collar (DISC) supplied with the Catheter is implanted subcutaneously, and is intended to assist with the direction of the catheter. The catheter is provided in three catheter lengths (32cm, 36cm and 40cm). The Catheter and DISC are packaged with accessories (stainless steel Tunneler and Sleeve, 0.038" Guidewire, 16Fr Introducer/Dilator, Coring Scalpel and Luer Caps) that are used to facilitate catheter placement.

The Modified NexSite HD, Hemodialysis Catheter for long term use is provided as a sterile, single-use device, and is sterilized using a validated ethylene oxide process. The Modified NexSite HD Hemodialysis Catheter for long term use is a blood contact device with greater than 30 days of exposure.

#### Comparison to Predicate Devices

Comparison testing was performed on pre-defined characteristics using proposed finished NexSite™ HD devices and commercial predicate devices (K121933). The test results support the substantial equivalence of the Modified NexSite™ HD device to the predicate devices.

#### Testing

Performance testing was performed on the predicate NexSite HD, Hemodialysis Catheter for long term use to assure reliable design and performance in accordance with ISO 10555-1. The materials and manufacturing processes used in the manufacture of the Modified NexSite HD catheters are identical to those used in the manufacture of cleared NexSite HD devices (K121933). Therefore most of the performance testing completed for the cleared NexSite HD catheters was applicable to the Modified NexSite HD catheters, and as such the testing was not repeated. The testing performed specifically on the proposed Modified NexSite HD catheters includes visual and dimensional analysis and pressure vs. flow testing. The test results demonstrate that the Modified NexSite™ HD, Catheter for long term use meets the requirements in the applicable standards and specifications, and is substantially equivalent to legally marketed predicate devices.

{2}------------------------------------------------

Special 510(k) Premarket Notification: Marvao Modified NexSite™ HD Hemodialysis Catheter for Long "ierm Use

#### Guidance

The FDA Guidance on Premarket Notification [(510(k)] Submission for Short-Term and Long-Term Intravascular Catheters, dated 3/16/95, was utilized in order to meet the FDA requirements for content and organization of this submission, as well as the Draft Guidance Industry and Food and Drug Administration Staff Implanted Blood Access Devices for Hemodialysis, June 28, 2013.

## Summary of Substantial Equivalence

Marvao Medical believes the Modified NexSite HD, Hemodialysis Catheter for long term use is substantially equivalent to the predicate NexSite HD Hernodialysis catheter (K121933). The indications for use, methods of operation, design and materials used are either identical or substantially equivalent to existing legally marketed predicate products.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three stripes forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 31, 2014

Marvao Medical Devices, Ltd. % Christine Nichols RAC Regulatory Affairs Manager Boston Biomedical Associates 386 West Main Street, Suite 7 Northborough, MA 01532

K133796 · Re:

> Trade/Device Name: Modified NexSite™ HD Hemodialysis Catheter for Long Term ીકદ Regulation Number: 21 CFR§ 876.5540 Regulation Name: Blood access device and accessories Regulatory Class: III Product Code: MSD Dated: January 10, 2014 Received: January 13, 2014

Dear Christine Nichols,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

{4}------------------------------------------------

# Page 2 - Christine Nichols RAC

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

**Herbert P. Lerner -S**

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health 

{5}------------------------------------------------

Special 510(k) Premarket Notification: Marvao Modified NexSite™ HD Hemodialysis Catheter for Lang Tem Use

#### ડ. Indications for Use Statement

510(k) Number (if known):

K133796

. .

Device Name:

·

Modified NexSite™ HD Hemodialysis Catheter for Long Term Use

Indications for Use:

Indicated for use in attaining long term vascular access for hemodialysis and apheresis. It may be inserted percutaneously and is primarily inserted in the internal jugular vein of an adult patient. Alternate insertion sites include the subclavian vein.

Prescription Use X (Per 21 CFR 801 Subpart D) Subpart C)

AND/OR

Over-The-Counter Use (Per 21 CFR 801

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Herbert P. Lerner -S 2014.01.31 13:59:10 -05'00'

Page 23 of 54

Attachment - Page 2 of 38

---

**Source:** [https://fda.innolitics.com/device/K133796](https://fda.innolitics.com/device/K133796)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
