K133724 · Medivators, Inc. · NIH · Aug 27, 2014 · Gastroenterology, Urology
Device Facts
Record ID
K133724
Device Name
MINNCARE HD
Applicant
Medivators, Inc.
Product Code
NIH · Gastroenterology, Urology
Decision Date
Aug 27, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5665
Device Class
Class 2
Indications for Use
Minncare HD is intended for the disinfection of water purification systems for hemodialysis. It should be diluted to a 1% concentration (1 part Minncare HD to 99 parts water) and used for a minimum contact time of 36 minutes at 20°C.
Device Story
Minncare HD is a liquid chemical disinfectant solution containing stabilized hydrogen peroxide, peracetic acid, and acetic acid. It is used to disinfect hemodialysis water purification systems. The operator dilutes the solution to 1% (1:99 ratio) and applies it to the system for a minimum of 36 minutes at 20°C. The operator confirms the concentration at the furthest point of the distribution loop using indicator test strips. After disinfection, the system is rinsed, and residual levels are verified to be 1 ppm or less using residual test strips. The device is supplied in high-density polyethylene bottles.
Clinical Evidence
No clinical data. Performance supported by non-clinical bench testing, including physical/chemical property analysis, sporicidal, tuberculocidal, fungicidal, virucidal, and bactericidal efficacy testing, simulated-use disinfection efficacy, material compatibility, test strip performance, and stability testing.
Technological Characteristics
Liquid chemical disinfectant; active ingredients: peracetic acid and hydrogen peroxide. Supplied in high-density polyethylene (HDPE) bottles. Requires dilution to 1% concentration. Monitored via indicator test strips for minimum recommended concentration (MRC) and residual PAA levels. Shelf-life: 12 months.
Indications for Use
Indicated for the disinfection of water purification systems used in hemodialysis.
Regulatory Classification
Identification
A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.
Special Controls
*Classification.* Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features the department's name in a circular arrangement around the perimeter. At the center of the seal is a stylized representation of a human figure, composed of three overlapping profiles facing to the right. The seal is rendered in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 27, 2014
MEDIVATORS, INC. Mrs. Kinnari Shah Senior Regulatory Affairs Specialist 14605 28th Ave. North Minneapolis, MN 55447
Re: K133724
Trade/Device Name: Minncare HD® Disinfectant Regulation Number: 21 CFR 876.5665 Regulation Name: Disinfectant, subsystem, water purification Regulatory Class: II Product Code: NIH Dated: July 30, 2014 Received: July 31, 2014
Dear Mrs. Shah:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{2}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Erin I. Keith -S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Medivators, a Cantel Medical Company. The logo consists of a blue circular graphic on the left, followed by the word "MEDIVATORS" in blue, sans-serif font. Below the company name is the text "A Cantel Medical Company" in a smaller, sans-serif font, also in blue. The logo is clean and professional, conveying a sense of trust and reliability.
## Indications for Use
510(k) number (if known): K133724
Device Name: Minncare® HD Disinfectant
#### Indications for Use:
Minncare HD is intended for the disinfection of water purification systems for hemodialysis. It should be diluted to a 1% concentration (1 part Minncare HD to 99 parts water) and used for a minimum contact time of 36 minutes at 20°C.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Medivators, a Cantel Medical Company. The logo consists of a blue circular graphic on the left, followed by the word "MEDIVATORS" in blue, sans-serif font. Below the company name, in a smaller font size, is the text "A Cantel Medical Company" also in blue. The overall design is clean and corporate, suggesting a professional and medical-related business.
### Minncare® HD Disinfectant
#### 510(k) Summary
| Manufacturer: | Medivators Inc., a Cantel Medical Company |
|----------------------|------------------------------------------------------------------------|
| Address: | 14605 28th Avenue North<br>Minneapolis, MN 55447 USA<br>(800) 328-3345 |
| Official Contact: | Kinnari Shah<br>Senior Regulatory Affairs Specialist |
| Date Prepared: | 13th August 2014 |
| Trade Name: | Minncare® HD Disinfectant |
| Common Name: | Water Purification System Disinfectant |
| Classification Name: | Disinfectant, subsystem, water purification |
| Product Code: | NIH |
| Device Class: | II |
| Classification Reg: | 876.5665 |
| 510k: | K133724 |
Medivators Inc. has supplied the following information to the US Food and Drug Administration to support substantial equivalence of the Minncare HD Disinfectants to other disinfectants currently cleared for sale in the U.S.
#### 1. Intended Use
Minncare HD is intended for the disinfection of water purification systems for hemodialysis. It should be diluted to a 1% concentration (1 part Minncare HD to 99 parts water) and used for a minimum contact time of 36 minutes at 20℃.
#### 2. Device Description
Minncare HD is a clear liquid disinfectant solution that consists of a stabilized mixture of hydrogen peroxide, peracetic acid and acetic acid. Minncare HD is intended to be used for hemodialysis water purification system disinfection when diluted to a 1% concentration (1 part Minncare HD to 99 parts water) and applied for a minimum contact time of 36 minutes at 20℃. The active microbicidal ingredient is peracetic acid (PAA), and the recommended use concentration of 1%. The concentration of the use solution should be confirmed with 1% Minncare HD indicator test strips to have reached the furthest point of the distribution loop during the system disinfection contact time. Upon completion of disinfection the system must be rinsed to remove residual levels of the disinfectant. Residual levels should be
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Medivators, a Cantel Medical Company. The logo consists of a blue circular graphic on the left, followed by the word "MEDIVATORS" in blue, block letters. Below the company name is the text "A Cantel Medical Company" in a smaller, blue font. The logo is simple and professional, conveying a sense of trust and reliability.
checked using Minncare HD residual test strips to ensure residual levels of 1 ppm PAA or less. Minncare HD is supplied in cases containing high density polyethylene plastic bottles. Individual bottles are labeled with all information necessary to use the device safely.
#### 3. Comparison to Other Device in Commercial Distribution Within the United States
MEDIVATORS Inc. believes Minncare HD is substantially equivalent in formulation, performance and indications to its predicate device - Hemoclean Disinfectant which is cleared under 510k - K023064. Hemoclean is a peracetic acid based liguid chemical germicide disinfectant solutions with similar chemistry, intended use and disinfectant performance to Minncare HD. Below is the comparison between the subject and the predicate device -
| Item | Minncare HD<br>MEDIVATORS Inc. | Hemoclean<br>Disinfectant<br>KRD CO., LTD<br>(K023064) |
|------------------------------------------------------------|------------------------------------------------|-----------------------------------------------------------------------|
| Active Ingredient | Peracetic Acid/ Hydrogen<br>Peroxide | Peracetic Acid/ Hydrogen<br>Peroxide |
| Device Use Indication | Water Purification<br>Systems for Hemodialysis | Hemodialysis machines,<br>Water Purification<br>Systems and Dialyzers |
| Form | Liquid solution | Liquid solution |
| Dilution Required | Yes | Yes |
| Minimum Contact Time | 36 minutes | 10 minutes |
| Use Temperature | 20°C | 20°C |
| Minimum Use<br>Concentration | 1% | 3% |
| Active Ingredient<br>Nominal Concentration | 5% / 22% | 1.7%/5.3% |
| Single Use | Yes | Yes |
| Sterilization Claim | No | No |
| Use and Compatibility<br>with Water Purification<br>System | Yes | Yes |
| Shelf-Life | 12 months | 12 months |
| Test Strip Used to<br>Monitor MRC | Yes | Yes |
{6}------------------------------------------------
#### 4. Summary of Non-Clinical Performance Data
FDA Guidance for content of premarket notification submissions for water purification system disinfectants (product code NIH) has not been established, and thus performance testing requirements were based on related FDA Guidance for Content and Format of Premarket Notification [510(k)] Submission for Liquid Chemical Sterilants/High Level Disinfectants (Jan 2000). However, it is important to note that Minncare HD is not intended to be used on critical or semi-critical devices and high-level disinfection is not claimed.
Performance testing has been conducted to show that Minncare HD is safe and effective for its intended use. The following types of data/information are provided in this submission in support of substantial equivalence to predicate device and to demonstrate that Minncare HD performs safely as intended.
- Detailed description of physical and chemical properties .
- Proposed labeling .
- Sporicidal, tuberculocidal, fungicidal, virucidal and bactericidal efficacy ●
- Simulated-use disinfection efficacy .
- . Material compatibility
- Test strip performance .
- Stability .
- Risk Analysis ●
#### 5. Conclusion
The performance testing data indicates that the subject device, Minncare HD, is substantially equivalent to the predicate device Hemoclean Disinfectant (K023064).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.