ATOVEN

K133774 · Bio-Med USA, Inc. · GEX · Jun 6, 2014 · General, Plastic Surgery

Device Facts

Record IDK133774
Device NameATOVEN
ApplicantBio-Med USA, Inc.
Product CodeGEX · General, Plastic Surgery
Decision DateJun 6, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

ATOVEN is a diode laser system designed to delivery of laser light to soft tissue during general surgery procedures. This device intended for treatment of reflux of the saphenous veins of thigh associated with varicose veins and variosities. Non-Contact mode

Device Story

ATOVEN is a diode laser system (1,470 nm wavelength) for soft tissue treatment in general surgery. System components include a CPU, LCD monitor, cooling system, laser controller, fiber optic pullback device, and foot pedal switch. The LCD monitor displays operating parameters, modes, time functions, and system status. Operated by clinicians in a surgical setting, the device delivers optical energy to treat saphenous vein reflux. The clinician monitors the procedure via the interface to control laser delivery. By inducing thermal effects in the target tissue, the device facilitates the treatment of varicose veins, potentially improving patient outcomes related to venous insufficiency.

Clinical Evidence

Bench testing only. The device was evaluated for safety and performance in accordance with IEC 60601-1 (General safety), IEC 60601-1-2 (Electromagnetic compatibility), IEC 60601-1-2-22 (Safety of therapeutic laser equipment), and IEC 60825-1 (Safety of laser products). No clinical data was provided.

Technological Characteristics

Diode laser system; 1,470 nm wavelength; semiconductor laser source. Components: CPU, LCD monitor, cooling system, laser controller, fiber optic pullback device, foot pedal. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1. No fiber, catheters, or guide wires included in submission.

Indications for Use

Indicated for treatment of reflux of the saphenous veins of the thigh associated with varicose veins and varicosities in patients requiring general surgery procedures.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ | | JUN 0 6 2014 | |---------|--------------| | | KJ33774 | | DIOTECH | ATOVEN | | | | # Summary of safety and effectiveness In accordance with Section 513 (1) of the SMDA as defined in 21CFR part 807.3 this summary is submitted to obtain Pre-Market 510 (K) notification. Prepared on May 28, 2014 #### 1. Manufacturer DIOTECH CO LTD will register after cleared #212, Ildongjiaenin, GeoJeong 4 dong, Saha gu, Busan, Rep of Korea Tel: 82 51 292 6236 F 82 51 292 6238 #### 2. Submitter, Correspondent Young Chi Bio-Med USA Inc. 111 Ellison Street, Paterson, NJ07505. U.S.A Tel: 973 278 5222 Fax 201 934 6030 e mail: biomedusa@msn.com #### 3. Name of Device | Trade name | : | ATOVEN | |----------------------|---|------------------------------------| | Classification name | : | Powered, Laser surgical instrument | | Classification panel | : | General & Plastic Surgery | | Common name | : | Diode laser system | | Regulation | : | 880.4810 Class II | | Product Code | : | GEX | #### 4. Legally marketed Predicate Device K110364 LiteBeam+1470 Dornier MedTech #### 5. Intended use ATOVEN is a diode laser system designed to delivery of laser light to soft tissue during general surgery procedures. This device intended for treatment of reflux of the saphenous veins associated with varicose veins and varicosities. Non-Contact mode {2}------------------------------------------------ #### 6. Device Description ATOVEN Diode laser systems are Laser Surgery Systems that use optical energy radiated from semiconductor laser in order to cure varices. with Wave Length 1,470mm, and consists of CPU, LCD monitor, Cooling Systems, laser controller, Fiber optic pullback device included foot pedal switch. LED monitor are displaying various graphic to shows Operating Parameter, Mode, Time function, system status to user. #### 7. Non-Clinical Data. No clinical standards have been established for Diode Laser, but did Safety and Performance test in accordance with below Standard. IEC 60601-1 Medical Electrical Equipment part 1 General Requirement for safety amend 2:1995 IEC 60601-1-2 Electromagnetic compatibility test ED 2007 IEC 60601-1-2-22: 1996 Particular requirement for the safety and Diagnostic of Therapeutic laser Equipment IEC60825_1 Safety of laser products Part 1, section 2, Manufacturing requirement. #### 8. Sterilization and Biocompatibility information This submission does not included any fiber, catheters, Guide Wire etc. accordingly no need any Sterilization and Biocompatibility information. #### 9. Conclusion. ATOVEN Diode laser system, in this submission, is substantially equivalent to several already cleared predicate device in respect to Intended use, Main function, Technology, Principal operation and performance. So, it does not raise any additional concerns regarding safety and effectiveness. Bio-Med USA will update and include in this summary any other information deemed seasonally necessary by the FDA {3}------------------------------------------------ Image /page/2/Picture/0 description: The image is a black and white circular logo for the U.S. Department of Health and Human Services. The logo features the department's symbol, which is a stylized representation of a human figure embracing a globe. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the top half of the circle. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 6. 2014 Bio-Med USA Incorporated Mr. Young Chi President 111 Ellison Street Paterson, New Jersey 07505 Re: K133774 Trade/Device Name: ATOVEN Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: May 2, 2014 Received: May 8, 2014 Dear Mr. Chi: We have reviewed your Scction 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {4}------------------------------------------------ Page 2 - Mr. Young Chi device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, # David Krause -S for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. 510(k) Number (if known) K133774 Device Name ATOVEN Indications for Use (Describe) ATOVEN is a diode laser system designed to delivery of laser light to soft tissue during general surgery procedures. This device intended for treatment of reflux of the saphenous veins of thigh associated with varicose veins and variosities. Non-Contact mode #### Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) # PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED. ### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) # Neil R Ogden -S 2014.06.06 11:53:13 -04'00' This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...