MEGA SOFT UNIVERSAL PATIENT RETURN ELECTRODE

K133726 · Megadyne Medical Products, Inc. · GEI · Jan 24, 2014 · General, Plastic Surgery

Device Facts

Record IDK133726
Device NameMEGA SOFT UNIVERSAL PATIENT RETURN ELECTRODE
ApplicantMegadyne Medical Products, Inc.
Product CodeGEI · General, Plastic Surgery
Decision DateJan 24, 2014
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic, Pediatric

Intended Use

The intended use of the MegaSoft Universal Patient Return Electrode is to conduct monopolar electrosurgical energy from target tissue of a patient back to an electrosurgical unit (ESU), or generator. This device is intended to be used whenever monopolar electrosurgy is indicated. Electrosurgical use is restricted to use with isolated monopolar electrosurgical generators.

Device Story

MegaSoft Universal Patient Return Electrode; reusable pad (36" x 20" x 0.135"); conductive layer encapsulated in viscoelastic Akton polymer and urethane film. Device placed on operating surface; patient lies on pad. Connects to monopolar electrosurgical unit (ESU) via two-conductor cable and DetachaCable. Functions as return path for electrosurgical energy via capacitive coupling; current-limiting design (<100 mA/cm2) prevents return electrode site burns. Used in clinical operating environments; operated by surgical staff. Benefits patient by providing safe return path for high-frequency surgical current, minimizing risk of thermal injury.

Clinical Evidence

Bench testing only. Device tested for conformance to IEC 60601-1 (general safety) and IEC 60601-2-2 (high-frequency surgical equipment safety). No clinical data provided.

Technological Characteristics

Conductive layer strain-relieved with urethane; dielectric layer of Akton viscoelastic polymer; encapsulated in urethane film. Dimensions: 36" x 20" x 0.135". Connectivity: two-conductor cable to ESU. Current-limiting design (<100 mA/cm2). Standards: IEC 60601-1:2005, IEC 60601-2-2:2009.

Indications for Use

Indicated for patients weighing ≥ 0.8 lb (350 grams) undergoing procedures requiring monopolar electrosurgery. Restricted to use with isolated monopolar electrosurgical generators. Not intended for RF ablation.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # K133726 Paqe 1 of 3 JAN 2 4 2014 #### Section 4 510(k) Summary December 4, 2013 ## A. Submitter's Name / Address Ronda K. Magneson Director, Regulatory Affairs Megadyne Medical Products, Inc: 11506 South State Street Draper, UT 84020 (801) 576-9669 (801) 576-9698 fax #### B. Contact Person Primary: Ronda K. Magneson Director of Regulatory Affairs Megadyne Medical Products, Inc. 11506 South State Street Draper, UT 84020 rmagneson@megadyne.com (801) 576-9669 (801) 576-9698 fax Alternate: Katie Hoff Regulatory Affairs Specialist II Megadyne Medical Products, Inc. 11506 South State Street Draper, UT 84020 khoff(@megadyne.com (801) 576-9669 (801) 576-9698 fax #### C. Megadyne's Manufacturing Facility Megadyne Medical Products, Inc. 11506 South State Street Draper, UT 84020 (801) 576-9669 (801) 576-9698 fax #### D. Device Name | Common Name: | Device, electrosurgical, cutting & coagulation &<br>accessories | |----------------------------|------------------------------------------------------------------------------------| | Trade Name: | MegaSoft Universal Patient Return Electrode | | Classification (if known): | 21 CFR 878.4400, Electrosurgical cutting and<br>coagulation device and accessories | Megadyne Medical Products, Inc. 510(k): Mega Soft Universal Reusable Patient Return Electrode Page 12 of 64 ﺗﮭ י {1}------------------------------------------------ #### E. Predicate Devices ﺗﮭ The predicate device is Megadyne's Mega Soft Return Electrode Pad which was cleared for marketing via 510(k) # K080741 by FDA's Office of Device Evaluation on December 12, 2008 ## F. Applicant Device Description The MegaSoft Universal Patient Return Electrode is constructed of a layer of conductive material strain-relieved with two sheets of urethane material, and sealed between two symmetric layers of a viscoelastic polymer called Akton® The Akton polymer is encapsulated by a layer of urethane film. One twoconductor cable connects the conductive layer of the device to a two conductor DetachaCable . The DetachaCable is connected to a standard monopolar electrosurgical unit (ESU). The device cable is insulated, strain-relieved, and connected well inside the device to prevent patient or user burns. In use, this device will lay on the operating surface with the patient lying on either side of the pad. The pad size is approximately 36" x 20" x 0.135" and is intended for patients weighing ≥ 0.8 lb (350 grams). #### G. Applicant Device Intended Use The intended use of the MegaSoft Universal Patient Return Electrode is to conduct monopolar electrosurgical energy from target tissue of a patient back to an electrosurgical unit (ESU), or generator. This device is intended to be used whenever monopolar electrosurgy is indicated. Electrosurgical use is restricted to use with isolated monopolar electrosurgical generators. #### H. Technological Characteristics The proposed device shares the same technological characteristics found in the Megadyne predicate devices. #### I. Safety information ﮟ The only current flow from the patient to the MegaSoft Universal Patient Return Electrode Pad is via capacitive coupling. This device is designed to be current limiting (<100 mA/cm2) to prevent the patient from getting return electrode site burns. Megadyne Medical Products, Inc. 510(k): Mega Soft Universal Reusable Patient Return Electrode Page 13 of 64 {2}------------------------------------------------ The Akton is used as a dielectric layer. The desirable attribute of this polymer is that it compresses but does not laterally move under pressure, thus maintaining dielectric protection. ー Questions of safety and effectiveness are the same for this device as they are for the predicate device and other patient return electrodes on the market. There are no new technologies incorporated into the device. Megadyne has conducted extensive testing to ensure conformance to the following voluntary standards: IEC 60601-1:2005, Medical electrical equipment, Part 1: General requirements for safety and essential performance IEC 60601-2-2:2009, Medical electrical equipment, Part 2: Particular requirements for the safety of high frequency surgical equipment Megadyne Medical Products, Inc. 510(k): Mega Soft Universal Reusable Patient Return Electrode Page 14 of 64 {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Megadyne Medical Products Incorporated Ms. Ronda K. Magneson Director of Regulatory Affairs 11506 South State Street Draper, Utah 84020 January 24, 2014 Re: K133726 Trade/Device Name: MegaSoft Universal® Patient Return Electrode Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: December 26, 2013 Received: December 27, 2013 Dear Ms. Magneson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {4}------------------------------------------------ Page 2 - Ms. Ronda K. Magneson device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Radiological Health # Joshua C. Nipper -S Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and For Enclosure {5}------------------------------------------------ ്കുന്നു. അവലംബം #### Indications for Use Statement Section 3 510(k) Number (if known): K133726 Device Name: # MegaSoft Universal® Patient Return Electrode Indications for use: This device is designed to be used whenever monopolar electrosurgery is indicated. The intended use of this device is to conduct monopolar electrosurgical energy from target tissue of a patient back to one or two electrosurgical units (ESU), or generators. Electrosurgical use is restricted to use with isolated monopolar electrosurgical generators. The device is not intended for RF ablation. Concurrence of CDRH, Office of Device Evaluation (ODE) | DSD-DIVISION SIGN-OFF | | |------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Division of Surgical Devices | | | 510(k) Number: | K133726 | | | Long H. Chen -A | | | Digitally signed by Long H. Chen-A : DN: c=US, o=US. Government, ou=HHS, ou=FDA, ou=People, cn=Long H. Chen-A, 0.9.2342.19200300.100.1.1=130036 9056 Date: 2014.01.23 08:43:53 -05'00' | | | for BSA | Prescription Use__X (Per 21 CFR 801.109) OR Over-The-Counter Use_ Megadyne Medical Products, Inc. Megadyne Medical Products, 210. 5 10(k): Mega Soft Universal Reusable Patient Return Electrode Page 11 of 64
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%