K133701 · Corventis, Inc. · QYX · Feb 6, 2014 · Cardiovascular
Device Facts
Record ID
K133701
Device Name
NUVANT MCT SYSTEM
Applicant
Corventis, Inc.
Product Code
QYX · Cardiovascular
Decision Date
Feb 6, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1025
Device Class
Class 2
Indications for Use
The NUVANT Mobile Cardiac Telemetry (MCT) System is intended to continuously measure, record and periodically transmit physiological data. The System is indicated for those patients who require monitoring for the detection of non-lethal cardiac arrhythmias such as, but not limited to, supraventricular tachycardias (e.g. atrial fibrillation, atrial flutter, paroxysmal SVTs), ventricular ectopy, bradyarrhythmias and conduction disorders. The NUVANT System monitors, derives and displays: - ECG - Heart Rate
Device Story
NUVANT MCT System provides ambulatory cardiac monitoring. System components: PiiX (wearable sensor) and zLink (handheld transceiver). PiiX continuously monitors ECG/heart rate; uses proprietary algorithms (rate, rhythm, morphology) to detect abnormalities. Upon detection, PiiX transmits data via Bluetooth to zLink; zLink transmits to secure Corventis server via cellular network. Patients use push button to trigger manual ECG transmission during symptoms. Certified cardiographic technicians at Corventis Monitoring Center review data, document symptoms, and prepare clinical reports for prescribing physicians via web portal. System aids diagnosis/identification of cardiac conditions/trends. Modifications include increased PiiX memory, two-part PiiX construction, and replacement of trigger magnet with push button.
Clinical Evidence
Bench testing only. No clinical data presented. Substantial equivalence supported by verification and validation testing, including biocompatibility (ISO 10993), mechanical, electrical, firmware, and system-level validation. Performance of cardiac rhythm and ST-segment measurement algorithms verified per AAMI/ANSI EC57.
Indicated for patients requiring monitoring for detection of non-lethal cardiac arrhythmias, including supraventricular tachycardias (atrial fibrillation, atrial flutter, paroxysmal SVTs), ventricular ectopy, bradyarrhythmias, and conduction disorders.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 6, 2023
Corventis, Inc. Cheryl Swanson Director, Quality And Regulatory Affairs 1410 Energy Park Drive, Suite 1 St. Paul. Minnesota 55108
Re: K133701
Trade/Device Name: Nuvant Mct System Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm) Regulatory Class: Class II Product Code: QYX, DSI
Dear Cheryl Swanson:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated February 6, 2014. Specifically, FDA is updating this SE Letter because FDA has created a new product code to better categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Jennifer Kozen, OHT2: Office of Cardiovascular Devices, 301-796-5813, Jennifer.Kozen(@fda.hhs.gov.
Sincerely.
Jennifer W. Shih -S
Jennifer Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Public Health Service
Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half of the circle. Inside the circle is an emblem featuring a stylized caduceus, with three human profiles facing to the right, representing health and human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 6, 2014
Corventis. Inc Cheryl Swanson Director, Quality and Regulatory Affairs 1410 Energy Park Drive, Suite 1 St. Paul, MN 55108
Re: K133701 Trade/Device Name: NUVANT MCT System Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: DSI Dated: January 7, 2014 Received: January 8, 2014
Dear Cheryl Swanson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Owen P. Faris -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### SECTION 4 - INDICATIONS FOR USE STATEMENT
Indications for Use
510(k) Number (if known): N/A
Device Name: NUVANT® Mobile Cardiac Telemetry System
Indications for Use:
The NUVANT Mobile Cardiac Telemetry (MCT) System is intended to continuously measure, record and periodically transmit physiological data. The System is indicated for those patients who require monitoring for the detection of non-lethal cardiac arrhythmias such as, but not limited to, supraventricular tachycardias (e.g. atrial fibrillation, atrial flutter, paroxysmal SVTs), ventricular ectopy, bradyarrhythmias and conduction disorders. The NUVANT System monitors, derives and displays:
- ECG
- · Heart Rate
Prescription Use _____________________________________________________________________________________________________________________________________________________________
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Digitally signed by
Owen P. Faris -S
Date: 2014.02.06
15:10:29-05'00'
{4}------------------------------------------------
# SECTION 5 - 510(K) SUMMARY
The following table provides background information regarding this Special 510(k) submission:
| Date Prepared: | December 2, 2013 |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submission Type: | Special 510(k): Device Modification |
| Device Trade Name: | NUVANT MCT System |
| Device Model Number: | MCTX-SY-nnn |
| Device Common Name: | Mobile Cardiac Telemetry (MCT) |
| Classification Regulation | Arrhythmia detector and alarm |
| Name: | (including ST-segment measurement<br>and alarm) |
| Product Code | DSI |
| Classification Regulation | 21 CFR 870.1025 |
| 510(k) Submitter/Owner<br>Name: | Corventis, Inc. |
| Submitter Address: | 1410 Energy Park Drive, Suite 1<br>St. Paul, MN 55108 |
| Contact Person: | Cheryl Swanson<br>Director, Quality and Regulatory Affairs<br>1410 Energy Park Drive, Suite 1<br>St. Paul, MN 55108<br>(651) 925-3803 (office)<br>(651) 762-1824 (cell)<br>(551) 389-3251 (fax) |
| Predicate Device: | NUVANT Mobile Cardiac Telemetry<br>System, cleared by FDA under 510(k)<br>number K113372 on March 7, 2012. |
| Prior Submissions for this<br>Change: | None |
{5}------------------------------------------------
## 5.1 DEVICE DESCRIPTION
#### Please note that the only changes in the Device Description relate to replacing the patient trigger magnet with a button.
The NUVANT MCT System consists primarily of the PiX monitoring device and the zLink data transmission device. Once activated, the wearable PiX sensor continuously monitors the heart and automatically collects ECGs. When rhythm abnormalities are detected, data are automatically transmitted from the PiiX device to the zLink, which then automatically transmits the data to the Corventis Monitoring Center. Patients can also trigger transmission of ECGs when they experience cardiac symptoms by using the Patient Trigger Button. Certified cardiographic technicians at the Corventis Monitoring Center review received data and document symptoms reported by patients. Clinical reports, prepared by the Corventis Monitoring Center, are delivered and made available at www.corventis.com to provide data to prescribing physicians for the diagnosis and identification of various clinical conditions, events and/or trends.
#### 5.1.1 HOW THE NUVANT MCT SYSTEM FUNCTIONS
#### Please note this is identical to the predicate NUVANT MCT System.
Once activated, the wearable PiiX sensor continuously monitors the heart and automatically transmits ECGs when rhythm abnormalities are detected.
Communication between the PiiX and zLink is enabled via BlueTooth™ technology, and communication between the zLink and the secure Corventis server is enabled via cellular technology.
### 5.1.2 SCIENTIFIC CONCEPTS OF THE DEVICE
The subject NUVANT MCT System uses proprietary algorithms based on rate, rhythm and morphology to continuously analyze rhythm abnormalities and to initiate automatic ECG transmission to the Server.
### 5.1.3 SIGNIFICANT PHYSICAL AND PERFORMANCE CHARACTERISTICS OF THE DEVICE
{6}------------------------------------------------
The subject NUVANT MCT System is comprised of the following non-sterile components:
- . PiiX
- zLink .
- Server ●
#### Please note these have always been cleared as a System, most recently under K113372.
### PIIX
The PiiX is a patient-worn device applied to the patient's torso. It contains the electrode assembly and electronics module for recording ECG and heart rate data. The primary battery is contained in the electrode assembly and is a non-rechargeable Lithium Manganese Dioxide Thin Cell. The electronics module has a small 3V Manganese Silicon Lithium Rechargeable battery to maintain the Real Time Clock of the electronics module.
The electrode assembly (skin contacting portion) contains the following components that directly contact patient skin:
- . Medical Grade Hydrogel Adhesive
- Medical Grade Clear Polyurethane Tape .
The PiiX device lasts 7.5 days, and each package can contain up to four PiiX devices for up to 30 days of uninterrupted wear, depending on prescription length.
The entire electrode assembly meets requirements of the ISO 10993 biocompatibility standard.
### ZLINK
### Please note there are no changes to the zLink in this submission.
The zLink is the patient hand-held transceiver that receives information from the PiiX and transmits it to the Corventis Server. It also interacts with the Corventis Server to receive system configuration data and other relevant hardware diagnostic information.
{7}------------------------------------------------
The zLink components are surrounded by a plastic housing and the zLink contains a rechargeable Lithium Ion battery.
#### SERVER
The Server receives information from the PiiX via the zLink. The secure server, among other things, derives, calculates and displays the patient's physiological parameters using the data collected by the PiiX.
### 5.2 INTENDED USE OF THE DEVICE
### Please note that the Intended Use of the Device is identical to the Predicate NUVANT MCT System.
The NUVANT MCT System is intended for the ambulatory recording and monitoring of physiological parameter(s).
### 5.3 INDICATIONS FOR USE STATEMENT
### Please note that the Indications for Use are identical to the predicate NUVANT MCT System.
The NUVANT Mobile Cardiac Telemetry (MCT) System is intended to continuously measure, record and periodically transmit physiological data. The System is indicated for those patients who require monitoring for the detection of non-lethal cardiac arrhythmias such as, but not limited to, supraventricular tachycardias (e.g. atrial fibrillation, atrial flutter, paroxysmal SVTs), ventricular ectopy, bradyarrhythmias and conduction disorders. The NUVANT System monitors, derives and displays:
- ECG .
- . Heart Rate
### 5.4 TECHNOLOGICAL CHARACTERISTICS OF THE SUBJECT NUVANT МСТ
The subject NUVANT MCT System operates identically to the predicate, based on the following fundamental scientific technology:
{8}------------------------------------------------
- . The collection of physiological parameters by a multi-sensor patientworn device (PiiX);
- The transmission of these parameters to a remote Server through a . transceiver (zLink); and
- The receipt of the parameters by the Server and subsequent derivation . into appropriate useful values for display.
# This fundamental scientific technology is identical to the predicate NUVANT MCT System.
The only differences are non-fundamental to the scientific technology, as explained below :
- 1. Increase memory capacity of the PiiX to allow for continuous ECG recording and storage in device memory
- 2. Convert the PiiX into two parts consisting of an electrode assembly and an electronics module that can only be disassembled in the factory setting
- 3. Replace the patient trigger magnet with a push button on the PiiX
These proposed changes will not impact the fundamental scientific technology of the device. The subject NUVANT MCT system uses the same sensors, in the same scientific methods (algorithms for detection and transmission methods), as the predicate NUVANT MCT System.
# 5.5 SUMMARY OF TESTING
All of the proposed changes to the subject NUVANT MCT System were fully verified and validated in accordance with design control requirements.
## 5.5.1 SUMMARY OF TESTS USED TO DEMONSTRATE SUBSTANTIAL EQUIVALENCE
The Subject NUVANT MCT System is supported by the following tests to demonstrate substantial equivalence. Note that the zLink is unchanged in this submission, but was used in system validation tests and is further supported by the same tests as the Predicate NUVANT MCT System.
# PiiX Verification and Validation Testing
{9}------------------------------------------------
K133701 Page 6 of 9
- Biocompatibility Testing .
- . Mechanical Verification Testing
- Electrical Verification Testing .
- Firmware Verification Testing .
- System Validation Testing .
### Server Verification and Validation Tests
- Software Verification Testing ●
- . System Validation Testing
## 5.5.2 GUIDANCE DOCUMENTS USED / CONSIDERED IN THESE DEVICE MODIFICATIONS
Guidance documents used/considered for these device modifications include, but are not limited to, the following:
- . Draft Guidance Document titled Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, issued June 14, 2013
- Draft Guidance Document titled Use of International Standard ISO . 10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing, issued on April 23, 2013
- Draft Guidance Document titled Design Considerations for Devices . Intended for Home Use, issued December 12, 2012
- Radio Frequency Wireless Technology in Medical Devices, issued on . August 14, 2013
- Guidance for the Content of Premarket Submissions for Software . Contained in Medical Devices, issued May 11, 2005
- Class II Special Controls Guidance Document, Arrhythmia Detector and . Alarm, issued on October 28, 2003.
### 5.5.3 STANDARDS USED TO DEMONSTRATE SUBSTANTIAL EQUIVALENCE
The following standards were also used, in whole or in part, to demonstrate substantial equivalence:
{10}------------------------------------------------
| Standards No. | Standards Title | FDA<br>Recognition<br>No. |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| IEC 60601-1 | Medical electrical equipment -<br>Part 1: General Requirements for<br>Basic Safety and Essential<br>Performance, 2006/A11:2011 | 5-4 |
| IEC 60601-1-2 | Medical electrical equipment -<br>Part 1-2: General requirements<br>for basic safety and essential<br>performance - Collateral<br>standard: Electromagnetic<br>compatibility - Requirements and<br>tests (Edition 3), 2007 | 5-53 |
| ANSI/AAMI/IEC<br>60601-2-47:2012 | Medical electrical equipment -<br>Part 2-47: Particular<br>requirements for the safety and<br>essential performance of<br>ambulatory electrocardiographic<br>systems | This<br>standard<br>replaces<br>EC38 (FDA<br>recognition<br>number 3-<br>65) |
| AAMI/ANSI/IEC<br>62304:2006 | Medical Device Software -<br>Software life cycle processes | 13-32 |
| AAMI/ANSI EC12 | Disposable ECG electrodes,<br>2000(R) 2010 | 3-52 |
| Standards No. | Standards Title | FDA<br>Recognition<br>No. |
| AAMI/ANSI EC57 | Testing and reporting<br>performance results of cardiac<br>rhythm and ST-segment<br>measurement algorithms,<br>1998/(R) 2008 | 3-73 |
| AAMI/ANSI/ISO<br>10993-1:2009 | Biological evaluation of medical<br>devices - Part 1: Evaluation and<br>testing within a risk management<br>process | 2-156 |
| AAMI/ANSI/ISO<br>10993-5:2009 | Biological evaluation of medical<br>devices - Part 5: Tests for in<br>vitro cytotoxicity | 2-153 |
| AAMI/ANSI/ISO<br>10993-10:2010 | Biological evaluation of medical<br>devices - Part 10: Tests for<br>irritation and skin sensitization | 2-174 |
{11}------------------------------------------------
The results of the testing indicate that the subject NUVANT MCT System performs as intended. Risk analysis and testing further confirmed that the changes did not introduce any new issues of safety or effectiveness, and the system performs as well as the predicate device.
### 5.6 CONCLUSION
The Indications for Use for the subject NUVANT MCT System are identical to our previous device (K113372). We have provided summary data to demonstrate reasonable assurance of the safety and effectiveness of the subject NUVANT MCT System and to demonstrate substantial equivalence to its predicate.
As supported by the descriptive information, verification, validation and standards testing, the modified NUVANT MCT System is as safe and effective, and performs as well as or better than the predicate device.
{12}------------------------------------------------
Therefore, the NUVANT MCT System is substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.