ACCU-CHEK Performa Blood Glucose Monitoring System

K133741 · Roche Diagnostics · NBW · Aug 29, 2014 · Clinical Chemistry

Device Facts

Record IDK133741
Device NameACCU-CHEK Performa Blood Glucose Monitoring System
ApplicantRoche Diagnostics
Product CodeNBW · Clinical Chemistry
Decision DateAug 29, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

The ACCU-CHEK Performa Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in venous whole blood or fresh capillary whole blood from the fingertips. The ACCU-CHEK Performa Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings as an aid in monitoring the effectiveness of glucose control. This system should only be used with single-use, auto-disabling lancing devices. The ACCU-CHEK Performa Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes. This system is also not for neonatal use. The ACCU-CHEK Performa test strips are for use with the ACCU-CHEK Performa meter to quantitatively measure glucose (sugar) in venous whole blood or fresh capillary whole blood samples from the fingertips.

Device Story

System measures glucose in venous or capillary whole blood; utilizes ACCU-CHEK Performa test strips and meter. Principle of operation: amperometric detection using Mut. Q-GDH enzyme. User inserts test strip into meter; applies blood sample to yellow window. Meter performs integrity check on strip resistance/conductivity before dosing; calculates glucose concentration; displays result. Used in professional healthcare settings; operated by clinicians. Provides quantitative glucose values to aid monitoring of glucose control. Benefits patient by facilitating effective glucose management.

Clinical Evidence

Bench testing only. Precision studies (repeatability/intermediate) performed with venous blood and linearity solutions (n=100-300 per level). Method comparison (n=105 capillary, n=427 venous) against PCA-HK reference method showed high correlation (r=0.995-0.997). User performance study (n=192 lay users) confirmed accuracy. Hematocrit (10-65%), altitude (up to 10,000 ft), temperature, humidity, and sample volume (0.6 μL) studies verified performance. Disinfection efficacy against HBV validated for multiple-patient use.

Technological Characteristics

Electrochemical biosensor; mutant quinoprotein glucose dehydrogenase (Mut. Q-GDH) reagent; Nitrosoaniline mediator. Dimensions: 3.7 x 2.1 x 0.9 in. Power: 3V CR2032 lithium coin cell. Connectivity: None (standalone). Calibration: Lot-specific code key. Standards: ISO 15197, CLSI EP5-A2, CLSI EP7-A2.

Indications for Use

Indicated for quantitative measurement of glucose in venous or capillary whole blood from fingertips in professional healthcare settings for monitoring glucose control. Contraindicated for diabetes diagnosis/screening and neonatal use. Not for use in patients with impaired peripheral circulation (e.g., severe dehydration, hypotension, shock, NYHA Class IV heart failure, peripheral arterial occlusive disease) or critically ill patients.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Auqust 29, 2014 ROCHE DIAGNOSTICS NATHAN CARRINGTON DIRECTOR OF REGULATORY AFFAIRS 9115 HAGUE ROAD INDIANAPOLIS MD 46250 Re: K133741 Trade/Device Name: ACCU-CHEK Performa Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: NBW. LFR Dated: July 25, 2014 Received: July 29, 2014 Dear Mr. Nathan Carrington: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Katherine Serrano -S For : Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) k133741 #### Device Name ACCU-CHEK Performa Blood Glucose Monitoring System #### Indications for Use (Describe) The ACCU-CHEK Performa Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in venous whole blood or fresh capillary whole blood from the fingertips. The ACCU-CHEK Performa Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings as an aid in monitoring the effectiveness of glucose control. This system should only be used with single-use, auto-disabling lancing devices. The ACCU-CHEK Performa Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes. This system is also not for neonatal use. The ACCU-CHEK Performa test strips are for use with the ACCU-CHEK Performa meter to quantitatively measure glucose (sugar) in venous whole blood or fresh capillary whole blood samples from the fingertips. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary ### Introduction According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence. ### 1. Submitter Name, Address, Contact Roche Diagnostic Corporation 9115 Hague Rd. Indianapolis, IN 46250 (317) 521-4793 Contact Person: Nathan Carrington Date Prepared: August 26, 2014 ### 2. Device Name | Proprietary names: | ACCU-CHEK Performa System | |--------------------|-------------------------------| | | ACCU-CHEK Performa Test Strip | | | ACCU-CHEK Performa Meter | Classification name: Glucose dehydrogenase, glucose test system (21 C.F.R. § 862.1345); Class II NBW, Blood Glucose Test System, Over-the-Counter LFR, Glucose Dehydrogenase ### 3. Predicate Device ACCU-CHEK Inform II System (k121679), concurrence received on 11 October 2012. The ACCU-CHEK Inform II blood glucose monitoring system is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings. {4}------------------------------------------------ ## 4. Device Description The ACCU-CHEK Performa System consists of the following: - ACCU-CHEK Performa Meter - ACCU-CHEK Performa test strips ● - ACCU-CHEK Inform II control solutions (k121679) ● - ACCU-CHEK Inform II Linearity Test Kit (k121679) The ACCU-CHEK Performa blood glucose monitoring system is a blood glucose monitoring system that makes use of the ACCU-CHEK Performa test strips, the ACCU-CHEK Performa meter, the ACCU-CHEK Inform II Control Solutions (cleared under k121679), and the ACCU-CHEK Inform II Linearity Test Kit (cleared under k121679). This system is a multi-patient use blood glucose monitoring system intended to be used in a professional environment to quantitatively measure glucose in venous whole blood and fresh capillary whole blood samples drawn from the fingertips. The components of the ACCU-CHEK Performa blood glucose monitoring system are shown below: Image /page/4/Picture/8 description: The image shows a blood glucose meter with labels pointing to different parts of the device. The front of the meter has a display screen that shows results, messages, and test results stored in memory. There are also right and left arrow buttons to enter memory, adjust settings, and scroll through results, as well as a test strip slot to insert the test strip. The back of the meter has an infrared window to transfer data to a computer, a power/set button to turn the meter on or off and set options, and a battery door that can be opened by pushing the tab in the direction of the arrow. Code Kev Slot -Insert code kev into this opening. 510(k) Summary {5}------------------------------------------------ Image /page/5/Figure/0 description: The image shows different components of the ACCU-CHEK blood glucose monitoring system. These components include a test strip container, a test strip, a code key, a battery, and control and linearity solutions. The test strip has a yellow window where blood or control solution is applied, and the battery should be inserted with the (+) symbol facing up. The control and linearity solutions include ACCU-CHEK Inform II control solution levels 1-2 and ACCU-CHEK Inform II Linearity Test Kits containing linearity solution levels 1-6. # 5. Intended Use The ACCU-CHEK Performa Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in venous whole blood or fresh capillary whole blood from the fingertips. The ACCU-CHEK Performa Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings as an aid in monitoring the effectiveness of glucose control. This system should only be used with single-use, auto-disabling lancing devices. The ACCU-CHEK Performa Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes. This system is also not for neonatal use. The ACCU-CHEK Performa test strips are for use with the ACCU-CHEK Performa meter to quantitatively measure glucose (sugar) in venous whole blood or fresh capillary whole blood samples from the fingertips. ## 6. Substantial Equivalence The ACCU-CHEK Performa System is substantially equivalent to the ACCU-CHEK Inform II System. Below is a table that provides a comparison between the ACCU-CHEK Performa System and its predicate device. {6}------------------------------------------------ # Similarities Table | System Feature/Claim | Detail | |-------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Test Strip | Same: The Performa and Inform II test strips are identical. | | Test Strip Production<br>Processes and Lot-Release<br>Criteria | Same: The Performa and Inform II test strips are identical. | | Meter firmware for<br>measurement sequence and<br>blood glucose calculation<br>algorithm | Same: The fundamental scientific technology for the<br>measurement of blood glucose has not changed from the<br>predicate. | | Meter Analog and Digital<br>device integrated circuits<br>used for glucose<br>measurement | Same | | Meter measurement engine<br>microprocessor | Same | | Underdose detection and<br>system fail-safes | Same | | Meter production site and<br>assembly methodology | Same | | Meter production test and<br>calibration methods | Same | | Integrity Check for Strip | Same: Early in the measurement sequence, the meter<br>measures the resistance of the gold on the un-dosed strip to<br>assure that it has been properly inserted and that the quality<br>is not compromised. The meter measures the background<br>conductivity and electrical current prior to dosing to assure<br>that the reagent quality is not compromised or that the strip<br>was not prematurely dosed. | {7}------------------------------------------------ # Similarities Table (continued) | System<br>Feature/Claim | ACCU-CHEK Performa Blood<br>Glucose Monitoring System | ACCU-CHEK Inform II Blood<br>Glucose Monitoring System<br>Predicate<br>(k121679) | |------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for<br>Use | Quantitative measurement of<br>glucose (sugar) in venous whole<br>blood or fresh capillary whole<br>blood samples. | Quantitative measurement of<br>glucose (sugar) in venous whole<br>blood, arterial whole blood,<br>neonatal heelstick, or fresh<br>capillary whole blood samples. | | Test Principle | Amperometric Detection | Amperometric Detection | | Enzyme | Mut. Q-GDH | Mut. Q-GDH | | Sample<br>Hematocrit | 10 to 65% | 10 to 65% | | Maximum<br>Altitude | 10,000 feet | 10,000 feet | | Sample Volume | 0.6 µL | 0.6 µL | | Test Time | 5 seconds | 5 seconds | | Operating<br>Temperature and<br>Relative<br>Humidity | 16 to 35°C<br>(61 to 95°F)<br>10 to 80% r.h. | 16 to 35°C<br>(61 to 95°F)<br>10–80% r.h. | | Coding | Lot-specific blood glucose<br>measurement parameters are<br>programmed into ROM code key | Lot-specific blood glucose<br>measurement parameters are<br>programmed into ROM key | | Precision | For response targets below 75<br>mg/dL, the SD is ≤ 5.0 mg/dL,<br>and for response targets ≥ 75<br>mg/dL, the CV is ≤ 5.0%. | For response targets below 75<br>mg/dL, the SD is ≤ 5.0 mg/dL,<br>and for response targets ≥ 75<br>mg/dL, the CV is ≤ 5.0%. | | Double Dosing | No | No | | Alternate Site<br>Testing | No | No | | System<br>Feature/Claim | ACCU-CHEK Performa Blood<br>Glucose Monitoring System | ACCU-CHEK Inform II Blood<br>Glucose Monitoring System<br>Predicate<br>(k121679) | | Closed and Open<br>Vial Shelf Life<br>Stability | 18 months | 18 months | | Control Solutions | Aqueous, 2 levels, uses ACCU-<br>CHEK Inform II Control<br>Solutions | Aqueous, 2 levels, uses ACCU-<br>CHEK Infrom II Control<br>Solutions | | Linearity Kit | Aqueous, 6 levels, uses ACCU-<br>CHEK Inform II Linearity Test<br>Kit | Aqueous, 6 levels, uses ACCU-<br>CHEK Inform II Linearity Test<br>Kit | | Primary<br>Packaging | Standard flip top vial | Standard flip top vial | | Limitations of<br>Procedure | Galactose >15 mg/dL will cause<br>overestimation of blood glucose<br>results. | Galactose >15 mg/dL will cause<br>overestimation of blood glucose<br>results. | | | Lipemic Samples >1800 mg/dL | Lipemic Samples >1800 mg/dL | | | Intravenous administration of<br>ascorbic acid which results in<br>blood concentrations of ascorbic<br>acid >3 mg/dL will cause<br>overestimation of blood glucose<br>results | Intravenous administration of<br>ascorbic acid which results in<br>blood concentrations of ascorbic<br>acid >3 mg/dL will cause<br>overestimation of blood glucose<br>results | {8}------------------------------------------------ # Similarities Table (continued) {9}------------------------------------------------ ## Differences Table | System<br>Feature/Claim | ACCU-CHEK Performa Blood<br>Glucose Monitoring System | ACCU-CHEK Inform II Blood<br>Glucose Monitoring System<br>Predicate<br>(k121679) | |-----------------------------------------------------------------|--------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------| | Meter Physical<br>Appearance<br>Including Size<br>and Weight | 3.7 in x 2.1 in x 0.9 in (LWH),<br>0.14 lbs<br>93 mm x 52 mm x 22 mm<br>(LWH), 62 g | 7.60 in x 3.74 in x 1.73 in<br>(LWH), 0.83 lbs<br>193 mm x 95 mm x 44 mm<br>(LWH), 376 g | | Ergonomics of<br>User Interface | Two buttons located on the face<br>of the meter; power button on top<br>edge of the meter. | Power button located on the face<br>of the meter; touch-sensitive<br>display. | | Code Key Port | Code key inserts directly into<br>code key slot in meter. | Code key inserts into code key<br>reader which transfers data to<br>meter via IR communication | | Battery | One 3-volt lithium type CR2032<br>coin cell | 3.7 V rechargeable battery pack<br>(lithium technology) | | Data<br>Management<br>Features | Simplified for basic blood<br>glucose testing. | Additional firmware for added<br>features, such as the tracking of<br>patient information | | Bar Code<br>Scanner | None | Present | | Transmission of<br>Retrospective<br>Data to External<br>Devices | None | Wirelessly to WLAN through RF<br>communication or to docking<br>station base unit through IR data<br>port | | Measuring Range | 20 – 600 mg/dL | 10 – 600 mg/dL | ## 7. Data demonstrating substantial equivalence Performance testing on the ACCU-CHEK Performa System demonstrated that the device meets the performance requirements for its intended use. The data demonstrate that the system is substantially equivalent to the predicate device. {10}------------------------------------------------ Below is the method comparison data for the system: Results for glucose concentrations less than 75 mg/dL | Within +5 mg/dL | Within +10 mg/dL | Within + 15 mg/dL | |-----------------|------------------|-------------------| | 17/19 (89.5%) | 18/19 (94.7%) | 19/19 (100%) | Results for glucose concentrations greater than or equal to 75 mg/dL | Within +5 % | Within +10 % | Within + 15 % | Within + 20 % | |---------------|---------------|---------------|---------------| | 50/86 (58.1%) | 78/86 (90.7%) | 84/86 (97.7%) | 85/86 (98.8%) | Below is the repeatability (wtihin lot) precision for the system: | Blood | 1 | 2 | 3 | 4 | 5 | |--------------|------|------|-------|-------|-------| | N | 100 | 100 | 100 | 100 | 100 | | Mean [mg/dL] | 37.1 | 81.1 | 134.6 | 216.2 | 345.3 | | SD [mg/dL] | 1.7 | 2.7 | 4.4 | 7.9 | 10.2 | | CV [%] | 4.7 | 3.4 | 3.3 | 3.6 | 3.0 | Below is the reproducibility (intermediate or day-to-day) precision for the system: | Control solutions | Low | Mid | High | |-------------------|------|-------|-------| | N | 100 | 100 | 100 | | Mean [mg/dL] | 45.8 | 118.8 | 310.8 | | SD [mg/dL] | 1.5 | 2.7 | 6.6 | | CV [%] | 3.2 | 2.3 | 2.1 |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...