BREAST-MED TISSUE MARKER

K133697 · Breast-Med, Inc. · NEU · Apr 10, 2015 · General, Plastic Surgery

Device Facts

Record IDK133697
Device NameBREAST-MED TISSUE MARKER
ApplicantBreast-Med, Inc.
Product CodeNEU · General, Plastic Surgery
Decision DateApr 10, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4300
Device ClassClass 2

Indications for Use

The Breast-Med Tissue Marker is indicated for use to radiographically mark soft tissue during a surgical procedure or for future surgical procedures.

Device Story

The Breast-Med Tissue Marker is a sterile, single-use implantable device consisting of a polymeric tube containing a desiccated sodium chloride solution with trace gadolinium chelate. It is placed into soft tissue during open, percutaneous, or endoscopic procedures to serve as a radiographic marker. The device is visible on standard radiographs (x-ray, mammography), ultrasound, and MRI. It is used by clinicians to mark surgical locations for future reference, aiding in the identification of specific tissue sites for subsequent procedures or monitoring.

Clinical Evidence

Bench testing only. Testing verified equivalent radiographic visualization to predicate devices under standard radiographs, ultrasound, and MRI in both tissue and phantom models. Safety testing conducted per ASTM F2182, ASTM F2052, ASTM F2213, and ASTM F2119.

Technological Characteristics

Polymeric tube containing desiccated sodium chloride and trace gadolinium chelate. Implantable clip (21 CFR 878.4300). Safety testing performed per ASTM F2182, ASTM F2052, ASTM F2213, and ASTM F2119. Sterile, nonpyrogenic, single-use.

Indications for Use

Indicated for patients requiring radiographic marking of soft tissue during or for future surgical procedures.

Regulatory Classification

Identification

An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked one behind the other, with a wing-like shape extending from the back of the heads. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 10, 2015 Breast-Med Incorporated Dr. Michael Nelson President 1745 Bridgewater Road Golden Valley, MN 55422 Re: K133697 Trade/Device Name: Breast-Med Tissue Marker Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable clip Regulatory Class: Class II Product Code: NEU Dated: April 6, 2015 Received: April 8, 2015 Dear Dr. Nelson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {2}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # David Krause -S - Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K133697 Device Name Breast-Med Tissue Marker Indications for Use (Describe) The Breast-Med Tissue Marker is indicated for use to radiographically mark soft tissue during a surgical procedure or for future surgical procedures. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ## 510(K) SUMMARY - K133697 #### In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the Breast-Med Tissue Marker is provided below. | Submitted by: | Breast-Med, Inc. | |-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Michael Nelson, M.D.<br>President<br>Breast-Med, Inc.<br>1745 Bridgewater Road<br>Golden Valley, MN 55422<br>Tel: 763-522-2121<br>Email: mtnelson@comcast.net | | Date of Summary: | April 10, 2015 | | Device Trade Name: | Breast-Med Tissue Marker | | 510(k) Number: | K133697 | | Product Code: | NEU | | Common or Usual Name: | Tissue Marker | | Classification Name: | Implantable clip (21 CFR 878.4300) | | Predicate Device(s): | BiomarC Tissue Marker, K032347<br>KDM-Mark1 Tissue Marker, K093473 | | Device Description: | The Breast-Med Tissue Marker is a sterile, nonpyrogenic, single<br>use tissue marker consisting of a polymeric tube filled with a<br>dessicated solution of sodium chloride with trace amounts of<br>gadolinium chelate that is visible on standard radiographs (x-<br>ray, mammography) as well as ultrasound and Magnetic<br>Resonance Imaging (MRI). The tissue marker is placed into<br>soft tissue during open, percutaneous, or endoscopic<br>procedures to radiographically mark a surgical location. | | Indication for Use: | The Breast-Med Tissue Marker is indicated for use to<br>radiographically mark soft tissue during a surgical procedure or<br>for future surgical procedures. | {5}------------------------------------------------ | Technological<br>Characteristics: | The technological characteristics of the Breast-Med Tissue<br>Marker are substantially equivalent to those of the predicates in<br>terms of the following:<br>Intended use Indications for use Target population Fundamental scientific technology Patient-contacting materials Method of delivery | |-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Performance Testing<br>Summary: | Bench testing was performed to verify that the subject device<br>has equivalent radiographic visualization to the predicates under<br>standard radiographs, ultrasound, and MRI in both tissue and<br>phantom models.<br><br>Safety testing of the device was performed utilizing the following<br>standards: ASTM F2182, ASTM F2052, ASTM F2213 and<br>ASTMF2119. | | Conclusion: | The Breast-Med Tissue Marker is substantially equivalent to the<br>named predicates based on technological comparison,<br>indications for use, and laboratory and other safety testing. It is<br>concluded that there are no new questions of safety and<br>effectiveness. |
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...