K133685 · Am USA · NGT · Aug 13, 2014 · General Hospital
Device Facts
Record ID
K133685
Device Name
0.9% SODIUM CHLORIDE FLUSH SYRINGE
Applicant
Am USA
Product Code
NGT · General Hospital
Decision Date
Aug 13, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Indications for Use
0.9% Sodium Chloride Flush Syringes are intended for use in flushing compatible intravenous administration sets and indwelling intravenous access devices. Use according to the recommendations of the manufacturer for the appropriate device.
Device Story
Device is a 20 mL non-sterile plastic syringe pre-filled with 20 mL of 0.9% Sodium Chloride USP; terminally sterilized via gamma radiation. Used in clinical settings by healthcare providers to flush intravenous administration sets and indwelling vascular access devices. Operates mechanically via manual plunger depression; no energy source. Provides a sterile fluid path for saline delivery. Benefits include maintenance of catheter patency and prevention of occlusion. Device is single-use; contains no preservatives, antimicrobial agents, plasticizers, or surfactants.
Clinical Evidence
Bench testing only. Testing included sterility, pyrogen (LAL) analysis, pH, water loss, iron, heavy metals, sodium/chloride content, and particulate matter. Results demonstrate performance equivalent to the predicate device.
Technological Characteristics
Materials: Polypropylene homopolymer (PF535), Dow Corning 360 Medical fluid, pharmaceutical-grade rubber plunger tips, Alpha Gray TPE G968B 4170 tip cap. Categorized per ISO 10993-1:2009 as external communicating device, blood path, indirect, limited contact (≤24 hours). Sterilization: Gamma radiation (SAL 10^-6). No energy source. Manual operation.
Indications for Use
Indicated for flushing compatible intravenous administration sets and indwelling intravenous access devices in patients requiring vascular access maintenance.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 13, 2014
AMUSA Mr. Alex Ferri VP/GM of Operations 5209 Linbar Dr. Suite 640 Nashville, TN 37211
Re: K133685
Trade/Device Name: 0.9% Sodium Chloride Flush Syringe Regulation Number: 21 CFR 880.5200 Regulation Name: Device, Flush, Vascular Access Regulatory Class: II Product Code: NGT Dated: July 11, 2014 Received: July 15, 2014
Dear Mr. Ferri:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Ferri
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mary S. Runner -
S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
Indications for Use
# Indications for Use Statement
510(k) Number (if known): ¿ 33 685
Device Name: 0.9% Sodium Chloride Flush Syringe
Indications for Use:
"0.9% Sodium Chloride Flush Syringes are intended for use in flushing compatible intravenous administration sets and indwelling intravenous access devices. Use according to the recommendations of the manufacturer for the appropriate device".
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
20mL Sterile 510(K) Attachment_ Page Le of 68 AMUSA. Nashville, TN
510(k) Summary
Image /page/3/Picture/4 description: The image shows the logo for AMUSA, a manufacturing company that produces quality products for healthcare. The logo features a stylized American flag on the left, followed by the company name in bold, sans-serif font. The "A" in AMUSA is stylized with a star in the middle. Below the company name is the tagline "MANUFACTURING QUALITY PRODUCTS FOR HEALTHCARE" in a smaller font.
# 510(k) Summary
As required by 809.92(a) (2).
. . . . . . .
510 (k) PREMARKET NOTIFICATION NUMBER: K133685
## Submitter and Owner of the 510(k)
AMUSA 5209 Linbar Dr., Suite 640 Nashville, TN 37211 Phone: 615-833-2633 Fax: 615-425-2772
## Official Correspondent
Alex Ferri VP/GM of Operations __________________________________________________________________________________________________________________________________________________________ AMUSA 5209 Linbar Dr., Suite 640 Nashville, TN 37211 Phone: 615-833-2633 Fax: 615-425-2772
### Date of Preparation: 5/15/14
## 510(k) Application Number
#### Trade/Proprietary Name
0.9% Sodium Chloride Flush Syringe Common Name/Usual Name Saline Flush Syringe Device Classification Name Device, Flush, Vascular Access Regulation Number 880.5200 Device Class Class II Device Classification Panel General Hospital Classification Product Code NGT
{4}------------------------------------------------
20mL Sterile 510(k) Attachment AMUSA, Nashville,
## THE PREDICATE DEVICE DESCRIPTION:
The Predicate Device, 510(k) Number: K120836 (AMUSA) consists of a non-sterile plastic 12 cc syringe filled with 3 mL, or 10 mL of 0.9% Sodium Chloride of a nor-selection that is terminally sterilized. The predicate device is sterile on the outside and has a sterile fluid path. The Sterility Assurance Level (SAL) is 10°. This is a single use device. The solution contains no preservatives, or antimicrobial agents. No plasticizers, additives, crosslink agents, reagents, or surfactants were used during the manufacture of the syringes. The syringe may be used on a sterile field if packaging is not damaged and aseptic technique is used.
#### Indications For Use:
Intended use: 0.9% Sodium Chloride Flush Syringes are intended for use in flushing compatible intravenous administration sets and indwelling intravenous access devices. Il se according to the recommendations of the manufacture for the appropriate device.
#### THE NEW DEVICE DESCRIPTION:
The new device, the subject of this 510(k), consists of a 20 mL, non-sterile plastic syringe filled with 20 mL of 0.9% Sodium Chloride Flush Solution USP that is terminally sterilized. The new devine is starle on the outside and has a sterile fluid path. The Sterility Assurance Level (SAL) is 10°. The syringe may be used on a sterile field if packaging is not damaged and aseptic technique is used. This is a ningle use device. The solution contains no preservatives, or antimicrobial agents. No plasticizers, additives, crosslink agents, reagents, colorants, inks, adhesives, surfactants, detergents, used during the manufacture of the syringes. The 20mL filled with 0.9% Sodium Chloride Flush Solution in a 20mL syringe is the only model which is subject to this 510(k). Other models previously cleared under 510K (K120836) consists of a non-sterile plastic 12 cc syringe filled with 3 mL, 5 mL, or 10 mL of 0.9% Sodium Chloride Flush Solution that is terminal sterilized.
#### Indications For Use:
Intended use: 0.9% Sodium Chloride Flush Syringes are intended for use in flushing compatible intravenous administration sets and indwelling intravenous access devices. Use according to the recommendations of the manufacture for the appropriate device.
#### TECHNILOGICAL/SUBSTANTIAL EQUIVALENCE SUMMARY: The technical
characteristics for the new device do not differ from those of the currently marketed device. These devices have the same design, the same fundamental scientific and chemical characteristics, the same labeling, and have the same intended use. The differences in the proposed new device are an increase in the size and fill volume of the syringe. All other aspects of the product design remain the same. This device is categorized in ISO 10993-1:2009 as "External communicating device – blood path, indirect per section 5.2.2(a). The device will have limited contact of less than or equal to 24 hours. All materials contact the fluid path and are treated as indirect patient contact. The syringe device is fabricated from the following raw materials:
- . Polypropylene homopolymer pellets (PF535)
- Dow Corning 360 Medical fluid .
- West Pharmaceutical Rubber Plunger Tips .
- . Tip Cap Non-sterile Basell PF535 Polypropylene, Alpha Gray TPE G968B 4170
No other materials including plasticizers, additives, crosslink agents, colorants, inks, adhesives, surfactants, detergents, etc., are used during the manufacture of the syringes. The device does not use an energy source.
{5}------------------------------------------------
20mL Sterile 510(K) Attachment Page_ र of (o AMUSA. Nashville, TN
## Technological/Substantial Equivalence Comparison Table:
| Technological/Substantial Equivalence Table | | |
|-----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Predicate Device | New Device |
| Intended Use | Intended use: 0.9% Sodium Chloride Flush Syringes<br>are intended for use in flushing compatible<br>intravenous administration sets and indwelling<br>intravenous access devices. | Intended use: 0.9% Sodium Chloride Flush Syringes are intended for<br>use in flushing compatible intravenous administration sets and<br>indwelling intravenous access devices. |
| Design | Predicate device and new device have identical<br>design. | New device has the same design as predicate. |
| Chemical<br>Composition<br>of Barrel | Non-sterile Polypropylene PF-535 Resin | Non-sterile Polypropylene PF-535 Resin |
| Chemical<br>Composition<br>of Plunger | Non-sterile Black Pharmaceutical Grade Rubber (Not<br>made with Natural Rubber Latex) | Non-sterile Black Pharmaceutical Grade Rubber (Not made with<br>Natural Rubber Latex) |
| Chemical<br>Composition<br>Tip Cap | Non-sterile Basell PF535 Polypropylene, Alpha Gray<br>TPE G968B 4170 | Non-sterile Basell PF535 Polypropylene, Alpha Gray TPE G968B 4170 |
| Environment | Positive Differential<br>Controlled and Monitored | Positive Differential<br>Controlled and Monitored |
| Energy<br>Source | There is not an energy source. The device is operated<br>mechanically. | There is not an energy source. The device is operated mechanically. |
| Non-<br>Pyrogenic | Entire device is nonpyrogenic | Entire device is nonpyrogenic |
| Non-Pyrogenic<br>Additives,<br>latex, or<br>preservatives | No additivies, natural rubber, or preservatives were<br>used in the manufacture of the syringes. | No additivies, natural rubber, or preservatives were used in the<br>manufacture of the syringes. |
| Sterile<br>Barrier<br>Package | Sterile barrier Pouch<br>polypropylene film 3-6 mil | Sterile barrier Pouch<br>polypropylene film 3-6 mil |
| Sterility<br>Assurance<br>Level | 10-6 | 10-6 |
| Sterility<br>Labeling of<br>Device | Fluid path and outside of device are sterile. | Fluid path and outside of device are sterile. |
| Sterilization<br>Method | Gamma | Gamma |
| Expiry Dating | 2 years | 2 years |
| Filling<br>Process | The filling process occurs in a positive differential,<br>monitored, and controlled environment. The<br>solution is contained in a sterile closed system. | The filling process occurs in a positive differential, monitored, and<br>controlled environment. The solution is contained in a sterile closed<br>system. |
| Equivalent<br>Bench and<br>Stability<br>Testing | The predicate device Bench and Stability tests<br>included:<br>• Sterility<br>• Pyrogen (LAL) analysis<br>• pH<br>• Water Loss<br>• Iron<br>• Heavy Metals<br>• Sodium, Chloride<br>• Particulate Matter | The same Bench and Stability tests were performed on the new<br>device as the predicate device. Bench and Stability tests included:<br>• Sterility<br>• Pyrogen (LAL) analysis<br>• pH<br>• Water Loss<br>• Iron<br>• Heavy Metals<br>• Sodium, Chloride<br>• Particulate Matter |
| Size and Fill<br>Volumes | 3 mL/12 mL; 5 mL/12mL, 10 mL/12mL | 20 mL/20 mL |
Conclusion: The new device is identical to the predicate device in plastic materials composition, the form, and function (e.g. Flushing of vascular access devices). The two changes to the new device involve 1. Increase in syringe size from a 12 mL syringe to a 20 mL syringe and 2. Increase in the solution volume. Non-clinical verification testing for the proposed change involved chemical-physical, functional, and product stability testing. The results of testing conducted verify the new device performed in an equivalent manner to the predicate device. Therefore, the new device is substantially equivalent to the predicate device. Other companies have FDA 510(k) clearance for similar devices.
8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.