← Product Code [JAK](/productcode/JAK) · K133589

# SOMATOM FORCE (K133589)

_Siemens Medical Solutions USA, Inc. · JAK · Apr 17, 2014 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K133589

## Device Facts

- **Applicant:** Siemens Medical Solutions USA, Inc.
- **Product Code:** [JAK](/productcode/JAK.md)
- **Decision Date:** Apr 17, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1750
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis.

## Device Story

Whole body X-ray Computed Tomography (CT) system; utilizes two continuously rotating tube-detector systems; operates on fan beam principle. Inputs: X-ray transmission data. Processing: Computer reconstruction of cross-sectional images; includes Iterative Beam Hardening Correction (IBHC) using 3D forward projection and two-compartment iodine/water model to reduce artifacts. Outputs: DICOM format CT images. Used in clinical settings by trained physicians/technicians. Output used by physicians as diagnostic aid; facilitates anatomical visualization and clinical decision-making. Benefits: Improved image quality via artifact reduction.

## Clinical Evidence

Bench testing only. Performance demonstrated through non-clinical integration, functional, and phantom testing. Verification and validation testing confirmed software specifications met acceptance criteria and device conforms to recognized electrical and mechanical safety standards (IEC/NEMA).

## Technological Characteristics

Whole body CT scanner; 2x Vectron X-ray tubes; 120kW generator; 96 detector rows; 192 slices/rotation. Connectivity: DICOM. Software: Command-based for scan control, reconstruction, and archive. Standards: IEC 60601-2-44, IEC 60601-2-28, IEC 61223-3-5, NEMA XR 25, ISO 10993-1.

## Regulatory Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- SOMATOM Definition Flash ([K121072](/device/K121072.md))

## Reference Devices

- SOMATOM Definition Flash ([K082220](/device/K082220.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

SIEMENS

K133589

APR 1 7 2014

510(K) SUMMARY FOR . SOMATOM Force

Submitted by: Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway Malvern, PA 19355

Date Prepared: November 21, 2013

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

- 1. General Information
Importer/Distributor: Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway Malvern, PA 19355

Establishment Registration Number: 2240869

Manufacturing Site: SIEMENS AG. MEDICAL SOLUTIONS Siemensstrasse 1 91301 Forchheim, GERMANY

Establishment Registration Number: 3004977335

#### 2. Contact Person:

Mrs. Kimberly Manqum Technical Specialist, Regulatory Affairs Submissions Siemens Medical Solutions, Inc. USA 51 Valley Stream Parkway D02 Malvern, PA 19355-1406 Phone: (610) 448-1772 Fax: (610) 448-1778 Email: kimberly mangum@siemens.com

#### Device Name and Classification 3.

SOMATOM Force Product Name: SOMATOM Force Propriety Trade Name: Computed Tomography X-ray System Classification Name: Radiology Classification Panel: 21 CFR §892.1750 CFR Section:

{1}------------------------------------------------

Device Class: Class II Product Code: 90JAK Legally Marketed Predicate Devices SOMATOM Definition Flash Trade Name: K21072 510(k)#: May 08, 2012 Clearance Date: System, X-Ray, Tomography, Computed Classification Name: Classification Panel: Radiology Classification Regulation:21 CFR § 892.1750 Device Class: 90 JAK Product Code:

#### 4. Substantial Equivalence:

Siemens SOMATOM Force is substantially equivalent to the following medical devices in commercial distribution:

| Predicate Device Name    | FDA Clearance<br>Number | FDA Clearance<br>Date |
|--------------------------|-------------------------|-----------------------|
| SOMATOM Definition Flash | K121072                 | May 08, 2012          |

#### Device Description: 5.

The SOMATOM Force is a whole body X-ray Computed Tomography System which features two continuously rotating tube-detector systems and functions according to the fan beam principle. The SOMATOM Force produces CT images in DICOM format, which can be used by postprocessing applications commercially distributed by Siemens and other vendors.

The system software is a command-based program used for patient management, data management, X-ray scan control, image reconstruction, and image archive/evaluation.

The computer system delivered with the CT scanner is able to run the post processing applications optionally.

- Summary of Technical Characteristics of the Subject Device as 6. Compared with the Predicate Device:
The SOMATOM Force is comparable in indications for use, design, material, functionality, technology, energy source and is considered substantially equivalent to the predicate device SOMATOM Definition Flash (K121072, clearance date May 08, 2012).

{2}------------------------------------------------

The differences between the legally marketed predicate device and the SOMATOM Force are as follows:

| Property                                    | SOMATOM Force                                    | SOMATOM Definition Flash                 |
|---------------------------------------------|--------------------------------------------------|------------------------------------------|
| X-Ray Assembly Tube                         | 2 × Vectron                                      | 2 × Straton MX P                         |
| Generator Power (max.)                      | 120 kW/120 kW                                    | 100 kW/100 kW                            |
| Number of Detector Rows                     | 96/96                                            | 64/64                                    |
| Maximum Number of<br>slices/rotation        | 192/192                                          | 128/128                                  |
| Number of measuring<br>channels in detector | 1840/1280                                        | 1472/960                                 |
| Beam Hardening<br>Correction                | Iterative Beam<br>Hardening Correction<br>(IBHC) | Posterior Fossa<br>Optimization<br>(PFO) |

The SOMATOM Force is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission.

The SOMATOM Force also supports Iterative Beam Hardening Correction (IBHC). The Iterative Beam Hardening Correction (IBHC) feature is an extension of the Posterior Fossa Optimization (PFO) feature. The PFO feature was cleared as part of the SOMATOM Definition Flash (K082220, clearance date October 10, 2008) and part of the predicate device SOMATOM Definition Flash (K121072, clearance date May 08, 2012).

IBHC is a raw data based beam hardening correction method designed to improve image quality. IBHC uses 3D forward projection, and an additional two-compartment iodine/water model to reduce beam hardening artifacts while maintaining anatomical structures.

The intended use, materials, energy source, and fundamental scientific technology are similar to the predicate device; therefore Siemens believes that they are substantially equivalent to the predicate device.

{3}------------------------------------------------

### 7. Nonclinical Testing:

The SOMATOM Force is designed to fulfill the requirements of following safety and performance standards:

| Recognition Product<br>Number | Area            | Title of Standard                                                                                                                                                                                 | Reference<br>Number and<br>Date        | Publication<br>Date | Standards<br>Development<br>Organization |
|-------------------------------|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|---------------------|------------------------------------------|
| 2-156                         | Biocomp         | Biological evaluation of<br>medical devices - Part 1:<br>Evaluation and testing within<br>a risk management process                                                                               | 10993-1:2009                           | 01/15/2013          | AAMI ANSI ISO                            |
| 12-120                        | Radiology       | Medical electrical equipment<br>- Part 2-44: Particular<br>requirements for the safety of<br>X-ray equipment for<br>computed tomography - Ed.<br>2.1                                              | 60601-2-44<br>(2002-11):               | 09/09/2008          | IEC                                      |
| 12-126                        | Radiology       | Medical electrical equipment -<br>Part 2-28: Particular<br>requirements for the safety of X-<br>ray source assemblies and X-ray<br>tube assemblies for medical<br>diagnosis - Ed. 1.0             | 60601-2-28:<br>1993                    | 10/31/2005          | IEC                                      |
| 12-127                        | Radiology       | Medical electrical equipment<br>- Part 2-32: Particular<br>requirements for the safety<br>of associated equipment of<br>X-ray equipment - Ed. 1.0                                                 | 60601-2-32:<br>1994                    | 10/31/2005          | IEC                                      |
| 12-199                        | Radiology       | Medical electrical equipment -<br>Part 1-3: General requirements<br>for basic safety 3. Collateral<br>Standard: General reguirements<br>for radiation protection in<br>diagnostic X-ray equipment | 60601-1-3 First<br>edition 1994-07     | 09/08/2009          | IEC                                      |
| 12-204                        | Radiology       | Medical electrical equipment<br>- Part 2-28: Particular<br>requirements for the basic<br>safety and essential<br>performance of X-ray tube<br>assemblies for medical<br>diagnosis                 | 60601-2-28<br>Edition 2.0<br>2010-03   | 08/05/2013          | IEC                                      |
| 12-210                        | Radiology       | Medical electrical equipment -<br>Part 1-3: General requirements<br>for basic safety and essential<br>performance - Collateral<br>Standard: Radiation protection in<br>diagnostic X-ray equipment | 60601-1-3<br>Edition 2.0<br>2008-01    | 08/05/2013          | IEC                                      |
| Recognition<br>Number         | Product<br>Area | Title of Standard                                                                                                                                                                                 | Reference<br>Number and<br>Date        | Publication<br>Date | Standards<br>Development<br>Organization |
| 12-223                        | Radiology       | Evaluation and routine<br>testing in medical imaging<br>departments - Part 3-5:<br>Acceptance tests - Imaging<br>performance of computed<br>tomography X-ray equipment<br>CORRIGENDUM 1           | 61223-3-5<br>(First edition<br>2004)   | 03/18/2011          | IEC                                      |
| 12-225                        | Radiology       | Computed Tomography Dose<br>Check                                                                                                                                                                 | XR 25                                  | 03/18/2011          | NEMA                                     |
| 12-226                        | Radiology       | Evaluation and routine testing in<br>medical imaging departments -<br>Part 2-6: Constancy tests -<br>Imaging performance of<br>computed tomography X-ray<br>equipment                             | 61223-2-6<br>Second Edition<br>2006-11 | 02/28/2011          | IEC                                      |
| 12-238                        | Radiology       | Digital Imaging and<br>Communications in Medicine<br>(DICOM) Set                                                                                                                                  | PS 3.1 - 3.18                          | 03/16/2012          | NEMA                                     |
| 13-8                          | Software        | Medical device software -<br>Software life cycle processes                                                                                                                                        | 62304 First<br>edition 2006-05         | 08/20/2012          | IEC                                      |
| 5-27                          | General         | Medical electrical equipment --<br>Part 1-1: General requirements<br>for safety - Collateral standard:<br>Safety requirements for medical<br>electrical systems                                   | 60601-1-<br>1:2000                     | 09/08/2009          | IEC                                      |
| 5-4                           | General         | Medical Electrical Equipment -<br>Part 1: General Requirements<br>for Safety, 1988; Amendment 1,<br>1991-11, Amendment 2, 1995                                                                    | 60601-1                                | 10/31/2005          | IEC                                      |
| 5-40                          | General         | Medical devices - Application of<br>risk management to medical<br>devices                                                                                                                         | 14971 Second<br>Edition 2007-<br>03-01 | 08/20/2012          | ISO                                      |
| 5-41                          | General         | Medical electrical equipment -<br>Part 1-4: General requirements<br>for safety --<br>Collateral standard:<br>Programmable electrical medical<br>systems, edition 1.1                              | 60601-1-<br>4:2000 Consol.<br>Ed. 1.1  | 09/08/2009          | IEC                                      |
| 5-54                          | General         | Medical electrical equipment -<br>Part 1-2: General requirements<br>for basic safety and essential<br>performance - Collateral<br>                                                                | 60601-1-<br>2:2007                     | 08/05/2013          | AAMI ANSI IEC                            |

{4}------------------------------------------------

This submission contains performance data to demonstrate continued conformance with special controls for medical devices containing software. Non clinical tests (integration and functional) and phantom testing were conducted for the SOMATOM Force during product development. The modifications described in this Premarket Notification were supported with verification/validation testing.

{5}------------------------------------------------

The Risk analysis was completed and risk control implemented to mitigate identified hazards. The testing results supports that all the software specifications have met the acceptance criteria. Testing for verification and validation of the device was found acceptable to support the claims of substantial equivalence.

Software Documentation for a Moderate Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005 is also included as part of this submission.

EMC/electrical safety was evaluated according to the IEC Standards. Siemens certify conformance to Voluntary Standards covering Electrical and Mechanical Safety. In conclusion, the identified risk of electrical hazards was mitigated and is substantially equivalent to the predicate device in terms of safety and effectiveness. All testing and validation has been completed.

### 8. Indications for Use:

This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis.

### 9. General Safety and Effectiveness Concerns:

The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device.

Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled during development, verification and validation testing. To minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practice and standards.

### 10. Conclusion as to Substantial Equivalence

In summary, Siemens is of the opinion that the SOMATOM Force does not introduce any new potential safety risk and is substantially equivalent to and performs as well as the predicate devices.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is a stylized symbol that resembles a person with outstretched arms, with three wavy lines extending from the person's arms.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 17, 2014

Siemens Medical Solutions USA, Inc. % Kimberly Mangum 51 Valley Stream Pkwy. MALVERN PA 19355

Re: K133589/S002

Trade/Device Name: SOMATOM Force Regulation Number: 21 CFR 892.1750 Regulation Name: System, X-Ray, Tomography, Computed Regulatory Class: II Product Code: JAK Dated: March 14, 2014 Received: March 18, 2014

Dear Ms. Mangum:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{7}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/McdicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

:

:

Sincerely yours,

Smh

for

Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

510(k) Number (if known)

K133589

Device Name SOMATOM Force

Indications for Use (Describe)

This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by a trained physician as an aid in diagnosis.

Type of Use (Select oner both, as applicable)

🇿 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

Image /page/8/Figure/12 description: The image shows a document with the heading "FOR FDA USE ONLY". Below the heading, there is a line that reads "Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)". An FDA logo is present, presumably indicating that the document is related to the Food and Drug Administration. The document appears to be an official form or record intended for internal use within the FDA.

---

**Source:** [https://fda.innolitics.com/device/K133589](https://fda.innolitics.com/device/K133589)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
