K133770 · Anchor Innovation Medical (A.I.M.) · GAT · Jun 10, 2014 · General, Plastic Surgery
Device Facts
Record ID
K133770
Device Name
A.I.M. KNOTLESS MENISCAL REPAIR DEVICE
Applicant
Anchor Innovation Medical (A.I.M.)
Product Code
GAT · General, Plastic Surgery
Decision Date
Jun 10, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5000
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The A.I.M. Knotless Meniscal Repair Device is an implantable suture retention device which facilitates percutaneous or endoscopic soft tissue repair, including the repair of meniscal tears.
Device Story
Device consists of two implants pre-threaded with USP Size 0 Ultra High Molecular Weight Polyethylene (UHMWPE) suture, pre-loaded into a needle delivery device. Used by surgeons in clinical settings for percutaneous or endoscopic soft tissue repair, including meniscal tears. Needle tips available in straight, curved, or reverse-curved configurations. Device facilitates tissue approximation by deploying implants and tensioning/locking the suture without requiring manual knot tying. Output is a secured suture construct within the tissue. Clinical benefit includes simplified repair procedure and effective tissue fixation.
Clinical Evidence
Bench testing only. Performance testing confirmed pull-out strength and cyclic loading performance are equivalent or superior to the predicate device. Biocompatibility testing conducted per ISO 10993 standards.
Technological Characteristics
Implantable suture retention device. Materials: USP Size 0 Ultra High Molecular Weight Polyethylene (UHMWPE) suture. Delivery: Needle-based delivery system (straight, curved, reverse-curved). Biocompatibility: ISO 10993 compliant. Sterilization: Validated process.
Indications for Use
Indicated for patients requiring percutaneous or endoscopic soft tissue repair, specifically for the repair of meniscal tears.
Regulatory Classification
Identification
Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
Predicate Devices
Smith & Nephew, ULTRA FAST-FIX AND ULTRA FAST-FIX AB MENISCAL REPAIR SYSTEMS (K072322)
Submission Summary (Full Text)
{0}------------------------------------------------
:
## 510(K) SUMMARY: A.I.M. KNOTLESS MENISCAL REPAIR DEVICE
| Submission Date: | December 11, 2013 | JUN 10 2014 |
|------------------------|-------------------|-------------|
| Submitter Information: | | |
| Company: | Anchor Innovation Medical (A.I.M.) |
|----------|------------------------------------|
| | 5410 Edson Lane |
| | Suite 308 |
| | Rockville, MD 20852 |
| Contact Person: | Cary Stalnecker |
|-----------------|-------------------------|
| | Chief Executive Officer |
## Correspondent Information:
Karen M. Becker, Ph.D. 5410 Edson Lane, Suite 308 Rockville, MD 20852
Phone: 301-770-7871 Fax: 301-770-7875 Email: kbecker@anivmed.com
## Device Information:
,
| Trade Name: | A.I.M. Knotless Meniscal Repair Device |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation Name: | Suture, non-degradable |
| Classification: | 21 CFR 878.5000 |
| Product Codes: | GAT |
| Device Class: | Class II |
| Predicate Devices: | Smith & Nephew, ULTRA FAST-FIX AND ULTRA FAST-FIX<br>AB MENISCAL REPAIR SYSTEMS (K072322) |
| Intended Use: | The A.I.M. Knotless Meniscal Repair Device is an implantable<br>suture retention device which facilitates percutaneous or<br>endoscopic soft tissue repair, including the repair of meniscal tears. |
{1}------------------------------------------------
.
1 .
ﺍﻟﻤﺴﺎﻋﺪ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘ
.
. .
1
. . . . . . . .
:
. .
:
.
,
:
| Device Description: | The A.I.M. Knotless Meniscal Repair Device is comprised of two implants pre-threaded with commercially available Ultra High Molecular Weight Polyethylene, USP Size 0 suture pre-loaded into a needle delivery device that will allow the surgeon to repair soft tissue, including a meniscal repair. The needle tip may be available straight, curved, and/or reverse-curved. |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Substantial Equivalence Summary: | The A.I.M. Knotless Meniscal Repair Device is substantially equivalent to the cited predicate, Smith & Nephew's Ultra Fast-Fix and Ultra Fast-Fix AB Meniscal Repair Systems (K072322), having the same, or similar, intended use, indications for use, and fundamental scientific technology. |
| | The A. I.M. Knotless Meniscal Repair Device employs a different method for deploying the implants, and for tensioning and locking the suture. These differences do not raise new questions of safety or effectiveness. |
| Performance Testing: | Bench testing was conducted to confirm that the A.I.M. Knotless Meniscal Repair Device performs as well or better than the predicate on pull-out performance and cyclic loading. |
| Safety Testing: | The A.I.M. Knotless Meniscal Repair Device was determined to be biocompatible based on the component materials, conformity with recognized standards, and testing consistent with the requirements of ISO 10993. Sterilization and packaging validation are consistent with product labeling. |
| Conclusion: | Based on the similarity in intended use, materials, and fundamental scientific technology, performance testing and safety testing, the A.I.M. Knotless Meniscal Repair Device is as safe and effective, and performs as well or better than, the predicate. |
.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized caduceus or a person with outstretched arms.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 10, 2014
Anchor Innovation Medical % Karen M. Becker, Ph.D. 5410 Edson Lane, Suite 308 Rockville, Maryland 20852
Re: K133770
Trade/Device Name: A.I.M. Knotless Meniscal Repair Device Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable poly(ethylene terephthalate) surgical suture Regulatory Class: Class II Product Code: GAT Dated: May 20, 2014 Received: May 21, 2014
Dear Dr. Becker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Karen M. Becker, Ph.D.
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Lori A. Wiggins
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a blank white canvas with two small, dark spots. The spots are irregularly shaped and appear to be randomly placed on the canvas. The spots are small and do not have any distinct features.
K133770 Page 1 of 1
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------510(k) Number (if known): K 133770
Device Name:
A.I.M. Knotless Meniscal Repair Device
Indications for Use:
The A.I.M. Knotless Meniscal Repair Device is an implantable suture retention device which facilitates percutaneous or endoscopic soft tissue repair, including the repair of meniscal tears.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Laurence D. Coyne -A
(Division Sign-Off) Division of Orthopedie Devices 510(k) Number: K133770
Page 1 of 1 _
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.